  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73142",
      "recalling_firm": "Zimmer Biomet, Inc.",
      "address_1": "1800 W Center St",
      "address_2": "N/A",
      "postal_code": "46580-2304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US (nationwide) and Internationally to   ARGENTINA               AUSTRALIA  AUSTRIA  BARBADOS  BELGIUM  BERMUDA  BOLIVIA  BRAZIL  CARIBBEAN  CHILE  CHINA  COLOMBIA  COST RICA  DENMARK  DOMINICAN REPUBLIC  DORAL FL  ECUADOR  EL SALVADOR  ENGLAND  FRANCE  GERMANY  GUATEMALA  HONDURAS  HONG KONG  INDIA  ISRAEL  ITALY  JAMAICA  JAPAN  KOREA  MALAYSIA  MEXICO  NEW ZEALAND  NICARAGUA  PANAMA  PERU  SINGAPORE  SPAIN  SWEDEN  SWITZERLAND  TAIWAN  THAILAND  VENEZUELA  WEST INDIES  Canary Islands  Croatia  Cyprus  Czech Republic  Canada",
      "recall_number": "Z-0840-2016",
      "product_description": "CR Flex GSF Porous Femoral  LPS Flex GSF Porous Femoral  LPS Flex Porous Femoral  prosthesis, knee, patello/femorotibial, semi-constrained, uncemented, porous, coated, polymer/metal/polymer  various sizes  This device is indicated for patients with severe knee pain and disability due to:   Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis.   Collagen disorders, and/or avascular necrosis of the femoral condyle.   Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy.   Moderate valgus, varus, or flexion deformities.   The salvage of previously failed surgical attempts or for a knee in which satisfactory stability in flexion cannot be obtained at the time of surgery. \" The CR-Flex GSF and LPS-Flex GSF uncoated (Option) and precoat styles are intended for cemented use only. \" The CR-Flex GSF and LPS-Flex GSF porous coated femoral components may be used cemented or uncemented (biological fixation). \" Specific uses with CR-Flex GSF or LPS-Flex GSF femorals:   Provides increased flexion capability for patients who have both the flexibility and desire to increase their flexion range.   The CR-Flex GSF femoral, when used with 90-prefix NexGen CR articular surfaces, 00-prefix 10, 12 or 14 mm or 90-prefix 17 or 20 mm Prolong¿ Highly Crosslinked Polyethylene CR articular surfaces, or with Gender Solutions Natural-Knee Flex Congruent articular surfaces, is designed for use with a functional posterior cruciate ligament and when load bearing range of motion (ROM) is expected to be less than or equal to 155 degrees.   The LPS-Flex GSF femoral, when used with LPS-Flex articular surfaces, is designed for use with both cruciate ligaments excised and when load bearing ROM is expected to be less than or equal to 155 degrees.\"",
      "product_quantity": "18482",
      "reason_for_recall": "LDPE bag containing the implant adheres to the highly polished implant surface.",
      "recall_initiation_date": "20160111",
      "center_classification_date": "20160222",
      "termination_date": "20170523",
      "report_date": "20160302",
      "code_info": "Item numbers:  00-5752-013-01  00-5752-013-02  00-5752-013-05  00-5752-013-06  00-5752-014-01  00-5752-014-02  00-5752-014-05  00-5752-014-06  00-5752-015-01  00-5752-015-02  00-5752-015-05  00-5752-015-06  00-5752-016-01  00-5752-016-02  00-5752-016-05  00-5752-016-06  00-5752-017-01  00-5752-017-02  00-5752-017-05  00-5752-017-06  00-5762-013-51  00-5762-013-52  00-5762-014-51  00-5762-014-52  00-5762-015-51  00-5762-015-52  00-5762-016-51  00-5762-016-52  00-5762-017-51  00-5762-017-52  00-5962-012-51  00-5962-012-52  00-5962-013-51  00-5962-013-52  00-5962-014-51  00-5962-014-52  00-5962-015-51  00-5962-015-52  00-5962-016-51  00-5962-016-52  00-5962-017-51  00-5962-017-52  00-5962-018-51  00-5962-018-52    Lot numbers:  11003002  11003328  11003397  11003526  11003606  11003637  11003639  11003667  11003972  11004037  11004739  11004835  11004913  11004915  11005025  11005124  11005180  11005183  11005286  11005287  11005288  11005389  11005454  11005613  11005669  11005756  11005843  11005917  11006049  11006065  11006143  11006207  11006225  11006244  11006412  11006591  11006650  11006688  11006696  11006700  11006733  11006807  11006834  11007020  11007021  11007087  11007099  11007174  11007191  11007205  11007274  11007372  11007386  11007388  11007411  11007457  11007477  11007542  11007660  11007662  11007669  11007681  11007683  11007703  11007741  11007752  11007821  11007866  11007872  11007896  11007936  11008030  11008283  11008308  11008313  11008324  11008378  11008402  61759051  61777352  61782446  61782447  61782448  61782449  61782450  61782451  61782452  61782453  61782454  61782455  61789737  61789738  61789747  61789748  61789749  61789759  61789761  61789762  61789763  61789764  61795241  61795242  61795243  61795245  61795246  61795248  61795249  61795250  61795251  61795252  61800466  61800467  61800478  61800479  61800480  61800481  61800482  61800483  61800484  61800485  61805217  61805218  61805219  61805220  61805221  61805226  61805227  61805228  61805229  61805230  61809915  61809916  61809917  61809918  61809919  61809921  61809922  61809923  61809924  61809925  61815341  61815342  61815345  61815346  61820888  61820889  61820890  61820891  61821984  61821985  61824117  61824118  61824119  61824128  61824129  61824130  61830853  61830854  61830855  61830858  61830859  61830860  61836622  61836623  61836624  61836625  61836626  61876068  61876069  61876070  61876071  61876072  61876073  61876074  61876075  61876076  61884862  61884864  61884865  61884866  61884894  61884895  61884896  61884897  61884898  61884899  61888624  61888625  61888626  61888628  61888629  61888630  61888631  61888636  61888637  61888638  61888639  61894296  61894297  61894298  61894299  61894301  61894302  61894303  61894304  61894305  61894306  61894307  61899369  61899370  61899371  61899372  61899373  61899374  61899375  61899377  61899378  61899379  61904017  61904018  61904019  61904020  61904021  61904022  61904023  61904024  61904025  61904026  61910455  61910456  61910457  61910458  61910459  61910460  61910461  61910462  61910463  61910464  61910465  61915857  61915858  61915859  61915860  61915861  61915862  61915863  61915864  61915865  61915866  61921164  61921165  61921166  61921167  61921168  61921169  61921170  61921171  61921172  61921173  61927621  61927622  61927623  61927624  61927625  61927626  61927627  61927628  61927630  61927631  61935802  61935803  61935804  61935806  61935807  61935811  61935812  61935813  61935814  61935815  61936656  61936657  61936658  61936659  61936660  61936661  61938890  61938891  61938892  61938893  61938894  61938895  61938898  61938899  61938901  61938902  61940220  61940221  61940222  61940224  61940225  61940226  61940227  61940228  61949483  61949484  61949485  61949486  61949487  61949488  61949496  61949497  61949498  61949499  61949501  61949503  61949504  61949505  61949506  61951516  61951517  61951518  61951519  61951520  61951522  61951523  61951524  61951525  61951526  61951527  61951531  61951532  61951533  61951534  61963768  61963769  61966302  61968860  61968861  61968862  61968863  61968864  61968865  61968866  61968867  61968869  61968870  61974774  61974775  61974776  61974777  61974778  61974779  61974780  61974781  61974782  61983343  61983344  61983345  61983351  61983352  61983364  61983365  61983366  61983368  61983369  61983370  61983371  61987919  61987920  61987921  61987922  61987923  61987925  61987926  61987927  61987930  61987931  61987932  61994881  61994882  61994883  61994884  61994888  61994889  61994891  61994892  61994893  61994894  61994895  62001350  62001351  62001352  62001353  62001354  62001355  62001356  62001357  62001358  62001359  62001368  62001369  62001371  62001372  62001373  62002453  62005701  62005702  62005704  62005705  62005706  62005707  62005708  62005709  62005710  62005711  62005712  62005713  62013400  62013401  62013402  62013403  62013410  62013411  62013412  62013413  62013425  62013426  62013427  62013428  62013429  62020187  62020188  62020189  62020190  62020191  62020192  62020193  62020194  62020195  62020196  62020197  62020198  62020199  62020200  62020201  62020202  62020216  62025637  62025638  62025639  62025640  62025641  62025642  62025643  62025644  62025645  62025646  62025647  62025648  62025649  62025650  62032544  62032545  62032546  62032547  62032548  62032549  62032550  62032551  62032552  62032553  62032554  62032565  62032566  62032567  62032568  62038799  62038801  62038802  62038803  62038804  62038805  62038812  62038813  62038831  62038832  62038833  62038834  62038835  62038836  62046073  62046074  62046075  62046076  62046083  62046084  62046085  62046086  62046087  62046088  62046089  62046090  62046091  62046092  62046093  62046094  62046095  62054543  62054544  62054545  62054546  62054547  62054560  62054562  62054563  62054564  62054565  62054566  62054567  62054571  62054572  62054573  62056823  62056824  62056825  62056826  62056827  62056836  62056837  62056838  62056839  62056892  62062038  62062039  62062040  62062041  62062042  62062043  62062044  62062045  62062046  62062047  62062048  62062049  62062050  62068140  62068141  62068142  62068150  62068151  62068152  62068153  62068154  62068155  62068163  62068165  62068166  62074861  62074862  62074863  62074864  62074865  62074866  62074867  62074868  62074869  62074870  62074871  62074872  62074873  62074874  62074880  62080577  62080578  62080579  62080581  62080582  62080583  62080584  62080585  62080586  62080587  62086010  62086011  62086777  62086778  62086779  62086780  62086781  62086782  62086783  62086784  62086785  62086786  62086787  62086788  62086789  62086790  62086791  62086792  62092417  62092418  62092419  62092420  62092421  62092431  62092432  62092433  62092434  62092435  62099026  62099031  62099032  62103526  62103527  62103529  62103530  62104132  62104133  62104134  62104135  62104136  62104137  62104138  62104139  62104140  62104141  62104142  62104143  62104155  62104849  62104850  62104852  62105365  62105367  62105371  62105372  62110567  62110568  62110569  62110570  62110575  62110576  62110577  62110578  62110579  62110580  62110581  62110582  62110590  62110591  62117104  62117105  62117106  62117107  62117108  62117109  62117110  62117111  62117112  62117125  62117126  62117127  62117128  62117129  62122753  62122754  62122755  62122756  62122760  62122761  62122762  62122764  62122765  62122766  62122767  62122768  62128145  62128146  62128147  62128148  62128149  62128150  62128151  62128152  62128153  62128154  62128155  62128156  62128157  62128158  62128159  62128160  62132132  62132133  62132134  62132135  62132136  62132137  62132138  62132139  62132140  62132146  62132165  62132168  62132169  62132170  62132171  62132172  62139317  62139318  62139320  62139321  62139339  62139340  62139341  62139342  62139343  62139370  62139371  62139372  62139373  62139374  62143711  62143712  62145113  62145114  62145115  62145116  62145117  62145131  62145132  62145133  62145134  62145135  62145136  62145140  62145141  62145142  62145143  62145144  62145145  62145146  62145147  62145148  62150722  62150723  62150724  62150725  62150726  62150746  62150747  62150748  62150749  62150750  62150751  62150752  62150753  62150754  62150755  62155182  62155183  62155184  62155199  62155201  62155202  62155203  62155238  62155239  62155240  62155250  62155251  62155252  62155253  62155254  62160512  62160513  62160514  62160515  62160516  62160517  62160518  62160519  62165137  62165138  62165139  62165140  62165141  62165146  62165147  62165148  62165150  62165151  62165152  62165153  62165154  62165155  62170481  62170482  62170494  62170496  62170534  62170535  62170536  62173422  62173423  62173424  62173425  62173426  62173427  62173428  62173429  62173431  62173432  62173433  62173434  62176043  62176044  62176045  62176046  62176047  62176048  62176049  62176050  62176051  62176052  62176053  62176054  62178778  62178779  62181371  62181372  62181373  62181374  62181375  62181376  62181384  62181385  62181386  62181387  62181388  62181389  62181390  62181391  62183503  62183504  62183505  62183506  62183509  62183510  62183511  62183512  62183513  62183519  62183520  62183521  62183522  62199820  62199836  62199837  62199838  62205158  62205159  62205160  62205161  62205162  62205163  62205164  62210402  62210403  62210404  62210405  62210406  62210407  62210408  62210409  62210410  62210411  62210412  62210413  62210414  62223977  62223979  62223980  62223981  62223985  62223986  62224002  62224003  62224004  62224005  62224006  62224007  62224008  62224009  62224013  62224014  62224017  62224018  62224019  62224020  62230253  62230254  62230255  62230256  62230257  62230258  62230290  62230291  62230292  62230293  62230294  62230330  62230331  62234279  62234280  62234281  62234282  62234288  62234296  62238035  62238036  62238037  62238038  62238039  62238040  62238118  62238119  62238120  62239690  62239691  62245122  62245123  62245125  62245126  62245127  62245150  62245162  62245163  62245174  62250268  62250269  62250271  62250274  77002019  61951531R  62001368R",
      "more_code_info": ""
    }
  ]
}