      "status": "Ongoing",
      "city": "Woburn",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84378",
      "recalling_firm": "Energetiq Technology Inc",
      "address_1": "7 Constitution Way",
      "address_2": "N/A",
      "postal_code": "01801-1024",
      "voluntary_mandated": "FDA Mandated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution.",
      "recall_number": "Z-0870-2020",
      "product_description": "EQ-400-RH-QZ and EQ-400-LH-BK products     Product Usage:  The device is a very broadband light source emitting incoherent non-laser light from the Deep UV through the visible wavelengths.",
      "product_quantity": "15",
      "reason_for_recall": "On September 27, 2019 Energetiq Technology (the manufacturer) discovered a problem with  their EQ-400-RH-QZ and EQ-400-LH-BK products during routine manufacturing functional  testing. It was noted that a safety interlock within the lamp does not function in a reliable manner  to shut down the system if the laser collimator is removed from the lamp while the system is  operating. This is a potential safety hazard for anyone working on or nearby the lamp.",
      "recall_initiation_date": "20190927",
      "center_classification_date": "20200129",
      "report_date": "20200205",
      "code_info": "The defect concerns a total of fifteen (15) of our EQ-400 products. manufactured between June 3, 2016 and September 26, 2019."
    }
  ]
}