  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Torrance",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89596",
      "recalling_firm": "Shimadzu Medical Systems",
      "address_1": "20101 S Vermont Ave",
      "address_2": "",
      "postal_code": "90502-1328",
      "voluntary_mandated": "FDA Mandated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "U.S.: AK, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, WA, and WV.    O.U.S.: N/A",
      "recall_number": "Z-0867-2022",
      "product_description": "MODEL: X-RAY R/F SYSTEM FLUOROspeed X1",
      "product_quantity": "25 systems",
      "reason_for_recall": "It was found that the irradiated x-ray may exceed the xray radiation dose rate specified in CFR 1020.32 in some specific case of fluoroscopic mode due to inadequate adjusting criteria in installation.",
      "recall_initiation_date": "20220110",
      "center_classification_date": "20220406",
      "report_date": "20220413",
      "code_info": "Model: FLUOROspeed X1  Generator Number: D150-40  Serial Numbers:  MQ493E19C001  MQ493E3B8002  MQ493E3B9001  MQ493E1B4001  MQ493E1A1003  MQ493E1A3001  MQ493E1B5001  MQ493E3B8001  MQ493E1A5001  MQ493E1A1005  MQ493E1A2001  MQ493E1A4001  MQ493E1B6002  MQ493E1A8001  MQ493E1B4002  MQ493E1A4003  MQ493E1A4002  MQ493E1B6001  MQ493E1A2002  MQ493E1AC001  MQ493E1A1002  MQ493E19C002  MQ493E1A3003  MQ493E1A1004  MQ493E1A3002",
      "more_code_info": ""
    }
  ]
}