  "results": [
    {
      "status": "Completed",
      "city": "San Diego",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89699",
      "recalling_firm": "Illumina, Inc.",
      "address_1": "5200 Illumina Way",
      "address_2": "N/A",
      "postal_code": "92122-4616",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US: PA, WA, CA, MI, FL, MD, AR, UT NC, Fl, IL, NE, NJ, TN, NY, IN, CO, GA, OK, TX, CT, SD, NM, VA, AL, LA, SC, WV, OR, IA, AZ, MN  OUS: Australia\tAustria\tBelgium\tChile\tChina\tDenmark\tFinland\tFrance\tGermany\tHungary\tIreland\tItaly\tJapan\tLatvia\tNetherlands\tNorway\tPoland\tPortugal\tRomania\tRussian Federation\tSaudi Arabia\tSingapore\tSlovakia\tSouth Africa\tSouth Korea\tSpain\tSweden\tSwitzerland\tTurkey\tUnited Arab Emirates\tUnited Kingdom\tVietnam",
      "recall_number": "Z-0854-2022",
      "product_description": "Illumina NextSeq 550Dx, REF: 20005715, CE, IVD",
      "product_quantity": "674 instruments",
      "reason_for_recall": "There is a potential that the DNA sequence analyzer may short circuit shortly after disconnection and reconnection of circuit board or imaging module due  to an insufficient reconnection of the ribbon cable. The short circuit results in the device being inoperable.",
      "recall_initiation_date": "20220222",
      "center_classification_date": "20220331",
      "report_date": "20220406",
      "code_info": "Lot #s NDX550110 - NDX550843 ; UDI: (01)00816270020125"
    }
  ]
}