{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Crumlin (North)",
      "state": "",
      "country": "Ireland",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "81746",
      "recalling_firm": "Randox Laboratories Ltd.",
      "address_1": "55 Diamond Road",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "WV, PR",
      "recall_number": "Z-0839-2019",
      "product_description": "Randox CALCIUM (Ca) Colorimetric Method RX Series  Cat. No. CA 3871,  R1. Arsenazo Reagent 9 x 51ml  GTIN: 05055273200904",
      "product_quantity": "225 kits",
      "reason_for_recall": "Update to the carryover avoidance technical bulletin to introduce additional steps for reagent carryover avoidance with the Calcium assay on RX instruments. Carryover to the Calcium reagent results may result in falsely elevated or decreased results to both patient and QC samples.",
      "recall_initiation_date": "20181204",
      "center_classification_date": "20190215",
      "termination_date": "20200618",
      "report_date": "20190227",
      "code_info": "All lot codes",
      "more_code_info": ""
    }
  ]
}