  "results": [
    {
      "status": "Terminated",
      "city": "Tuusula",
      "state": "",
      "country": "Finland",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84544",
      "recalling_firm": "Palodex Group Oy",
      "address_1": "Nahkelantie 160",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US: WI and PA    OUS: ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BRAZIL, CANADA, CHILE, CHINA, COSTA RICA, CZECH REPUBLIC, DENMARK, ECUADOR, FINLAND, FRANCE, GERMANY, GREECE, HUNGARY, INDIA, ISRAEL, ITALY, JAPAN, KOREA, REPUBLIC OF, KUWAIT, MALAYSIA, MALTA, NETHERLANDS, NEW ZEALAND, NORWAY, OMAN, PAKISTAN, POLAND, QATAR, ROMANIA, RUSSIAN FEDERATION, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVENIA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, REPUBLIC OF CHINA, THAILAND, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, VIETNAM",
      "recall_number": "Z-0842-2020",
      "product_description": "Instrumentarium Dental Express SOLID STATE X-RAY IMAGER (FLAT PANEL/DIGITAL IMAGER)",
      "product_quantity": "4,563 Total Globally (all products)",
      "reason_for_recall": "Slow degradation of plastic insulation used in the PSU DC plug of the DIGORA Optime, Scan eXam, Express, GXPS-500 systems may result in short circuiting which may cause heating and melting. The melting may generate smoke and eventually cause the PSU to fail.",
      "recall_initiation_date": "20200107",
      "center_classification_date": "20200123",
      "termination_date": "20240415",
      "report_date": "20200129",
      "code_info": "Serial Number: TBD (range from IJ1208217 to IJ1313309)    PSU's having nine-digit serial numbers beginning with four digits as follows:  1143XXXXX  1221XXXXX  1224XXXXX and  1229XXXXX",
      "more_code_info": ""
    }
  ]
}