  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87023",
      "recalling_firm": "GE Healthcare, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "N/A",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide",
      "recall_number": "Z-0857-2021",
      "product_description": "Discovery 670 DR (Model H3100BT), NM/CT 870 DR (Model 3100Q and H3100AS), Discovery NM670 Pro (Model 3100XB), Discovery 670 ES (Model H3100XF) - Product Usage: intended for use by appropriately trained healthcare professionals to aid in detecting, localizing, diagnosing of diseases and organ function for the evaluation of diseases, trauma, abnormalities, and disorders such NM System: General Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body, using a variety of scanning modes supported by various acquisition types and imaging features designed to enhance image quality.",
      "product_quantity": "52 untits",
      "reason_for_recall": "GE Healthcare has become aware that rotor bearing screws were found loose on one detector in the field, leading to the release of one of the four rotor bearings in the detector.",
      "recall_initiation_date": "20201210",
      "center_classification_date": "20210113",
      "report_date": "20210120",
      "code_info": "Discovery 670 DR (Model H3100BT), Serial Numbers: DRGY56033,  DRGY56037, DRGY56035, DRGY56030, DRGZ56043, DRGZ56042, DRGY56032, DRGY56031, DRGY56034, DRGY56036      NM/CT 870 DR (Model 3100Q), Serial Numbers: 870Y64066, and (Model H3100AS), Serial Numbers:  870Y64077, 870Y64064, 870Y64081, 870Y64079, 870Y64067, 870Y64074, 870Z64117, 870Y64078, 870Y64080, 870Y64069, 870Y64085, 870Z64121, 870Z64119, 870Z64122, 870Y64056, 870Y64063, 870Z64114, 870Z64115, 870Y64068, 870Y64082, 870Z64116, 870Z64120, 870Y64065, 870Y64084, 870Y64061, 870Y64083, 870Z64113, 870Y64070, 870Z64123, 870Y64058, 870Z64118, 870Y64057     Discovery NM670 Pro (Model 3100XB), Serial Numbers: PRGY52027,   PRGY52031, PRGZ52047, PRGZ52046, PRGY52028, PRGZ52048, PRGY52029, PRGY52030       Discovery 670 ES (Model H3100XF), Serial Numbers:  ESGZ53005"
    }
  ]
}