  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Raynham",
      "address_1": "325 Paramount Drive",
      "reason_for_recall": "Codman Certas Programmable Valves used for hydrocephalus may not operate properly",
      "address_2": "",
      "product_quantity": "1460 units",
      "code_info": "Lot Codes:   CLNCB1  CMBBP1  CMBCPN  CMCBWG  CMCC2V  CMDCN8  CMDDJN  CMGB8H  CMJCL8  CMJCPW  CMKB21  CMMCGR  CMNBFB  CMNBZJ  CMNC8T  CMNCYL  CMPB4V  CNCCPF  CNDBCW  CNDBRH  CNDCJ2  CNFCK7  CNGBG0  CNGBK2  CNHCKH  CNHCZN  CNJBW5  CNJBWZ  CNJC38  CNJC4K  CNJCD3  CNJCFR  CNJCMN  CNKBWP  CNKCC8  CNKCL3  CNLBR6  CNLCL3  CNLCW3  CNNCWF  CNPB1R  CNPBRH  CNPCLR  CPBC96  CPBCCR  CPBCRV  CPCBG9  CPCCMK  CPDBYC  CPDC07  CPDCC7  CNBBCV  CNCB6W  CNGBRY  CNHB95  CNNDCK  CNPB1T  CNPBML  CPBB3G  CPCBG8  CPCCBH  CPCCM1  CPCCMZ  CPFBKP",
      "center_classification_date": "20140128",
      "distribution_pattern": "Worldwide Distribution - US Nationwide, Australia, Austria, Belgium, Brazil, Switzerland, Czech Republic, Germany, Spain, Finland, France, Great Britain, Israel, Italy, Netherlands, Norway, Portugal, Russia, Saudi Arabia, Sweden, Slovenia, Slovakia, Turkey, South Africa, and UK.",
      "state": "MA",
      "product_description": "Codman Certas Programmable Valve In Line Valve Only  Product Code: 82-8800\t    Product Usage:  The CODMAN CERTAS Programmable Valve is an implantable device that provides constant intraventricular pressure and drainage of CSF (cerebrospinal fluid) for the management of hydrocephalus. The valve can be set to a choice of eight opening pressure settings for constant intraventricular perssure drainage of CSF. The CODMAN CERTAS Therapy Management System (TMS) allows the non-invasive reading or adjustment of the valve setting.",
      "report_date": "20140205",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Codman & Shurtleff, Inc.",
      "recall_number": "Z-0858-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "65179",
      "termination_date": "20150602",
      "more_code_info": "",
      "recall_initiation_date": "20130516",
      "postal_code": "02767",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}