  "results": [
    {
      "status": "Terminated",
      "city": "Raritan",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "81767",
      "recalling_firm": "Ortho Clinical Diagnostics Inc",
      "address_1": "1001 US Highway 202",
      "address_2": "",
      "postal_code": "08869-1424",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distributed nationwide domestically. Foreign distribution to Australia, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, Singapore, UK, France, Germany, Italy, Spain, Portugal, Poland, Russia, Denmark, Norway, Sweden, Belgium, and the Netherlands.",
      "recall_number": "Z-0844-2019",
      "product_description": "Vitros Myoglobin Calibrators, Product Code 6801043 For use in the calibration of the VITROS ECi/ECiQ Immunodiagnostic Systems, the VITROS 3600 Immunodiagnostic System and the VITROS 5600 Integrated System for the quantitative measurement of myoglobin in human serum and plasma",
      "product_quantity": "4030",
      "reason_for_recall": "A stability issue in the calibrators may cause calibration failures or atypical Quality Control results following a successful calibration.",
      "recall_initiation_date": "20181129",
      "center_classification_date": "20190215",
      "termination_date": "20240206",
      "report_date": "20190227",
      "code_info": "Lots 1271 through 1310 (expired)  Lots 1315 & Above (in-date)",
      "more_code_info": ""
    }
  ]
}