  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Salt Lake City",
      "address_1": "384 N Wright Brothers Dr",
      "reason_for_recall": "mA Accuracy may exceed the design and labeling specification of+ 10%. Per 1020.32(f): Deviation of x-ray tube potential and current from the indicated values shall not exceed the maximum deviation as stated by the manufacturer in accordance with 1020.30(h)(3).",
      "address_2": "",
      "product_quantity": "61 units installed in US",
      "code_info": "GE OEC Brivo 865 Plus Serial Numbers:    B4S13011  B4S13012  B4S13016  B4S13018  B4S13020  B4S13029  B4S13031  B4S13035  B4S13038  B4S13043  B4S13054  B4S13056  B4S13059-  B4S13061  B4S14002-  B4S14004  B4S14007  B4S14015  B4S14016  B4S14022  B4S14022  B4S1403 3  B4S1403 4",
      "center_classification_date": "20150112",
      "distribution_pattern": "Worldwide Distribution",
      "state": "UT",
      "product_description": "GE OEC Brivo 865 Plus Mobile C-Arm X-Ray Product    Used for general surgical applications and musculoskeletal procedures to visualize.",
      "report_date": "20150121",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "GE OEC Medical Systems, Inc",
      "recall_number": "Z-0841-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "69660",
      "termination_date": "20150520",
      "more_code_info": "",
      "recall_initiation_date": "20140915",
      "postal_code": "84116-2862",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}