      "country": "United States",
      "city": "Irvine",
      "address_1": "33 Technology Dr",
      "reason_for_recall": "Advanced Sterilization Products is recalling the STERRAD System cassettes  because the the leak indicator on the STERRAD System cassettes packaging may not function properly after exposure to hydrogen peroxide.",
      "address_2": "",
      "product_quantity": "19,978 units total (9,792 units in US)",
      "code_info": "33090405  33101190  33120219  33100496  33110562  33130219  33110366  33130465  33111520  33130282  10033071044  33110944  33111953  33120721  33090217  33101176  10033070871  33102083  10033050639  10033060586  33102226  33111924  33101329  33100417  33101016  33111073  33120793  33110111  33090419  33100454  10033060228  10033060960  33111103  10033070692  33101139  33101386  33101053  10033070360  33120921  33120191  10033050160  10033060708  33101575  33130237  10033050205  33120742  33110536  1033141141  33110622  33090535  33090549  10033060991  33120620  33130291  10033060648  33110286  33101395  33120795  10033060743  33101375  33102015  33110760  33100411  33101406  10033060266  10033070410  10033050565  33081285  33110575  33090340  33111511  33101110  33081415  33111660  10033060407  33100166  33100318  33130097  33110043  33081367  33101197  33101343  33100592  33111108  33111413  33120560  33110415  33102140  33130353  33100423  10033060173  33081029  33110958  10033070162  10033080326  33111513  10033080222  33090107  33090352  33090118  33111583  33090560  10033060942  33111555  33110481  33101447  10033070646  33111790  33100545  33111940  33102257  33111983  33111904  33120723  33110218  10033080304  33120450  33120497  33120540  33120543  33130027  33100483  33100418  33100422  33100032  33110838  33111228  10033060274  33120838  33101567  33101002  10033070254  33111685  33050422  33110188  33110544  1033141029  1033142008  33111618  33111915  33090506  10033070765  33090385  33111025  10033060205  33111516  33110553  33111488  33110588  33130181  33101526  33090480  33130031  33101163  33130273  10033060638  33090455  33140028  33111965  33090245  10033050201  33081422  10033070645  33130458  33090005  33090567  33110340  10033070524  33140008  33111653  10033060608  33120840  33090451  33111796  10033050455  33090470  33100173  10033060658  10033070457  33100275  33110628  33111066  33110653  10033050566  33102338  33120616  33101008  10033050270  10033060591  33090016  33102332  10033060636  10033060899  33111051  33120677  33081144  10033080293  10033060506  33101042  33101142  33102166  1033141046  10033070587  33100270  33101533  33130454  33140103  33090337  33100360  33140001  33101358  33090412  10033070688  10033070341  33110300  33111082  10033070903  33110100  10033050693  33110133  33111955  33110108  33111167  33110938  33110956  33110860  33111134  33111148  33111458  33111429  33110131  33111043  33101412  33110532  33090264  33101372  33102387  33120144  1033142038  33090553  10033060455  33110714  33111844  33110753  33101075  33102065  33100552  10033070256  10033070309  33081452  33102118  33110129  33110006  10033060211  33120099  33100009  33110559  33101553  33120237  33140029  33120141  33111046  33110355  33110797  33111906  33120374  33111826  33111220  33120387  33102079  33110721  33120464  33110008  33110560  33081091  33100319  33081149  33090527  33090530  33111860  33110624  10033060735  33090059  33102025  1033141128  33110527  33102030  33110531  10033060827  10033050415  33111475  33120648  33120150  33090019  33081264  33110303  10033060862  33120232  10033080232  33111024  10033080284  10033080317  10033060706  33111892  10033070685  33111437  33111503  10033080335  33100329  33100175  10033070604  33110222  33100047  33111883  33081292  33111903  33100437  10033060923  10033050591  33101382  33101550  10033060609  33120562  33111629  33110748  10033070376  33101166  33101509  33110097  33111812  33120508  10033071010  33120278  33120940  33120155  33111299  10033060257  10033050645  33110810  33110912  10033050665  33110409  33101431  33111471  10033060239  33110154  33111393  33111407  33100055  10033080070  33120066  10033050467  10033070844  33110505  10033050530  33111166  33102286  10033050627  10033050628  10033070484  33101423  33101123  10033060291  33120924  33110565  33111224  33101072  33101192  33110976  33111028  33110239  33101571  33111619  33120328  33120891  33090301  33110421  33120726  33101135  10033080229  33100070  33090489  33110891  33081420  33102061  33120348  10033070804  33110264  33101572  33101094  33090184  10033060710  33111150  33111581  33090400  1033141075  33110882  33111528  10033060800  10033050180  33100224  10033050249  10033070387  33120880  10033070625  33090206  33110501  10033060714  33111484  33111491  33111452  33111823  33130235  33120053  33111055  33130515  10033060998  33090087  33100091  33102376  33120937  10033070886  10033070899  10033060500  33120426  33101168  33101330  33100337  33120069  33101405  33101161  33101351  10033050181  33111231  33081402  33130468  33110063  33100373  33100323  33111806  10033050432  10033070581  33101622  33100346  33110060  10033050219  33102127  33100106  33100113  33100001  33090073  33090554  33100087  33100119  33111200  33110699  33111164  33111462  33111172  10033050179  33111658  33111499  33111737  33110898  33111036  33110886  33110260  33111777  10033060466  10033060797  33120311  33101366  33101408  33111993  33111543  33090010  10033060227  33101555  33130327  33120010  33130135  10033050280  10033070255  10033060961  33090452  10033070547  33090490  10033080216  33120220  10033060565  33110473  10033050429  33101418  33120209  33111049  33130226  1033141077  33100618  33130076  33120288  10033060148  10033070133  10033060426  10033080218  33111683  33110445  33140073  33100109  33120358  33110964  33110181  33110617  10033050419  33100338  33120076  10033070057  10033060159  10033060489  33111385  33120103  33101476  33130199  33101370  10033070980  10033070578  33110592  33111705  33100084  33100078  33111382  33120894  33110258  33100245  33102060  33110496  33101000  33120154  33111115  10033071016  10033060258  33100448  33120251  10033070493  33120929  33101554  33100607  33100604  33110608  33102290  33102274  33102280  10033060968  33101188  1033141022  33120481  1033141034  33110660  33101305  1033141050  33120224  33111397  1033141020  33100095  33100096  33081117  10033050258  33111728  10033070086  10033060849  10033060359  33120519  10033060930  10033080316  33111858  33101640  33110657  33101334  33111016  10033070983  33110468  33110106  33110541  10033050573  10033070215  1033141017  33110356  10033070346  33120310  10033060202  10033060391  33120020  33110024  33110800  33110169  33100535  33100272  33101353  10033050534  33111918  33110772  33100602  10033060441  10033070624  33120476  33101662  33110163  33111780  33111243  33111669  33090015  33110615  33111383  10033080203  33111815  33120794  33111916  33110498  33110172  33120132  33110142  10033050490  33120605  33101432  33111720  33111563  1033141125  33081365  33111734  33111662  33120041  33081297  33111736  33111283  33130117  33100002  10033070370  10033050609  33090235  33090175  33081080  33120511  33090053  33101385  33102241  10033060770  33130001  33111237  33090205  33111507  33110749  1033142025  33111943  33120245  33111836  33100053  33081133  33120610  33101429  10033070815  33101035  33110817  33120927  33100573  33120569  10033060501  10033070544  33090171  33110687  33110821  33101171  1033141104  33100105  33090322  33100605  33100137  10033060244  10033070051  33100183  33120425  10033050183  10033060971  10033060154  10033070881  33110016  33110019  1033141102  33081082  33100061  33110506  33101081  33130396  33101118  33120116  10033050582  33111495  33111514  10033070571  10033060789  33101479  33102307  33100511  33090167  33101058  33102077  33130183  33110794  33111930  33120240  33111663  33100394  33100250  1033141106  33130078  33101355  33090021  33110525  33111029  33100185  33110512  33110288  33120503  33120505  10033070243  33110153  33110210  33081211  33101508  33140038  33090426  33110630  33102315  33100238  33110907  33111828  10033080138  33102126  33101627  33120377  33100155  33110338  33111084  33120489  10033050409  10033070988  33100287  10033060997  10033070394  33110438  33110279  33090339  33110662  1033141001  33120776  10033070619  33102182  10033070076  33110198  33102142  33120176  33120887  10033070661  33111446  33102189  33090504  33090159  33130514  33130088  33100567  33120462  33110551  33110465  33111391  33100339  33120268  33100430  33100206  10033050504  10033060760  33100226  33100419  33101564  10033060286  33102235  33090108  33111225  33100102  10033060531  33110839  10033060865  33111277  33090511  33081118  1033141134  10033070244  10033060358  33120313  33110674  33081382  10033080241  33081467  33130557  33100572  10033080024  33110658  33111064  10033060984  33111438  33100505  33100066  33110492  33110580  33130313  33111431  33102310  10033070292  33130186  33111641  10033050246  10033070379  33110139  10033060527  33111696  33111949  33101536  10033070735  33100386  33110859  33110507  33100370  33102209  33111749  33090266  33120710  33090090  33102277  33120613  10033060985  33111911  33120771  10033050621  33111538  33110418  33111072  33111861  33100583  33081059  33110500  33100616  33081145  10033060564  33110489  33081009  33101359  33101666  33111089  33120792  33101079  33111369  33120479  33102181  33110689  33120803  33090531  33120614  10033060433  10033060128  33100409  10033070444  33110996  33130386  33120587  10033070259  33111052  33110652  33111015  33111508  33102325  33110827  33101004  33111494  33120760  33110853  33110213  10033070689  10033070422  10033070206  33110225  1033141126  33110228  33081302  33110079  33090330  10033060884  33100427  33081380  33110052  33120728  33101595  33120385  33120491  33130490  33110422  33110989  33100633  10033060581  10033060762  33111958  33110274  33130026  10033070236  10033060109  33120203  33110656  33102262  33110862  33090443  33110516  33081403  10033060175  33110278  10033070428  33102321  33110132  33110270  33120280  10033060438  10033080287  33110308  10033071027  33090483  33120524  33111464  33100099  33090397  33090143  10033050261  10033050416  33101445  10033060813  10033070231  33081457  33081132  33111674  33090260  33102293  33120223  33101487  33111000  33100215  10033070062  10033080236  33120951  33100348  33130172  10033060640  33101021  10033070513  33110273  33111133  33120040  33101495  10033060957  33100570  10033060765  33110214  33081399  33120720  33111809  33111406  10033060344  33081227  33090479  33101472  33110842  33101007  33100089  33081323  33101159  33090536  33110591  33130387  33111763  10033060293  10033060567  10033060485  33101427  33111606  33100426  33130368  33102330  33101156  10033070726  33140034  33090384  33100512  33100544  33120777  33111739  33081408  33090334  33111600  33130151  33110243  33130266  33111557  33110363  33120621  33100168  33090210  33120215  33090374  33110700  33081130  33100509  33090502  33081406  10033080243  10033060470  10033060252  10033060480  33130071  33101115  10033060161  33110876  33120337  33090024  33111276  33101198  33081095  33110138  33120400  33102354  33111621  33120595  33090332  10033060101  33120386  33102011  10033060775  33100201  10033060560  33120006  10033070401  10033060920  33101194  33130504  33100524  10033050208  33130087  33130511  33081126  10033060380  33140017  10033060265  33110439  10033050230  33130146  33110215  33100309  33130145  33101623  33101667  33101655  33100313  33100130  33100143  33100146  33100402  33130535  33130336  10033070081  33111349  33100069  33110947  33102370  33111767  33102132  33120090  33101537  33111821  33120603  33100161  33100431  33102021  33110336  33100029  33111991  33111219  33090344  10033080224  10033080083  33111679  33110745  33110762  33110715  1033141115  33120359  33120797  10033060461  33110238  10033060463  33130491  10033060342  33110834  33111434  33110466  33111123  33100205  33100128  33111441  33100521  33111394  33102388  33111687  33110315  33100181  33111744  10033060337  33130022  33110320  33111510  33081127  33120516  33111316  33111615  10033080115  10033070289  33100088  33100504  33111401  33111439  33111474  33111517  33111789  33130466  10033060497  10033060815  10033070612  33120885  33120736  33100265  33100218  33110479  10033060807  33090083  33110058  33101131  33130444  10033070944  33111960  33110767  33110895  33140069  33102384  10033070213  33110262  33110342  33110616  10033050595  33081045  33110542  33130049  33111936  33100387  33100369  33081312  33110639  33101345  33120052  10033060592  33130503  10033050440  10033060547  10033070605  33120946  33110103  33090414  33111994  10033070083  33101450  33100295  33120948  33111588  33110330  33102063  33101175  33081423  33090471  33120272  33111512  33120085  33111731  33100098  33111116  33110322  33111544  10033060237  33110925  33111260  33110618  33090061  33120139  33070361  33102212  33110464  33110472  33130095  10033060377  10033060379  10033070185  33120881  33111409  33111416  33102276  33120717  33100575  33102392  33120070  33090488  33110207  33111927  33102243  33110102  10033050579  10033070882  33110391  33111604  33120681  10033060794  10033070989  33120542  33130014  33120285  33120226  33090409  10033070753  33081147  33101551  33120496  33111688  33102323  33110347  33110603  10033080128  33100288  33130393  33120487  33120436  33101041  10033060986  10033080186  33120455  33100613  33111419  33111724  10033060897  33100176  33110339  33130286  33111675  33120583  33081275  33111532  33081389  33110905  33110452  33130437  33090183  33100125  10033060215  10033060404  10033060424  33102357  33110493  33102167  33101100  10033080033  10033070437  10033070438  10033050426  10033080182  33111395  33111613  33101659  33102155  33120437  33110855  33111873  33140074  33130206  33110387  33100479  33100517  33120420  33120725  33101043  33090117  33111109  10033070838  33111271  33120098  33110299  1033142002  33090161  33111492  33120149  33110185  33111715  33100040  33120827  33120128  10033070117  33100590  33102316  33100554  33100486  33101346  33110483  33110799  33130051  10033071034  10033060822  33081022  33101084  33101435  33100297  33120913  33111896  33111897  33111881  33111701  33111948  33120315  33101383  33111964  10033050575  33130401  10033050217  33101457  33140043  33110359  33110537  10033060907  33120118  33090242  33111554  33102103  33100494  33130497  33111751  33110045  33111700  33111323  33101347  33090207  10033050491  10033070063  10033060898  33111710  33111392  33081387  10033060801  33130174  33120102  33081391  33111686  10033070528  33090219  33100216  33111569  33110152  33100506  33111321  10033060559  33111976  33110490  33101581  33102048  33102378  33120574  33090494  33110511  33090199  33102215  33090261  33111568  33111601  33111610  10033060571  33100071  33120035  33101364  33090368  33111718  33111258  10033050412  33100395  33111363  33111336  10033050482  33090376  1033141079  33081040  33111067  33131001  33120589  10033070155  33081090  33120179  33120515  33081038  33100037  10033080009  33110758  10033070549  33110900  33110979  33101493  33120054  33111371  33120592  33081310  33110086  33100007  33120528  33100145  33120367  33110626  33102204  10033080247  33081181  33111564  33120896  33102344  33130407  33111722  33090109  33090119  33090142  33111902  33110836  10033070618  33102301  33100157  33130227  33110272  10033080324  33120286  10033050301  33120320  33102143  33110921  33081386  33100447  33130371  10033060217  33101354  10033080300  33120789  10033050681  33120034  33111791  33110150  10033070981  10033070583  33102185  33130023  33101387  33130338  33101636  10033071017  10033050611  33101012  33100400  33120217  33120143  33111650  33120242  10033070104  33081413  33101531  33100048  33130238  33111772  33100351  33111138  33090146  33111399  10033060905  33101421  33081196  33081193  33110877  33081015  33081187  33081216  33081219  33081224  33090464  10033050547  10033060343  10033070008  10033070087  33100492  33081166  33081178  33081286  33101625  33101628  33102266  33110085  10033070991  10033071000  33081229  33110040  33111905  33120064  33102353  10033070996  33130502  33130346  10033060465  33101630  33081217  33110977  33110881  33130110  33081188  33081220  33081225  33081233  33102064  33111208  33120023  33130355  33130513  10033050605  10033070443  10033070824  10033070058  33111142  33081308  10033060561  10033070385  33111306  1033141021  10033060292  10033050633  33081138  10033070670  10033080254  33081274  10033050492  10033070084  33102152  10033070013  33120264  33102141  33120429  33120555  33090256  10033050676  10033060207  33081174  10033070488  33090048  33120642  33081222  33102291  10033070067  10033080173  33081232  33100056  33130154  10033060604  33120380  33102096  33110348  33071033  33101580  33101030  33120199  33120057  33081303  33081328  33090532  33110812  10033050439  33110447  33090124  33111335  10033050688  33100305  33110206  33090458  10033070987  10033070447  33090555  10033070467  33100556  10033060248  33101307  33120862  10033050451  33100538  33120096  10033070816  33101086  10033070398  33090189  33110046  33101095  33111506  33101558  33110545  10033050304  33130105  10033070393  33102084  33110178  33111013  33120922  10033071013  33130224  33111888  33111379  10033060369  33100485  33090263  33100278  1033141118  33111682  33101540  33110140  33111529  33102347  33110235  33130428  33111545  33111741  33111935  33110485  33110004  33130331  33120454  33102339  33120722  33102115  33111847  33111748  1033141117  33101650  33101106  33111966  33110184  33102317  10033070537  10033060600  33110316  1033141085  10033050188  10033060264  33120634  33110823  33120252  33101388  33130249  33111890  10033060136  10033060933  10033050496  33101562  33120506  33081051  33101446  33081236  1033141103  33111337  33102133  10033070978  33120189  33120746  33110350  33120604  10033080244  33100551  33100581  33101323  33102250  33110313  33090297  33081253  33120306  33130096  33110712  33102389  10033070494  33102377  10033070147  33081366  10033050654  10033070431  33100475  33102003  33102008  10033070551  10033050571  10033060913  33090566  10033070973  33090429  10033060492  33100008  33100006  33111235  10033050672  33120617  33120945  33130029  33101039  10033070862  10033070626  33081050  33110571  33102327  33101326  33110296  33090243  33130188  33110259  33090148  33111117  33100232  33111288  33120192  33102107  33090380  33100606  33090512  33102275  33111666  10033060672  33111303  33111914  10033060259  33101482  33090294  33111859  33100516  10033070984  10033060796  33090493  33110187  33081444  10033060357  33090080  33110534  10033070451  10033050606  33110798  10033050577  33090045  33090233  10033060105  33090276  33120048  33110941  33130103  33081075  10033070767  33100249  33101407  33102345  33102341  33102263  33111711  33111530  33111410  33101017  33111456  33111535  33110461  10033070594  33111747  33111074  10033070837  33110080  33120745  33110346  33100619  33101169  10033080117  33110693  33110164  33102313  10033050556  33111761  33110864  33110736  33101574  33120338  10033060593  33100465  10033070352  33081008  10033070070  33102348  10033070293  33081277  10033080154  33130113  33130175  33101065  33120101  33120458  10033070025  33101578  33120164  10033080062  1033141095  33120844  33110650  33111414  33100242  33110683  33130400  33111961  33081197  10033060119  33130293  10033070014  33100378  33101481  33111608  33101337  33110739  33120469  33120017  33120716  33130248  33130147  33110936  33090357  33090548  33120729  10033060938  33090274  33102134  10033080043  33090043  10033070456  33120084  33102367  10033070651  10033050438  33100005  33130426  33120151  10033060234  33110048  33100547  33120326  33110162  33101511  33110203  33110778  33102298  33111607  33111553  10033080219  33100314  10033070771  33111689  33102179  33101464  33120201  10033060872  33100634  33090406  10033050618  33111386  33100579  33110360  10033060472  33110680  33110406  33110416  33111898  33120534  33120134  33101444  33101397  10033070751  33111368  33102053  33120346  33090008  10033060190  10033081260  10033050220  33111575  10033060495  10033070575  10033060490  10033080069  33081428  33120548  10033070402  33120309  33130341  33100079  33090356  33111365  33110351  33111069  33110084  33110425  10033070783  33111590  10033070074  33110436  33110923  10033060117  33101098  33101466  10033060512  10033060482  10033050206  33101621  33130028  33111673  33120121  33120512  10033070716  33110276  33102383  33081324  33100187  33110894  33100439  33100340  33100342  33120234  33110450  33090306  33100237  10033070287  33111664  33100474  33100532  33110064  33100420  33102360  10033050649  33120275  33100497  33100140  33100629  33111640  33110236  33090355  33100555  10033080060  33120147  33110955  10033060294  33100488  33110557  10033070189  10033080003  33120353  33102236  33102343  33120609  33111398  33090111  33100367  33110649  33120468  10033060831  10033050583  10033060839  10033060829  10033071045  33102186  10033060383  33120062  33102382  33111100  33081084  33102299  33111879  1033141083  10033050544  33100256  33101477  33111633  33111570  10033050671  33120600  33120068  33101324  10033050594  33110695  33110567  33120130  10033050536  33111257  33120233  33102279  10033070860  33101475  33102322  33110267  10033060663  10033060931  10033070221  10033070599  33111818  33111114  33100459  33110331  10033080141  10033070017  33110470  10033060553  33111846  33102288  33111424  33111677  33140052  33120344  33110735  33120837  33100227  33110200  33101374  33110980  10033070088  33130187  33100594  33100557  10033070898  33120173  33110430  10033070600  33090116  33110808  10033060314  33120293  10033070673  33101411  33130019  33110754  33111586  33110232  10033060476  33120117  33100229  33101499  1033141053  33100450  33130005  33130500  33081388  33110546  33090290  10033070171  33111031  10033050430  10033080280  33110641  33090365  10033070204  33101300  33101341  10033060450  33110390  33120750  33130554  33102157  33110918  33130507  33130318  10033070080  33130279  33120119  10033050182  33110435  10033080250  33111444  33111647  33120784  33102221  33110866  10033060629  33110376  33110378  33090058  33140022  10033070037  10033050279  33120663  10033080087  33120222  33101448  33081429  33111162  33100325  33120954  33100462  33111776  33110863  33110305  10033060783  33110082  10033070436  33110205  10033080032  10033060784  10033050289  33090571  33100164  10033050172  33111810  33110595  33111110  33090558  33100585  33120612  33081321  33120303  10033060243  33111485  33102264  33111161  10033050642  33101164  10033070383  33100406  33110256  10033080042  33090227  33102062  33110395  33090342  33111865  33090030  33111969  33090418  33120451  10033060888  33101639  33100240  33090102  33111753  33081086  33120210  33110771  33111551  33130112  33101473  33110635  33101312  33090084  33100603  10033060731  33101023  33100169  33100063  33100064  33111248  33111794  33102268  10033060178  33090538  33111827  33120148  33120160  33120283  33110520  33120734  33120770  10033080114  10033070109  33110822  33111380  33110884  33110937  10033060607  10033060753  10033050651  33090529  10033070182  33100534  10033080206  33120800  33111350  33100263  33102194  33090517  10033060751  33110607  33100054  33110009  10033050483  33081396  33120112  10033070660  33102232  33102351  1033141101  10033050248  33120944  33102205  33120321  33100344  33120520  33120573  33110830  33101356  33100355  33101560  10033060635  33120590  33111717  33081421  33101082  10033050448  33101303  33101510  10033050442  10033060736  33110851  33101193  10033080104  10033080101  33111691  33110999  33100500  33130317  33130389  33102239  33130107  33111081  33102178  33081390  33110722  10033050237  33081025  33110146  33110229  33101170  33111757  33111531  33110190  33130320  33111445  33120714  33100065  33081442  33130323  10033050564  33111909  33102302  33120942  10033070555  33130120  33111907  33120431  33090564  33100025  33120393  33100129  33081134  33101336  10033070273  10033050423  33111030  33130129  33110920  33130537  33110807  33090393  33101518  33102193  33101557  33111678  10033070389  33110328  33110095  33102386  33101044  33111122  33110297  33110634  33110091  10033060443  33111314  33102369  33110549  10033070740  33120871  10033070576  33120216  33100158  10033070418  33081434  33140061  33120473  33111712  33120483  33090277  10033070452  33111632  33110433  33110871  33111759  33120284  1033142058  33111032  10033060866  33111450  10033060352  33110333  33111785  33101441  33101420  33110448  33110570  33110250  33101419  33111053  33090232  33120063  33110193  10033050658  10033070275  33100135  33111370  1033141109  33111007  33110197  10033071012  33110432  33110151  33111045  33120773  33111143  33111661  33090071  33110317  33081123  33101155  33100282  33102289  33102335  33110101  10033070072  33090283  33100076  33120641  33120916  33102278  33111819  33120649  33110818  33081243  33130021  33130052  33110869  33110703  33110204  33120111  33130012  33110165  33110411  10033070887  33110558  33110323  10033060218  10033060473  33100330  10033070035  33110196  10033050446  33090012  33101598  10033060247  33120109  33090499  33081020  10033050459  33130325  33140004  33110966  33100085  33101490  33120184  33102271  10033050589  33090492  1033142024  10033080214  33101315  33081412  33110833  10033050669  33120430  33110298  33101492  10033050660  33111268  33111233  33110018  33111180  33110011  33110685  33110015  33101668  10033070397  33111671  33100162  10033060707  33120626  33102020  33102027  33102049  33130526  10033080035  33090020  33111952  33120123  33110783  33090547  33120081  10033070187  10033060347  10033060419  33101342  33100035  33100111  33100300  33120168  33050413  33130053  10033070516  33100375  10033050444  33090287  10033080058  33130191  10033050568  33101022  33110782  33110706  33130013  33110224  33090280  10033060711  10033050562  33111941  33140076  33081076  33081440  33120593  33111719  33090568  33111804  33110440  10033070343  33081004  33130149  33100399  33090354  10033080189  33120901  33111648  33111630  33110388  33111173  33110837  33110474  10033070736  33110383  33100171  1033141063  33081005  33111825  33120484  33120015  10033060610  10033070873  33100134  33110135  33111387  33101069  33111832  33110924  33102284  33081404  33090424  33120536  10033050260  10033060232  33110651  10033050590  33081028  33120585  33111567  33101150  33110766  33130106  33110691  33110806  33100151  33110370  33120629  33131000  10033070115  10033050629  33111920  10033050452  10033070121  1033142001  10033060103  10033050668  33120796  33111515  1033141016  33111784  33120138  33111631  33090007  33090465  33101673  33090113  33111037  33090060  10033050559  33120578  33120790  33110307  33140056  33120276  33110513  33100482  33110922  33120172  33110354  33110417  10033060440  33110372  33120142  10033060368  10033080252  33090132  10033080272  33111655  33102032  33101637  10033070643  33110948  33110098  33081237  33101456  10033060886  33101325  33101516  33090526  10033060572  33090141  33100303  33081296  33110578  33081443  10033070024  33120686  33120507  33130125  33090518  33090522  33090017  33120345  33090178  33130176  33101073  33100039  33120783  33130201  10033070477  1033141144  1033142006  33090190  33101371  33120439  33100052  33120877  33101590  33111750  33101400  33111709  10033050570  10033050407  33110134  33102309  33120488  10033080052  33090396  33100068  33101134  10033080306  33120801  33110969  33101626  33102160  33110047  33081047  33102254  33110414  10033050456  10033060195  10033060622  33130104  10033070304  33120004  33110362  33130209  33090331  33090278  10033050622  33081393  33100027  33110494  10033071020  33111487  33110609  33120213  33120207  33102159  10033060675  33102210  33102292  33110517  33102208  33102173  33120317  33100214  33111505  33101332  33110988  33101610  10033070763  10033070772  10033070050  33110849  33101538  33101360  33110757  33120952  33102131  33090044  1033141028  33111102  33102246  10033060220  33102014  33111684  33111417  33081307  33111353  33081027  33081354  10033060755  33111269  33111270  10033060966  33110543  33110424  10033070474  33110419  33090028  33101020  10033070089  33120658  10033060941  33081153  33111490  33110208  33110374  33120097  33110209  33120158  33130505  10033070582  33081119  33110246  10033060892  33120625  33081276  10033070468  33111091  33081273  33110231  33111922  10033060300  33110309  33110769  33110352  33090259  10033080142  10033070515  33100090  33100382  1033142032  33110051  10033050434  33111047  10033050687  10033060170  10033050472  33120471  33101032  10033050424  10033070455  33090425  33110826  33081267  10033070969  33110568  33120156  33102195  33110973  33081272  33090029  33090520  33100136  33120646  33100154  33111817  33110645  33111781  33120279  10033070095  33110105  10033070118  33140053  10033060634  33101428  33110710  33120114  33102267  33130044  33120852  33110583  33110569  10033070342  10033080332  10033070775  33120923  33100019  10033050463  33110514  10033060262  33111743  33100023  33111603  10033050587  33101565  10033060702  10033060467  33111170  33110226  33130567  33110510  33100333  33140020  33110130  33110282  10033070414  33101160  33111877  33090147  33090022  33111704  10033060507  33110943  33130315  33101077  33130011  33100230  10033060922  33081000  33120122  33120093  33130133  33111893  10033060417  33110186  10033050574  33100170  33111175  10033060401  10033060475  33111854  33120816  33081247  33100510  33102282  33120046  33090467  33100141  33101067  33110495  33130314  1033142045  33110843  33102033  33111124  33110384  33111556  33100432  33111752  33111714  33111396  33110036  10033050408  33110950  10033060947  33130100  33100326  10033060514  10033080313  33101532  33111378  33130470  33101340  33090257  33111801  33101111  33130374  33110586  10033060488  33101055  33120355  33110059  33111620  33090559  33110793  33110090  33111085  33120382  33090360  33101582  33120829  33090528  33101545  33120120  33111764  33090027  33111934  33140066  33130324  33101025  33100507  33090508  10033060995  33110763  33110963  33110443  33100443  33130476  33111092  33111829  10033070867  33130178  33110407  10033060569  33081180  33090466  10033050617  33110577  33102139  33111652  1033141078  33130308  10033070967  33120033  33100536  33130126  33110903  33100610  33102368  33110926  33081439  33120086  10033070623  33081089  33081309  33120044  33111524  33111451  33111428  33110789  33110675  33100408  33120421  33090164  33110442  33111426  33111774  33102305  33110711  33110361  10033070226  33101057  33120371  33120352  33081081  33081124  10033060508  33111742  10033050514  33111842  33110780  33100515  33102202  33102191  33100453  33111635  33100491  33090140  33081150  33111830  10033050213  33101045  33101078  33101036  10033070369  33081395  33111668  33110540  10033060345  10033050406  33100413  33110701  33101544  10033071022  33081001  33111769  33111656  33090040  33110713  33130163  33110166  33120307  10033050245  33130363  33111470  33081043  33090198  33101489  33102390  33110717  33111792  10033070108  33110027  33120563  33101331  33110148  10033070658  1033141008  33100301  33110326  33110756  10033060388  33120336  10033050529  10033070207  10033060100  33090454  33081231  33110631  33101310  33120171  10033070872  33120530  33101003  33102161  33100589  33111525  33101552  33120826  33100213  33100425  33101663  33110572  10033050638  33102087  33120627  33101010  33111099  33100558  33101120  33110795  33111372  33120767  10033060405  33110854  10033070308  10033080100  33101014  10033050614  33100574  33101653  33102285  33110397  33111856  33130417  10033060621  10033070038  33110026  33100451  33111430  33111755  33090103  33101046  33100153  33110521  33110729  33110742  10033070400  10033060962  33110951  33110266  33090208  10033050523  10033070054  10033080139  33110177  33102281  33111762  33101101  33100139  33100231  33130531  10033060932  10033050613  33111895  33120660  33120738  33100075  10033070191  33111565  33101589  33111587  33090562  33110761  10",
      "center_classification_date": "20141222",
      "distribution_pattern": "Worldwide Distribution.",
      "state": "CA",
      "product_description": "STERRAD NX, Product Code 10133    The STERRAD Cassettes are used on STERRAD Sterilization System which is a low temperature general purpose sterilizer used to sterilize heat and moisture sensitive reusable medical devices.",
      "report_date": "20141231",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Advanced Sterilization Products",
      "recall_number": "Z-0845-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "69692",
      "termination_date": "20150804",
      "more_code_info": "033070548  33100208  33101006  1033141107  10033070946  33111667  10033050517  10033070449  33090120  33111509  33090514  33101471  33130321  33130329  10033060166  10033060787  10033080122  33111207  33120423  33111442  33130334  33110915  10033070224  10033070470  33110268  33111919  33102223  33120673  10033060158  10033080111  33111954  33100455  33110335  33101535  33110857  33100446  33111833  33111824  33110310  33110314  10033050630  33090127  33110621  33110182  33090569  33102170  33120146  33110928  33110740  33110589  33110373  33110393  33120406  33111176  33111083  10033070413  33110933  33110954  33110858  33081042  1033141026  33111326  33110619  33111244  10033080242  33110637  10033050400  33100192  33100380  33101063  10033071023  33111800  33111642  33102230  33111868  10033070432  33110802  10033071014  10033060493  33111238  33090427  33110261  33102312  10033070459  33110523  33120212  10033080191  33120897  10033050505  33111496  33102358  33140019  33102259  10033070482  33081039  33110719  33100304  33120190  33101191  10033060918  33120899  33120670  33130339  33110777  33090459  10033050428  10033060509  33090078  33101485  33100259  1033141133  1033141132  33100247  33101631  33110708  33100576  33111693  33120525  33120356  33100365  33110811  33111853  33101539  33111297  33110705  33100262  10033070235  33111217  33130162  33120073  10033060242  33120474  33110934  33101141  33110426  10033070079  33081167  10033060952  33081325  33110816  10033070301  10033070760  33110155  10033060460  33110724  10033060260  33090226  33111636  33111894  33111889  33120791  33110020  33102303  33110462  33110353  33110034  33110919  10033070729  33100612  33110223  33110327  33100328  33130132  33111291  10033070448  33100180  33110389  33111760  33111246  33090282  33111080  33111939  33110365  33110856  33101013  10033060654  33110636  33120510  33110191  33102283  10033070463  33120238  10033070892  10033070846  10033060613  33081414  33120347  10033070748  33110295  10033060970  10033060983  33110987  33110499  33090023  33140049  33110702  10033070060  33111963  33120133  33130080  33130512  33110831  33111838  1033141014  10033080041  33100124  33130225  33102385  33100559  33102010  10033080089  33102379  33120186  1033141114  33102213  33130528  10033060612  33130474  33120180  33081405  33120018  33120077  33140027  33111708  33130484  10033070492  10033070744  33102116  33120218  10033050236  10033060767  33101529  33120907  33102055  33081088  33111364  33101302  33101380  10033060756  10033060144  33101009  33110875  10033070601  33102086  10033050585  33111886  33100012  10033080204  33111807  33120071  33100320  33102122  33111079  33100527  33111501  33110195  33090402  33111725  33090225  10033070030  33081131  10033070423  33101054  33100518  33090416  33110202  33110716  33110904  10033070433  10033060192  33130516  33101585  10033080258  33110033  33100484  33111740  33120502  33102120  10033070465  33111758  33111729  33120460  33090348  33090101  33110932  33100520  33120092  33110556  10033050161  33111006  33120895  10033080067  33102371  33111304  33081369  33120091  10033060826  10033070977  33111322  33100225  33111415  10033070211  33090187  33130397  33100530  33111469  33120124  33111651  33100550  33100317  33102009  33100528  33100424  1033142010  10033070055  33101603  10033080202  33101038  33100508  33130433  33081044  10033060185  33110790  33111183  33111232  10033050479  10033050624  33110456  10033070390  33110576  33120005  10033060251  10033070068  33081456  10033060302  10033050433  10033060226  10033060363  33081041  33100478  33130467  33100220  33081383  10033080336  10033080051  33111560  33102174  33110953  33120129  33110394  10033080103  33130167  33100072  33120832  33111347  33111145  33120106  10033060532  1033142049  33102176  33111558  33102328  33111559  33111694  33081353  33102081  33130494  33120227  33110233  10033050667  10033060580  33110497  33081177  33110159  33110399  33110792  10033060317  10033060958  10033050404  33120302  33120294  10033060350  10033050603  33101568  33111723  33110524  10033050554  33120920  33101189  33110526  33120893  33110564  33120208  33110720  33081346  33110089  33101488  33130330  33120402  33120586  10033050596  33110013  33120339  33130268  33100523  33102231  33110092  33120926  33102153  33111561  33111850  33110167  33101376  33111504  33110194  33090284  33130456  1033141042  33110548  33100104  33110554  10033070120  33110049  33111094  33100254  33102165  33111968  33100368  33111617  33110519  10033070205  33120882  33120493  33111433  10033070059  1033141086  33120449  33111502  33102164  33081438  33111891  33111111  10033070473  10033050215  33120297  33100352  10033080282  10033070806  33081186  33081239  10033070682  33120558  33090193  10033080074  33100630  33120943  33101119  33110796  33110908  33111562  33100296  33110434  33100366  10033070478  10033060579  33130009  33100291  33120465  33100374  33100281  33100359  33100244  33100343  33100236  33102220  33120628  10033060573  10033060188  33111862  33120575  10033070950  10033070141  10033070399  33120167  33120552  10033070864  33130509  33101586  33111716  10033050535  33101587  33081351  33081416  10033060599  33090049  33100277  33090211  33120082  33100470  33120135  33110629  33110611  33120105  33111135  33120319  33130039  10033070472  33100401  33090163  33110451  33130560  33100092  33120415  33100046  10033060548  10033060439  33111582  33120690  33101083  10033080226  10033070621  33090533  10033070100  33140003  33102197  33102196  33102201  33102177  33111249  33130128  33120845  33111721  33102199  33090222  33102180  33130119  33081419  33100428  10033080120  10033050664  33081234  33110688  33100398  33110961  33120691  33111589  33120323  33111497  33100435  33102175  33110170  33130391  33120334  33101525  33101338  33130255  33120162  33101425  33120108  10033060201  33140037  33090292  10033071015  10033060980  10033070667  33130561  33111840  33111343  33111327  33130269  33110083  33130354  33120622  33100132  33081037  33110344  33120853  33081002  1033141031  33100235  33090173  10033060664  33111222  10033070966  33102129  10033070979  33120912  33111139  10033060633  33120577  33120396  33120561  10033060766  1033141159  10033060876  10033070526  33101394  33120080  33081426  1033141048  33101559  33100042  33102101  33130232  33110673  33110946  10033060385  10033070580  33090475  33111384  33100163  33100495  33111093  33090561  10033060328  10033080105  33100267  10033060312  10033080118  10033060631  10033060272  10033060716  33090315  33120299  33120175  33120365  10033070396  33120930  10033061020  33120163  10033060688  10033070947  33101396  10033080221  33110604  33111831  10033070404  10033060769  33111098  33102156  33111014  33111876  33130047  33102113  33100050  33111872  10033050447  33111465  33120269  10033060209  33090509  10033070098  10033050494  33110522  10033070134  33110385  10033070222  10033060330  33120089  33081032  33081282  1033141018  33110212  33130427  33102169  33111950  33110623  33111178  10033060975  33120395  33111086  33090540  33130360  33140036  1033141012  33120094  33081034  33081409  33101665  10033060758  10033050287  33120290  33120039  33081176  33102214  33130431  33120533  33120466  33111706  33090351  33120298  10033060130  33101604  10033070776  10033050486  33110369  33120928  33101449  33120170  10033080314  10033080064  33090006  33130498  33081266  33111884  33101179  33110118  33101019  33120718  10033060176  33120492  33090495  33090570  33081427  33101507  33110220  33120354  33100358  33111808  33111908  33100293  33111799  33100356  10033070297  10033050632  33101059  33090089  33102171  33111573  33100414  10033050470  33130024  33102163  33111880  33120526  33100577  10033070668  33102249  33102069  10033060963  33111703  33120047  33111253  33081030  10033060256  33100563  33101613  33130082  33101144  33102245  33102306  33090367  33130239  33110672  33101399  33100013  33111500  33100246  33090082  33101534  33081230  33111151  33110062  33101618  33100203  33130565  1033141030  33081019  33102017  33102296  33110392  10033060617  33120061  33120477  33110805  10033060836  33110828  33110861  33110840  33110759  33090338  10033060582  33110677  33101125  33111875  33120013  33111622  33102057  10033050202  33111917  33101195  33110396  33110819  33130307  10033070875  10033080171  10033060478  33081035  33120381  33101566  33100043  10033050302  33120095  33130257  33120906  33130563  33081241  33081466  33102219  33110696  33102229  33110939  33110993  33081148  33111899  33101112  33120615  10033060746  33111851  33111867  33090128  33090275  33090353  33100519  33100553  33101649  33102098  33110846  33110942  33110960  33111050  33120414  33120889  10033070434  10033070557  10033070758  10033070298  33081006  33130352  33140047  33090358  33111422  33111778  33090350  33090293  33090295  10033070372  33130075  10033070757  33110161  33100354  10033070795  33111420  33110733  33060910  33100377  33111136  33130359  33120611  33090346  10033060575  10033070378  33100626  33111071  10033060671  33120835  33130111  33081281  33130430  10033050610  33090543  10033050673  33090498  1033141072  33120527  1033141032  33102093  33111041  33100334  33100178  33110911  33110539  33120666  33140033  33130004  33090234  33120107  33111313  33130214  33130332  1033141007  33102222  33102255  33120678  33101576  10033050647  33101556  33111137  33120500  33110917  33090218  33110446  33110292  33120644  33102102  33110585  33081165  33110217  33120602  33102304  33111699  10033060376  33101015  33111727  33100172  33090438  33081023  33111882  33111967  33120301  33100271  33120422  33140039  33130043  33110880  33111566  33111203  33120221  33120537  10033050644  33100549  33101068  33090269  33100436  33111454  33120707  10033080180  10033060817  33111293  33100391  10033060709  10033080135  33100635  33120671  10033060434  33102334  33111611  33101162  33111874  33101543  33111938  10033060477  33120758  33081418  1033141015  33110743  33111523  33081201  33110061  33090121  33130361  33111518  33120905  33110801  33111947  33100542  33110427  33102110  33110312  33120918  33081087  1033141067  33102355  33120366  10033050569  33101157  33111841  10033070785  33100350  33100331  33130070  33102261  33120580  33111574  10033070214  33101465  33101124  33111654  33120333  33111797  10033070545  10033070741  33100239  33111096  33120868  33101173  33090001  33111459  33110293  33110306  33130156  33111467  33090130  33120196  10033070759  33100526  10033060479  33111286  33110211  33101103  33111302  33111461  33110632  10033070265  33100433  33120517  33111063  33102073  33101500  10033080053  33120316  33110057  33100219  33120434  33081048  33090177  33101138  33111274  33111910  33130173  33101088  33111730  10033050431  33111160  33111869  10033070695  33090074  1033142009  33110730  33102362  33111106  33111320  10033080156  33111408  33101333  33090523  33120839  10033071005  33120544  33081373  33100142  33100123  33090194  33102158  33090428  33100429  33110081  33110957  33111026  33100397",
      "recall_initiation_date": "20141104",
      "postal_code": "92618-2346",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}