{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Bronshoj",
      "state": "N/A",
      "country": "Denmark",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84496",
      "recalling_firm": "Radiometer Medical ApS",
      "address_1": "Akandevej 21",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) to states of: AL, AR, AZ, CA, FL, GA, IA, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NH, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI and WV; and countries of: \u0013 Australia, Austria, Canada, China, Czech Republic, Denmark, Finland, France, Germany, Hungary, India, Italy, Japan (Tokyo-not a country but it was listed), Netherlands, Norway, Poland, Singapore, South Korea, Spain, Sweden, Switzerland, Turkey and United Kingdom.",
      "recall_number": "Z-0864-2020",
      "product_description": "AQURE basic sytem, FLEXLINK, REF 933-599,  Version: 2.3.3 (17), CE IVD, Rx Only, UDI: (01)05700699335999    Intended use: The AQURE system is intended to let you manage analytical devices and operator profiles.",
      "product_quantity": "947",
      "reason_for_recall": "The firm has become aware that there is a potential problem relating to the blood gas and immunoassay analyzer Systems that may result in patient mix-up when connected to some third-party devices. The error may lead to serious adverse health consequences for the patient caused by patient data mix-up.",
      "recall_initiation_date": "20190808",
      "center_classification_date": "20200124",
      "termination_date": "20220606",
      "report_date": "20200205",
      "code_info": "Software version 2.3.5 and below"
    }
  ]
}