  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Waltham",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86939",
      "recalling_firm": "Fresenius Medical Care Holdings, Inc.",
      "address_1": "920 Winter St Bld 950",
      "address_2": "N/A",
      "postal_code": "02451-1521",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-0839-2021",
      "product_description": "Naturalyte Liquid Acid concentrate drums, Product Code 13-2251-0 - Product Usage: used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine.",
      "product_quantity": "224 drums",
      "reason_for_recall": "The deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously communicated.  Calcium concentration is below manufacturing specification.",
      "recall_initiation_date": "20201123",
      "center_classification_date": "20210111",
      "report_date": "20210120",
      "code_info": "Lot Numbers: 20LXAC025  20LXAC081  20LXAC082  20LXAC083    Additional Affected lots as of 3/17/21: 20LXAC029  20NXAC021  20LXAC026  20LXAC065  20LXAC028  20NXAC029  20LXAC071  20LXAC103  20LXAC027  20LXAC070  20LXAC067  20NXAC019  20LXAC082  20LXAC025  20LXAC081  20LXAC083"
    }
  ]
}