  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Whitland",
      "state": "N/A",
      "country": "United Kingdom",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87063",
      "recalling_firm": "The Magstim Company Limited",
      "address_1": "Spring Gardens",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "MN, NC, and NY",
      "recall_number": "Z-0869-2021",
      "product_description": "Neurosign 100, Product Number 9883-01. Intraoperative Nerve Monitor - Product Usage: indicated for use in stimulating and monitoring cranial motor nerves, especially the facial nerve (VII), during ENT surgical procedures.",
      "product_quantity": "7 units",
      "reason_for_recall": "A small number of Neurosign 100 Intraoperative Nerve Monitors may have been dispatched without being configured appropriately for the supply mains within the target Country. The specific deficiency includes:  1. Onboard voltage selector set to 240V (European Standard), versus 120V requirement for the United States and Canada.  2. Fuses installed in the Power Entry Module (PEM) 2xT315mAL (240V Standard), versus the 2xT630mAL requirement for a 120V mains supply.",
      "recall_initiation_date": "20201208",
      "center_classification_date": "20210115",
      "termination_date": "20230216",
      "report_date": "20210127",
      "code_info": "Serial Numbers:  2507, 2511, 2514, 2515, 2522, 2527, 2533, 2534, 2537",
      "more_code_info": ""
    }
  ]
}