  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Elgin",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "75716",
      "recalling_firm": "Northgate Technologies, Inc.",
      "address_1": "1591 Scottsdale Ct",
      "address_2": "N/A",
      "postal_code": "60123-9361",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US including FL and MA.  Internationally to Canada",
      "recall_number": "Z-0866-2017",
      "product_description": "Insufflator, laparoscopic    The ConMed GS2000, Smith & Nephew 500, Novadaq 50L and Nebulae I insufflators are devices that provide CO2 gas distention of surgical cavities for diagnostic and/or operative endoscopy. The insufflator has multiple operating modes which can be used for the following procedures: General Laparoscopic, Pediatric Laparoscopic, Bariatric Laparoscopic and Minimally Invasive Vessel Harvesting procedures",
      "product_quantity": "49 devices",
      "reason_for_recall": "Device could cause a patient overpressure situation without any visual or auditory indication or warning",
      "recall_initiation_date": "20161121",
      "center_classification_date": "20161221",
      "termination_date": "20171127",
      "report_date": "20161228",
      "code_info": "Serial Numbers: #89170HXB, #89556DYD, #89652FYA, #89661FYA, #89663FYA, #89664FYA, #89670FYA, #89672FYA, #89675FYA, #89677FYA, #89685FYB, #89697FYB, #89708GYA, #89713GYA, #89719GYB, #89720GYB, #89729GYB, #89732GYB, #89734GYB, #89737GYB, #89889IYD, #90165AZB, #90285BZD, #90387DZB, #90391DZB, #90415DZD, #90438DZD, #90469EZB, #90528EZD, #90555FZB, #90556FZB, #90557FZB, #90558FZB, #90561FZB, #90583GZC, #90621GZB, #90654HZB, #90739IZD, #90869KZD, #90931LZD, #90936LZD, #90954LZD, #90962LZD, #90972LZD, #91105CAD, #91108CAD, #91114CAD, #91120CAD, #91380FAC, #91406FAD, #91471GAC, #91544IAC"
    }
  ]
}