      "country": "United States",
      "city": "Saint Louis",
      "address_1": "3365 Tree Court Industrial Blvd",
      "reason_for_recall": "A software anomaly was identified where the system may automatically transition from \"Standby\" to \"Ready to Fire\" state while in surgical use.",
      "address_2": "",
      "product_quantity": "106",
      "code_info": "Serial numbers:   SPC01790, SPC01791, SPC01792, SPC01793, SPC01794, SPC01795, SPC01796, SPC01797, SPC01798, SPC01799, SPC01800, SPC01801, SPC01802, SPC01803, SPC01806, SPC01807, SPC01808, SPC01809, SPC01810, SPC01811, SPC01899, SPC01900, SPC01901, SPC01902, SPC01903, SPC01909, SPC01910, SPC01911, SPC01920, SPC01921, SPC01922, SPC01923, SPC01927, SPC01928, SPC01929, SPC01930, SPC01931, SPC01932, SPC01939, SPC01943, SPC01944, SPC01945, SPC01946, SPC01947, SPC01948, SPC01950, SPC01951, SPC01952, SPC01953, SPC01954, SPC01955, SPC01956, SPC01957, SPC01958, SPC01961, SPC01963, SPC01968, SPC01969, SPC01975, SPC01980, SPC01981, SPC01982, SPC01983, SPC01995, SPC01996, SPC01997, SPC02004, SPC02005, SPC02006, SPC02007, SPC02008, SPC02009, SPC02010, SPC02011, SPC02012, SPC02013, SPC02014, SPC02020, SPC02029, SPC02030, SPC02037, SPC02043, SPC02051, SPC02052, SPC02053, SPC02056, SPC02057, SPC02059, SPC02061, SPC02067, SPC02068, SPC02073, SPC02074, SPC02075, SPC02076, SPC02077, SPC02099, SPC02108, SPC02109, SPC02110, SPC02111, SPC02112, SPC02113, SPC02114, SPC02115 and SPC02120",
      "center_classification_date": "20141223",
      "distribution_pattern": "Worldwide Distribution - USA Distribution including to the states of  AL, CA, CT, FL, GA, HI, IL, IN, KS, MI, MN, MO, MS, NH, NY, OH, OK, PA, TX, WI and WV., and to the countries of  Austria, Australia, Belgium, China, Czech Republic, France, Germany, Great Britain, Greece, Hong Kong, Hungary, India,  Ireland, Italy, Jordan, Malaysia, Morocco, Netherlands, Norway, Portugal, Singapore, Spain, Sweden, Switzerland, Turkey and the United Arab Emirates.",
      "state": "MO",
      "product_description": "Bausch & Lomb Stellaris PC w/ Laser, BL14304  The responsible firm name is Bausch & Lomb Incorporated Rochester, NY",
      "report_date": "20141231",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Bausch & Lomb Inc",
      "recall_number": "Z-0862-2015",
      "initial_firm_notification": "Telephone",
      "product_type": "Devices",
      "event_id": "69717",
      "termination_date": "20170323",
      "more_code_info": "",
      "recall_initiation_date": "20141104",
      "postal_code": "63122-6615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}