{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Tuusula",
      "state": "",
      "country": "Finland",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84544",
      "recalling_firm": "Palodex Group Oy",
      "address_1": "Nahkelantie 160",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US: WI and PA    OUS: ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BRAZIL, CANADA, CHILE, CHINA, COSTA RICA, CZECH REPUBLIC, DENMARK, ECUADOR, FINLAND, FRANCE, GERMANY, GREECE, HUNGARY, INDIA, ISRAEL, ITALY, JAPAN, KOREA, REPUBLIC OF, KUWAIT, MALAYSIA, MALTA, NETHERLANDS, NEW ZEALAND, NORWAY, OMAN, PAKISTAN, POLAND, QATAR, ROMANIA, RUSSIAN FEDERATION, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVENIA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, REPUBLIC OF CHINA, THAILAND, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, VIETNAM",
      "recall_number": "Z-0840-2020",
      "product_description": "Gendex GXPS-500 SOLID STATE X-RAY IMAGER (FLAT PANEL/DIGITAL IMAGER)",
      "product_quantity": "101",
      "reason_for_recall": "Slow degradation of plastic insulation used in the PSU DC plug of the DIGORA Optime, Scan eXam, Express, GXPS-500 systems may result in short circuiting which may cause heating and melting. The melting may generate smoke and eventually cause the PSU to fail.",
      "recall_initiation_date": "20200107",
      "center_classification_date": "20200123",
      "termination_date": "20240415",
      "report_date": "20200129",
      "code_info": "Serial Number: GJ1209488, GJ1209489, GJ1209491, GJ1209492, GJ1209493, GJ1209494, GJ1209495, GJ1209496, GJ1209498, GJ1209499, GJ1209504, GJ1209505, GJ1209506, GJ1209507, GJ1209509, GJ1209510, GJ1209511, GJ1209513, GJ1209514, GJ1209516, GJ1211287, GJ1211288, GJ1211289, GJ1211290, GJ1211291, GJ1211292, GJ1211293, GJ1211294, GJ1211295, GJ1211296, GJ1211297, GJ1211298, GJ1211299, GJ1211300, GJ1211301, GJ1211302, GJ1211303, GJ1211304, GJ1211305, GJ1211307, GJ1311619, GJ1311621, GJ1311622, GJ1311623, GJ1311624, GJ1311625, GJ1311626, GJ1311627, GJ1311628, GJ1311629, GJ1311630, GJ1311631, GJ1311632, GJ1311633, GJ1311634, GJ1311635, GJ1311636, GJ1311637, GJ1311638, GJ1311639, GJ1311640, GJ1311641, GJ1311642, GJ1311643, GJ1311644, GJ1311645, GJ1311646, GJ1311647, GJ1311648, GJ1311650, GJ1311651, GJ1311653, GJ1311656, GJ1311658, GJ1311659, GJ1311660, GJ1311661, GJ1311662, GJ1311663, GJ1311664, GJ1311665, GJ1311666, GJ1311669, GJ1311670, GJ1311671, GJ1311672, GJ1311673, GJ1311674, GJ1311675, GJ1311676, GJ1311677, GJ1311678, GJ1311679, GJ1311680, GJ1311681, GJ1311682, GJ1311683, GJ1311684, GJ1311685, GJ1311686, GJ1311722    PSU's having nine-digit serial numbers beginning with four digits as follows:  1143XXXXX  1221XXXXX  1224XXXXX and  1229XXXXX",
      "more_code_info": ""
    }
  ]
}