      "status": "Terminated",
      "city": "Webster",
      "state": "TX",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87057",
      "recalling_firm": "Ansh Labs, LLC",
      "address_1": "445 W Medical Center Blvd",
      "address_2": "N/A",
      "postal_code": "77598-4217",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution including in the states of TX, MO, OK, MD, FL, NY, OR, GA, NC. OUS: None",
      "recall_number": "Z-0842-2021",
      "product_description": "AnshLabs - SARS-CoV2 IgM ELISA (u-Capture), IVD, REF: AL-1002, kit contents, Ab PLATE 1 X ! Plate, IgM Sample DIL 1 x 100 mL, ENZ CONJ CONC 1 x 1.5 mL, Calibrators A-C 3 x 1.2 mL, CONJ DIL 1 X 12 mL, Stop SOLN 1 X 12 mL, TMB SOLN 1 X 12 mL, WASH CONC A 1 X 60 mL, (01)00853592008219 - Product Usage: intended for the qualitative detection of IgM antibodies in serum collected from the individuals suspected with signs and symptoms of COVID-19 infection by their healthcare provider.",
      "product_quantity": "327 kits (93 test per  kit)",
      "reason_for_recall": "Labeling update; To reduce the risk of false positive/false negative results the IFU for the adaptive immune response to SRS-CoV-2 testing is being updated  to remove Plasma as a \"sample type\" and to update the \"limitation\" section to include potential cross-reactivity with non-SARS-CoV2 strains.",
      "recall_initiation_date": "20201218",
      "center_classification_date": "20210111",
      "termination_date": "20220111",
      "report_date": "20210120",
      "code_info": "Lot #/Expiration Date: 040720/2021-10-06, 041620/2021-10-15, 041720/2021-10-18, 042420/2021-10-15, 042720-A/2021-10-26"
    }
  ]
}