  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "San Diego",
      "address_1": "5200 Illumina Way",
      "reason_for_recall": "It was discovered a software limitation with MiSeq Reporter software (version 2.2.31) when using the MiSeqDx Universal Kit 1.0 on the MiSeqDx instrument.  The software cannot report insertions or deletions at specific genomic locations with  respect to the primer.",
      "address_2": "",
      "product_quantity": "13 units",
      "code_info": "Serial No. M70130, M70189, M70108, M70165, M70166, M70175, M70172, M70103.",
      "center_classification_date": "20141223",
      "distribution_pattern": "US Distribution to the states of  AZ, GA, CA, OH, and VA., and one Internationally to France.",
      "state": "CA",
      "product_description": "Illumina MiSeqDx Universal Kit 1.0, PN 15039608    The Illumina MiSeqDx Universal Kit 1.0 is a set of reagents and consumables used in the processing of human genomic DNA samples derived from peripheral whole blood, and in the subsequent targeted sequencing of the resulting sample libraries. User-supplied analyte specific reagents are required for the preparation of libraries targeting specific genomic regions of interest. The MiSeqDx Universal Kit 1.0 is intended for use with the MiSeqDx instrument.",
      "report_date": "20141231",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Illumina Inc",
      "recall_number": "Z-0849-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "69766",
      "termination_date": "20150318",
      "more_code_info": "",
      "recall_initiation_date": "20141113",
      "postal_code": "92121",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}