  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Andover",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "78849",
      "recalling_firm": "Philips Electronics North  America Corporation",
      "address_1": "3000 Minuteman Rd",
      "address_2": "N/A",
      "postal_code": "01810-1032",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide",
      "recall_number": "Z-0850-2018",
      "product_description": "Philips Network Firewall (Cisco ASA 5506), Model # 867098",
      "product_quantity": "332 units",
      "reason_for_recall": "Firewall installed with Philips IntelliVue Information Center iX or Information Center Classic may have a defective component, which may result in loss of connection to the Information Center iX.  The defect involves the clock signal component within the firewall. This component has a high probability  of failing in appliances that have been running for greater than 18 months. If the clock signal component were to fail, the firewall will stop functioning, will not boot, and is not recoverable.  This failure will result in loss of communication between devices that are separated by the firewall, which may cause the Information Center to reboot.",
      "recall_initiation_date": "20171005",
      "center_classification_date": "20180302",
      "termination_date": "20190701",
      "report_date": "20180314",
      "code_info": "Serial Numbers for use in U.S.A.:  JMX2030Y2B8  JMX2030Y2CR  JMX2030Y2BN  JMX2030Y2AY  JMX2040Y2NZ  JMX2022Y1U4  JMX2040Y2NU  JMX2045Y2L0  JMX2030Y2D9  JMX2030Y2B1  JMX2035Y0XC  JMX2035Y0XT  JMX2030Y2DY  JMX2035Y0XA  JMX2040Y2R2  JMX2031Y0EV  JMX2039Y1FZ  JMX2030Y2BT  JMX2030Y2EE  JMX2030Y2DM  JMX194840WA  JMX2014419Y  JMX201441AB  JMX201441A4  JMX2014Z155  JMX2014Z15C  JMX2014Z15L  JMX2039Y1H7  JMX2039Y1H5  JMX2039Y1HR  JMX2039Y1JM  JMX2022Y194  JMX2018Z009  JMX2022Y15F  JMX2040Y2QP  JMX2030Y2PZ  JMX2045Y2E4  JMX20184033  JMX2030Y2A8  JMX2039Y1GB  JMX2040Y2P9  JMX2030Y2E6  JMX2030Y2EG  JMX2022Y181  JMX2022Y1A7  JMX2040Y2QW  JMX2030Y2B6"
    }
  ]
}