{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Indianapolis",
      "address_1": "7736 Zionsville Rd",
      "reason_for_recall": "One brand of batteries have a small dimensional difference in the negative button terminal, which, when the batteries are inserted incorrectly, can create the possibility for the battery terminals to make reverse contact.  This reverse contact can cause a potential for overheating.",
      "address_2": "",
      "product_quantity": "44,448 total",
      "code_info": "All Serial Numbers shipped during the 11/1/2016 to 10/25/2019 timeframe.  UDI: 00304040037713",
      "center_classification_date": "20200121",
      "distribution_pattern": "The products were distributed US Nationwide.    The products were distributed to the following foreign countries:  Australia, Austria, Belgium, Bermuda, Canada, China, Chile, Costa Rica, Cayman Islands, Czech Republic, England, Finland, France, Germany, Greece, Hong Kong, India, Iraq, Italy, Jordan, Lebanon, Lithuania, Mexico, Morocco, Panama, Philippines, Poland, Qatar, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Trinidad, UAE.",
      "state": "IN",
      "product_description": "Henry Schein CardioChek Plus Professional Analyzer, New Analyzer, boxed, with carrying case, Reference Number 4060 HS, 570-0411",
      "report_date": "20200129",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Polymer Technology Systems, Inc.",
      "recall_number": "Z-0835-2020",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "84274",
      "termination_date": "20200706",
      "more_code_info": "",
      "recall_initiation_date": "20191202",
      "postal_code": "46268-2175",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}