      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73142",
      "recalling_firm": "Zimmer Biomet, Inc.",
      "address_1": "1800 W Center St",
      "address_2": "N/A",
      "postal_code": "46580-2304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US (nationwide) and Internationally to   ARGENTINA               AUSTRALIA  AUSTRIA  BARBADOS  BELGIUM  BERMUDA  BOLIVIA  BRAZIL  CARIBBEAN  CHILE  CHINA  COLOMBIA  COST RICA  DENMARK  DOMINICAN REPUBLIC  DORAL FL  ECUADOR  EL SALVADOR  ENGLAND  FRANCE  GERMANY  GUATEMALA  HONDURAS  HONG KONG  INDIA  ISRAEL  ITALY  JAMAICA  JAPAN  KOREA  MALAYSIA  MEXICO  NEW ZEALAND  NICARAGUA  PANAMA  PERU  SINGAPORE  SPAIN  SWEDEN  SWITZERLAND  TAIWAN  THAILAND  VENEZUELA  WEST INDIES  Canary Islands  Croatia  Cyprus  Czech Republic  Canada",
      "recall_number": "Z-0835-2016",
      "product_description": "NexGen Knee CR Option Fem  NexGen Knee CR Porous Fem  NexGen Knee CR Precoat Fem  NexGen Knee LPS Porous Fem  NexGen Knee LPS Precoat Fem  prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer/metal/polymer  various sizes  This device is indicated for patients with severe knee pain and disability due to:   Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis.   Collagen disorders, and/or avascular necrosis of the femoral condyle.   Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy.   Moderate valgus, varus, or flexion deformities.   The salvage of previously failed surgical attempts or for a knee in which satisfactory stability in flexion cannot be obtained at the time of surgery. \" CR and LPS porous coated femoral and tibial baseplate components may be used cemented or uncemented (biological fixation). The CR Hydroxyapatite/tricalcium phosphate [HA/TCP] coated femoral or tibial baseplate components may only be used uncemented. All other femoral, tibial baseplate and all-polyethylene patella components are indicated for cemented use only.\"",
      "product_quantity": "32508",
      "reason_for_recall": "LDPE bag containing the implant adheres to the highly polished implant surface.",
      "recall_initiation_date": "20160111",
      "center_classification_date": "20160222",
      "termination_date": "20170523",
      "report_date": "20160302",
      "code_info": "Item numbers:  00-5966-013-01  00-5966-013-02  00-5966-014-01  00-5966-014-02  00-5966-015-01  00-5966-015-02  00-5966-016-01  00-5966-016-02  00-5966-017-01  00-5966-017-02  00-5970-011-01  00-5970-011-02  00-5970-012-01  00-5970-012-02  00-5970-013-01  00-5970-013-02  00-5970-013-03  00-5970-013-04  00-5970-014-01  00-5970-014-02  00-5970-014-03  00-5970-014-04  00-5970-015-01  00-5970-015-02  00-5970-015-03  00-5970-015-04  00-5970-016-01  00-5970-016-02  00-5970-017-01  00-5970-017-02  00-5970-018-01  00-5970-018-02  00-5972-011-01  00-5972-011-02  00-5972-012-01  00-5972-012-02  00-5972-013-01  00-5972-013-02  00-5972-013-03  00-5972-013-04  00-5972-014-01  00-5972-014-02  00-5972-014-03  00-5972-014-04  00-5972-015-01  00-5972-015-02  00-5972-015-03  00-5972-015-04  00-5972-016-01  00-5972-016-02  00-5972-017-01  00-5972-017-02  00-5972-018-01  00-5972-018-02  00-5990-011-01  00-5990-011-02  00-5990-011-31  00-5990-011-32  00-5990-012-01  00-5990-012-02  00-5990-012-31  00-5990-012-32  00-5990-013-01  00-5990-013-02  00-5990-013-31  00-5990-013-32  00-5990-014-01  00-5990-014-02  00-5990-014-31  00-5990-014-32  00-5990-015-01  00-5990-015-02  00-5990-015-31  00-5990-015-32  00-5990-016-01  00-5990-016-02  00-5990-016-31  00-5990-016-32  00-5990-017-01  00-5990-017-02  00-5990-017-31  00-5990-017-32  00-5990-017-91  00-5990-017-92  00-5990-018-01  00-5990-018-02  00-5990-018-31  00-5990-018-32  00-5992-012-01  00-5992-012-02  00-5992-012-31  00-5992-012-32  00-5992-013-01  00-5992-013-02  00-5992-013-31  00-5992-013-32  00-5992-014-01  00-5992-014-02  00-5992-014-31  00-5992-014-32  00-5992-015-01  00-5992-015-02  00-5992-015-31  00-5992-015-32  00-5992-016-01  00-5992-016-02  00-5992-016-31  00-5992-016-32  00-5992-017-01  00-5992-017-02  00-5992-017-31  00-5992-017-32  00-5992-018-01  00-5992-018-02  00-5992-018-31  00-5992-018-32    Lot numbers:  11002771  11003380  11003415  11003450  11003451  11003469  11003477  11003574  11003575  11003576  11003583  11003586  11003593  11003597  11003601  11003602  11003612  11003613  11003626  11003643  11003644  11003666  11003673  11003825  11003832  11003870  11003871  11003889  11003890  11003893  11003941  11003942  11003954  11003956  11004035  11004043  11004058  11004080  11004101  11004189  11004302  11004500  11004508  11004543  11004566  11004597  11004683  11004766  11004806  11004817  11004868  11004869  11004870  11004894  11004914  11004946  11004950  11004952  11004954  11004955  11004957  11004958  11004959  11004971  11004972  11004980  11005018  11005034  11005054  11005055  11005057  11005068  11005096  11005099  11005100  11005112  11005232  11005233  11005248  11005320  11005328  11005367  11005374  11005425  11005444  11005461  11005489  11005491  11005511  11005530  11005561  11005594  11005602  11005629  11005632  11005647  11005674  11005677  11005678  11005685  11005697  11005703  11005707  11005732  11005747  11005752  11005764  11005795  11005796  11005820  11005891  11005893  11005911  11005912  11005922  11005966  11006019  11006035  11006055  11006059  11006084  11006113  11006116  11006160  11006194  11006202  11006245  11006248  11006249  11006262  11006316  11006319  11006332  11006362  11006365  11006403  11006458  11006471  11006538  11006588  11006598  11006600  11006604  11006605  11006607  11006631  11006640  11006670  11006695  11006718  11006719  11006728  11006736  11006738  11006740  11006773  11006792  11006810  11006812  11006828  11006877  11006926  11006962  11006963  11007031  11007045  11007047  11007058  11007063  11007070  11007072  11007078  11007080  11007097  11007112  11007160  11007161  11007163  11007172  11007179  11007286  11007296  11007324  11400177  11400178  11400200  11400224  11400230  11400233  11400234  11400249  11400289  11400296  11400459  11400473  61750273  61758922  61768186  61771840  61771842  61776962  61776964  61776966  61777574  61777575  61777578  61777590  61777594  61777595  61777597  61782540  61782541  61782542  61782543  61782544  61782545  61782585  61782586  61782587  61782588  61782659  61782662  61782669  61782670  61782671  61782672  61782673  61782674  61782675  61782676  61782677  61782678  61782679  61782680  61782833  61782834  61782835  61782836  61782837  61782838  61788717  61788718  61788719  61788720  61788721  61788722  61788723  61788724  61788726  61788727  61788728  61788729  61788737  61788740  61788745  61788746  61788747  61788748  61788750  61788751  61788752  61789025  61789026  61789028  61789029  61789030  61789031  61789032  61789444  61789445  61789446  61789447  61789448  61789449  61795143  61795144  61795145  61795791  61795792  61795793  61795794  61795795  61795796  61795797  61795798  61795807  61795808  61795809  61795814  61795815  61795816  61795817  61795818  61795819  61795820  61795821  61795826  61795827  61795828  61795829  61795830  61795833  61795834  61795835  61795836  61795837  61795921  61795922  61795923  61795924  61795925  61799903  61799904  61799905  61800641  61800642  61800643  61800644  61800645  61800646  61800647  61800648  61800717  61800718  61800719  61800720  61800721  61800722  61800723  61800724  61800725  61800726  61800727  61800728  61800736  61800737  61800738  61800739  61800740  61800741  61800742  61800743  61800745  61800746  61800747  61800748  61801091  61801093  61801094  61801096  61801097  61805238  61805239  61805240  61805241  61805242  61805243  61805526  61805527  61805528  61805529  61805530  61805531  61805532  61805533  61805534  61805535  61805536  61805537  61805538  61805539  61805540  61805541  61805542  61805543  61805544  61805545  61805546  61805547  61805548  61805549  61805550  61805551  61805658  61805659  61805660  61805661  61805671  61805672  61805673  61810029  61810030  61810031  61810032  61810123  61810124  61810125  61810126  61810127  61810128  61810131  61810132  61810133  61810134  61810135  61810136  61810137  61810138  61810139  61810140  61810141  61810143  61810144  61810145  61810146  61810147  61810149  61810150  61810151  61810152  61810153  61810154  61810155  61810156  61810157  61810213  61810214  61810215  61810216  61810218  61815445  61815446  61815447  61815449  61815450  61815451  61815453  61815454  61815455  61815456  61815457  61815458  61815459  61815460  61815461  61815462  61815463  61815464  61815465  61815466  61815472  61815473  61815474  61815511  61815512  61815513  61815514  61822056  61822057  61822058  61822292  61822293  61822296  61822298  61822299  61822301  61822302  61822338  61822339  61822340  61822341  61822342  61822343  61822344  61822345  61822356  61822357  61824009  61824010  61824011  61824518  61824519  61824520  61824521  61824522  61824578  61824579  61824580  61824581  61824582  61824583  61824588  61824589  61824590  61824591  61824592  61824593  61824594  61830484  61830626  61830627  61831582  61831583  61831585  61831586  61831587  61831603  61831604  61831605  61831606  61831607  61831608  61831651  61831653  61831655  61831656  61831657  61831658  61836498  61836499  61836501  61836502  61836504  61836528  61836529  61836530  61836531  61836567  61836568  61836569  61836570  61836628  61836629  61836630  61836631  61846049  61846050  61846051  61846052  61846123  61846124  61846125  61846126  61846127  61846150  61846151  61846152  61846153  61846154  61846155  61846156  61846157  61846158  61846160  61846161  61846162  61846163  61846176  61846177  61846178  61846179  61846180  61846181  61848399  61848485  61848486  61848487  61848488  61848489  61848521  61848522  61848523  61848524  61848525  61848526  61848527  61848528  61848529  61848530  61848531  61848532  61848533  61848534  61848535  61848536  61848537  61848538  61848539  61848540  61850923  61850927  61851721  61853646  61855382  61855383  61855384  61855501  61855502  61855503  61855529  61855530  61855531  61855532  61855533  61855534  61855535  61855536  61855537  61855538  61855539  61855540  61855541  61855542  61855543  61855544  61855545  61855546  61855806  61855961  61857319  61857320  61857321  61857322  61857323  61857325  61857326  61857328  61857329  61857352  61857354  61857355  61857356  61857357  61857390  61857391  61857690  61857691  61857692  61860218  61861350  61861359  61861360  61861699  61861701  61861702  61861726  61861727  61861728  61861729  61861730  61861731  61861733  61862136  61862137  61862138  61862170  61862171  61862172  61862173  61862174  61862175  61862176  61862177  61862178  61863945  61864875  61864880  61866042  61866043  61866044  61866627  61866628  61866712  61868866  61869848  61870729  61870730  61870731  61870903  61870907  61870908  61870910  61870911  61870912  61870914  61871734  61871735  61871861  61871862  61871863  61871864  61871865  61871866  61871867  61874460  61875443  61875444  61875445  61875928  61875929  61875930  61876140  61876141  61876155  61876156  61876157  61876158  61878276  61879230  61879232  61881046  61881667  61881668  61881669  61881670  61881673  61883267  61884185  61884186  61884188  61884829  61885035  61885036  61885037  61885038  61885213  61885214  61885215  61885216  61885217  61885218  61885219  61885220  61885221  61886269  61886304  61886305  61886324  61886336  61886338  61888038  61888039  61888040  61889119  61889222  61889223  61889224  61889225  61889226  61889227  61889228  61889229  61889230  61889231  61889232  61890022  61890028  61890516  61890693  61890695  61890756  61890764  61890766  61892368  61892370  61892372  61893618  61893620  61893623  61893624  61893725  61893726  61893727  61893728  61893730  61893731  61893732  61893784  61893785  61893786  61893787  61893788  61893789  61893933  61893934  61893936  61893937  61893940  61893944  61893947  61893957  61893963  61893964  61893966  61893974  61894404  61894405  61894406  61894407  61894408  61894409  61894410  61894414  61894415  61894416  61894417  61896064  61896065  61899325  61899470  61899471  61899472  61899473  61899474  61899475  61899476  61899477  61899478  61899479  61899480  61899481  61899482  61899493  61899494  61899495  61899496  61901220  61901221  61902948  61902950  61904098  61904099  61904101  61904102  61904117  61904124  61904125  61904126  61904192  61907443  61907444  61908305  61909336  61909337  61909338  61909341  61909342  61910130  61910508  61912160  61912162  61912163  61912165  61912166  61912167  61912169  61912170  61912177  61912178  61912179  61912180  61912181  61912182  61912183  61912184  61912185  61912186  61912187  61912188  61912189  61912190  61912191  61912192  61912260  61912261  61912262  61912263  61912264  61912265  61912266  61912267  61912268  61912269  61912270  61912271  61912272  61912273  61912274  61912275  61912276  61912277  61912278  61912279  61912280  61912281  61912282  61912283  61912352  61912353  61912354  61912548  61912549  61912550  61912551  61913187  61913195  61915853  61915854  61915975  61915976  61915977  61915978  61915981  61915982  61915983  61915984  61915985  61915986  61915987  61915988  61915989  61915990  61915991  61915992  61915993  61915994  61915995  61915996  61915997  61915998  61915999  61916001  61916002  61916003  61916004  61916005  61916010  61916011  61921141  61921142  61921284  61921285  61921286  61921287  61921290  61921291  61921292  61921293  61921294  61921295  61921296  61921297  61921298  61921299  61921301  61921302  61921303  61921304  61921305  61921306  61921307  61921308  61921309  61921310  61921311  61921312  61923572  61923576  61924452  61927503  61927504  61927818  61927819  61927820  61927821  61927822  61927823  61927824  61928560  61928561  61928562  61928563  61928564  61928565  61928566  61930157  61932688  61932689  61933315  61933316  61933317  61933318  61933319  61933321  61933322  61933323  61933324  61933325  61933326  61933327  61933328  61933329  61933807  61933833  61935981  61936667  61936691  61936692  61936693  61936694  61936695  61936696  61936699  61937795  61937796  61937798  61937801  61937802  61937803  61938536  61939183  61939710  61939711  61939712  61939713  61939714  61939715  61939716  61940001  61942839  61942840  61943605  61944766  61944888  61946983  61948591  61948592  61948706  61948707  61948710  61948711  61949202  61949203  61949204  61949205  61949206  61949207  61949208  61949209  61949210  61949211  61949340  61949341  61949543  61949544  61949545  61949546  61949547  61949548  61949549  61949550  61951404  61951405  61951544  61951545  61951546  61951547  61952590  61952591  61952592  61952593  61952594  61952595  61952596  61952597  61952598  61954297  61954298  61954299  61954301  61954302  61954303  61954304  61955894  61955895  61955896  61955897  61955898  61958204  61958205  61958206  61958427  61958428  61958429  61959154  61959155  61959165  61959166  61959167  61959168  61959169  61959170  61959171  61959172  61962617  61962618  61963770  61963771  61963772  61963773  61963774  61963775  61963776  61963777  61963778  61963781  61963782  61963783  61963784  61963785  61963787  61963788  61963789  61963790  61966301  61967967  61968155  61968156  61968157  61968158  61968159  61968160  61968161  61968162  61968164  61968165  61968166  61968167  61968168  61968169  61968170  61968334  61968335  61968336  61968337  61968338  61968339  61968875  61968876  61968877  61968878  61974742  61974743  61974864  61974868  61974871  61974874  61975156  61975157  61975158  61975159  61975160  61975161  61975169  61975170  61975171  61975172  61975189  61975193  61975207  61975208  61975209  61975211  61975213  61983389  61983390  61983391  61983411  61983412  61983414  61983415  61983416  61983417  61983422  61983424  61983425  61983427  61983428  61983429  61983430  61983431  61983432  61983433  61983434  61983435  61987842  61987843  61987844  61987845  61987846  61987847  61987848  61987849  61987850  61987851  61987852  61987853  61987854  61987855  61987856  61987857  61987858  61987859  61987915  61987916  61987917  61994386  61994387  61994862  61994863  61994864  61994865  61994867  61994868  61994871  61994874  61994875  61994880  61994910  61994911  61994912  62000748  62000749  62001280  62001282  62001283  62001284  62001285  62001286  62001288  62001289  62001312  62001313  62001314  62001315  62001316  62001317  62001318  62001319  62001320  62001379  62001380  62001381  62002485  62002486  62005486  62005487  62005596  62005597  62005598  62005599  62005600  62005601  62005602  62005603  62005604  62005605  62005606  62005607  62005608  62005609  62005610  62005611  62005714  62005715  62005716  62005717  62005718  62012463  62012464  62013324  62013325  62013326  62013328  62013329  62013330  62013331  62013332  62013333  62013334  62013335  62013336  62013337  62013338  62013339  62013340  62013341  62013395  62013396  62013398  62013399  62019250  62019251  62019252  62019253  62019254  62019255  62020210  62020211  62020212  62020279  62020280  62020281  62020282  62020283  62020284  62020285  62020286  62020287  62020288  62020289  62020290  62020291  62020292  62020293  62020294  62020295  62020296  62020297  62020298  62020299  62025539  62025540  62025541  62025542  62025543  62025616  62025617  62025620  62025621  62025622  62025623  62025624  62025625  62025626  62025627  62025629  62025630  62025631  62025632  62025633  62025634  62025635  62025636  62025662  62025665  62025666  62031845  62031846  62031847  62032484  62032485  62032486  62032487  62032488  62032489  62032509  62032510  62032511  62032512  62032513  62032514  62032516  62032517  62032570  62032571  62032572  62032573  62038287  62038288  62038838  62038839  62038840  62038841  62041099  62041100  62041101  62041102  62041103  62041104  62041105  62041106  62041107  62041108  62041109  62041110  62041111  62041112  62041113  62045621  62045622  62045624  62045625  62045626  62045627  62045628  62045629  62045630  62045631  62045632  62045633  62045634  62045635  62045636  62045637  62045638  62045639  62045640  62046096  62046097  62046098  62047526  62054455  62054522  62054526  62054527  62054528  62054529  62054530  62054531  62054532  62054533  62054534  62054535  62054536  62054537  62054538  62054568  62054569  62056354  62056357  62056358  62056360  62056361  62056362  62056363  62056364  62056365  62056366  62056371  62056372  62056375  62056376  62056377  62056378  62056379  62056380  62056381  62056382  62056383  62056521  62056522  62056523  62056524  62056525  62056851  62056854  62056855  62061978  62061979  62061980  62062051  62062052  62062053  62062082  62062083  62062084  62062085  62062086  62062087  62062088  62062089  62062090  62062091  62062092  62062093  62062094  62062095  62062096  62068042  62068043  62068046  62068047  62068048  62068049  62068050  62068051  62068052  62068053  62068054  62068055  62068056  62068057  62068058  62068059  62068060  62068061  62068062  62068063  62068064  62068065  62068066  62068156  62068157  62068158  62074747  62074748  62074838  62074839  62074840  62074841  62074842  62074843  62074844  62074845  62074846  62074847  62074848  62074849  62074850  62074852  62074853  62074854  62074855  62074856  62074857  62074881  62074882  62074883  62074884  62080597  62080598  62080599  62080623  62080624  62080625  62080627  62080628  62080629  62080631  62080632  62080633  62080634  62080636  62080638  62080639  62080640  62080641  62080642  62086047  62086056  62086063  62086064  62086065  62086066  62086635  62086744  62086745  62086746  62086802  62086840  62086841  62086842  62086846  62086847  62086848  62086849  62086850  62086851  62086852  62086853  62086854  62086855  62086856  62086857  62086858  62086859  62087257  62087258  62087529  62087532  62087585  62087597  62092286  62092341  62092342  62092403  62092404  62092405  62092406  62092407  62092408  62092409  62092410  62092411  62092412  62092414  62092415  62092425  62092426  62092436  62099045  62103515  62103516  62103517  62103518  62103519  62103520  62104066  62104067  62104068  62104069  62104093  62104094  62104095  62104096  62104097  62104098  62104099  62104101  62104102  62104103  62104104  62104105  62104106  62104107  62104153  62104154  62104808  62104811  62104813  62104814  62104816  62104853  62105168  62105205  62105206  62105207  62105208  62105211  62105213  62105216  62105220  62105221  62105222  62105230  62105373  62106699  62107203  62110545  62110546  62110547  62110553  62110554  62110555  62110556  62110592  62110646  62110647  62110648  62110649  62110650  62110651  62110652  62110653  62110654  62110655  62110656  62110657  62111364  62115657  62115660  62117011  62117012  62117013  62117014  62117092  62117093  62117094  62117095  62117096  62117097  62117098  62117099  62117101  62117102  62117103  62117113  62117114  62117284  62117285  62117330  62118584  62121776  62122744  62122745  62122786  62122788  62122789  62122793  62123042  62123057  62123887  62123888  62127116  62127117  62127442  62127446  62127473  62128107  62128108  62128109  62128110  62128111  62128112  62128114  62128115  62128116  62128117  62128118  62128119  62128120  62128171  62128172  62128362  62128364  62128365  62128378  62129223  62129224  62131028  62131253  62132152  62132153  62132154  62132257  62132258  62132259  62132394  62132395  62132396  62132397  62132398  62132401  62132402  62132403  62132404  62132405  62132406  62132407  62132408  62132409  62132410  62132411  62132412  62133159  62133829  62135901  62135902  62135903  62135904  62135905  62135907  62135910  62135911  62135912  62135913  62135914  62135916  62135919  62135922  62135923  62135924  62135925  62135926  62135927  62135928  62135929  62135933  62135934  62135935  62135937  62135939  62135942  62135943  62135944  62135945  62135946  62135947  62137538  62138334  62138942  62138943  62139461  62139463  62139464  62139465  62139466  62139467  62139996  62140116  62140751  62140752  62140755  62140757  62140758  62140759  62140760  62140761  62140762  62140763  62141725  62144706  62144707  62144835  62144836  62144838  62144839  62144840  62144841  62144842  62144843  62144852  62144860  62145124  62150461  62150462  62150463  62150464  62150465  62150466  62150467  62150468  62150469  62150470  62150471  62150472  62150692  62150695  62150696  62150705  62150706  62150709  62150710  62150711  62152295  62155160  62155161  62155162  62155163  62155164  62155165  62155167  62155169  62155172  62155176  62155177  62156940  62156941  62160282  62160499  62160678  62160694  62160695  62160697  77000861  77000862  77000863  77000864  77000865  77000866  77000867  77000868  77000869  77001050  77001051  77001052  77001053  77001054  77001055  77001056  77001057  77001058  77001088  77001109  77001110  77001111  77001112  77001326  77001327  77001328  77001329  77001330  77001331  77001332  77001333  77001334  77001335  77001336  77001337  77001338  77001367  61810124R  61824009R  61855541R  61862177R  62001286R  62068049R",
      "more_code_info": ""
    }
  ]
}