  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Raritan",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87019",
      "recalling_firm": "Ortho Clinical Diagnostics Inc",
      "address_1": "1001 US Highway 202",
      "address_2": "N/A",
      "postal_code": "08869-1424",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution and the countries of AR, AU, BM, BR, CA, CL, CN, CO, CR, DO, EC, FR, FR, GT, HT, IN, JP, MX, PA, PE, PR, PY, SV.",
      "recall_number": "Z-0845-2021",
      "product_description": "VITROS XT Chemistry Products ALB-TP Slides (ALB is not affected) - IVD quantitatively measures total protein (TP) concentration in serum and plasma using VITROS XT 3400 Chemistry Systems and VITROS XT 7600 Integrated Systems    Product Code:\t6844293",
      "product_quantity": "US=9802; OUS: 1298 Total: 11,100",
      "reason_for_recall": "Eltrombopag Interference  for VITROS Chemistry Products TP Slides and VITROS XT Chemistry Products ALB-TP Slides",
      "recall_initiation_date": "20201215",
      "center_classification_date": "20210112",
      "termination_date": "20210727",
      "report_date": "20210120",
      "code_info": "All Current, Future and Past Lots    UDI: 10758750031771"
    }
  ]
}