  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Sunnyvale",
      "address_1": "1266 Kifer Rd Bldg 101",
      "reason_for_recall": "During an internal inspection, Intuitive found that the sterile packaging of the da Vinci S/Si Harmonic ACE Inserts may become damaged during transit.  The potential damage includes pinholes and /or small openings (less than a few millimeters) on the pouch leading to a potential breach in sterility.",
      "address_2": "",
      "product_quantity": "43755 inserts",
      "code_info": "All Lots",
      "center_classification_date": "20190215",
      "distribution_pattern": "US and Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Denmark, Finland, France, Germany, India, Israel, Italy, Japan, Kuwait, Lebanon, Mexico, Monaco, Netherlands, Pakistan, Qatar, Romania, Russian Federation, Saudi Arabia, Singapore, South Korea, Spain, Taiwan, Turkey, United Kingdom",
      "state": "CA",
      "product_description": "daVinci Harmonic ACE Curved Shears",
      "report_date": "20190227",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Intuitive Surgical, Inc.",
      "recall_number": "Z-0841-2019",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "81972",
      "more_code_info": "",
      "recall_initiation_date": "20190116",
      "postal_code": "94086-5304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Ongoing"
    }
  ]
}