  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "West Chester",
      "state": "PA",
      "country": "United States",
      "classification": "Class III",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "64307",
      "recalling_firm": "Animas Corporation",
      "address_1": "200 Lawrence Dr",
      "address_2": "N/A",
      "postal_code": "19380-3428",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) including Washington, DC, Puerto Rico, US Virgin Island and the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY, and the  countries of Austria, Australia, Bermuda, Canada, New Zealand, Czech Republic, Italy, France, Finland, Hungary, Israel, United Kingdom, Spain, Sweden, Norway, Denmark, Germany, Mexico and Ireland.",
      "recall_number": "Z-0836-2013",
      "product_description": "Animas(R) 2020, Animas(R) IR 1250 (1200 plus), Animas(R) IR 1200 Insulin Infusion Pumps  Indicated for continuous subcutaneous infusion of insulin.",
      "product_quantity": "37,998",
      "reason_for_recall": "Due to an internal calendar date limit (December 31, 2015) these pumps will cease to operate on January 1, 2016, and will result in a call service alarm that is displayed on the pump display.",
      "recall_initiation_date": "20121217",
      "center_classification_date": "20130217",
      "termination_date": "20131224",
      "report_date": "20130227",
      "code_info": "All Animas 2020, IR1250 and IR1200 Pump Models - Part Numbers 101370-00, 101371-00, 101372-00, 101374-00, 101375-00, 100380-00, 100381-00, 100382-00, 100384-00, 100385-00, 100450-00, 100210-00, 100211-00, 100212-00, 100250-00, 100251-00, 100252-00, 100200-00, 100201-00, 100202-00, 100170-00, 100171-00 and 100172-00."
    }
  ]
}