  "results": [
    {
      "status": "Ongoing",
      "city": "Hazelwood",
      "state": "MO",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "91148",
      "recalling_firm": "Biomerieux Inc",
      "address_1": "595 Anglum Rd",
      "address_2": "N/A",
      "postal_code": "63042-2320",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-0861-2023",
      "product_description": "VITEK 2 REAGENT AST-XN15 TEST KIT 20 CARDS, CATALOG 423829",
      "product_quantity": "58 units",
      "reason_for_recall": "As the temperature and time out of range was exceeded, product performances cannot be guaranteed",
      "recall_initiation_date": "20221109",
      "center_classification_date": "20230104",
      "report_date": "20230111",
      "code_info": "UDI/DI 03573026620288, Batch Numbers:  8312243203"
    }
  ]
}