      "country": "United States",
      "city": "Mounds View",
      "address_1": "8200 Coral Sea St Ne",
      "reason_for_recall": "There is a potential for an electrical short circuit to develop in a subset of 479 Model 26907 power supply cable plugs that connect to the Medtronic Model 29901 CareLink Encore programmer.  A failure mode of the power supply has been found where an electrical short occurs on the low voltage direct current connector and results in the connector heating up, generating smoke, melting and in some cases disconnecting from the programmer.",
      "address_2": "",
      "product_quantity": "731 units",
      "code_info": "UPN:  00643169121973  Lot/Serial Numbers:  JVD000376P  JVD000415P  JVD000242P  JVD000262P  JVD000264P  JVD000272P  JVD000278P  JVD000283P  JVD000289P  JVD000291P  JVD000293P  JVD000295P  JVD000300P  JVD000311P  JVD000316P  JVD000320P  JVD000330P  JVD000332P  JVD000333P  JVD000334P  JVD000337P  JVD000338P  JVD000344P  JVD000346P  JVD000348P  JVD000350P  JVD000351P  JVD000352P  JVD000354P  JVD000356P  JVD000357P  JVD000359P  JVD000360P  JVD000361P  JVD000362P  JVD000363P  JVD000364P  JVD000367P  JVD000369P  JVD000373P  JVD000377P  JVD000379P  JVD000381P  JVD000382P  JVD000383P  JVD000384P  JVD000388P  JVD000390P  JVD000392P  JVD000394P  JVD000407P  JVD000408P  JVD000409P  JVD000410P  JVD000411P  JVD000412P  JVD000413P  JVD000417P  JVD000419P  JVD000420P  JVD000421P    UPN: 0643169166974  Lot/Serial Numbers:  JVD000331P  JVD000426P  JVD000430P    UPN: 00643169166981  Lot/Serial Numbers:  JVD000273P  JVD000275P  JVD000299P  JVD000307P  JVD000339P    UPN: 00643169166998  Lot/Serial Numbers:  JVD000245P  JVD000265P  JVD000297P  JVD000301P  JVD000313P  JVD000387P  JVD000391P  JVD001090P  JVD001137P    UPN: 00643169167001  Lot/Serial Numbers:  JVD000353P  JVD000395P    UPN: 00643169167018  Lot/Serial Numbers:  JVD000386P  JVD000389P  JVD000425P  JVD000432P  JVD000269P  JVD000290P  JVD000326P    UPN: 00643169167025  Lot/Serial Numbers:   JVD000439P  JVD000308P  JVD000385P  JVD001089P  JVD000370P    UPN: 00643169167049  Lot/Serial Numbers:  JVD000322P  JVD000378P  JVD000418P  JVD000434P    UPN: 00643169167063  Lot/Serial Numbers:  JVD000298P  JVD000305P  JVD000318P  JVD000342P  JVD000355P    UPN: 00643169167070  Lot/Serial Numbers:  JVD000374P  JVD000292P  JVD000324P  JVD000343P  JVD000371P  JVD000247P  JVD000257P  JVD000261P  JVD000287P  JVD000372P  JVD000399P  JVD000405P  JVD000414P    UPN: 00643169167087  Lot/Serial Numbers:  JVD000253P  JVD000263P  JVD000284P  JVD000304P  JVD000317P  JVD000319P  JVD000393P  JVD000398P  JVD000402P  JVD001031P  JVD001091P    UPN: 00643169167094  Lot/Serial Numbers:  JVD000282P  JVD000271P  JVD000397P  JVD000404P    UPN: 00643169167100  Lot/Serial Numbers:  JVD000267P  JVD001004P  JVD000294P  JVD000314P    UPN: 00643169225732  Lot/Serial Numbers:  JVD001067P  JVD001068P  JVD001098P  JVD001112P  JVD001146P  JVD001147P  JVD001008P  JVD001016P  JVD001063P  JVD001064P  JVD001065P  JVD000525P  JVD001002P  JVD000474P  JVD000489P  JVD000501P  JVD000523P  JVD000463P  JVD000473P  JVD000484P  JVD000459P    UPN: 00643169225749  Lot/Serial Numbers:  JVD000251P  JVD000266P  JVD000285P    UPN: 00643169225756  Lot/Serial Numbers:  JVD000347P  JVD000288P    UPN: 00643169225763  Lot/Serial Numbers:   JVD000259P  JVD000260P    UPN: 00643169225770  Lot/Serial Numbers:  JVD000244P  JVD000254P  JVD000256P  JVD000258P  JVD000276P  JVD000309P  JVD000310P    UPN: 00643169369030  Lot/Serial Numbers:  JVD001032P  JVD001271P  JVD000279P  JVD000437P  JVD001027P  JVD002984P  JVD002985P  JVD002986P  JVD002988P  JVD002989P  JVD002991P  JVD002992P  JVD002996P  JVD002997P  JVD002999P  JVD003001P  JVD003002P  JVD003004P  JVD003006P  JVD001389P  JVD001465P  JVD000255P  JVD001024P  JVD001186P  JVD001212P  JVD001226P  JVD001235P  JVD001242P  JVD001292P  JVD001406P  JVD001454P  JVD001481P  JVD001502P  JVD001345P  JVD001185P  JVD001267P  JVD001347P  JVD000401P  JVD001026P  JVD001110P  JVD001152P  JVD001203P  JVD001247P  JVD001311P  JVD001507P  JVD000321P  JVD000336P  JVD000424P  JVD000428P  JVD000461P  JVD000487P  JVD000493P  JVD000528P  JVD001000P  JVD001102P  JVD001104P  JVD001106P  JVD001115P  JVD001138P  JVD001157P  JVD001171P  JVD001178P  JVD001179P  JVD001187P  JVD001191P  JVD001202P  JVD001204P  JVD001208P  JVD001214P  JVD001221P  JVD001246P  JVD001260P  JVD001298P  JVD001299P  JVD001306P  JVD001308P  JVD001314P  JVD001349P  JVD001387P  JVD001419P  JVD001431P  JVD001434P  JVD001447P  JVD001617P  JVD001018P  JVD001448P  JVD001289P  JVD001415P  JVD001023P  JVD001253P  JVD001466P  JVD001467P  JVD001283P  JVD001512P  JVD001057P  JVD001430P  JVD001571P  JVD001254P  JVD001309P  JVD001468P  JVD001470P  JVD001182P  JVD001375P  JVD001161P  JVD000496P  JVD001162P  JVD001380P  JVD001085P  JVD001441P  JVD000423P  JVD001426P  JVD000328P  JVD001164P  JVD001172P  JVD001087P  JVD001453P  JVD001218P  JVD000427P  JVD000524P  JVD001007P  JVD001061P  JVD001386P  JVD001417P  JVD001491P  JVD001055P  JVD001094P  JVD001378P  JVD000274P  JVD000502P  JVD000503P  JVD001052P  JVD001070P  JVD001176P  JVD001461P  JVD001483P  JVD000368P  JVD001078P  JVD001265P  JVD001450P  JVD001009P  JVD001416P  JVD001047P  JVD000329P  JVD001370P  JVD000277P  JVD000323P  JVD001194P  JVD001363P  JVD001036P  JVD000480P  JVD001167P  JVD001328P  JVD001359P  JVD001001P  JVD000464P  JVD001056P  JVD001425P  JVD001046P  JVD001005P  JVD000491P  JVD001132P  JVD001219P  JVD001291P  JVD001071P  JVD001072P  JVD001352P  JVD001492P  JVD000280P  JVD000472P  JVD000531P  JVD001021P  JVD001042P  JVD001045P  JVD001082P  JVD001183P  JVD001301P  JVD001364P  JVD000519P  JVD000466P  JVD001356P  JVD001059P  JVD001159P  JVD000494P  JVD001030P  JVD001109P  JVD001116P  JVD001455P  JVD001077P  JVD001486P  JVD000286P  JVD000312P  JVD001438P  JVD001510P  JVD001160P  JVD001262P  JVD001270P  JVD001284P  JVD000513P  JVD001107P  JVD001108P  JVD001540P  JVD001213P  JVD001230P  JVD001158P  JVD001233P  JVD001193P  JVD001224P  JVD001288P  JVD001290P  JVD001407P  JVD000438P  JVD000504P  JVD001256P  JVD000488P  JVD001125P  JVD001232P  JVD001245P  JVD001278P  JVD001279P  JVD001281P  JVD001432P  JVD001261P  JVD001268P  JVD001493P  JVD001096P  JVD001080P  JVD000508P  JVD001049P  JVD001197P  JVD001251P  JVD001274P  JVD001484P  JVD001442P  JVD001495P  JVD000462P  JVD001105P  JVD001128P  JVD001225P  JVD001294P  JVD001422P  JVD000515P  JVD000516P  JVD001180P  JVD001383P  JVD001457P  JVD000526P  JVD001041P  JVD001083P  JVD001111P  JVD001354P  JVD001396P  JVD001471P  JVD001220P  JVD001286P  JVD001139P  JVD001413P  JVD001472P  JVD000302P  JVD000396P  JVD000400P  JVD000475P  JVD000483P  JVD000486P  JVD000490P  JVD000498P  JVD000514P  JVD000517P  JVD000532P  JVD001003P  JVD001019P  JVD001022P  JVD001033P  JVD001034P  JVD001043P  JVD001048P  JVD001051P  JVD001053P  JVD001079P  JVD001084P  JVD001092P  JVD001095P  JVD001097P  JVD001099P  JVD001103P  JVD001131P  JVD001133P  JVD001134P  JVD001135P  JVD001150P  JVD001151P  JVD001154P  JVD001163P  JVD001168P  JVD001169P  JVD001170P  JVD001181P  JVD001188P  JVD001189P  JVD001190P  JVD001192P  JVD001196P  JVD001198P  JVD001199P  JVD001200P  JVD001201P  JVD001210P  JVD001217P  JVD001227P  JVD001229P  JVD001244P  JVD001248P  JVD001249P  JVD001257P  JVD001263P  JVD001264P  JVD001272P  JVD001273P  JVD001282P  JVD001293P  JVD001303P  JVD001307P  JVD001312P  JVD001323P  JVD001343P  JVD001351P  JVD001353P  JVD001358P  JVD001362P  JVD001365P  JVD001367P  JVD001371P  JVD001372P  JVD001373P  JVD001379P  JVD001381P  JVD001382P  JVD001384P  JVD001385P  JVD001391P  JVD001392P  JVD001394P  JVD001395P  JVD001405P  JVD001420P  JVD001421P  JVD001427P  JVD001433P  JVD001436P  JVD001451P  JVD001480P  JVD001482P  JVD001494P  JVD001504P  JVD001020P  JVD001240P  JVD001313P  JVD001410P  JVD001165P  JVD001148P  JVD000499P  JVD000500P  JVD000505P  JVD001445P  JVD000380P  JVD000520P  JVD001006P  JVD001100P  JVD001175P  JVD001252P  JVD001460P  JVD001477P  JVD001404P  JVD001485P  JVD000465P  JVD001231P  JVD000375P  JVD000436P  JVD000485P  JVD001144P  JVD001402P  JVD000303P  JVD000315P  JVD000325P  JVD000327P  JVD000335P  JVD000345P  JVD000349P  JVD000365P  JVD000366P  JVD000406P  JVD000416P  JVD000422P  JVD000433P  JVD000435P  JVD000442P  JVD000443P  JVD000458P  JVD000468P  JVD000469P  JVD000476P  JVD000477P  JVD000512P  JVD001011P  JVD001017P  JVD001040P  JVD001044P  JVD001060P  JVD001073P  JVD001074P  JVD001081P  JVD001088P  JVD001113P  JVD001117P  JVD001119P  JVD001120P  JVD001121P  JVD001122P  JVD001123P  JVD001140P  JVD001141P  JVD001153P  JVD001184P  JVD001195P  JVD001206P  JVD001211P  JVD001215P  JVD001216P  JVD001228P  JVD001243P  JVD001259P  JVD001280P  JVD001285P  JVD001287P  JVD001302P  JVD001319P  JVD001322P  JVD001324P  JVD001350P  JVD001357P  JVD001361P  JVD001374P  JVD001376P  JVD001388P  JVD001390P  JVD001393P  JVD001411P  JVD001418P  JVD001423P  JVD001429P  JVD001440P  JVD001443P  JVD001449P  JVD001462P  JVD000306P  JVD000460P  JVD000471P  JVD000495P  JVD000506P  JVD000507P  JVD001066P  JVD001086P  JVD001222P  JVD001223P  JVD001236P  JVD001239P  JVD001275P  JVD001276P  JVD001277P  JVD001295P  JVD001296P  JVD001310P  JVD001377P  JVD001444P  JVD001446P  JVD001496P  JVD000243P  JVD000246P  JVD000248P  JVD000249P  JVD000250P  JVD000252P  JVD000270P  JVD000281P  JVD000340P  JVD000341P  JVD000403P  JVD000429P  JVD000431P  JVD000440P  JVD000441P  JVD000456P  JVD000457P  JVD000470P  JVD000492P  JVD000510P  JVD000527P  JVD000530P  JVD001010P  JVD001012P  JVD001013P  JVD001015P  JVD001035P  JVD001037P  JVD001039P  JVD001050P  JVD001054P  JVD001058P  JVD001075P  JVD001076P  JVD001093P  JVD001101P  JVD001114P  JVD001118P  JVD001124P  JVD001126P  JVD001127P  JVD001129P  JVD001142P  JVD001143P  JVD001149P  JVD001155P  JVD001156P  JVD001166P  JVD001173P  JVD001174P  JVD001177P  JVD001205P  JVD001207P  JVD001209P  JVD001234P  JVD001237P  JVD001238P  JVD001241P  JVD001250P  JVD001255P  JVD001258P  JVD001266P  JVD001269P  JVD001297P  JVD001300P  JVD001304P  JVD001305P  JVD001355P  JVD001360P  JVD001366P  JVD001369P  JVD001399P  JVD001412P  JVD001414P  JVD001424P  JVD001428P  JVD001435P  JVD001439P  JVD001452P  JVD001456P  JVD001458P  JVD001459P  JVD001463P  JVD001469P  JVD001497P  JVD001498P  JVD001501P  JVD001515P",
      "center_classification_date": "20190216",
      "distribution_pattern": "US and Armenia, Bulgaria, Croatia, Georgia, Hungary, India, Iran, Jordan, Kazakhstan, Latvia, Lebanon, Oman, Pakistan, Qatar, Romania, Saudi Arabia,   Syrian Arab Republic",
      "state": "MN",
      "product_description": "Medtronic Power Supply 26907, Medtronic CareLink Encore(TM) 29901 Programmer",
      "report_date": "20190227",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)",
      "recall_number": "Z-0845-2019",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "81950",
      "termination_date": "20210330",
      "more_code_info": "",
      "recall_initiation_date": "20190117",
      "postal_code": "55112-4391",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}