      "country": "United States",
      "city": "Parsippany",
      "address_1": "10 Pomeroy Rd",
      "reason_for_recall": "During cleaning, two separate units of the NexGen Tibial Base plates did not fit with the Tibial Base Plate handle.",
      "address_2": "",
      "product_quantity": "8 units",
      "code_info": "Part number 00-5887-045-00 Lot number 62592928",
      "center_classification_date": "20141223",
      "distribution_pattern": "Worldwide Distribution - US Distribution and the countries of Australia and Denmark.",
      "state": "NJ",
      "product_description": "NexGen Complete Knee Solution Monoblock Tibial Provisional / Drill Guide Tibial Size 5 Femoral Size Green / CH  Zimmer U.K. Ltd., SN3 4FP, UK  Zimmer Trabecular Metal Technology, Inc  10 Pomeroy Road, Parsippany, NJ 07054 USA",
      "report_date": "20141231",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Zimmer Trabecular Metal Technology, Inc.",
      "recall_number": "Z-0851-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "69866",
      "termination_date": "20150828",
      "more_code_info": "",
      "recall_initiation_date": "20141202",
      "postal_code": "07054-3722",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}