  "results": [
    {
      "status": "Ongoing",
      "city": "Cincinnati",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89666",
      "recalling_firm": "Meridian Bioscience Inc",
      "address_1": "3471 River Hills Dr",
      "address_2": "N/A",
      "postal_code": "45244-3023",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, ME, MI, MN, MO, MS, MT, NC, NE, NJ, NV, NY, OH, OR, PA, PR, RI, SD, TN, TX, UT, VA, VT, WI, WV and the countries of Australia, Chile, Italy, Switzerland.",
      "recall_number": "Z-0836-2022",
      "product_description": "Revogene, For In Vitro Diagnostic use in performing nucleic acid testing in clinical laboratories.   Meridian branded Revogene: Catalog Number 610210  GenePOC branded Revogene: Catalog Number 130840",
      "product_quantity": "452 units (398 US, 54 OUS)",
      "reason_for_recall": "The photomultiplier tube, may have compromised functionality in some instruments. As a result, the user is likely to observe an increase in false-positive or indeterminate results.",
      "recall_initiation_date": "20220215",
      "center_classification_date": "20220330",
      "report_date": "20220406",
      "code_info": "UDI: 00840733102318 All units in the field, all serial numbers."
    }
  ]
}