  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Brea",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84268",
      "recalling_firm": "Beckman Coulter Inc.",
      "address_1": "250 S Kraemer Blvd",
      "address_2": "",
      "postal_code": "92821-6232",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) including Puerto Rico and countries of: Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Brunei, Bulgaria, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Ecuador, France, Germany, Ghana, Greece, Guatemala, Hong Kong, Hungary, India, Indonesia,  Israel, Italy, Japan, Jordan, Kazakhstan, Kuwait, Lebanon, Libya, Macau, Malaysia, Maldives, Mexico, Moldova, Montenegro, Morocco, Netherlands, Netherlands Antillean, New Zealand, Nigeria, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Switzerland, Taiwan, Thailand, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Vietnam.",
      "recall_number": "Z-0843-2020",
      "product_description": "iChem VELOCITY Urine Chemistry Analyzer",
      "product_quantity": "4314 units",
      "reason_for_recall": "Beckman Coulter has become aware of an issue affecting Specific Gravity chemistry settings for the iChemVELOCITY Urine Chemistry System that prevents the trigger of a chemistry system flag  S3  and may result in erroneous Specific Gravity (SG) results.",
      "recall_initiation_date": "20191107",
      "center_classification_date": "20200123",
      "termination_date": "20230911",
      "report_date": "20200129",
      "code_info": "All serial numbers and SW versions    V01257  V02416  V01444  V02619  V02959  V03933  V03934  V02022  V02027  V04322  CV04002  V03288  CV04113  V04236  V03472  V04250  V02367  V02287  V01134  V03796  V02453  V03186  V02317  V04256  V03114  V01360  V01655  V02065  V04528  CV04005  V02330  V02323  V02329  V02326  V04511  V02642  V03571  V03428  V02909  V01251  V01115  V02081  V02248  V03575  CV04036  V03403  V04388  V04387  V04389  V04328  V04390  V04141  CV04076  V01214  V04227  V02438  V02449  V02485  V02423  V02432  V02614  v01838  V02649  V02502  V02652  V02464  V02634  V02635  V02450  V02534  V02436  V02536  V02459  V04312  V02530  V02463  V02535  V02435  V04284  CV04083  CV04047  CV04084  V03409  V03413  V01689  V01753  V01754  V03854  V03722  V04753  V02243  V02242  V03851  V03925  V04369  V02026  CV04165  V03848  V04540  V04516  V04515  V01404  V02430  V03565  V04706  V01335  V02234  V02373  V02892  V01829  V02446  V01188  V02782  V03456  V01429  V02021  V02020  V02966  V02949  V04767  V04674  V04770  CV04003  V04769  V04740  CV04004  V02899  V01977  V04538  V04309  V04607  V01318  V01546  V04257  V03566  V01548  V01547  V02753  V01541  V01542  V03922  V02369  V01555  V02070  V02181  V01720  CV04086  V03834  V04291  V04549  V02913  V04590  V03798  V01911  V01998  V03567  V04258  CV04125  V03238  V04329  V03327  V04592  V03271  V01776  V01712  V04702  CV04163  V04519  V02086  V01824  V01467  V03100  V04669  V04464  V04471  V04584  CV04229  CV04228  CV04196  CV04227  CV04265  V04654  V01872  V04138  CV04132  V04732  V02210  V03003  V03283  V03285  V04432  CV04144  V01182  V01179  V03033  V02361  V04433  V01842  V04377  V01717  V03776  V01831  V03890  V01567  V01568  V03237  V03000  V02998  CV04067  V04775  V03464  V02046  V01581  V01493  V03842  CV04209  V04379  V01460  V03113  CV04043  V01146  V01781  V04545  V04547  V04606  V04302  V04434  V04689  V04195  V04559  V03251  V04153  V02650  V02121  CV04085  CV04082  V02752  V02645  CV04121  V02641  V02640  CV04116  V03152  V01264  V03462  CV04109  V04675  V02084  V01109  V01950  V01944  V01443  V01207  V01209  V02132  V03465  V02439  V03417  V03797  CV04060  V02185  V01432  V02150  V02078  V02194  V02127  V02217  V02810  V04655  V03748  CV04140  V01533  V02957  V01864  V04318  V04385  V01632  V01633  V03011  V03008  V03047  V04467  V01281  V02492  V01135  V01161  V02602  V02934  V04514  V01482  CV04234  V02186  V02884  V02411  V01200  V03245  V01627  V01877  V03090  V01092  V01094  V01332  V02962  V04594  V04341  V03953  V03977  V01876  V01147  V01871  V03467  CV04102  CV04107  CV04115  V03700  V03357  V03775  V04335  V02180  V01376  V01981  V01978  V01985  V04349  V03438  V04498  CV04103  V01486  V01616  V01502  V01642  V03427  V03889  V03106  V03105  CV04183  V03122  V03107  V02269  CV04208  V01713  V04641  V03774  V03935  V04550  V03384  CV04124  V03720  V03716  V02808  V04420  V01620  V03858  V04556  V02295  V02306  V02468  V01059  V04372  V03441  V03376  V02874  V03145  V04348  V01375  V04436  V04117  V03853  V04242  V01919  V01236  V03187  V03825  V01389  V02398  V02129  V03382  V03055  V02195  V02201  V02093  V02097  V02085  V04587  V01237  V04443  V01505  V02018  V02706  V04457  V01400  V01401  V01834  V03388  V02656  V02674  V02414  V04535  CV04251  CV04188  V04311  CV04191  V04555  V04597  V02861  CV04054  V01641  V03402  V02419  V02558  V02239  V01612  V03474  V04566  V01442  V01388  V04475  CV04044  CV04040  V01294  V01298  V02069  V01159  V01393  V04279  CV04026  V01730  V01046  V01274  V04553  V03470  V01326  V02717  V04683  V04625  V04714  V04716  V03957  V03958  V04493  V01117  V01118  V01752  V02868  V02964  V02587  V02965  V02517  V04313  V04658  CV04091  V03439  V01534  V01664  V02505  V01250  V04615  V04375  V03576  V01770  V04413  V01431  V01072  V01862  V02881  V04391  V02796  V02327  V02958  V02852  V03819  V03827  V04278  V04757  V03818  V03826  CV04032  V03423  V04248  V04906  V01314  V01076  V02144  V02992  V01244  V01889  V01893  V01361  CV04053  V03434  V04705  V03419  V01459  CV04117  V03133  V01437  V04704  V03463  V03481  V04193  V03163  V02875  V04345  V02193  V04635  V01755  V04537  V03046  V03978  V03086  V02983  CV04221  V03088  V03085  V03087  V02988  V02995  V03089  CV04106  V01599  V02756  V02123  V01625  V02568  V03444  V01929  V03859  V04353  V02067  V04903  V03319  V04548  V04441  V01610  V04401  V03864  V01329  CV04009  V02697  V02727  V02509  V03849  V01132  V02036  V02033  V02054  V03560  V01540  V02849  V04762  CV04062  V02646  V02665  V01551  V02460  V01645  CV04061  CV04070  V03698  V01989  V02529  V03356  V01313  V04634  V02956  V02952  V02396  V03049  V03860  V03794  V03793  CV04101  V04137  V02110  CV04195  CV04194  V01788  V01170  V03303  V03254  V03275  V03274  V03331  V03316  V01765  V04463  CV04092  V03201  V04651  V03043  V02669  CV04192  V02062  V02038  V01895  V02041  V01830  V01213  V03813  V01570  CV04217  V03082  V03440  V02190  V04909  CV04049  V01874  V01210  V01462  V03918  V03786  V03422  V04600  V03057  V02689  V02337  V03415  V04619  V03370  V03391  V03281  V03250  V03915  V01077  V02679  V01594  CV04052  CV04051  V04644  V01901  V04297  CV04154  V03101  V04505  V03379  V03790  V02087  V02501  V04596  V04308  CV04235  V02664  V04611  V02491  V02448  V01138  V03040  V03386  V02451  V01173  V02951  V04340  V03167  V01461  CV04139  V03079  V03702  CV04130  V04319  V04238  V04735  V04381  V02147  V02457  V02452  V01183  V01421  V02955  V01934  V01936  V01941  V01945  V02551  V02933  V01958  V04724  V04710  CV04020  V04508  V04571  V01980  V04241  V04339  V01571  V01572  CV04114  CV04104  V03815  V03803  V03824  V03888  V01160  V01769  V02143  V03425  V02854  V02299  V02298  V02172  V02171  V02251  V02257  V02284  V02289  V02259  V02263  V02252  V02256  V04496  V04497  V02233  V02383  V02384  V01942  V01938  V02381  V02380  V02254  V02249  V02023  V02009  V02409  V02019  V04771  V04310  V04768  V01184  V04143  V04142  CV04170  CV04179  V04247  V04246  V04694  V04633  V01996  V02051  V04636  V04480  V04479  V02285  V01852  V03780  V03781  V04240  V04239  V04521  V04522  V04263  V04264  V04469  V04468  V02250  CV04112  CV04111  V04575  V04582  V04614  V04612  V04270  V04272  V02286  V02140  V02139  V02170  V02091  V02245  V02231  V02138  V02099  V04599  V04292  V04253  V02146  V02188  V03812  V03804  V02003  V02258  V01853  V02288  V02322  V04224  V04223  V02729  V04266  V03872  CV04245  V04576  V04574  V03749  V03779  CV04233  V02961  V04534  V02313  V03955  V04759  V04360  V04362  V03111  V02533  V01686  V01304  CV04149  V01305  CV04150  V04552  V01545  CV04038  V02789  V01430  V03874  V03758  V04666  V03473  V01169  CV04088  V04512  CV04087  CV04089  V02948  CV04048  V04093  V03866  V02167  CV04093  V02014  V03919  V04288  CV04108  V01857  V01856  V02082  V01233  V04527  V02984  V01198  V04678  V02166  V02412  V02158  V02169  V01719  V04286  V02117  CV04007  V01714  V04902  V04622  V02954  V01666  V01199  V04734  V03102  V01850  V02910  V04742  V01903  V01563  V01957  CVO4030  CV04042  V03387  CV04028  V03817  V01564  V03380  V03828  CV04018  V02004  V02017  V01395  V01061  CV04197  V03433  V02685  V02893  V02699  V01293  CV04006  CV04001  V02580  V04267  V02267  V04453  V03451  V03920  V03401  V01386  V02319  V04688  V01423  V01586  V01951  V01544  V02620  V01935  V01255  V03932  V01413  V02102  V02131  CV04159  V01967  V04509  V02470  V01196  V03189  CV04135  V03258  V04504  V03373  CV04176  V03156  V03723  V03833  V04262  V04407  V01994  V02010  V01367  V01858  V04336  V01205  V04624  V04294  V04295  V01405  V03830  V02310  CV04024  V01055  V01639  V02630  V02497  V01075  V04255  CV04247  CV04063  V04673  V04542  V01107  V01409  V01463  V01789  CV04097  V01279  V01315  V04376  V01192  V01849  V01873  V04593  CV04219  V01427  V01252  CV04064  V04307  CV04012  V04298  V02356  V03279  v02901  V01316  V03928  V03929  V01759  V02226  V01325  V01081  V03364  V01387  V02151  V04446  V03891  V04448  V04447  V04354  V01122  V04252  V03421  CV04065  V04743  V03400  V03855  V01189  V02985  V02931  V02859  V02886  V03058  V02929  V04736  V01176  V03130  V02608  V02539  V03759  V02723  V02725  V02275  V01615  V01333  V01726  V01909  V04703  V04358  V03367  V01439  V02804  V03699  V03975  V03841  V02503  V04727  V03777  CV04010  V01940  V04520  V03351  V01504  V02421  V01501  V01503  V02354  V03282  V03273  V03476  V02094  V04280  V02066  V03466  V01477  V02472  V04323  V01954  V01870  CV04027  V01841  V01410  V04304  CV04232  V01525  V01334  V01778  V02903  V02904  V02308  V01786  V01785  V04523  V02247  V04096  CV04204  V02107  V03383  V04747  V01216  V02454  V04546  V03411  V01565  V02098  V04359  V02927  V01270  V03873  V04908  V03381  CV04050  V04409  V04657  V04418  CV04014  V04745  V04752  CV04057  V01905  V04533  V03450  V03458  V01402  V02192  V04419  CV04236  V04460  V03244  V04331  V01249  V04261  V04426  V04410  CV04123  V03108  V01623  V01509  V01243  V02960  V02136  V03767  V03153  V01223  CV04158  V03369  V03358  V02715  V01306  V01307  V03365  V03371  V04503  V02538  V02537  V02149  V04454  CV04222  V04111  V03954  V02035  V04667  V02783  V02494  V04605  V04602  V04101  V02519  V04541  V04393  V01296  V03852  V03847  V02527  V03110  V04455  V04474  V04513  V04565  V04525  V02896  V03176  V02581  V01716  V01101  V03078  V04772  V04343  V01767  V03483  V03329  V04265  V03432  V03437  V04334  V01640  V04570  V04324  V04744  V03856  V01913  V03454  V04618  V04586  V04585  V04901  V04562  V03442  V04543  V02703  V04639  CV04119  CV04118  CV04011  CV04202  V03249  V01636  V03725  V03901  V04273  V02888  V04529  V04623  V03009  V04276  CV04138  V02202  V04254  V04099  V04098  V04656  CV04021  V03561  V02836  V02944  V04483  CV04071  V04365  V04364  V04462  V04616  V02154  V01762  V03372  V02963  V01590  V01239  V01411  V02191  V04659  V01149  V03143  V03276  V04495  V04518  V02894  V01038  V01129  V01465  V04628  V02063  V02100  V02096  V02428  V03416  V03460  V02126  V01576  V04717  V01037  V02490  V01142  CV04058  CV04073  V02071  V04140  CV04033  V04500  V04900  V04321  V04439  V02680  V02235  V04368  V02716  V03921  V01440  V03917  V01403  V02673  V02404  V02420  V01190  V01860  V01308  V02973  V04114  V04115  CV04180  V03926  V04608  V02801  V02607  V02050  CV04178  V01904  V02092  V02134  V01569  V01425  CV04098  V01136  V01722  V04665  V03426  V04112  V01772  V02426  V02161  V02176  V03260  V03142  V02596  V02375  V02731  V03385  V01227  V02830  V02028  V02628  V02124  V02638  V02996  V02219  V02552  V04627  V03836  V01619  V03475  V01180  V01869  V01163  v03791  V02858  V04610  V04684  V02336  V01475  V03001  V01271  V02877  V01736  V01372  V02883  V04116  V02880  V04415  V04305  V02867  CV04151  V01235  V02179  V02427  V02183  V04285  V02182  V04230  V04231  V04232  V04234  V04228  V04283  V04293  V04357  V01956  V02244  V03412  V02583  V04754  V04118  V03876  V02946  CV04226  V02709  V01507  V01297  V01476  V03746  CV04077  V04681  V02812  V03857  V03290  V02072  CV04152  V01601  V04494  V04672  V01832  V04748  V01771  V03304  V01984  V01208  V01201  V03765  V03377  V04094  V03206  V01768  V03161  V04435  V02212  V03200  CV04015  CV04081  V03850  V03816  V03846  V03199  V03198  V03875  V01892  V01744  V04303  V03076  V01099  V01095  V02891  V03139  V01757  V04466  V04249  V02895  CV04017  V04452  V04564  V04551  V04445  V04563  V04544  V04487  V04580  V04572  V04492  V04502  V04539  V04526  V04491  V04488  V01261  V04613  V04604  V04444  V04438  V04450  V02189  V01855  V04361  V04347  V03019  CV04013  V04314  V01584  CV04243  V03745  V04399  V02080  V01818  V01843  V02758  V04486  V03005  V01529  V01418  V04229  V04620  V01847  V04687  V02216  CV04078  V03004  V02197  V02923  V02749  V01186  CV04055  CV04056  V03072  V03399  V01217  V04629  CV04122  V01397  V02467  V03715  V03436  V02208  V02056  V04301  V03792  V03789  V03773  V02402  V01979  V01782  V02012  V01330  CV04240  V04290  V02878  V04728  V04773  V04774  V04289  V03443  V03443  V03952  V04682  V02350  V04524  V02885  V03568  V04237  V04219  V04630  V01999  V04591  V04595  V03788  CV04022  V02601  V04578  V04554  V02349  V04476  V02876  V01970  V04461  V01878  V01550  V01178  V04429  V02073  V03461  V04532  V04701  V04408  V04398  V04449  V01562  V01103  V02168  V02897  V03353  V03270  V01707  V04473  V04383  V04384  V04646  V02076  V01299  CV04000  V01614  V02422  V02767  V04113  V02871  V04135  V04905  V04458  V03719  V03923  V03916  V02024  V03404  V02921  V02924  V02922  CV04160  V04766  V04764  CV04164  V04713  V01535  CV04268  CV04224  V02187  V02431  V04760  V03930  V01727  V02211  V03900  V04506  V02553  V02547  CV04045  CV04095  V03084  V03482  V03471  V03144  V04274  V04320  V01141  V03410  V04588  V04421  V04386  V01394  CV04126  V03562  V03563  V03418  V03564  V03453  V02826  V01303  V02809  V04560  V03264  V01780  V03902  V03724  V02559  V04299  V04423  V01215  V04738  V04707  V02390  V04100  V02722  V02104  V04222  V01224  V01229  V02643  V03477  V02887  V01428  V03870  V03116  V03420  V01588  V02207  V04134  V02095  V02526  V02074  V01613  V03269  CV04166  V04459  V02882  V04275  V04277  V01531  CV04257  CV04187  V04712  V03956  V03865  CV04133  V01952  V04645  V03924  V04693  V01512  V02521  V02626  CV04212  V04720  V04722  V04726  V04739  V04733  V01888  V03747  V04746  V01436  V04751  V01914  V01319  V03128  V02866  V01317  V01805  v01837  V03863  V01763  V03801  V02970  V03480  V03136  V02418  V02413  V02403  V02406  CV04264  V02765  V02761  V01803  V04609  V03032  V03026  V02543  V04306  V04315  V01203  V01204  V01825  CV04161  CV04254  V02371  V02368  CV04255  V01426  V02342  V03366  V02730  CV04128  CV04203  V01798  V04698  V01698  V01699  CV04246  V01378  V01379  V01543  V04755  V01108  V03744  V02675  V03732  V01634  V04095  CV04041  CV04008  V04660  CV04074  V04664  V01058  V01048  V01073  V01043  V03829  V04756  V01669  V04577  V04583  CV04072  CV04069  V04510  V04482  V04481  V01097  V04567  V01683  V04758  V04680  V04679  V04531  V02644  V04626  V02556  V03123  V02586  V04696  V01750  V04695  V01718  V01670  V03886  CV04169  CV04193  CV04206  CV04172  CV04037  V02407  V01986  V01983  V01965  V01962  CV04039  V01598  CV04262  CV04261  V02372  CV04034  CVO4025  V02382  V01057  V01964  V01228  V02890  V01963  CV04260  CV04270  V01193  V01194  CV04258  CV04186  CV04174  V04715  V04725  V04719  V02748  V04316  V04652  V01358  CV04244  V02757  V04729  V04370  V04355  V02915  V04400  V01611  V01191  CV04205  V01125  V04404  V04412  V01056  V01084  V01008  V01083  V01734  V04671  V01766  V01779  V04730  V01743  V02676  V04661  V04691  V04692  V01738  V01733  V04718  V04721  V04663  V02029  V03766  V02773  V02914  V01074  V01054  V02991  V01123  V04351  CV04213  V04416  V04414  CV04190  V04330  V04417  V04425  V02942  V04676  V02705  V02743  V02763  V02764  V04649  V04650  V01674  V04763  V01764  V01470  V04907  V01466  V01804  V01731  V01884  V01481  CV04171  CV04185  CV04141  CV04137  CV04157  CV04155  CV04207  V01513  CV04199  CV04216  CV04198  CV04134  CV04142  CV04143  CV04136  CV04237  CV04239  CV04148  V01524  CV04129  V03892  V01449  V01575  V04268  V04269  V01064  V01042  V01044  V03081  V01071  V03785  V02037  V01248  V02793  V02762  V03867  V03355  V03219  V03308  V03868  V03309  V03837  V01668  V04300  V04690  V01760  V03138  V03141  V02566  V02582  V02805  V02813  CV04263  V04617  V04589  V04568  V04741  V04761  V04677  V04731  V04411  V04378  V04402  V04424  CV04210  V02052  V01100  V01167  V04139  V02294  V04109  V04110  V03869  V02318  V01080  V01086  V02869  CV04211  V01070  V02292  V03974  V03750  V02042  V04136  V03302  V02040  V01729  V04711  CV04252  CV04253  CV04189  CV04201  CV04181  CV04182  CV04096  V01185  CV04080  V01277  CV04220  CV04223  V03408  V04427  V04442  V03051  V02057  V01320  V03802  CV04153  V04102  V02508  V01399  V01416  V03253  V02755  V04366  V03484  V04648  V04642  V04643  V03406  V03406  V03799  V04637  V02125  V01254  V04403  V02862  V03927  V04484  CV04075  V02240  V04737  V04557  V02821  V04431  V04499  V04581  V04697  V03884  V04507  V04653  V04561  V04638  V04579  CV04031  V02227  V04456  V02982  V02055  V02061  V04647  V04338  V04337  V02465  V04662  V04708  V02651  V02647  V02846  V02845  V01197  V03574  V02916  V02733  V01165  V03435  V03701  V01561  V04603  V01310  V01311  CV04066  CV04068  V03731  CV04079  V03730  V02109  V01844  V01875  V01240  V01242  V04296  V03903  V03424  V01174  V03469  CV04099  V02918  V02932  V03931  CV04256  CV04238  V02374  CV04230  CV04231  V01728  CV04127  V03132  V04686  V01284  V03914  V03787  V02106  V02274  V02476  V02272  V04367  V02512  V04685  V02953  V03543  CV04090  V03479  V01756  V01746  V01747  V01356  V03478  V02077  V03893  CV04177  V04451  V04428  V03080  V04325  V03714  V01438  V03717  V03718  V03721  V04558  V01638  V01637  V04346  V01003  V03104  V03109  CV04110  V04573  V04621  V04470  V04601  V04776  V03278  V02068  V02079  V02567  V04350  V02986  V02334  V04472  V02855  V04344  CV04046  V01222  CV04147  CV04167  V03784  V01087  CV04146  CV04145  V04260  V02711  V01643  V03795  V04371  V03048  V01538  V03894  V03272  V01732  CV04100  V02999  V03885  CVO4035  V01282  V04530  V01458  V02089  V02064  V03800  V04342  V01096  V02721  V03448  V03445  V03449  V03446  V03150  V01036  V02560  V02555  V02462  CV04016  V02395  V02393  V01412  V03570  V03012  V04536  V02307  V03468  V04374  V04422  CV04059  V04275  V04277  V01531  CV04257  CV04187  V04712  V03956  V03865  CV04133  V01952  V04645  V03924  V04693  V01512  V02521  V02626  CV04212  V04720  V04722  V04726  V04739  V04733  V01888  V03747  V04746  V01436  V04751  V01914  V01319  V03128  V02866  V01317  V01805  v01837  V03863  V01763  V03801  V02970  V03480  V03136  V02418  V02413  V02403  V02406  CV04264  V02765  V02761  V01803  V04609  V03032  V03026  V02543  V04306  V04315  V01203  V01204  V01825  CV04161  CV04254  V02371  V02368  CV04255  V01426  V02342  V03366  V02730  CV04128  CV04203  V01798  V04698  V01698  V01699  CV04246  V01378  V01379  V01543  V04755  V01108  V03744  V02675  V03732  V01634  V04095  CV04041  CV04008  V04660  CV04074  V04664  V01058  V01048  V01073  V01043  V03829  V04756  V01669  V04577  V04583  CV04072  CV04069  V04510  V04482  V04481  V01097  V04567  V01683  V04758  V04680  V04679  V04531  V02644  V04626  V02556  V03123  V02586  V04696  V01750  V04695  V01718  V01670  V03886  CV04169  CV04193  CV04206  CV04172  CV04037  V02407  V01986  V01983  V01965  V01962  CV04039  V01598  CV04262  CV04261  V02372  CV04034  CVO4025  V02382  V01057  V01964  V01228  V02890  V01963  CV04260  CV04270  V01193  V01194  CV04258  CV04186  CV04174  V04715  V04725  V04719  V02748  V04316  V04652  V01358  CV04244  V02757  V04729  V04370  V04355  V02915  V04400  V01611  V01191  CV04205  V01125  V04404  V04412  V01056  V01084  V01008  V01083  V01734  V04671  V01766  V01779  V04730  V01743  V02676  V04661  V04691  V04692  V01738  V01733  V04718  V04721  V04663  V02029  V03766  V02773  V02914  V01074  V01054  V02991  V01123  V04351  CV04213  V04416  V04414  CV04190  V04330  V04417  V04425  V02942  V04676  V02705  V02743  V02763  V02764  V04649  V04650  V01674  V04763  V01764  V01470  V04907  V01466  V01804  V01731  V01884  V01481  CV04171  CV04185  CV04141  CV04137  CV04157  CV04155  CV04207  V01513  CV04199  CV04216  CV04198  CV04134  CV04142  CV04143  CV04136  CV04237  CV04239  CV04148  V01524  CV04129  V03892  V01449  V01575  V04268  V04269  V01064  V01042  V01044  V03081  V01071  V03785  V02037  V01248  V02793  V02762  V03867  V03355  V03219  V03308  V03868  V03309  V03837  V01668  V04300  V04690  V01760  V03138  V03141  V02566  V02582  V02805  V02813  CV04263  V04617  V04589  V04568  V04741  V04761  V04677  V04731  V04411  V04378  V04402  V04424  CV04210  V02052  V01100  V01167  V04139  V02294  V04109  V04110  V03869  V02318  V01080  V01086  V02869  CV04211  V01070  V02292  V03974  V03750  V02042  V04136  V03302  V02040  V01729  V04711  CV04252  CV04253  CV04189  CV04201  CV04181  CV04182  CV04096  V01185  CV04080  V01277  CV04220  CV04223  V03408  V04427  V04442  V03051  V02057  V01320  V02953  V03543  CV04266  V05118  V02609  V05079  V02408  V03966  V05568  V02262  V0291  V05224  VEL10010579  V03336  CV05017  V01997  V03494  V05309  V03522  CV05411  V03601  V03605  CV05288  V05580  V05573  CV05292  V05579  V05577  V05031  V05431  CV05289  CV05287  V05429  V05474  V05557  V05574  V05575  V05576  V05029  V05389  V05562  V05032  V03599  V03603  V05475  V05423  V05430  V05470  V05428  V03608  CV05284  CV05286  V03507  CV05128  CV05302  CV05068  V05350  V01807  V01926  V01806  V01991  V01928  V02499  V03220  V03223  V03224  V04196  V02788  V03544  V02532  V05161  V05158  V05177  V01266  V01896  V05132  V05382  V05381  V05365  V05363  V05395  V03910  V05211  V03944  V03209  V03146  V03168  V02941  10060539  V01537  CV05248  10050524  V01289  CV05234  CV05232  V02684  CV05247  10050520  10040512  V01263  V05493  V05494  V03242  V02271  V02268  V04162  V02277  V02279  V02270  V02280  V04163  V05162  V05264  CV05356  V02276  V02681  V02624  V03390  V0472  V03292  V03096  V03099  V03053  V03069  V03062  V03061  V03094  V03064  V03067  V03024  V03050  V05434  V05542  V05541  V05543  V0541  V0301  V0383  CV05361`  CV05362  V03119  V03118  V03092  V03994  V03980  V03030  V02415  CV05160  CV05159  CV05161  CV05120  CV05158  CV05316  CV05317  CV05308  CV05298  CV05296  V03821  V03823  V03822  V03760  V05427  V01033  V02043  V05322  V03552  V05290  V0465  V05060  10030497  V03240  V03103  V01301  V01300  V01396  V02260  V01748  V01219  V02228  V01045  V01010  V0346  V0413  V0332  V0369  V0452  V03644  V03735  CV05347  CV05006  CV05250  V02237  V02088  V05442  V05445  V05444  V05446  CV05211  V05058  V01454  V01455  V01456  V01457  V05280  V0221  V02873  V02829  V02863  V03124  V01063  V02900  V0287  V04037  V02785  V03243  V02870  V01384  V03696  V03558  V05255  V05234  V05117  V03491  V03547  V02702  CV05040  V0337  CV05001  V02746  V03052  V0382  10030503  CV05196  V02572  V02926  V02917  V02701  V02174  V02174  V0327  V0460  CV05283  V0388  V05233  V03736  V05092  V05014  CV05114  CV05123  CV05113  V05112  V05369  V05113  V03289  V02489  V03912  V02816  V05278  V02939  V02304  V0317  V02629  V02612  V05157  V02990  CV05039  V05013  V03742  V01605  V01342  V013426160  6160  V04168  V01419  V05016  V03838  V01291  V05185  V04014  V01897  V02657  9020348  V02975  V0269  V0376  V0410  V0531  V0252  V02879  V03301  V02744  V05108  V03354  V03342  V01035  V0270  V02112  V02515  V03347  V03343  V05107  V05326  V03346  V03345  V01715  V02507  V02514  CV05052  V05496  CV05056  CV05071  CV05047  CV05073  CV05074  CV05070  CV05055  CV05063  CV05054  CV05057  V05345  CV05051  V01153  V02113  V05106  V03348  V01420  V01453  V02950  V05471  CV05260  V05287  CV05041  CV05395  V01787  V01626  V03878  CV05023  V05017  V03704  V05300  V05301  CV05204  CV05024  V03861  V03862  V05041  CV05297  CV05295  V01723  V01812  V01815  V0357  CV05028  CV05029  V02720  V03772  V03964  V02827  V03013  V05100  V05101  CV05115  V01988  V03589  V01946  V01943  V01939  V01721  V02834  CV05030  V05006  V05612  V0467  V03609  V02198  V02199  V05038  V05269  V05191  V05273  V05271  V05244  V02853  V02433  V02889  V02844  V01923  V0389  V05289  V05402  V05487  VEL09010322  V05286  V05401  V05317  V05323  V02768  CV05410  V01391  V02141  V02561  V03497  V05588  V02159  V02058  V02200  V0285  CV05152  CV05153  V02778  V03690  V0450  V02905  V0344  V03205  CV05182  V01417  V01488  V02002  V03091  CV05277  V01128  V03557  V02726  V03604  V05056  V02524  V03965  V02479  V02461  V05443  V02780  V02777  V05219  V02455  V0535  V03431  V03042  V03757  V03882  V0542  V0538  V02347  VEL09100440  V0381  V02976  V0260  V02989  V02495  V02496  V05359  V05361  CV05018  V04209  V03054  V03610  V03537  V03538  V06690  V01690  V05135  V05141  CV05265  CV05014  CV05407  CV05009  V03280  V03528  V03526  V03525  V01578  V01577  CV05321  CV05386  CV05365  V05466  CV05385  CV05320  V05008  V05476  V03619  V03582  V03546  V03625  V03628  V05467  V03627  V03624  V03617  V03621  V03616  V05465  V03623  V03620  V03622  V03629  V03545  V02936  V02597  V03162  V02599  V02045  V03095  V03363  V01839  V01851  V02481  V01796  V01758  V02938  V02617  V05090  V05367  V04041  V04132  V04103  V04104  V04048  V04122  V04081  V04088  V04079  V04148  V04206  V04124  V04128  V0273  V05222  V03165  V02622  V01027  V01028  V01029  V0286  V01040  V01088  V01260  V01268  V05320  V02707  V03997  V03227  V02802  V05279  V05268  V03905  V03906  V05318  V02394  V01740  V0241  V05265  V05267  V05569  V05582  V01882  V0361  VEL09080419  CV05393  V02296  V03606  CV05285  CV05282  V05030  V03607  V03969  V03710  V03695  V03995  V03711  V03556  V02719  V02352  V02222  CV05082  V02039  V03075  V03540  CV05083  V02355  V01113  V02565  V02563  V02297  V0468  V0288  V02978  V03756  V02366  CV05091  V03498  V02987  V0220  V05151  V03529  V03549  V02273  V05011  V04023  CV05098  V0513  V02937  V03171  V03027  V05455  CV05021  V05339  CV05002  CV05007  CV05104  CV05195  CV05183  CV05148  V05095  V02364  CV05346  V01971  V02779  V02974  V03252  V02542  CV05433  V03246  CV05367  V0550  V01408  V03148  V02979  V05572  V05514  V05225  V0249  V05003  CV05426  V05270  CV05166  CV05299  CV05377  V01648  V03712  V02632  V02610  V02698  V03277  V02774  V02786  V02787  V03340  V02800  V02781  V03341  V02772  V02814  V02775  V03344  V0397  V05150  V02339  V0222  CV05037  CV05045  V02564  V05066  V02473  V02471  V03970  V05426  V03971  CV05388  V05067  CV05394  V05425  CV05251  V0298  V0421  V02266  V02266  V02128  V05046  V0571  V0420  V02137  V05400  V05399  V03492  V03503  CV05072  V03493  V05069  V03504  V05513  CV05322  V02341  V01795  V01790  V02548  V02545  V03261  V02516  V02518  V02605  V02600  V02603  V01794  V01791  V01792  V02510  V02677  V03262  V02520  V02678  V01879  CV05392  V01922  V05602  CV05035  CV05376  V01880  V01868  V01959  CV05143  V03808  V02754  V03121  CV05427  V0387  V02850  V0296  V0495  V05057  V03536  V02357  V03266  V03263  CV05290  CV05293  V05036  V05021  V05034  V02771  V02569  V03154  V03083  V05241  V03125  V03203  V03913  V02236  V03879  V02478  V01927  V02392  V02625  CV05119  V05160  V05435  V05432  CV05258  V02738  V02469  V03539  CV05194  CV05327  CV05097  V04024  V05407  V05362  CV05263  CV05264  V02405  V03007  V02469  V01152  V0243  V01032  V05040  V03642  V05055  V0504  V05159  V01172  V01490  V02032  V02358  V02441  V01366  V05246  CV05389  CV05110  V03183  V03184  V02291  V01187  CV05425  V0536  V0345  V03337  CV05259  V03510  V01452  V01451  V01514  V03600  V05393  V02759  V02511  V02856  V02594  V05360  V03648  CV05236  V01947  V03542  V03809  V01955  V03334  V05068  V01657  V03551  V05291  V05033  V05436  V02370  V03523  CV05238  CV05216  CV05101  CV05102  CV05100  V05595  V05598  CV05046  V05599  CV05224  V05236  V05387  CV05241  V05229  V05228  V05561  V05560  V05243  CV05237  CV05225  CV05217  CV05227  CV05226  CV05099  V05356  V05227  CV05228  V05178  V01473  V01474  V04000  V02525  V01840  V01276  V01489  V01527  V01848  V0453  V01833  V01273  V01005  V01846  V01487  V0297  V0544  V01680  V0295  V01269  V01275  V01681  V02670  V01867  V02666  V02661  V02672  V01828  V02667  V02671  V01835  V01823  V01881  V01836  V02659  V02658  V0360  V0356  V0507  V0435  V01676  V01677  V01678  v01814  V01062  V0276  V0393  V0437  V0548  V01528  V02218  V0425  V03999  V0448  CV05078  V05119  V01030  V0265  CV05048  V01346  CV05043  V02290  V01891  V02264  CV05218  V05404  V04002  V02223  V02214  CV05369  V02265  V02283  V01124  V05479  V04392  V02794  V03414  CV05262  V05245  V03317  V05604  V05603  CV05065  CV05062  V03495  V05044  V05062  CV05060  V03904  V04034  CV05069  V03496  V03626  V05283  CV05414  V04029  V03612  V03899  V05284  V05285  V03655  V03651  V04028  V03896  V03897  V04033  V05180  V03898  V03631  V03630  CV05413  V04030  V05600  V05115  V03500  V04031  V04032  V04036  CV05066  V03654  V05179  V05114  V03374  V03375  V03378  CV05008  V03015  V02550  V02692  V02691  V05002  V02688  V02683  V02693  V02728  V02736  V0352  V03590  V01972  V03368  V04213  V05383  V02740  V02739  V03733  V02075  CV05257  CV05254  CV05188  CV05187  CV05336  CV05337  V01774  V03025  V01357  V01355  V02967  V02969  V05530  V03362  VEL08110290  V0258  V02708  V02947  V05567  V02119  V02232  V01992  V02621  V02458  V01811  V01799  V01654  V01810  V02442  V02493  V02864  V02440  V01910  V01907  V02444  V01773  V01631  V02445  V02011  V02338  CV05301  CV05397  CV05086  CV05363  V05485  CV05084  CV05364  V05563  V05564  V05565  V05165  CV05240  CV05242  CV05087  CV05291  V02700  V02841  V02847  V02848  V03294  V03293  V03307  V03295  V05072  V05071  V05070  CV05000  V0306  V01390  V01221  V05023  V05022  V03115  V01312  V01339  V01415  V01380  V01709  V0391  V05061  V05126  V05144  V05143  V05142  V03014  V02820  V0454  V0492  V05305  CV05390  CV05360  CV05359  CV05149  CV05106  CV05150  CV05384  V03763  V03769  V05307  V03771  V03770  V02770  V02776  V02549  V02557  V01797  V03041  V05181  CV05095  V05154  V05183  V03248  CV05049  V05015  CV05271  V05380  V05379  V05373  V05372  V0350  VEL10325  V0217  CV05350  V03555  CV05038  V03553  V02362  V02425  V02718  CV05005  V01751  V03210  CV05231  V05282  V05091  CV05116  CV05117  V02704  V01604  V01603  V03693  V03843  V03840  V02311  CV05398  CV05022  CV05015  V05587  CV05230  CV05313  CV05061  CV05059  V02365  V05051    V03805  V04200  V01327  V01328  V03112  V0392  V0372  V0390  V01110  V0426  CV05025  V03226  V03056  V01121  V01280  V01933  V03596  V03593  V03594  CV05221  V03204  CV05431  V03649  V03753  V02857  CV05253  CV05191  CV05192  CV05189  CV05328  CV05329  V01345  V01351  V01349  V01596  V01343  CV05193  CV05256  CV05255  V03020  V01353  V01347  V01348  CV05338  V03349  V01350  V05605  V05010  V05275  V05155  V01022  V05042  CV05339  CV05280  CV05279  V01177  V01091  V05192  V05292  CV05064  CV05067  V05164  CV05075  V05366  V05343  V02769  V02696  V02695  V02351  V0336  V02340  V03259  V03972  V03158  V03073  V03018  V01974  V01953  V03178  CV05050  CV05112  CV05111  V03968  CV05103  V02590  V02544  V03216  V03880  V0234  V05515  V05520  V05517  V05377  V05376  V03029  V03360  CV05345  CV05294  CV05382  V01930  V01066  CV05176  V02907  V02815  CV05004  V02902  V0377  V02616  V03306  V01621  V05328  V03578  V02593  V0396  V0237  V03550  V0555  V0230  V0305  V02400  V05020  V03643  V02577  V01041  V01065  V05063  V02588  V01025  V01024  V04021  V03592  V02345  V05298  V05299  V05321  V03831  CV05353  V02410  V0525  CV05077  CV05076  CV05081  V05327  V01890  CV05428  V02540  V02528  V02522  V01031  V03170  V03181  V02360  V03023  V00277  2760  V00449  V00408  V01660  V03211  V02799  V05064  CV05381  V02803  V05147  V02397  V05054  V0335  V0459  CV05179  V05498  CV05357  V0302  V03297  V02654  V02376  CV05243  CV05244  CV05079  V03713  V03694  V03583  CV05019  CV05020  V03591  V04035  V02333  V05167  V02344  CV05032  CV05031  CV05036  V05519  CV05044  V05116  CV05042  V05325  V05324  V05408  V02653  CV05368  CV05412  V02790  V03506  V02523  V02797  V05216  V05218  V03035  V03034  V03509  V03505  V03658  V05121  V04001  V03296  V02466  V05526  V02742  CV05311  V03291  V05346  V03520  CV05223  CV05400  CV05358  V05368  CV05351  V05403  CV05213  CV05212  V04022  V05556  V03131  V05545  CV05343  CV05342  CV05366  V03305  V05281  V05375  V02977  V04167  V05131  V03656  V05184  V02791  V05344  V03993  V03516  V03511  V05217  V05497  V03895  CV05222  CV05210  V05310  CV05373  CV05371  V05073  V03684  V03685  CV05396  V03683  V05043  CV05372  V02806  V02807  V05374  V02811  V02483  V05499  CV05215  V05555  V05558  V05125  V01937  CV05219  CV05220  V05529  V03151  V02391  V02278  CV05164  CV05163  V05534  V03761  V03762  V02637  V03129  V02792  V02997  V02531  CV05419  V03533  V02627  V02633  V02817  V03697  V03652  V02606  V05388  V02993  V03513  V05527  V03501  CV05162  V03159  CV05121  V03255  V02745  V03256  V03228  V05129  V05406  V03512  V03430  CV05199  CV05198  CV05202  CV05303  V05052  V02156  V05319  CV05261  V02636  V05053  V02690  V05127  V02840  V03534  V01931  V05544  V01925  V05456  V02980  V02798  V05288  V05232  V03597  V02315  V03530  V05566  CV05249  V02480  V02488  V02475  V05163  V03508  V02484  V02487  V02477  V02482  V02486  V02474  V03134  V03521  V03535  V05535  V02346  CV05352  CV05122  V03998  V05128  V03157  V02694  V03936  V02498  V01976  V02714  V05533  V02994  V03657  CV05229  cv05107  V02589  V02686  V02001  V05559  V0376",
      "more_code_info": "4  V02385  V03532  V03517  V03907  CV05332  V03515  V03807  V05047  V01924  V05199  V03514  V01508  CV05374  CV05003  V03519  V04164  V05351  V03708  V02585  CV05378  V02795  CV05266  V05148  CV05309  CV05344  V05459  V05478  V05477  V05473  V05508  V05293  CV05319  V02925  CV05334  V02173  CV05416  V05546  V05548  CV05341  CV05340  V05528  V05552  V05551  V05549  V05550  V05554  CV05354  V02928  V02865  V05093  CV05415  V02591  V01363  V05539  V02832  V05097  V05000  V05050  V05139  V05140  V05138  V05276  V02105  VV05274  V02060  V05215  V05202  CV05269  V03174  V05263  CV05245  CV05239  CV05165  CV05246  CV05174  CV05276  V05037  V03988  V03236  V05331  CV05275  V05313  V04009  V03175  CV05146  CV05185  V05492  V05491  CV05399  CV05312  CV05205  V05332  V04016  V039039  V03740  V05420  V05584  V05578  V03222  V05035  V05137  V02135  V03963  V03961  V03065  V02241  V02302  V05102  V05104  V05272  V05385  V05424  V01364  V01341  V01608  V03754  V05524  CV05186  CV05147  V05311  V05532  CV05108  V03937  V03668  V03660  V03681  V03098  V05110  10050522  V02301  V03666  V01433  V01691  V01801  V05149  V03983  V03208  V05146  V03973  V03981  V05226  V01800  V03667  V02713  V05277  V04169  V05076  V01340  CV05157  CV05170  CV05307  CV05092  V05335  CV05180  V03284  V05230  V03962  V01337  V05295  V05296  V05302  V03310  V05085  V05464  CV05151  CV05155  CV05130  V01585  CV05169  V05337  CV05173  V02833  V04017  V05525  V02930  V02324  V02735  V02504  V05019  V05059  CV05403  CV05404  V03949  V02388  V03662  V05510  V05511  CV05422  CV05306  CV05300  CV05326  CV05324  CV05325  V01498  V01500  V01106  CV05323  V01495  V02424  V02053  V04015  V04025  V05133  V01468  V03359  V01020  V05463  V05441  V05201  V04011  V03039  CV05268  V03326  V03946  V03328  V02044  V05089  CV05315  CV05267  V01006  CV05058  V01434  VEL10020485  V01574  V01472  V01703  V01344  V03232  V05509  CV05080  CV05085  CV05094  CV05090  CV05088  CV05089  V03231  CV05093  V05330  V03247  V03689  V02133  V03674  CV05172  CV05118  V05488  V04217  V04027  V04008  V04007  CV05145  CV05144  V03325  150010422  V03664  V04216  CV05171  V05386  V05251  V05018  V03673  0233  V01499  V05260  CV05318  V03734  V03960  V03948  V02387  V03169  V03323  V05342  V05341  V02083  V01496  V05304  V03676  V02300  V01651  V01650  V02320  V03164  V05523  V05355  10050527  V05316  V05103  V03212  V02825  V03987  V05538  V05522  V02325  V05460  V04012  V03992  V30992  V02316  V03324  V01352  V05438  V05437  V05512  VEL10020483  V01354  CV05129  9090430  V05418  V02908  V02911  V02912  V05410  V05490  V05489  V05397  V05396  V05415  V05414  V05411  CV05270  V01414  10100577  V05252  CV05272  V03739  V03938  V04006  V03939  V05314  CV05156  V03541  V03611  V03036  V05469  V05501  V01802  V01817  V01609  V03659  V03663  V03665  V05258  V05259  CV05420  CV05418  CV05417  V05303  CV05401  CV05310  CV05405  CV05406  CV05402  CV05273  V03661  V03940  V05223  V03160  V05253  V03361  V03989  V05221  V03990  V03984  CV05125  CV05096  CV05168  V05419  CV05167  V05220  V04004  V04003  V04005  V03670  V03332  V03741  V05398  V05422  V05257  V03985  V03986  V05256  CV05274  V01290  V02389  V05136  V05169  V05170  V05152  V05098  V05124  V05123  V05080  V05075  V05134  V05122  CV05154  V05439  V05500  V05338  V05336  V05204  V05168  V05334  V05274  V05262  V01195  V03692  V03982  V02828  V05507  V04018  V03941  V05333  V01658  V01813  V01435  1004051  V05370  V02822  V03672  V03429  V03669  V04170  V03942  V05081  V05261  V05084  V01171  V04026  V02823  V01993  V03234  V03233  V01448  V05315  CV05053  V03389  V01112  V03097  0244  V01497  V05078  V03155  V05297  V01845  V05203  V03653  V03783  V03782  V03908  V05096  CV05383  V03909  V03943  V03945  V03728  V03820  V03021  V03911  V03806  V03707  V03705  V03950  V05461  9090427  V03093  V03991  V02386  V03117  V03063  CV05206  V05294  V04013  V05094  V03074  V05171  V04010  V03031  CV05305  CV05304  V05077  V03149  V01865  V02824  V01692  V05231  V04019  V05082  CV05207  V02378  V03680  V02500  CV05314  V05174  V03947  V05536  V03671  V03239  V05495  V05468  V03709  V03743  VEL09050374  VEL09010321  V02613  V02615  V03286  V03287  V02118  V03235  V03951  V05308  CV05355  V03193  V03191  V02562  V02851  V02750  V02611  V03330  V02751  V03333  V03322  V03321  V03320  V03077  V03066  V03313  V03312  V02940  V02631  V02623  V03071  V03070  V03068  V03195  V03339  V02618  V05109  V05200  V05065  V02108  V05088  V04087  V04090  V04212  V04203  V04202  V04082  V04086  V04147  V04042  V04144  V04145  V04149  V04150  V04159  V04201  V04039  V04040  V04155  V04180  V04063  V04158  V04108  V04038  V04071  V04065  V04064  V04060  V04059  V04057  V04044  V04068  V04062  V04069  V04061  V04067  V04151  V04045  V04046  V04211  V03706  V03703  V04084  V04107  V04123  V04089  V04174  V04171  V04175  V04160  V04083  V04177  V04173  V04172  V04176  V03729  V04157  V04058  V04050  V04056  V04152  V04156  V04055  V04043  V04052  V04049  V04053  V04091  V04047  V04051  V04072  V04073  V04070  V04074  V04204  V04205  V05182  V04154  V04207  V04127  V04146  V04121  V04119  V04133  V04130  V04131  V04125  V04092  V05001  V04129  V04120  V04178  V04085  V04105  V04080  V04078  V04077  V04075  V04076"
    }
  ]
}