      "status": "Terminated",
      "city": "Sunnyvale",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "75891",
      "recalling_firm": "Accuray Incorporated",
      "address_1": "1310 Chesapeake Ter",
      "address_2": "N/A",
      "postal_code": "94089-1100",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US including CA. MN, OK. FL. MO, TN, KS, TX, AZ, MS, IA, MA. NC, NY. IN, NJ, VA. LA, WA, WI, PA. CT, G,. MT, MD, LA., WV, OH, and DC.  Foreign distribution to Taiwan, Japan, Korea, Turkey. Malaysia, Vietnam, Greece, China, Spain, France, Italy, Pakistan, and Georgia,",
      "recall_number": "Z-0865-2017",
      "product_description": "CyberKnife Robotic Radiosurgery Systems, models: G3, G4, VSI.    Radiology: The CyberKnife Robotic Radiosurgery System is indicated for treatment planning and image guided stereotactic Radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when  radiation treatment is indicated.",
      "product_quantity": "115 affected devices.",
      "reason_for_recall": "Accuray has become aware of a potential safety issue involving the possibility of electric shock during  maintenance activities from the lon Pump HV Power Supply at the back of the modulator cabinet. This issue is limited to certain G3, G4, and VSI models of the CyberKnife¿ Robotic Radiosurgery System.",
      "recall_initiation_date": "20161206",
      "center_classification_date": "20161221",
      "termination_date": "20170609",
      "report_date": "20161228",
      "code_info": "models: G3, G4, VSI.  Serial numbers:  C0023  C0030  C0045  C0046  C0047  C0006  C0012  C0040  C0048  C0049  C0013  C0011  C0052  C0055  C0022  C0053  C0042  C0057  C0017  C0059  C0056  C0060  C0062  C0061  C0067  C0068  C0070  C0041  C0063  C0065  C0074  C0071  C0079  C0078  C0080  C0064  C0090  C0024  C0087  C0091  C0095  C0094  C0092  C0093  C0097  C0098  C0099  C0100  C0101  C0102  C0103  C0106  C0089  C0108  C0107  C0105  C0112  C0114  C0118  C0119  C0113  C0123  C0126  C0124  C0121  C0132  C0130  C0131  C0125  C0133  C0135  C0117  C0134  C0038  C0136  C0137  C0127  C0143  C0144  C0014  C0138  C0146  C0140  C0145  C0150  C0033  C0151  C0152  C0147  C0149  C0148  C0158  C0155  C0159  C0007  C0163  C0157  C0165  C0008  C0162  C0085  C0161  C0086  C0073  C0115  C0076  C0153  C0077  C0111  C0066  C0027  C0075  C0084  C0096  C0154."
    }
  ]
}