{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Munchen",
      "state": "",
      "country": "Germany",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "67199",
      "recalling_firm": "Sorin Group Deutschland GmbH",
      "address_1": "Lindberghstrasse 25",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US (Nationwide) and PR; and Internationally to:BE, ES, FI, GB, NL, SE, AR, AT, AU, CA, CH, CN, CR, EG, HK, ID, JO, JP, KR, KW, LB, LY, MM, PL, RO, RS, RU, SG. TH, TN, TR, TW, UA, VN, and ZA.",
      "recall_number": "Z-0856-2014",
      "product_description": "S5/C5 Heart-lung machine    Product Usage:  The StOckert S5/ Sorin C5 System is intended to be used during cardiopulmonary bypass for procedures lasting six (6) hours or less.",
      "product_quantity": "323",
      "reason_for_recall": "Sorin is initiating a field correction on S5/C5 Heart-lung machine and CP5 centrifugal pump system due to reports of pump speed control knob failures resulting in no response.",
      "recall_initiation_date": "20131204",
      "center_classification_date": "20140128",
      "termination_date": "20151228",
      "report_date": "20140205",
      "code_info": "S5 Item Number: 10-80-00, 10-85-00, 28-9S-80, 28-95-8S. C5 Item Number: S8-00-00",
      "more_code_info": ""
    }
  ]
}