      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73142",
      "recalling_firm": "Zimmer Biomet, Inc.",
      "address_1": "1800 W Center St",
      "address_2": "N/A",
      "postal_code": "46580-2304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US (nationwide) and Internationally to   ARGENTINA               AUSTRALIA  AUSTRIA  BARBADOS  BELGIUM  BERMUDA  BOLIVIA  BRAZIL  CARIBBEAN  CHILE  CHINA  COLOMBIA  COST RICA  DENMARK  DOMINICAN REPUBLIC  DORAL FL  ECUADOR  EL SALVADOR  ENGLAND  FRANCE  GERMANY  GUATEMALA  HONDURAS  HONG KONG  INDIA  ISRAEL  ITALY  JAMAICA  JAPAN  KOREA  MALAYSIA  MEXICO  NEW ZEALAND  NICARAGUA  PANAMA  PERU  SINGAPORE  SPAIN  SWEDEN  SWITZERLAND  TAIWAN  THAILAND  VENEZUELA  WEST INDIES  Canary Islands  Croatia  Cyprus  Czech Republic  Canada",
      "recall_number": "Z-0839-2016",
      "product_description": "NKII CSTI Porous Femoral  NKII NP Femoral  prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer/metal/polymer  variou sizes  Product Usage\t\"\" The Natural-Knee II System with Cancellous-Structured Titanium (CSTi) Porous Coating is indicated for uncemented or cemented use in skeletally mature individuals with intact medial and lateral collateral ligaments undergoing primary surgery for rehabilitating knees damaged as a result of Noninflammatory Degenerative Joint Disease (NIDJD) or Inflammatory Joint Disease (IJD). \" The Natural-Knee II primary components without CSTi Porous Coating, all posterior stabilized and revision components, the N-K II Constrained Knee System, the modular cemented tibial baseplate, and the revision stem and revision fluted stem are indicated for cemented use only in skeletally mature individuals with conditions of Noninflammatory Degenerative Joint Disease (NIDJD) or Inflammatory Joint Disease (IJD), correctable valgus-varus deformity and moderate flexion contracture, or failed previous surgery where pain, deformity or dysfunction persists. When the primary or N-K II Constrained Knee System components are used, the medial and collateral ligaments must be intact. \" The Natural-Knee II Durasul polyethylene components are indicated for patient conditions of noninflammatory degenerative joint disease (NIDJD), e.g., avascular necrosis, osteoarthritis, and inflammatory degenerative joint disease (IJD), e.g., rheumatoid arthritis; correctable valgus-varus deformity and moderate flexion contracture; those patients with failed previous surgery where pain, deformity, or dysfunction persists; and revision of previously failed knee arthroplasty. These devices are intended for cemented use only in the United States.\"",
      "product_quantity": "13483",
      "reason_for_recall": "LDPE bag containing the implant adheres to the highly polished implant surface.",
      "recall_initiation_date": "20160111",
      "center_classification_date": "20160222",
      "termination_date": "20170523",
      "report_date": "20160302",
      "code_info": "Item numbers:  6212-00-007  6212-00-008  6212-00-009  6212-00-010  6212-00-011  6212-00-020  6212-00-021  6212-00-030  6212-00-031  6212-00-040  6212-00-041  6212-00-050  6212-00-051  6307-00-006  6307-00-007  6307-00-008  6307-00-009  6307-00-010  6307-00-011  6307-00-020  6307-00-021  6307-00-030  6307-00-031  6307-00-040  6307-00-041  6307-00-050  6307-00-051    Lot numbers:  11002695  11003148  11003157  11003159  11003165  11003167  11003170  11003181  11003190  11003192  11003193  11003195  11003196  11003201  11003212  11003214  11003216  11003223  11003228  11003230  11003231  11003232  11003235  11003236  11003240  11003241  11003244  11003248  11003249  11003251  11003252  11003253  11003255  11003257  11003258  11003259  11003261  11003267  11003981  11004011  11004072  11004206  11004207  11004208  11004247  11004248  11004286  11004310  11004330  11004367  11004383  11004384  11004385  11004386  11004410  11004411  11004426  11004427  11004473  11004475  11004541  11004550  11004551  11004588  11004589  11004590  11004592  11004593  11004603  11004604  11004610  11004636  11004637  11004644  11004664  11004783  11004784  11004785  11004800  11004802  11004803  11004839  11004841  11004844  11004846  11004876  11004879  11004937  11004938  11004939  11004940  11004960  11004967  11004968  11004969  11005012  11005013  11005014  11005015  11005039  11005040  11005041  11005042  11005178  11005267  11005295  11005299  11005300  11005380  11005512  11005514  11005590  11005591  11005603  11005604  11005619  11005620  11005713  11005745  11005774  11005777  11005876  11005923  11005924  11005951  11006034  11006104  11006123  11006124  11006125  11006147  11006148  11006216  11006217  11006218  11006293  11006294  11006326  11006393  11006394  11006395  11006426  11006427  11006446  11006475  11006502  11006503  11006528  11006532  11006533  11006559  11006560  11006568  11006569  11006606  11006636  11006637  11006643  11006678  11006679  11006680  11006681  11006724  11006725  11006757  11006766  11006785  11006787  11006805  11006806  11006811  11006842  11006860  11006861  11006862  11006866  11006867  11006897  11006899  11006903  11006904  11006905  11006906  11006907  11006908  11006923  11006928  11006929  11006937  11006942  11006952  11006953  11006980  11007011  11007012  11007013  11007022  11007023  11007034  11007052  11007054  11007094  11007117  11007176  11007182  11007223  11007227  11007261  11007262  11007278  11007283  11007284  11007320  11007329  11007330  11007345  11007371  61518075  61755278  61760575  61760615  61766534  61772811  61772812  61772815  61772816  61772817  61772837  61778702  61778704  61778705  61778708  61778709  61780764  61784350  61784351  61784425  61784426  61784427  61784456  61784457  61784462  61784467  61784468  61784516  61784518  61786347  61789301  61789302  61789303  61789304  61789305  61789306  61789307  61789308  61789309  61789310  61789311  61789312  61789313  61789314  61789315  61789316  61789317  61789318  61789321  61789322  61789323  61793614  61793615  61793616  61793617  61794567  61794568  61794569  61794570  61794571  61794572  61794573  61794574  61794575  61794576  61794577  61794578  61794579  61794580  61794581  61794582  61794583  61794584  61794585  61794586  61800274  61800276  61800290  61800291  61800292  61800293  61800294  61800302  61800304  61800308  61800313  61800315  61800317  61800318  61800324  61803298  61803299  61803301  61803302  61804475  61806904  61806905  61806906  61806907  61806908  61806909  61806910  61806911  61806912  61806913  61806914  61806915  61806916  61806917  61806919  61806920  61806921  61806922  61812444  61812446  61812447  61812448  61812450  61812451  61812453  61812472  61812474  61812476  61812486  61812488  61812492  61817673  61817674  61817675  61817676  61817677  61817678  61817679  61817680  61817683  61817684  61817685  61822664  61822665  61822666  61822667  61822668  61822669  61822670  61822671  61822672  61822673  61822674  61822676  61822677  61822678  61825263  61826637  61826638  61826639  61826640  61826641  61826642  61826643  61826644  61826645  61826646  61826648  61826649  61826650  61826651  61831170  61831172  61831173  61831174  61831175  61831191  61831204  61831213  61831214  61831228  61831236  61831243  61831250  61831252  61831263  61837795  61837797  61837798  61837799  61837801  61837802  61837803  61837821  61837822  61837823  61837824  61837825  61837826  61837827  61845509  61845510  61845511  61845512  61845513  61845514  61845515  61845516  61845520  61845521  61845522  61845523  61845524  61845525  61846166  61846167  61846168  61846169  61846170  61846171  61846173  61846174  61846175  61846182  61846183  61846184  61846185  61846186  61846187  61854117  61854119  61854120  61854121  61854122  61854123  61854124  61854125  61854126  61854128  61854129  61854130  61854131  61854132  61859311  61859312  61859313  61859314  61859315  61859316  61859317  61859318  61859319  61859320  61859321  61859322  61859323  61859324  61859325  61859326  61859327  61864352  61864353  61864354  61864355  61864356  61864357  61864358  61864359  61864360  61864361  61864362  61864363  61864365  61864366  61864369  61864370  61864371  61864372  61864373  61869495  61869496  61869497  61869498  61869499  61869501  61869502  61869503  61869504  61869505  61869506  61869507  61869508  61869509  61869510  61873353  61873354  61873355  61873356  61873357  61873358  61873359  61873360  61873361  61873362  61873363  61873364  61873365  61873368  61873369  61873370  61878651  61878652  61878653  61878654  61878656  61878658  61878660  61878661  61878666  61878667  61878670  61878671  61878673  61878678  61878680  61878682  61878684  61878686  61882794  61882795  61882796  61882797  61882798  61882799  61882801  61882803  61882804  61882822  61882824  61882825  61882826  61882828  61886773  61886774  61886775  61886776  61886789  61886792  61886795  61886796  61886798  61886801  61886804  61886805  61886808  61886809  61887035  61887036  61887037  61887039  61887040  61887042  61887046  61891598  61891599  61891601  61891605  61891607  61891609  61891614  61891617  61891619  61891620  61891630  61891631  61891635  61891636  61891642  61891644  61891645  61891655  61891656  61891658  61897563  61897564  61897565  61897567  61897568  61897569  61897570  61897571  61897572  61897573  61897574  61897575  61897576  61897577  61897578  61897579  61897580  61897582  61897583  61897584  61902287  61902288  61902289  61902290  61902291  61902292  61902293  61902294  61902295  61902296  61902297  61902298  61902299  61902301  61902302  61902303  61902305  61902306  61902307  61902308  61902309  61902310  61907607  61907610  61907611  61907612  61907613  61907614  61907616  61907617  61907618  61907619  61907620  61907621  61907622  61907623  61907624  61907625  61907626  61907627  61907628  61907629  61907630  61911930  61911931  61911932  61911933  61911934  61911935  61911936  61911937  61911938  61911939  61911940  61911941  61911942  61912980  61912981  61912984  61912985  61912986  61912987  61912988  61912989  61912990  61917583  61917584  61917585  61917586  61917587  61917588  61917589  61917590  61917591  61917592  61917593  61917594  61917595  61917596  61917597  61917598  61917599  61917601  61917602  61917603  61917604  61917605  61922058  61922059  61922060  61922061  61922062  61922063  61922064  61922065  61922066  61922067  61922068  61922069  61922070  61922071  61922072  61922073  61922074  61922075  61922076  61922077  61922078  61922079  61927511  61927512  61927513  61927514  61927515  61927516  61927517  61927518  61927520  61927522  61927523  61927525  61927526  61927527  61927528  61927529  61927530  61927531  61927532  61927533  61927552  61933139  61933140  61933141  61933142  61933143  61933144  61933145  61933146  61933147  61933148  61933149  61933150  61934098  61934099  61934101  61934103  61934104  61934105  61934106  61934107  61934108  61936935  61938580  61938581  61938583  61938584  61938585  61938586  61938587  61938588  61938589  61938590  61938591  61943791  61943792  61943793  61943794  61943795  61943796  61943797  61943798  61943799  61943800  61943801  61943802  61943803  61943804  61943805  61943806  61943807  61943808  61943809  61943810  61943811  61943812  61948460  61948461  61948462  61948463  61948464  61948465  61948466  61948467  61948468  61948469  61948470  61948471  61948472  61948473  61948474  61948475  61948476  61948477  61948478  61948479  61948480  61948481  61953923  61953926  61953927  61953928  61953929  61953931  61953932  61953933  61953934  61953937  61953940  61953942  61953943  61953944  61953945  61953946  61953947  61953950  61953951  61953952  61953953  61953954  61959429  61959430  61959431  61959434  61959435  61959436  61959439  61959440  61959441  61959444  61959445  61959446  61959449  61959450  61964992  61964993  61964994  61964995  61964996  61964997  61964998  61964999  61965000  61965001  61965002  61965003  61965004  61965005  61965006  61965007  61965010  61965011  61965012  61965013  61965014  61970480  61970481  61970482  61970483  61970484  61970485  61970486  61970487  61970488  61970489  61970490  61970491  61970492  61970493  61970494  61970495  61970496  61970497  61970498  61970499  61970501  61970502  61975739  61975740  61975741  61975742  61975744  61975745  61975747  61975748  61975749  61975750  61975751  61975752  61975753  61975754  61975755  61975756  61975757  61975765  61975766  61975768  61975769  61975785  61975788  61982638  61982639  61982640  61982641  61982642  61982643  61982644  61982645  61982646  61982648  61982649  61982650  61982651  61982652  61982653  61982654  61982660  61982661  61982662  61982663  61982664  61982665  61982666  61982667  61982668  61982669  61982670  61987860  61987861  61987862  61987863  61987864  61987865  61987866  61987867  61987868  61987869  61987870  61987871  61987874  61987875  61987876  61987877  61987878  61987880  61987881  61987882  61987883  61987884  61987885  61987886  61987887  61987894  61989840  61994033  61994034  61994035  61994036  61994037  61994038  61994039  61994040  61994041  61994042  61994043  61994044  61994045  61994046  61994047  61994048  61994049  61994050  61994051  61994052  61994053  61994054  61994055  61994056  61994057  61994058  61994059  61994060  61994061  61994062  62002501  62002502  62002503  62002504  62002505  62002506  62004451  62004452  62004453  62004454  62004455  62004456  62005330  62005331  62005335  62005336  62005337  62005339  62006599  62006601  62006602  62006603  62006605  62006608  62007028  62007029  62007030  62007031  62008616  62008627  62008628  62008629  62008630  62008631  62009634  62009635  62009636  62009637  62009638  62010988  62010989  62010990  62010991  62010992  62010993  62011191  62012224  62012225  62012226  62012227  62012228  62012229  62013016  62013017  62013022  62013026  62013028  62013031  62013832  62013833  62013834  62013835  62013836  62013837  62013838  62016085  62016086  62016087  62016846  62016847  62016850  62016860  62016861  62016862  62016863  62017458  62017467  62017490  62017491  62017492  62017493  62017494  62018567  62018568  62018569  62018571  62018572  62018573  62019766  62019767  62019768  62019769  62019770  62021786  62021788  62021789  62021812  62021823  62021824  62023222  62023223  62023224  62023225  62023226  62023227  62024224  62024225  62024229  62024231  62024234  62024236  62025282  62025283  62025285  62025286  62025287  62025289  62025290  62026987  62026988  62026989  62026990  62026991  62026992  62026993  62027656  62027658  62027660  62027661  62027662  62027663  62027666  62027667  62027668  62027669  62027670  62027671  62027672  62027673  62027674  62027675  62027676  62027677  62027678  62027679  62027680  62027681  62027682  62027684  62027685  62027687  62027689  62027691  62027692  62027694  62027695  62027696  62029618  62029619  62029620  62029621  62029622  62029623  62029624  62029625  62029626  62029627  62029628  62029631  62029632  62029633  62029634  62029635  62029636  62029637  62029638  62029644  62029646  62029647  62029648  62029649  62029650  62029651  62029652  62039858  62039859  62039860  62039862  62039863  62039864  62039865  62041390  62041391  62041405  62041406  62041407  62041408  62041409  62042547  62042548  62042549  62042550  62042551  62042552  62042553  62043787  62043788  62043789  62043790  62043791  62043794  62043795  62045277  62045278  62045279  62045280  62045281  62045282  62045283  62046892  62046893  62046894  62046895  62046896  62046897  62046899  62048260  62048262  62048263  62048264  62048266  62048267  62048268  62049266  62049271  62049272  62049273  62049275  62050808  62050809  62050810  62050811  62050812  62050813  62052226  62052227  62052228  62052229  62052231  62052232  62052233  62053515  62053518  62053520  62053521  62053524  62053526  62053528  62054726  62054727  62054731  62054732  62054733  62054734  62054735  62055187  62055188  62055189  62055190  62055191  62056281  62056282  62056283  62056284  62056285  62056367  62056368  62056369  62060829  62060831  62060833  62060834  62060838  62063121  62063122  62063123  62064573  62064578  62064580  62064581  62064583  62066054  62066065  62066078  62066080  62066081  62067042  62067075  62067079  62067080  62067081  62067809  62067810  62067811  62067812  62069167  62069169  62069170  62069171  62069172  62070153  62070155  62070156  62070157  62070159  62071471  62071481  62071482  62071484  62072453  62072454  62072455  62072458  62073968  62074013  62074014  62074015  62074017  62076025  62076035  62076037  62076038  62076664  62076665  62076666  62076667  62076668  62077967  62077973  62077975  62077984  62077986  62077988  62079186  62079191  62079193  62079195  62079196  62079198  62080386  62080391  62080392  62080393  62080394  62080395  62081452  62081454  62081455  62081456  62081457  62082460  62082462  62082463  62082464  62082465  62083954  62083957  62083958  62083959  62084018  62084020  62084021  62084022  62084023  62084024  62084902  62084903  62084904  62084905  62084906  62084907  62086260  62086261  62086262  62086263  62086264  62086265  62087814  62087818  62087819  62087820  62087821  62088244  62088245  62088246  62088247  62088248  62089273  62089286  62089287  62089288  62089289  62089290  62089869  62089870  62089871  62091787  62091788  62091789  62091790  62091791  62091792  62092924  62092929  62092930  62092931  62092932  62094795  62094796  62094797  62094798  62098196  62098201  62098202  62098203  62099521  62099522  62099523  62099524  62099525  62100745  62100746  62100747  62100748  62100749  62100750  62102278  62102279  62102280  62102281  62102282  62102283  62103397  62103398  62103399  62103401  62103402  62104922  62104923  62104924  62104925  62104926  62105915  62105916  62105917  62105919  62107224  62107227  62107228  62107230  62107231  62107234  62108801  62108809  62108810  62108811  62108812  62108813  62110062  62110063  62110064  62110065  62110066  62110067  62111510  62111511  62111512  62111513  62111514  62111515  62112094  62112095  62112096  62112097  62112098  62112103  62113619  62113620  62113621  62113622  62113623  62113624  62114886  62114887  62114888  62114889  62114890  62114891  62116650  62116651  62116652  62116653  62116655  62118049  62118050  62118051  62118052  62118053  62118677  62118678  62118679  62118680  62118682  62125569  62125576  62125577  62125579  62125581  62126474  62126475  62126476  62126477  62126478  62126479  62127209  62127210  62127211  62127212  62127213  62127214  62129309  62129310  62129311  62129312  62129313  62130498  62130499  62130501  62130505  62130506  62130507  62131507  62131508  62131509  62131510  62131511  62131512  62131514  62131515  62131516  62131517  62137144  62142191  62142199  62142201  62142202  62143065  62143068  62143070  62143071  62143072  62143923  62143927  62143928  62143929  62143930  62143931  62144950  62144953  62144954  62144955  62144956  62145943  62145946  62145948  62145950  62145951  62147883  62147887  62147888  62147889  62147891  62147892  62147893  62148616  62148629  62148631  62148632  62148633  62149391  62149392  62149393  62149394  62149395  62149396  62150720  62150758  62150759  62150760  62150761  62151573  62151574  62151575  62151576  62151577  62153080  62153081  62153082  62153083  62153084  62153085  62154415  62154416  62154417  62154418  62154419  62154420  62155317  62155318  62155320  62155321  62156238  62156242  62156243  62156244  62156245  62156964  62156965  62156966  62156967  62158307  62158310  62158311  62158313  62158314  62159847  62159849  62159850  62159851  62159854  62160814  62160817  62160818  62160820  62161662  62161663  62162359  62164085  62164088  77001577  77001578  77001579  77001580  77001582  77001584  77001619",
      "more_code_info": ""
    }
  ]
}