      "status": "Terminated",
      "city": "Snoqualmie",
      "state": "WA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "67243",
      "recalling_firm": "Spacelabs Healthcare, Llc",
      "address_1": "35301 Se Center St",
      "address_2": "",
      "postal_code": "98065-9216",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) including states of: AZ, CA, FL, GA, IA, IL, LA, MA, ME, MI, MO, MS, NY, PA, SC, TN, TX, VA, and Puerto Rico; and Internationally to: Canada, Chile, and Panama.",
      "recall_number": "Z-0871-2014",
      "product_description": "ARIATELE TELEMETRY TRANSMITTER, Model 96281, with SpO2 Option C.    Model 96281 is intended for use with either adult or neonatal patient  populations in a hospital environment.",
      "product_quantity": "1057 units (988 in the US and 69 outside of the US).",
      "reason_for_recall": "The AriaTele Model 96281 with SpO2 monitoring Option C is recalled because the AriaTele (transmitter) display and the Telemetry Central Station display may show a SpO2 value when the sensor is not connected to the patient or in certain sensor failed conditions.",
      "recall_initiation_date": "20131219",
      "center_classification_date": "20140130",
      "termination_date": "20140612",
      "report_date": "20140205",
      "code_info": "Serial Numbers IN the US:    6281- 002002, 6281- 002003, 6281- 002004, 6281- 002005, 6281- 002006, 6281- 002007, 6281- 002008, 6281- 002009, 6281- 002010, 6281- 002011, 6281- 002012, 6281- 002013, 6281-000295,  6281-000401,  6281-000402,  6281-000403,  6281-000404,  6281-000405,  6281-000406,  6281-000407,  6281-000408,  6281-000503,  6281-000504,  6281-000505,  6281-000506,  6281-000507,  6281-000508,  6281-000512,  6281-000513,  6281-000514,  6281-000515,  6281-000516,  6281-000517,  6281-000518,  6281-000519,  6281-000520,  6281-000521,  6281-000522,  6281-000523,  6281-000524,  6281-000526,  6281-000527,  6281-000528,  6281-000529,  6281-000530,  6281-000531,  6281-000532,  6281-000533,  6281-000534,  6281-000535,  6281-000536,  6281-000537,  6281-000538,  6281-000539,  6281-000540,  6281-000541,  6281-000542,  6281-000543,  6281-000544,  6281-000545,  6281-000546,  6281-000547,  6281-000549,  6281-000550,  6281-000551,  6281-000552,  6281-000553,  6281-000554,  6281-000555,  6281-000556,  6281-000557,  6281-000558,  6281-000559,  6281-000560,  6281-000561,  6281-000562,  6281-000563,  6281-000564,  6281-000565,  6281-000566,  6281-000567,  6281-000570,  6281-000574,  6281-000575,  6281-000576,  6281-000577,  6281-000578,  6281-000579,  6281-000580,  6281-000581,  6281-000582,  6281-000583,  6281-000584,  6281-000585,  6281-000586,  6281-000587,  6281-000588,  6281-000589,  6281-000590,  6281-000591,  6281-000592,  6281-000593,  6281-000594,  6281-000595,  6281-000596,  6281-000597,  6281-000598,  6281-000599,  6281-000600,  6281-000601,  6281-000602,  6281-000603,  6281-000604,  6281-000605,  6281-000606,  6281-000607,  6281-000608,  6281-000609,  6281-000610,  6281-000611,  6281-000612,  6281-000613,  6281-000614,  6281-000615,  6281-000616,  6281-000617,  6281-000618,  6281-000619,  6281-000620,  6281-000621,  6281-000622,  6281-000623,  6281-000624,  6281-000625,  6281-000626,  6281-000627,  6281-000628,  6281-000629,  6281-000630,  6281-000631,  6281-000632,  6281-000633,  6281-000634,  6281-000635,  6281-000636,  6281-000637,  6281-000638,  6281-000639,  6281-000640,  6281-000641,  6281-000642,  6281-000643,  6281-000644,  6281-000645,  6281-000646,  6281-000647,  6281-000648,  6281-000649,  6281-000650,  6281-000651,  6281-000652,  6281-000653,  6281-000654,  6281-000655,  6281-000656,  6281-000657,  6281-000658,  6281-000659,  6281-000660,  6281-000661,  6281-000662,  6281-000663,  6281-000664,  6281-000665,  6281-000666,  6281-000667,  6281-000668,  6281-000669,  6281-000670,  6281-000671,  6281-000672,  6281-000673,  6281-000674,  6281-000675,  6281-000676,  6281-000677,  6281-000678,  6281-000679,  6281-000680,  6281-000681,  6281-000682,  6281-000683,  6281-000684,  6281-000685,  6281-000686,  6281-000687,  6281-000688,  6281-000689,  6281-000690,  6281-000691,  6281-000692,  6281-000693,  6281-000694,  6281-000695,  6281-000696,  6281-000697,  6281-000698,  6281-000699,  6281-000700,  6281-000701,  6281-000702,  6281-000703,  6281-000704,  6281-000705,  6281-000706,  6281-000707,  6281-000708,  6281-000709,  6281-000710,  6281-000711,  6281-000712,  6281-000713,  6281-000714,  6281-000715,  6281-000716,  6281-000717,  6281-000718,  6281-000719,  6281-000720,  6281-000721,  6281-000722,  6281-000723,  6281-000724,  6281-000725,  6281-000853,  6281-000854,  6281-000855,  6281-000856,  6281-000857,  6281-000858,  6281-000859,  6281-000860,  6281-000861,  6281-000862,  6281-000863,  6281-000864,  6281-000865,  6281-000866,  6281-000867,  6281-000868,  6281-000869,  6281-000870,  6281-000871,  6281-000872,  6281-000938,  6281-000939,  6281-000940,  6281-000941,  6281-000942,  6281-000943,  6281-000944,  6281-000945,  6281-000946,  6281-000947,  6281-000948,  6281-000949,  6281-000950,  6281-000951,  6281-000952,  6281-000953,  6281-000954,  6281-000955,  6281-000956,  6281-000957,  6281-000958,  6281-000959,  6281-000960,  6281-000961,  6281-000962,  6281-000963,  6281-000964,  6281-000965,  6281-000966,  6281-000967,  6281-000968,  6281-000969,  6281-000970,  6281-000971,  6281-000972,  6281-000973,  6281-000974,  6281-000975,  6281-000976,  6281-000977,  6281-000978,  6281-000979,  6281-000981,  6281-000982,  6281-000983,  6281-000984,  6281-000985,  6281-000986,  6281-000987,  6281-000988,  6281-000989,  6281-000990,  6281-000991,  6281-000992,  6281-001005,  6281-001006,  6281-001007,  6281-001008,  6281-001009,  6281-001010,  6281-001011,  6281-001012,  6281-001013,  6281-001014,  6281-001015,  6281-001016,  6281-001017,  6281-001018,  6281-001019,  6281-001020,  6281-001021,  6281-001022,  6281-001023,  6281-001024,  6281-001025,  6281-001026,  6281-001027,  6281-001028,  6281-001029,  6281-001030,  6281-001031,  6281-001036,  6281-001037,  6281-001080,  6281-001081,  6281-001082,  6281-001083,  6281-001084,  6281-001085,  6281-001086,  6281-001087,  6281-001088,  6281-001089,  6281-001090,  6281-001091,  6281-001092,  6281-001093,  6281-001094,  6281-001095,  6281-001096,  6281-001097,  6281-001098,  6281-001099,  6281-001100,  6281-001101,  6281-001102,  6281-001103,  6281-001104,  6281-001105,  6281-001106,  6281-001107,  6281-001108,  6281-001109,  6281-001110,  6281-001111,  6281-001112,  6281-001113,  6281-001114,  6281-001115,  6281-001116,  6281-001117,  6281-001118,  6281-001119,  6281-001120,  6281-001121,  6281-001122,  6281-001123,  6281-001124,  6281-001125,  6281-001126,  6281-001127,  6281-001128,  6281-001129,  6281-001130,  6281-001131,  6281-001132,  6281-001133,  6281-001134,  6281-001135,  6281-001136,  6281-001137,  6281-001138,  6281-001139,  6281-001140,  6281-001141,  6281-001142,  6281-001143,  6281-001144,  6281-001145,  6281-001146,  6281-001147,  6281-001148,  6281-001149,  6281-001150,  6281-001151,  6281-001152,  6281-001153,  6281-001154,  6281-001155,  6281-001156,  6281-001157,  6281-001158,  6281-001159,  6281-001160,  6281-001161,  6281-001162,  6281-001163,  6281-001164,  6281-001165,  6281-001166,  6281-001167,  6281-001168,  6281-001169,  6281-001170,  6281-001171,  6281-001178,  6281-001179,  6281-001180,  6281-001181,  6281-001182,  6281-001183,  6281-001184,  6281-001185,  6281-001186,  6281-001187,  6281-001188,  6281-001189,  6281-001190,  6281-001191,  6281-001192,  6281-001193,  6281-001194,  6281-001195,  6281-001196,  6281-001197,  6281-001222,  6281-001223,  6281-001224,  6281-001225,  6281-001226,  6281-001227,  6281-001228,  6281-001229,  6281-001230,  6281-001231,  6281-001232,  6281-001233,  6281-001234,  6281-001235,  6281-001236,  6281-001237,  6281-001238,  6281-001239,  6281-001240,  6281-001241,  6281-001242,  6281-001243,  6281-001244,  6281-001245,  6281-001246,  6281-001247,  6281-001248,  6281-001249,  6281-001250,  6281-001251,  6281-001252,  6281-001253,  6281-001254,  6281-001255,  6281-001256,  6281-001257,  6281-001258,  6281-001259,  6281-001260,  6281-001261,  6281-001262,  6281-001263,  6281-001264,  6281-001265,  6281-001266,  6281-001267,  6281-001268,  6281-001269,  6281-001270,  6281-001271,  6281-001272,  6281-001273,  6281-001274,  6281-001275,  6281-001276,  6281-001277,  6281-001278,  6281-001279,  6281-001280,  6281-001281,  6281-001282,  6281-001283,  6281-001284,  6281-001285,  6281-001286,  6281-001287,  6281-001288,  6281-001289,  6281-001290,  6281-001291,  6281-001292,  6281-001293,  6281-001294,  6281-001295,  6281-001296,  6281-001297,  6281-001298,  6281-001299,  6281-001300,  6281-001301,  6281-001302,  6281-001303,  6281-001304,  6281-001305,  6281-001306,  6281-001307,  6281-001308,  6281-001309,  6281-001310,  6281-001311,  6281-001312,  6281-001313,  6281-001314,  6281-001315,  6281-001316,  6281-001317,  6281-001318,  6281-001319,  6281-001320,  6281-001321,  6281-001322,  6281-001323,  6281-001331,  6281-001332,  6281-001337,  6281-001338,  6281-001390,  6281-001391,  6281-001392,  6281-001393,  6281-001394,  6281-001395,  6281-001396,  6281-001397,  6281-001398,  6281-001399,  6281-001400,  6281-001401,  6281-001402,  6281-001403,  6281-001404,  6281-001405,  6281-001406,  6281-001407,  6281-001408,  6281-001409,  6281-001410,  6281-001412,  6281-001413,  6281-001414,  6281-001415,  6281-001416,  6281-001417,  6281-001418,  6281-001419,  6281-001420,  6281-001421,  6281-001422,  6281-001423,  6281-001424,  6281-001425,  6281-001426,  6281-001427,  6281-001428,  6281-001429,  6281-001430,  6281-001431,  6281-001432,  6281-001433,  6281-001434,  6281-001435,  6281-001436,  6281-001437,  6281-001438,  6281-001439,  6281-001440,  6281-001441,  6281-001449,  6281-001450,  6281-001451,  6281-001452,  6281-001453,  6281-001454,  6281-001459,  6281-001468,  6281-001469,  6281-001470,  6281-001471,  6281-001472,  6281-001595,  6281-001596,  6281-001597,  6281-001598,  6281-001610,  6281-001611,  6281-001615,  6281-001616,  6281-001648,  6281-001802,  6281-001803,  6281-001804,  6281-001805,  6281-001806,  6281-001807,  6281-001808,  6281-001809,  6281-001810,  6281-001811,  6281-001812,  6281-001813,  6281-001814,  6281-001815,  6281-001816,  6281-001817,  6281-001818,  6281-001819,  6281-001820,  6281-001914,  6281-001915,  6281-001916,  6281-001917,  6281-001918,  6281-001919,  6281-001920,  6281-001921,  6281-001922,  6281-001923,  6281-001924,  6281-001925,  6281-001926,  6281-001927,  6281-001928,  6281-001929,  6281-001930,  6281-001931,  6281-001932,  6281-001933,  6281-001934,  6281-001935,  6281-001936,  6281-001937,  6281-001938,  6281-001939,  6281-001940,  6281-001941,  6281-001942,  6281-001943,  6281-001944,  6281-001945,  6281-001946,  6281-001947,  6281-001948,  6281-001949,  6281-001950,  6281-001951,  6281-001952,  6281-001953,  6281-001954,  6281-001955,  6281-001956,  6281-001957,  6281-001958,  6281-001959,  6281-001960,  6281-001961,  6281-001962,  6281-001963,  6281-001964,  6281-001965,  6281-001966,  6281-001967,  6281-001968,  6281-001969,  6281-001970,  6281-001971,  6281-001972,  6281-001973,  6281-001974,  6281-001975,  6281-001976,  6281-001977,  6281-001978,  6281-001979,  6281-001980,  6281-001998,  6281-002014,  6281-002015,  6281-002016,  6281-002017,  6281-002018,  6281-002019,  6281-002020,  6281-002021,  6281-002022,  6281-002023,  6281-002024,  6281-002025,  6281-002026,  6281-002027,  6281-002028,  6281-002029,  6281-002030,  6281-002031,  6281-002032,  6281-002033,  6281-002034,  6281-002035,  6281-002036,  6281-002037,  6281-002038,  6281-002039,  6281-002040,  6281-002041,  6281-002042,  6281-002043,  6281-002044,  6281-002045,  6281-002046,  6281-002047,  6281-002048,  6281-002049,  6281-002050,  6281-002051,  6281-002052,  6281-002053,  6281-002054,  6281-002055,  6281-002056,  6281-002057,  6281-002058,  6281-002059,  6281-002060,  6281-002061,  6281-002062,  6281-002063,  6281-002064,  6281-002065,  6281-002066,  6281-002067,  6281-002068,  6281-002069,  6281-002070,  6281-002071,  6281-002072,  6281-002073,  6281-002074,  6281-002075,  6281-002076,  6281-002077,  6281-002078,  6281-002079,  6281-002080,  6281-002081,  6281-002082,  6281-002083,  6281-002084,  6281-002085,  6281-002086,  6281-002087,  6281-002088,  6281-002089,  6281-00209 ,  6281-002090,  6281-002091,  6281-002092,  6281-002093,  6281-002094,  6281-002095,  6281-002096,  6281-002100,  6281-002101,  6281-002102,  6281-002103,  6281-002104,  6281-002105,  6281-002106,  6281-002107,  6281-002108,  6281-002109,  6281-002110,  6281-002111,  6281-002112,  6281-002113,  6281-002114,  6281-002115,  6281-002116,  6281-002117,  6281-002118,  6281-002119,  6281-002120,  6281-002121,  6281-002122,  6281-002123,  6281-002124,  6281-002125,  6281-002126,  6281-002127,  6281-002128,  6281-002129,  6281-002130,  6281-002131,  6281-002132,  6281-002133,  6281-002134,  6281-002135,  6281-002136,  6281-002137,  6281-002138,  6281-002139,  6281-002140,  6281-002141,  6281-002142,  6281-002143,  6281-002144,  6281-002145,  6281-002146,  6281-002147,  6281-002148,  6281-002149,  6281-002150,  6281-002151,  6281-002152,  6281-002153,  6281-002154,  6281-002155,  6281-002156,  6281-002157,  6281-002158,  6281-002159,  6281-002160,  6281-002161,  6281-002162,  6281-002163,  6281-002164,  6281-002165,  6281-002166,  6281-002167,  6281-002168,  6281-002169,  6281-002170,  6281-002171,  6281-002172,  6281-002173,  6281-002174,  6281-002175,  6281-002176,  6281-002177,  6281-002178,  6281-002179,  6281-002180,  6281-002181,  6281-002182,  6281-002183,  6281-002184,  6281-002185,  6281-002186,  6281-002187,  6281-002188,  6281-002189,  6281-002190,  6281-002191,  6281-002375,  6281-002376,  6281-002377,  6281-002378,  6281-002379,  6281-002380,  6281-002381,  6281-002382,  6281-002383,  6281-002384,  6281-002507,  6281-002545,  6281-002546,  6281-002547,  6281-002548,  6281-002549,  6281-002550,  6281-002551,  6281-002552,  6281-002553,  6281-002554,  6281-002555,  6281-002556,  6281-002557,  6281-002558,  6281-002559,  6281-002560,  6281-002561,  6281-002562,  6281-002563,  6281-002564,  6281-002565,  6281-002566,  6281-002567,  6281-002568,  6281-002569,  6281-002570,  6281-002571,  6281-002572,  6281-002573,  6281-002574,  6281-002575,  6281-002576,  6281-002577,  6281-002578,  6281-002579,  6281-002580,  6281-002581,  6281-002582,  6281-002583,  6281-002584,  6281-002585,  6281-002586,  6281-002587,  6281-002588,  6281-002589,  6281-002590,  6281-002591,  6281-002592,  6281-002593,  6281-002594,  6281-002595,  6281-002596,  6281-002597,  6281-002598,  6281-002599,  6281-002600,  6281-002601,  6281-002602,  6281-002603,  6281-002604,  6281-002605,  6281-002606,  6281-002607,  6281-002608,  6281-002609,  6281-002610,  6281-002611,  6281-002612,  6281-002613,  6281-002614,  6281-002615,  6281-002616,  6281-002617,  6281-002618,  6281-002619,  6281-002620,  6281-002621,  6281-002622,  6281-002623,  6281-002624,  6281-002625,  6281-002626,  6281-002627,  6281-002681,  6281-002682,  6281-002683, and 6281-002684.    SERIAL NUMBERS OUTSIDE THE US:    6281-000451, 6281-000771,  6281-000772,  6281-000773,  6281-000774,  6281-000775,  6281-000776,  6281-000993,  6281-000994,  6281-000995,  6281-000996,  6281-000997,  6281-000998,  6281-000999,  6281-001000,  6281-001001,  6281-001002,  6281-001003,  6281-001004,  6281-001038,  6281-001039,  6281-001040,  6281-001041,  6281-001042,  6281-001043,  6281-001044,  6281-001045,  6281-001046,  6281-001047,  6281-001048,  6281-001049,  6281-001050,  6281-001051,  6281-001052,  6281-001053,  6281-001054,  6281-001055,  6281-001056,  6281-001057,  6281-001058,  6281-001059,  6281-001060,  6281-001061,  6281-001062,  6281-001063,  6281-001064,  6281-001065,  6281-001066,  6281-001067,  6281-001068,  6281-001069,  6281-001070,  6281-001071,  6281-001172,  6281-001173,  6281-001174,  6281-001175,  6281-001475,  6281-001476,  6281-001514,  6281-001515,  6281-001516,  6281-001517,  6281-001705,  6281-001706,  6281-001994,  6281-001995,  6281-001996,  and 6281-001997.     Spacelabs is currently conducting a internal audit and discovered that there are additional lot numbers for this recall.  6681-003056, 6281-002500, 6281-002501, 6281-002659, 6281-002661, 6281-002662, 6281-002663, and 6281-002664.",
      "more_code_info": ""
    }
  ]
}