  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Cleveland",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79134",
      "recalling_firm": "Philips Medical Systems (Cleveland) Inc",
      "address_1": "595 Miner Rd",
      "address_2": "N/A",
      "postal_code": "44143-2131",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide",
      "recall_number": "Z-0840-2018",
      "product_description": "Brilliance Big Bore Oncology CTwith 4.2.0 software version. Model #' 728243     Product Usage:  Computed Tomography X-ray systems intended to produce cross-sectional Images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. These devices may include signal analysts and display equipment, patient and equipment supports, components, and accessories.",
      "product_quantity": "11",
      "reason_for_recall": "Software issues including:  Software errors that may result in Extended Field of View check message not displayed when Field of View is set >600 mm for a paused scan;  Request for change to automatically turn on the center x/y 0 setting and the disable result rotation setting when turning on the CT simulation Exam Card;  Software errors that may result in Tumor LOC allowing POI (Point of Interest) locking when a non-isocenter slice is displayed;  Software errors that may result in partial set of images or inability to generate CT images.",
      "recall_initiation_date": "20180103",
      "center_classification_date": "20180302",
      "termination_date": "20190702",
      "report_date": "20180314",
      "code_info": "Brilliance Big Bore Oncology CT, Model # 728243  Serial #s:  76025  76022  76024  76023  76016  76021  76008  76010  76002  76020  76011"
    }
  ]
}