  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Salt Lake City",
      "address_1": "384 N Wright Brothers Dr",
      "reason_for_recall": "mA Accuracy may exceed the design and labeling specification of+ 10%. Per 1020.32(f): Deviation of x-ray tube potential and current from the indicated values shall not exceed the maximum deviation as stated by the manufacturer in accordance with 1020.30(h)(3).",
      "address_2": "",
      "product_quantity": "61 units installed in US",
      "code_info": "GE OEC Brivo 715 Prime Serial Numbers:    B2S1301 6  B2S1301 8  B2S1302 1  B2S13022  B2S13025  B2S13039  B2S13060  B2S13063  B2S13065  B2S14001  B2S14007  B2S14022  B2S14039  B2S1404 0  B2S1404 5  B2S1406 5  B2S1406 6",
      "center_classification_date": "20150112",
      "distribution_pattern": "Worldwide Distribution",
      "state": "UT",
      "product_description": "GE OEC Brivo 715 Prime Mobile C-Arm X-Ray Product    Used for general surgical applications and musculoskeletal procedures to visualize.",
      "report_date": "20150121",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "GE OEC Medical Systems, Inc",
      "recall_number": "Z-0840-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "69660",
      "termination_date": "20150520",
      "more_code_info": "",
      "recall_initiation_date": "20140915",
      "postal_code": "84116-2862",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}