{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73142",
      "recalling_firm": "Zimmer Biomet, Inc.",
      "address_1": "1800 W Center St",
      "address_2": "N/A",
      "postal_code": "46580-2304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US (nationwide) and Internationally to   ARGENTINA               AUSTRALIA  AUSTRIA  BARBADOS  BELGIUM  BERMUDA  BOLIVIA  BRAZIL  CARIBBEAN  CHILE  CHINA  COLOMBIA  COST RICA  DENMARK  DOMINICAN REPUBLIC  DORAL FL  ECUADOR  EL SALVADOR  ENGLAND  FRANCE  GERMANY  GUATEMALA  HONDURAS  HONG KONG  INDIA  ISRAEL  ITALY  JAMAICA  JAPAN  KOREA  MALAYSIA  MEXICO  NEW ZEALAND  NICARAGUA  PANAMA  PERU  SINGAPORE  SPAIN  SWEDEN  SWITZERLAND  TAIWAN  THAILAND  VENEZUELA  WEST INDIES  Canary Islands  Croatia  Cyprus  Czech Republic  Canada",
      "recall_number": "Z-0832-2016",
      "product_description": "NexGen LPS-FLEX Option Femoral  NexGen LPS-FLEX Precoat Femoral  prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer/metal/polymer  various sizes  Product Usage\t\"\" This device is indicated for patients with severe knee pain and disability due to:   Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis.   Collagen disorders, and/or avascular necrosis of the femoral condyle.   Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy.   Moderate valgus, varus, or flexion deformities.   The salvage of previously failed surgical attempts or for a knee in which satisfactory stability in flexion cannot be obtained at the time of surgery. \" CR-Flex and LPS-Flex porous coated femoral and tibial baseplate components may be used cemented or uncemented (biological fixation). CR-Flex HA/TCP porous coated femoral and tibial baseplate components may only be used uncemented. All other femoral, tibial baseplate and all-polyethylene patella components are indicated for cemented use only. \" Specific uses with CR-Flex or LPS-Flex femorals:   Provides increased flexion capability for patients who have both the flexibility and desire to increase their flexion range.   The CR-Flex femoral, when used with 90-prefix NexGen CR articular surfaces, 00-prefix 10, 12 or 14 mm or 90-prefix 17 or 20 mm Prolong¿ Highly Crosslinked Polyethylene CR articular surfaces, is designed for use with a functional posterior cruciate ligament and when load bearing range of motion (ROM) is expected to be less than or equal to 155 degrees.   The LPS-Flex femoral, when used with the LPS-Flex articular surfaces, is designed for use with both cruciate ligaments excised and when load bearing ROM is expected to be less than or equal to 155 degrees. \" Specific Uses with NexGen LPS or LCCK femorals:   The LPS femoral is designed for use with both cruciate ligaments excised and when load bearing ROM is expected to be less than or equal to 130 degrees.   The usage of the LCCK femoral is the same as the LPS with the exception that femoral bone loss can be accommodated by femoral stem extensions and augmentation. No varus/valgus constraint or other stability other than that provided by the NexGen LPS is provided when the LCCK femoral component is used with the LPS-Flex articular surfaces.\"",
      "product_quantity": "117359",
      "reason_for_recall": "LDPE bag containing the implant adheres to the highly polished implant surface.",
      "recall_initiation_date": "20160111",
      "center_classification_date": "20160222",
      "termination_date": "20170523",
      "report_date": "20160302",
      "code_info": "Item numbers:  00-5960-011-31  00-5960-011-32  00-5960-012-31  00-5960-012-32  00-5960-013-31  00-5960-013-32  00-5960-014-31  00-5960-014-32  00-5960-015-31  00-5960-015-32  00-5960-016-31  00-5960-016-32  00-5960-017-31  00-5960-017-32  00-5964-012-01  00-5964-012-02  00-5964-013-01  00-5964-013-02  00-5964-014-01  00-5964-014-02  00-5964-015-01  00-5964-015-02  00-5964-016-01  00-5964-016-02  00-5964-017-01  00-5964-017-02    Lot numbers:  11003819  11003822  11003824  11003955  11003973  11003997  11004030  11004042  11004076  11004086  11004103  11004104  11004105  11004160  11004166  11004173  11004174  11004188  11004190  11004223  11004252  11004255  11004271  11004280  11004300  11004336  11004373  11004377  11004469  11004503  11004546  11004552  11004559  11004565  11004598  11004628  11004629  11004630  11004654  11004675  11004680  11004682  11004706  11004716  11004721  11004722  11004723  11004755  11004756  11004759  11004767  11004768  11004793  11004795  11004821  11004871  11004899  11004901  11004902  11004947  11004949  11004973  11005024  11005026  11005029  11005058  11005063  11005064  11005085  11005098  11005102  11005114  11005123  11005133  11005146  11005154  11005198  11005205  11005213  11005223  11005241  11005271  11005278  11005319  11005321  11005331  11005335  11005336  11005340  11005344  11005377  11005378  11005379  11005382  11005438  11005445  11005477  11005488  11005556  11005562  11005572  11005575  11005578  11005580  11005592  11005596  11005614  11005631  11005664  11005665  11005671  11005679  11005681  11005689  11005706  11005714  11005716  11005717  11005726  11005730  11005733  11005739  11005751  11005754  11005765  11005770  11005778  11005780  11005784  11005785  11005798  11005802  11005803  11005818  11005819  11005861  11005866  11005867  11005869  11005892  11005909  11005915  11005964  11005979  11005981  11005983  11006022  11006047  11006053  11006057  11006064  11006066  11006079  11006080  11006120  11006152  11006156  11006177  11006184  11006185  11006195  11006208  11006241  11006246  11006247  11006260  11006261  11006317  11006318  11006330  11006349  11006358  11006369  11006370  11006387  11006390  11006405  11006440  11006454  11006460  11006461  11006463  11006464  11006469  11006473  11006474  11006484  11006487  11006500  11006505  11006524  11006549  11006563  11006574  11006577  11006579  11006580  11006581  11006592  11006616  11006648  11006672  11006687  11006711  11006734  11006741  11006742  11006754  11006762  11006794  11006801  11006808  11006829  11006838  11006853  11006865  11006889  11006934  11006948  11006950  11006985  11007016  11007041  11007042  11007059  11007069  11007071  11007075  11007076  11007077  11007089  11007091  11007102  11007110  11007111  11007129  11007131  11007132  11007134  11007162  11007178  11007230  11007264  11007265  11007269  11007302  11007304  11007325  11007349  11007366  11007377  11007387  11007395  11007396  11007408  11007412  11007437  11007456  11007458  11007480  11007496  11007506  11007515  11007682  11400098  11400100  11400115  11400153  11400154  11400166  11400171  11400175  11400184  11400191  11400243  11400263  11400273  11400276  11400277  11400285  11400288  11400290  11400294  11400308  11400309  11400320  11400324  11400328  11400333  11400344  11400346  11400349  11400350  11400366  11400380  11400385  11400387  11400392  11400405  11400429  11400435  11400443  11400447  11400462  11400463  11400467  11400469  11400474  11400480  11400485  11400491  11400498  11400501  61417634  61417635  61417636  61417637  61417981  61417986  61417987  61418002  61418003  61418004  61776947  61782238  61782262  61782267  61782275  61782296  61782304  61782305  61782315  61782330  61782346  61782348  61789179  61789181  61789185  61789187  61789189  61789209  61789211  61789212  61789213  61789216  61789217  61789218  61789219  61789220  61789221  61789222  61789223  61789224  61789225  61789227  61789228  61789229  61789230  61789231  61789232  61789233  61789234  61789236  61789239  61789241  61789242  61789243  61789246  61789249  61789250  61789251  61789252  61789253  61789256  61789257  61789259  61789260  61789261  61789262  61789263  61789264  61789265  61789266  61789267  61795041  61795042  61795043  61795044  61795045  61795046  61795047  61795048  61795071  61795072  61795073  61795074  61795075  61795076  61795078  61795079  61795080  61795081  61795082  61795083  61795084  61795085  61795086  61795087  61795088  61795089  61795090  61795091  61795092  61795093  61795094  61795095  61795096  61795097  61795098  61795099  61795101  61795102  61795103  61795104  61795105  61795106  61795107  61795108  61795109  61795110  61795111  61795112  61795113  61795114  61795115  61795116  61795117  61795118  61795119  61795120  61795121  61795122  61795123  61795124  61795125  61795126  61795127  61795128  61795129  61795139  61795140  61795142  61795994  61795995  61795996  61795997  61795998  61795999  61796001  61796002  61796003  61796365  61796366  61796367  61796368  61798635  61798636  61799345  61799346  61799347  61799348  61799351  61799352  61799353  61799354  61799355  61799356  61799357  61799359  61799399  61799401  61799404  61799406  61799407  61799408  61799409  61799410  61799411  61799412  61799413  61799414  61799415  61799416  61799417  61799418  61799419  61799421  61799422  61799423  61799424  61799425  61799426  61799427  61799428  61799429  61799430  61799431  61799432  61799438  61799439  61799440  61799441  61799442  61799443  61799444  61799445  61799446  61799447  61799448  61799449  61799450  61799451  61799452  61799460  61799461  61799467  61799468  61799470  61799471  61799474  61799475  61799476  61799477  61799478  61799490  61799491  61799492  61799493  61799494  61799495  61799496  61799497  61799498  61799499  61799500  61799501  61799502  61800624  61800625  61800626  61800627  61801489  61801490  61801491  61801492  61802098  61802099  61802101  61802102  61802109  61802110  61803559  61803560  61803561  61803562  61803573  61803574  61804266  61805083  61805084  61805169  61805171  61805172  61805174  61805175  61805176  61805178  61805179  61805180  61805181  61805199  61805776  61805777  61805778  61805779  61805780  61805781  61805782  61805784  61805785  61805786  61805787  61805788  61805789  61805790  61805791  61805792  61805793  61805805  61805806  61805807  61805808  61805809  61805810  61805811  61805812  61805813  61805815  61805816  61805817  61805818  61805819  61805822  61805823  61805824  61805825  61805826  61805827  61805829  61805830  61805831  61805832  61805833  61805834  61805835  61805836  61805841  61805842  61805843  61805844  61805845  61805846  61805847  61805848  61805857  61805858  61805864  61805865  61805866  61805867  61805868  61805869  61805870  61806264  61806266  61806267  61806268  61806474  61806475  61806476  61806477  61806478  61806479  61806480  61806481  61806482  61807818  61807819  61807820  61807821  61807822  61808586  61808587  61808588  61808589  61808590  61808591  61808592  61808593  61808594  61808595  61808794  61808795  61808796  61808868  61808874  61809702  61809703  61809704  61809706  61809708  61809709  61809710  61809711  61809712  61809713  61809780  61809926  61809927  61809932  61809933  61809934  61809935  61809936  61809937  61809938  61809939  61809940  61809941  61809946  61809947  61809948  61809949  61809950  61809951  61809952  61809953  61809954  61809955  61809956  61809957  61809962  61809963  61809964  61809965  61809967  61809968  61809969  61809970  61809971  61809972  61809974  61809975  61809976  61809977  61809978  61809987  61809988  61809989  61809990  61810005  61810006  61810007  61810008  61810010  61810011  61810012  61810013  61810014  61810015  61810016  61810019  61810020  61810021  61810022  61810023  61810025  61810026  61810027  61810028  61810397  61810399  61810401  61810402  61810403  61810404  61810405  61810406  61810407  61810408  61810765  61810766  61810767  61810768  61810769  61810770  61810771  61810772  61810773  61810774  61810775  61810776  61810777  61810778  61810779  61810780  61810781  61810782  61810783  61810784  61810849  61810850  61810851  61810852  61810853  61810854  61810855  61810856  61810857  61810859  61811763  61811764  61811765  61811766  61811769  61811770  61811771  61811772  61811773  61811774  61812983  61813659  61813660  61813661  61813662  61813663  61813685  61813686  61813687  61813690  61813691  61813692  61814740  61814741  61814742  61814743  61814745  61814748  61815393  61815394  61815471  61815547  61815548  61815549  61815550  61815551  61815552  61815553  61815554  61815555  61815556  61815559  61815560  61815561  61815562  61815563  61815564  61815565  61815566  61815567  61815568  61815569  61815570  61815571  61815626  61815627  61815628  61815629  61815630  61815631  61815632  61815633  61815634  61815691  61815701  61815742  61815787  61815791  61815799  61815802  61815804  61815805  61815807  61816394  61817743  61817745  61818491  61819926  61819927  61819928  61819929  61819930  61819955  61820632  61820633  61820634  61820635  61820637  61820638  61820649  61820667  61821122  61821123  61821124  61821126  61821127  61821405  61821406  61821716  61821719  61821720  61821721  61821722  61821723  61821742  61821743  61821848  61821849  61821851  61821853  61821855  61821856  61821857  61821866  61821867  61821868  61821869  61821870  61821871  61821873  61821874  61821888  61821889  61821890  61821894  61821895  61821896  61821925  61821926  61821927  61821928  61821929  61821930  61821932  61821933  61821934  61821935  61821936  61822967  61823876  61823878  61823879  61823881  61823914  61823915  61823916  61823917  61823918  61823960  61823966  61823967  61823970  61823971  61823972  61823973  61823974  61823975  61823976  61823978  61823979  61823980  61823981  61823982  61823983  61823984  61823985  61823986  61823987  61823988  61823989  61823990  61823991  61823992  61823993  61823994  61823995  61823996  61823997  61823998  61823999  61824001  61824002  61824003  61824005  61824606  61824607  61824766  61824767  61824768  61824777  61824778  61824784  61824785  61824786  61824935  61826919  61826920  61826921  61829047  61829048  61829049  61829050  61829055  61829056  61829821  61829822  61829823  61829825  61829826  61829827  61829828  61829829  61829830  61829832  61829845  61829847  61829848  61829849  61829850  61830362  61830364  61830365  61830376  61830377  61830378  61830729  61830732  61830733  61830734  61830735  61830736  61830762  61830808  61830809  61830810  61830811  61830812  61830813  61830814  61830815  61830816  61830817  61830818  61830819  61830820  61830821  61830822  61830823  61830824  61830825  61830826  61830827  61830828  61830829  61830830  61830831  61830833  61830834  61830835  61830836  61830838  61830839  61830840  61830841  61832271  61832305  61833583  61833586  61833587  61833614  61833615  61833616  61833617  61833618  61834846  61834847  61834857  61835355  61835359  61835745  61835821  61835822  61835823  61835911  61835912  61836573  61836574  61836576  61836577  61836578  61836579  61837162  61837316  61837317  61837318  61838278  61838279  61838280  61838281  61838282  61838283  61838284  61838285  61838286  61838287  61838288  61838289  61838290  61838291  61838292  61838293  61838294  61838295  61838296  61838297  61838298  61838299  61838301  61838302  61838303  61838304  61838305  61838306  61838307  61839396  61839397  61839398  61839399  61839401  61839408  61839409  61839410  61839411  61839412  61839413  61839414  61839415  61839416  61839417  61839418  61839419  61839420  61839421  61839422  61839423  61839424  61839425  61839426  61839427  61839428  61839429  61839690  61839692  61839731  61839732  61840455  61840457  61840458  61840461  61841267  61841268  61841270  61841272  61841274  61841275  61842335  61842337  61842338  61842339  61842340  61842341  61842342  61842344  61842345  61843118  61843120  61843121  61843122  61843124  61843125  61843126  61843144  61843146  61844226  61844229  61844232  61844579  61844580  61844581  61844582  61844585  61844886  61844890  61844891  61844927  61844928  61844930  61844932  61845904  61845905  61845906  61845907  61845908  61845909  61845910  61845911  61845914  61845915  61845925  61845926  61845927  61845928  61845929  61845930  61845931  61845932  61845933  61845934  61845935  61845936  61845937  61845938  61845939  61845940  61845941  61845942  61845943  61845944  61845945  61845946  61845949  61845950  61845951  61845952  61845953  61845954  61845955  61845956  61845957  61845958  61845959  61845960  61846250  61847656  61848272  61848273  61848274  61848275  61848277  61848278  61848279  61848280  61848281  61848282  61848283  61848329  61848330  61848331  61848333  61848334  61848335  61848336  61848337  61848339  61848340  61848341  61848342  61848344  61848345  61848346  61848347  61848348  61848350  61848351  61848352  61848353  61848356  61848357  61848358  61848359  61848360  61848362  61848363  61848364  61848365  61848366  61848367  61848368  61848369  61848370  61848371  61848372  61848373  61848374  61848375  61848376  61848377  61848378  61848379  61848380  61848510  61848511  61848512  61848516  61849032  61849033  61849034  61849035  61849036  61849037  61850206  61850207  61850915  61850918  61851226  61851227  61851228  61851710  61851711  61851712  61851714  61851715  61851716  61851717  61851718  61851719  61852846  61852847  61852848  61853685  61853686  61853687  61853688  61853689  61853690  61853691  61854809  61854810  61854811  61854812  61854813  61854838  61854840  61854841  61855238  61855239  61855240  61855241  61855242  61855243  61855244  61855245  61855246  61855247  61855248  61855251  61855252  61855253  61855254  61855255  61855256  61855257  61855353  61855354  61855355  61855356  61855357  61855359  61855360  61855361  61855362  61855363  61855364  61855365  61855366  61855367  61855368  61855369  61855370  61855953  61855954  61856378  61856380  61856381  61856382  61856383  61856732  61856733  61856734  61856737  61856740  61856741  61856742  61856913  61856914  61856917  61856918  61856919  61856920  61856921  61856928  61856929  61856930  61856931  61856932  61856933  61856934  61856935  61856936  61856937  61856939  61856940  61856941  61856942  61856943  61856944  61856945  61856946  61856947  61856948  61856949  61856950  61856951  61856952  61856953  61856972  61856973  61856974  61856975  61856979  61856980  61856981  61856982  61856983  61856984  61856985  61856986  61856987  61856988  61856989  61856990  61856991  61857883  61857884  61857885  61857889  61857890  61858465  61858466  61858468  61858469  61858470  61858471  61859210  61859211  61859213  61859216  61859219  61860207  61860208  61860209  61860210  61860212  61861340  61861341  61861345  61861347  61861348  61861349  61861507  61861508  61861509  61861510  61861512  61861513  61861514  61861517  61861518  61861519  61861520  61861521  61861522  61861523  61861524  61861525  61861526  61861527  61861528  61861529  61861530  61861531  61861532  61861533  61861538  61861539  61861540  61861541  61861542  61861543  61861544  61861545  61861546  61861547  61861548  61861549  61861550  61861551  61861552  61861553  61861554  61861555  61861556  61861557  61861558  61861559  61861560  61861885  61861886  61861887  61861888  61862122  61862124  61862125  61862127  61862129  61863934  61863935  61863936  61863937  61863938  61863940  61863941  61863985  61864864  61864865  61864867  61864869  61864870  61864871  61864872  61864873  61864874  61865894  61865895  61865896  61865897  61865898  61865899  61865901  61865902  61865903  61865904  61866014  61866015  61866016  61866017  61866018  61866019  61866020  61866157  61866158  61866159  61866160  61866161  61866162  61866163  61866164  61866165  61866166  61866167  61866168  61866169  61866170  61866171  61866172  61866173  61866174  61866175  61866176  61866177  61866178  61866179  61866180  61866181  61866182  61866183  61866184  61866185  61866186  61866187  61866188  61866189  61866190  61866191  61866192  61866193  61866194  61866195  61866196  61866197  61866198  61866694  61866695  61866696  61866697  61866698  61866699  61866701  61866702  61866704  61867594  61867595  61867596  61867597  61867598  61867599  61867601  61867602  61867603  61868855  61868856  61868857  61868858  61868859  61868860  61869840  61869841  61869842  61869845  61869847  61870712  61870714  61870715  61870716  61870717  61870718  61870719  61870809  61870810  61870811  61870812  61870813  61870814  61870815  61870816  61870817  61870818  61870819  61871104  61871105  61871106  61871107  61871108  61871109  61871110  61871136  61871137  61871138  61871139  61871140  61871141  61871142  61871143  61871144  61871145  61871146  61871147  61871148  61871149  61871150  61871151  61871152  61871153  61871158  61871162  61871172  61871192  61871193  61871194  61871195  61871196  61871197  61871199  61871200  61871201  61871202  61871203  61871204  61871205  61871681  61871682  61871683  61871684  61871685  61871686  61871687  61871688  61871689  61871690  61871691  61871692  61871693  61871694  61871695  61871696  61871697  61871702  61871703  61871704  61871705  61871706  61871707  61871708  61871709  61871710  61871711  61871712  61871713  61871714  61871715  61871716  61871717  61871718  61871719  61871720  61871721  61871722  61871723  61871724  61871725  61871726  61871727  61871728  61871729  61871730  61871731  61871732  61871733  61872322  61872324  61872325  61872327  61872328  61872329  61872330  61873427  61873429  61873430  61873431  61873432  61873433  61873434  61873435  61874449  61874451  61874452  61874453  61874454  61874457  61874458  61875436  61875437  61875438  61875889  61875890  61875891  61875892  61875894  61875895  61875896  61875898  61875899  61875903  61875904  61875906  61875907  61875908  61875909  61875921  61875922  61875923  61875924  61875925  61875926  61875927  61876903  61876906  61878264  61878265  61878266  61878267  61879218  61879219  61879220  61879222  61880091  61881029  61881035  61881036  61881662  61881664  61881677  61883259  61884183  61884191  61884746  61884748  61884749  61884750  61884751  61884752  61884753  61884754  61884755  61884756  61884757  61884759  61884760  61884761  61884762  61884763  61884764  61884765  61884766  61884767  61884768  61884769  61884770  61884771  61884772  61884773  61884774  61884775  61884776  61884777  61884778  61884779  61884780  61884781  61884784  61884785  61884786  61884787  61884788  61884789  61884790  61884791  61884792  61885170  61885171  61885766  61885767  61885769  61886257  61886306  61886307  61886316  61888049  61888050  61888586  61888589  61888590  61888591  61888592  61888593  61888594  61888595  61888596  61888597  61888598  61888599  61888601  61888602  61888603  61888604  61888605  61888606  61888607  61888608  61888609  61888610  61888611  61888612  61888613  61888614  61888615  61888616  61888617  61888618  61888619  61888620  61888621  61888622  61888898  61888899  61888901  61888902  61888903  61888904  61888905  61888906  61888907  61890009  61890033  61890034  61890035  61890037  61890038  61890114  61890115  61890116  61890117  61890689  61890749  61890750  61892357  61892358  61892361  61892381  61893238  61893743  61893744  61893745  61893746  61893747  61893748  61894035  61894037  61894038  61894039  61894085  61894086  61894090  61894091  61894092  61894093  61894094  61894095  61894096  61894119  61894120  61894262  61894263  61894264  61894265  61894266  61894267  61894268  61894269  61894270  61894271  61894272  61894273  61894274  61894275  61894276  61894277  61894278  61894279  61894280  61894281  61894282  61894283  61894284  61894285  61894286  61894287  61894288  61894289  61894290  61894291  61894292  61894293  61894294  61894295  61894530  61895301  61895302  61895307  61895309  61896040  61897282  61897283  61898484  61898486  61899279  61899280  61899281  61899282  61899284  61899285  61899286  61899287  61899288  61899289  61899291  61899292  61899293  61899294  61899295  61899296  61899297  61899298  61899299  61899301  61899302  61899304  61899305  61899306  61899307  61899308  61899309  61899310  61899311  61899312  61899313  61899314  61899315  61899316  61899317  61899318  61899319  61899320  61899321  61899322  61899323  61899324  61899326  61899327  61899328  61899329  61899330  61899331  61899332  61899333  61899334  61899335  61899336  61899337  61899338  61899614  61899615  61899642  61899643  61899644  61900452  61900453  61900471  61901205  61901411  61902938  61902940  61902962  61903914  61903915  61903916  61903917  61903918  61903919  61903920  61903921  61903922  61903923  61903924  61903925  61903926  61903927  61903928  61903929  61903930  61903931  61903932  61903933  61903934  61903935  61903936  61903937  61903938  61903939  61903940  61903944  61903945  61903946  61903947  61903948  61903949  61903950  61903951  61903952  61903953  61903954  61903955  61903956  61903957  61903958  61903959  61903960  61903961  61903971  61903972  61903973  61903974  61903975  61903976  61903977  61903978  61903979  61905090  61905148  61905153  61905154  61905248  61905249  61905250  61905251  61905252  61905253  61905254  61905255  61905257  61905258  61905259  61905260  61905261  61905262  61906065  61907439  61907440  61907441  61908190  61908191  61908192  61908193  61908194  61908195  61908196  61908197  61908198  61908199  61908201  61908202  61908203  61908204  61908205  61908206  61908207  61908208  61908209  61908210  61908211  61908212  61908213  61908214  61908215  61908216  61908217  61908218  61908219  61908220  61908221  61908222  61908223  61908225  61908226  61908227  61908228  61908229  61908230  61908234  61908235  61908262  61908263  61908264  61908265  61908266  61908267  61908268  61908269  61908281  61908282  61908283  61908316  61908317  61908318  61909326  61909327  61909328  61909329  61909330  61909331  61909332  61909334  61909462  61910187  61910190  61910191  61910192  61910445  61910858  61910859  61910860  61910861  61910863  61914476  61914477  61915296  61915297  61915299  61915303  61915304  61915789  61915790  61915791  61915792  61915793  61915794  61915795  61915796  61915797  61915798  61915799  61915801  61915802  61915803  61915804  61915805  61915806  61915807  61915808  61915809  61915810  61915811  61915812  61915813  61915814  61915815  61915816  61915817  61915818  61915819  61915820  61915821  61915822  61915823  61915824  61915825  61915826  61915827  61915828  61915829  61915830  61915831  61915832  61915833  61915834  61916142  61916143  61916145  61917745  61918960  61918962  61918963  61918964  61918965  61918966  61918983  61918984  61918985  61918986  61918987  61920423  61920429  61920430  61920431  61920432  61920433  61921093  61921094  61921095  61921096  61921097  61921098  61921099  61921101  61921102  61921103  61921104  61921105  61921106  61921107  61921108  61921109  61921110  61921111  61921112  61921113  61921114  61921115  61921116  61921117  61921118  61921119  61921120  61921121  61921122  61921123  61921124  61921125  61921126  61921127  61921128  61921129  61921130  61921131  61921132  61921133  61921134  61921135  61921137  61921139  61921140  61921436  61922243  61922248  61922276  61923555  61923556  61923557  61923558  61923562  61923563  61923564  61923586  61923588  61924436  61924438  61924439  61924440  61924442  61924443  61924444  61924445  61925142  61926118  61926123  61926124  61927556  61927557  61927558  61927559  61927796  61927797  61927798  61927799  61927801  61927802  61927803  61927804  61927805  61927806  61927807  61927808  61927809  61927810  61927811  61927812  61927813  61927814  61927815  61927816  61927817  61927855  61927856  61927857  61927858  61927859  61927860  61927861  61927862  61927863  61927864  61927865  61927866  61927867  61927868  61927869  61927870  61927871  61927872  61927873  61927874  61927875  61927952  61927953  61927954  61927956  61927957  61927958  61927960  61927961  61927962  61929394  61929395  61929396  61929397  61929398  61929424  61930142  61930143  61930144  61930145  61930146  61930147  61930148  61930149  61931655  61931656  61931657  61931658  61931659  61931660  61931661  61931936  61931937  61931938  61932270  61932271  61932272  61932273  61932274  61932275  61932276  61932277  61932278  61932279  61932280  61932629  61932630  61932631  61932632  61932633  61932634  61932635  61932636  61932637  61932638  61932639  61932640  61932641  61932642  61932643  61932644  61932645  61932646  61932647  61932648  61932649  61932650  61932651  61932652  61932653  61932654  61932655  61932656  61932657  61932658  61932659  61932660  61932661  61932662  61932663  61932664  61932665  61932666  61932667  61932668  61932669  61932670  61932671  61932709  61932710  61932711  61932712  61933495  61933828  61933830  61933831  61933832  61934951  61934952  61934953  61934954  61935889  61935890  61935971  61935972  61935974  61935975  61935977  61935979  61936370  61936371  61936374  61936479  61936480  61936481  61936482  61936483  61936484  61936485  61936486  61936487  61936488  61936490  61936491  61936492  61936493  61936494  61936495  61936496  61936497  61936498  61936499  61936501  61936502  61936507  61936508  61936512  61936780  61936781  61936782  61936783  61936784  61936785  61936786  61936788  61936789  61936790  61936791  61937274  61937276  61937278  61937280  61937283  61937753  61937754  61937786  61937787  61937788  61937789  61937790  61937791  61937793  61938464  61938465  61938466  61938467  61938468  61938469  61938470  61938471  61938472  61938473  61938474  61938475  61938476  61938477  61938478  61938479  61938481  61938482  61938483  61938485  61938486  61938488  61938489  61938490  61938491  61938492  61938493  61938494  61938495  61938496  61938497  61938498  61938499  61938500  61938502  61938514  61938515  61938518  61938519  61938520  61938521  61938526  61938528  61938529  61938530  61938734  61938737  61938752  61939173  61939174  61939176  61939179  61939180  61939788  61939790  61939792  61939793  61939794  61939796  61939797  61939925  61939927  61940215  61940216  61940217  61940218  61940343  61940346  61940947  61940948  61940949  61940951  61940952  61940953  61940954  61940955  61941378  61941381  61941386  61941574  61941575  61941576  61941577  61941578  61941579  61941580  61943178  61943179  61943180  61943184  61943437  61943438  61943439  61943440  61943441  61943442  61943444  61943445  61943450  61943477  61943488  61943492  61943493  61943494  61943499  61943504  61943511  61943515  61943516  61943518  61944656  61944657  61944658  61944878  61944879  61944880  61944881  61944882  61944883  61944884  61944885  61946454  61946477  61946478  61946479  61946480  61946481  61946482  61946483  61946484  61946485  61946487  61946869  61946870  61946946  61946947  61946948  61946949  61946950  61946951  61946952  61946953  61946976  61946977  61946978  61946979  61947004  61947005  61947007  61947008  61947009  61947010  61947184  61947185  61947186  61947187  61947189  61948133  61949269  61949270  61949271  61949272  61949273  61949274  61949275  61949276  61949277  61949278  61949279  61949280  61949281  61949283  61949284  61949285  61949289  61949290  61949291  61949293  61949294  61949295  61949296  61949297  61949298  61949299  61949301  61949302  61949304  61949305  61949306  61949307  61949308  61949309  61949310  61949311  61949312  61949313  61949314  61949315  61949316  61949317  61949318  61949319  61949345  61949346  61949347  61949348  61949349  61949359  61949360  61949361  61949362  61949363  61949364  61949693  61949696  61949698  61950689  61950697  61950703  61950707  61951101  61951102  61951106  61951334  61951335  61951338  61951339  61951340  61951346  61951347  61951348  61951349  61951350  61951351  61951352  61951353  61951354  61951355  61951356  61951357  61951358  61951359  61951360  61951361  61951362  61951363  61951364  61951365  61951366  61951367  61951368  61951369  61951370  61951371  61951372  61951373  61951374  61951375  61951376  61951377  61951378  61951379  61951380  61951381  61951382  61951383  61951384  61951385  61951386  61951387  61951645  61951653  61951657  61951659  61952874  61954536  61954542  61954544  61955559  61955560  61958173  61958174  61958175  61958176  61958177  61958178  61958179  61958180  61958181  61958182  61958183  61958184  61958185  61958186  61958188  61958189  61958190  61958191  61958192  61958193  61958194  61958195  61958196  61958197  61958198  61958199  61958201  61958202  61958209  61958210  61960815  61961797  61961798  61961799  61962231  61962235  61962238  61962239  61962301  61962304  61962306  61962548  61962549  61962550  61962551  61962553  61962554  61962555  61962556  61962557  61962558  61962559  61962560  61962561  61962562  61962564  61962566  61962567  61962568  61962569  61962570  61962571  61962572  61962573  61962574  61962575  61962576  61962577  61962578  61962579  61962580  61962581  61962853  61962856  61962860  61962861  61962862  61962863  61962864  61962875  61962877  61962878  61962987  61962988  61962989  61962990  61962991  61962992  61962999  61963001  61963002  61963003  61963004  61963005  61963397  61963398  61963401  61963402  61963403  61963404  61963842  61963843  61963844  61963845  61963846  61963848  61963883  61963884  61963885  61963886  61963887  61963898  61963899  61963901  61964111  61964112  61964113  61964114  61964431  61964432  61964433  61964440  61964441  61965570  61965571  61965572  61965931  61966232  61966233  61966234  61966235  61966803  61966804  61966805  61966809  61967914  61967915  61967916  61967917  61967918  61967919  61967920  61967921  61967922  61967923  61967924  61967925  61967926  61967927  61968027  61968028  61968029  61968030  61968031  61968032  61968033  61968034  61968035  61968036  61968037  61968038  61968039  61968040  61968041  61968042  61968043  61968044  61968045  61968046  61968047  61968048  61968049  61968050  61968051  61968052  61968053  61968054  61968055  61968056  61968057  61968058  61968059  61968060  61968061  61968062  61968063  61968064  61968065  61968066  61968067  61968068  61968883  61968884  61970674  61970675  61970676  61970677  61970678  61970689  61971149  61971154  61971155  61971188  61971719  61971722  61972155  61972157  61972158  61972162  61972163  61972164  61972165  61972728  61972729  61972730  61972731  61972733  61972734  61972735  61972743  61972745  61973056  61973057  61973226  61973227  61973230  61973231  61973232  61973233  61973234  61973235  61973248  61974114  61974116  61974118  61974119  61974120  61974121  61974456  61974459  61974462  61974464  61974466  61974467  61974468  61974470  61974473  61974475  61974476  61974477  61974479  61974480  61974481  61974483  61974484  61974486  61974487  61974488  61974489  61974491  61974492  61974493  61974494  61974495  61974497  61974498  61974499  61974500  61974501  61974503  61974504  61974506  61974507  61974508  61974557  61974558  61974559  61974560  61974561  61974562  61974563  61974564  61974565  61974566  61974567  61974568  61974569  61974570  61976410  61976412  61976413  61976433  61977245  61977258  61977543  61977544  61977545  61977547  61977548  61977549  61978249  61978251  61978254  61978258  61978259  61978262  61978494  61978497  61978501  61978505  61978510  61978519  61978551  61978552  61979096  61979102  61979104  61979152  61979153  61979765  61979769  61979770  61979771  61979772  61980304  61980343  61980566  61980567  61980568  61980594  61980595  61982274  61982293  61982294  61982450  61982455  61982456  61982457  61982718  61982719  61982724  61982725  61982749  61982750  61982755  61982756  61982823  61982824  61982825  61982826  61982827  61982828  61982857  61982858  61982859  61982860  61982861  61982862  61982863  61982864  61982865  61982866  61982867  61982868  61982869  61982870  61982871  61982872  61982873  61982874  61982875  61982877  61982888  61982889  61982890  61982891  61982892  61982893  61982894  61982895  61982896  61982897  61982898  61982899  61982901  61982902  61982903  61982904  61984204  61984207  61984833  61984845  619",
      "more_code_info": "84857  61984862  61985019  61985020  61985021  61985038  61985042  61986242  61986412  61986413  61986414  61986415  61986417  61986427  61986428  61986689  61986693  61986696  61986697  61986698  61986734  61986735  61986853  61986854  61987056  61987057  61987058  61987059  61987074  61987408  61987410  61987537  61987538  61987539  61987540  61987541  61987542  61987543  61987544  61987568  61987739  61987741  61987742  61987797  61987798  61987799  61987800  61987801  61987802  61987803  61987804  61987805  61987806  61987807  61987808  61987809  61987810  61987811  61987812  61987813  61987814  61987815  61987816  61987817  61987818  61987819  61987820  61987821  61987822  61987824  61987825  61987826  61987827  61987828  61987829  61987830  61987831  61987832  61988117  61988127  61988128  61988445  61988447  61988448  61988449  61988461  61988462  61988554  61989033  61989034  61989036  61989088  61989663  61989664  61989666  61989667  61989672  61989673  61989674  61990659  61990907  61990908  61990910  61992033  61992050  61992702  61992977  61992978  61992979  61992980  61992982  61993578  61993579  61993580  61993581  61993583  61993585  61993586  61993588  61993593  61993599  61993601  61993602  61993603  61993604  61993605  61993606  61993607  61993608  61993611  61993613  61993614  61993615  61993620  61993621  61993649  61993650  61993652  61993653  61993654  61993655  61993656  61993657  61993923  61993924  61995663  61995683  61995687  61996342  61996343  61996351  61996967  61997335  61998616  61998617  61998798  61998837  61999551  61999552  61999553  61999554  61999574  61999925  61999973  62000378  62000379  62000380  62000386  62000387  62000388  62000389  62000390  62000391  62000392  62000393  62000394  62000395  62000396  62000397  62000399  62000400  62000401  62000402  62000403  62000404  62000405  62000406  62000407  62000408  62000411  62000413  62000414  62000415  62000416  62000417  62000418  62000419  62000420  62000421  62000422  62000446  62000447  62000969  62000970  62000985  62000990  62002011  62002012  62002013  62002022  62002024  62002050  62002052  62002061  62002062  62002064  62002078  62002251  62002480  62002481  62002482  62002483  62002484  62003646  62003654  62003763  62003764  62003766  62003767  62003775  62003961  62003963  62004486  62004487  62004488  62004489  62004490  62004491  62004492  62004493  62004494  62004495  62004496  62004498  62004499  62004500  62004501  62004502  62004504  62004505  62004506  62004507  62004508  62004509  62004510  62004511  62004512  62004513  62004791  62004837  62004839  62004904  62004905  62004906  62004907  62004908  62004909  62004910  62004911  62004913  62004914  62004915  62004916  62004917  62004918  62004919  62004920  62004921  62004922  62004923  62004924  62004994  62005002  62005850  62005851  62005897  62005948  62005949  62005956  62005957  62006457  62006727  62006728  62006992  62006997  62007005  62008259  62008260  62008261  62008269  62008270  62008275  62008276  62008277  62008278  62009299  62009300  62009301  62009302  62009307  62009309  62010099  62010101  62010197  62010326  62010334  62010335  62010336  62011221  62011226  62011302  62011303  62011304  62011555  62011556  62011557  62011558  62011565  62011566  62011567  62012054  62012055  62012056  62012311  62012316  62012393  62012418  62012419  62012420  62012421  62012422  62012424  62012425  62012426  62012427  62012428  62012429  62012430  62012431  62012655  62012656  62012661  62013197  62013198  62013207  62013208  62013209  62013210  62013211  62013212  62013213  62013214  62013215  62013216  62013217  62013218  62013219  62013222  62013223  62013224  62013225  62013226  62013227  62013228  62013229  62013230  62013231  62013232  62013233  62013234  62013235  62013236  62013237  62013238  62013239  62013240  62013241  62013242  62013244  62013468  62013469  62013555  62013556  62014072  62014081  62014082  62014386  62014395  62014798  62014801  62015841  62015842  62015843  62015844  62015845  62015846  62015853  62015854  62016678  62016679  62016806  62016827  62017252  62017254  62017255  62017596  62017910  62017914  62017915  62018046  62018049  62018136  62018138  62018141  62018158  62018502  62018584  62018586  62018840  62018841  62018842  62018843  62018851  62018852  62018856  62018857  62018861  62018862  62018863  62018864  62018865  62018866  62018867  62018868  62018869  62018870  62018871  62018872  62018873  62018874  62018875  62018876  62018877  62018878  62018879  62018880  62018881  62018882  62018883  62018884  62018885  62018886  62018887  62018888  62018889  62018890  62018891  62018892  62019284  62019285  62019483  62019484  62019485  62019488  62020635  62020640  62020645  62020796  62020799  62020800  62021627  62021635  62021641  62022294  62022295  62022306  62022307  62022309  62022928  62023280  62023281  62023782  62023785  62023786  62023787  62023788  62023808  62023809  62024563  62024566  62024591  62024951  62025523  62025524  62025525  62025526  62025735  62025736  62025737  62025738  62025739  62025740  62025741  62025742  62025743  62025744  62025745  62025746  62025747  62025748  62025749  62025750  62025751  62025752  62025753  62025754  62025755  62025756  62025757  62025758  62025759  62025760  62025761  62025762  62025763  62025764  62025765  62025766  62025767  62025768  62025769  62025770  62025771  62025772  62025773  62025774  62025775  62025776  62025777  62025778  62025779  62025780  62025781  62025782  62025783  62025784  62025785  62026090  62026091  62026827  62026830  62026852  62026853  62027076  62027077  62027187  62027188  62027190  62027192  62027193  62027194  62027215  62027216  62028014  62028440  62028441  62028444  62028445  62028446  62028463  62028464  62029262  62029265  62029451  62029857  62029859  62029860  62029876  62030271  62030272  62030273  62030481  62030484  62030582  62030584  62030586  62030621  62030624  62030631  62030632  62031793  62031799  62031801  62031879  62031880  62031881  62031882  62031883  62031884  62031885  62031886  62031887  62031888  62031889  62031890  62031891  62031892  62031893  62031895  62031896  62031897  62031898  62031899  62031900  62031901  62031902  62031903  62031904  62031905  62031906  62031907  62031908  62031909  62031910  62031911  62031912  62031913  62031914  62031915  62031916  62031919  62031920  62031921  62031922  62031923  62031925  62031926  62031927  62031928  62031929  62031930  62031931  62031932  62032628  62032630  62033084  62033085  62033831  62033832  62034590  62034591  62034592  62034593  62034594  62034595  62034602  62034606  62034608  62034609  62035833  62035835  62035836  62035838  62036142  62036143  62036144  62036525  62036670  62036677  62036678  62038011  62038012  62038035  62038036  62038037  62038164  62038165  62038167  62038168  62038169  62038170  62038171  62038172  62038173  62038174  62038175  62038176  62038177  62038178  62038179  62038180  62038181  62038182  62038183  62038184  62038185  62038186  62038187  62038188  62038189  62038190  62038191  62038192  62038193  62038194  62038195  62038196  62038197  62038198  62038199  62038201  62038202  62038203  62038204  62038205  62038206  62038207  62038208  62038209  62038210  62038211  62038212  62038213  62038214  62038215  62038216  62038217  62038218  62039081  62039082  62039092  62040761  62040762  62040763  62041495  62041497  62041518  62042283  62042336  62043133  62043134  62043135  62043150  62043151  62043161  62043162  62043164  62043282  62043312  62043313  62043324  62043325  62043939  62043946  62043950  62044611  62044621  62044622  62044628  62046068  62046069  62046103  62046104  62046105  62046106  62046107  62046108  62046109  62046110  62046111  62046112  62046113  62046114  62046115  62046116  62046117  62046118  62046119  62046120  62046121  62046122  62046123  62046124  62046125  62046126  62046127  62046128  62046129  62046130  62046157  62046158  62046159  62046160  62046161  62046162  62046218  62046219  62046220  62046234  62046467  62046468  62046469  62046876  62046877  62046879  62046905  62046962  62047519  62047520  62047522  62047529  62047538  62048209  62048212  62048434  62048452  62050593  62050594  62052656  62052709  62054023  62054024  62054025  62054026  62054027  62054028  62054029  62054030  62054031  62054032  62054033  62054034  62054035  62054036  62054037  62054038  62054039  62054040  62054041  62054042  62054043  62054044  62054045  62054046  62054047  62054048  62054049  62054050  62054051  62054052  62054053  62054054  62054055  62054056  62054057  62054058  62054059  62054060  62054061  62054062  62054063  62054074  62054099  62054101  62054102  62054103  62054104  62054649  62054652  62054821  62054822  62054826  62054828  62055032  62055033  62055034  62055158  62055159  62055160  62055161  62055162  62055163  62055164  62055165  62055537  62055604  62055605  62055606  62056201  62056202  62056294  62056295  62056296  62056297  62056298  62056299  62056301  62056302  62056303  62056304  62056305  62056306  62056307  62056308  62056309  62056311  62056312  62056313  62056314  62056315  62056316  62056317  62056318  62056319  62056320  62056321  62056322  62056323  62056324  62056325  62056350  62056791  62057224  62057225  62057226  62058206  62058207  62058208  62058209  62058210  62058231  62058740  62058752  62059368  62059373  62060475  62060476  62060501  62060502  62060503  62060514  62060516  62060518  62061687  62061871  62061889  62061890  62061947  62061948  62061949  62061950  62061951  62061952  62061953  62061954  62061955  62061956  62061957  62061958  62061959  62061960  62061961  62061962  62061963  62061964  62061965  62061966  62061967  62061968  62061969  62061970  62061971  62061972  62061973  62061974  62061975  62061976  62061977  62062237  62062243  62063281  62063708  62063709  62063715  62063716  62064891  62065288  62065290  62065294  62065297  62066541  62066542  62066557  62066558  62066729  62066732  62066734  62066737  62066832  62066833  62066834  62066835  62067343  62067344  62067359  62067360  62067730  62067731  62067767  62067771  62067778  62067788  62067791  62067919  62067921  62067922  62067948  62067949  62067950  62067951  62067957  62067958  62067960  62067961  62067999  62068000  62068001  62068002  62068003  62068004  62068005  62068006  62068007  62068008  62068009  62068010  62068011  62068012  62068013  62068014  62068015  62068016  62068017  62068018  62068020  62068021  62068022  62068023  62068024  62068025  62068026  62068027  62068028  62068029  62068030  62068031  62068033  62068034  62068035  62068036  62068037  62068307  62068308  62068309  62068310  62068315  62068316  62069325  62069326  62069328  62069329  62069330  62069331  62070238  62070821  62070822  62070823  62072140  62072148  62072194  62072205  62072508  62072509  62072510  62073096  62073098  62073099  62073101  62073102  62074153  62074156  62074157  62074158  62074159  62074161  62074170  62074178  62074693  62074694  62074697  62074698  62074699  62074701  62074704  62074705  62074706  62074707  62074711  62074712  62074714  62074715  62074717  62074719  62074720  62074722  62074723  62074724  62074753  62074754  62074755  62075012  62075013  62076312  62076317  62076319  62076320  62077279  62077280  62077283  62077285  62077287  62077289  62077836  62077855  62078813  62078814  62078815  62078840  62078841  62078842  62079813  62079814  62079816  62079817  62079828  62079849  62079850  62079852  62079853  62079854  62079867  62079868  62080297  62080298  62080302  62080481  62080482  62080483  62080484  62080485  62080486  62080487  62080488  62080489  62080490  62080491  62080492  62080493  62080494  62080495  62080496  62080497  62080498  62080499  62080501  62080502  62080503  62080504  62080505  62080506  62080507  62080508  62080509  62080510  62080511  62080512  62080513  62080514  62080784  62080806  62080807  62080808  62080809  62080810  62080811  62080812  62081420  62081421  62081422  62081423  62081424  62081863  62081864  62081866  62081867  62082373  62082374  62083058  62083060  62083061  62083065  62083080  62083084  62083085  62083086  62083087  62083116  62083119  62083120  62083122  62083151  62083152  62083153  62083154  62083162  62084423  62084424  62084431  62084432  62085526  62085527  62085942  62085943  62086332  62086352  62086354  62086357  62086607  62086624  62086625  62086627  62086628  62086672  62086673  62086674  62086675  62086684  62086685  62086686  62086692  62086693  62086694  62086695  62086696  62086697  62086698  62086699  62086700  62086701  62086702  62086703  62086704  62086705  62086706  62086707  62086708  62086709  62086710  62086711  62086712  62086713  62086714  62086715  62086716  62086717  62086718  62086719  62086979  62086980  62086982  62086984  62086985  62086986  62087011  62087181  62087187  62087932  62087933  62088272  62088273  62088274  62088275  62088276  62088855  62088857  62088865  62088866  62088868  62088869  62088870  62088871  62088872  62089669  62089670  62089671  62090130  62090131  62090139  62090371  62090385  62090386  62090387  62090388  62090389  62090390  62090391  62091086  62091110  62091381  62091382  62091389  62091537  62091835  62091836  62091856  62091858  62091985  62091986  62092004  62092255  62092256  62092258  62092293  62092294  62092305  62092306  62092366  62092367  62092368  62092369  62092370  62092371  62092372  62092373  62092374  62092375  62092376  62092377  62092378  62092379  62092380  62092381  62092382  62092383  62092384  62092385  62092517  62092522  62092523  62094378  62094379  62094380  62094993  62094997  62094998  62094999  62095003  62095004  62095005  62095006  62095827  62095828  62095829  62095831  62095999  62096001  62096002  62096032  62096034  62096615  62096616  62096698  62096699  62096701  62096702  62097840  62097841  62098048  62098050  62098054  62098055  62098277  62098279  62098781  62098836  62098856  62098976  62098979  62098983  62098984  62098991  62098992  62099053  62099056  62099057  62099079  62099261  62099279  62099280  62099281  62099861  62099942  62100207  62100325  62100458  62100459  62100462  62100463  62100464  62100479  62101240  62101289  62102050  62102051  62102052  62102054  62102058  62102081  62102082  62102083  62102084  62102679  62102694  62102696  62102874  62103205  62103225  62103226  62103510  62104012  62104013  62104014  62104015  62104016  62104017  62104018  62104019  62104020  62104021  62104022  62104023  62104024  62104025  62104026  62104027  62104028  62104029  62104030  62104031  62104032  62104033  62104034  62104035  62104036  62104037  62104038  62104039  62104040  62104041  62104042  62104043  62104044  62104046  62104156  62104157  62104158  62104159  62104160  62104161  62104162  62104163  62104164  62104165  62104822  62104823  62104836  62104837  62104841  62104842  62104843  62104844  62105264  62105271  62105312  62105313  62105314  62105315  62105318  62105319  62105320  62105321  62105322  62105323  62105324  62105325  62105326  62105327  62105328  62105329  62105724  62105725  62105757  62106689  62106690  62107108  62107959  62107978  62107979  62107980  62108017  62108024  62109150  62109155  62109158  62109189  62109651  62109652  62109653  62110227  62110228  62110229  62110244  62110245  62110509  62110510  62110511  62110512  62110513  62110514  62110515  62110516  62110517  62110518  62110519  62110520  62110521  62110522  62110523  62110524  62110525  62110526  62110527  62110528  62110529  62110530  62110531  62110532  62110533  62110534  62110535  62110536  62110537  62110538  62110539  62110540  62110541  62110542  62110543  62110544  62110559  62110560  62110561  62110562  62110797  62110803  62110804  62110805  62110806  62110807  62110808  62110825  62111837  62111943  62111944  62111955  62111980  62111981  62111983  62111984  62112033  62112034  62112035  62112537  62112538  62112539  62112540  62112683  62112764  62112766  62112768  62112970  62112973  62112986  62112987  62112990  62112991  62112992  62112993  62113014  62113015  62113016  62114022  62114102  62114103  62114363  62115230  62115233  62115651  62115654  62115849  62115850  62115956  62115957  62115993  62115994  62115995  62115999  62117018  62117019  62117020  62117021  62117022  62117023  62117024  62117025  62117026  62117027  62117028  62117029  62117030  62117031  62117032  62117033  62117034  62117035  62117036  62117037  62117038  62117039  62117040  62117041  62117042  62117043  62117044  62117045  62117046  62117047  62117048  62117049  62117050  62117051  62117052  62117053  62117293  62117295  62117298  62117310  62117312  62117424  62117429  62118378  62118379  62118380  62118381  62118382  62118401  62118403  62118551  62118552  62118553  62118554  62118563  62118569  62118570  62118572  62119157  62119531  62119533  62119535  62119536  62119537  62119538  62119539  62119559  62119560  62120602  62121190  62121192  62121193  62121225  62121226  62121227  62121552  62121553  62121554  62121769  62121771  62121786  62121787  62121788  62121791  62121793  62121808  62121809  62122526  62122534  62122535  62122628  62122629  62122631  62122654  62122656  62122657  62122658  62122659  62122719  62122720  62122721  62122722  62122723  62122724  62122725  62122726  62122727  62122728  62122729  62122730  62122731  62122732  62122733  62122734  62122735  62122736  62122737  62122858  62122859  62124285  62124287  62124288  62124293  62124295  62124296  62124609  62124757  62124762  62124771  62124778  62124887  62125118  62125120  62125261  62125512  62125514  62125516  62126092  62126093  62126094  62126095  62126097  62126108  62126119  62126120  62126125  62126126  62126681  62126682  62126685  62126825  62126827  62126829  62127081  62127082  62127083  62127084  62127097  62127114  62127115  62127405  62127409  62127563  62128193  62128194  62128195  62128196  62128197  62128198  62128199  62128201  62128202  62128203  62128204  62128205  62128206  62128207  62128208  62128209  62128210  62128211  62128222  62128223  62128224  62128225  62128226  62128227  62128231  62128232  62128233  62128234  62128235  62128236  62128237  62128238  62128239  62128240  62128241  62128359  62128360  62128372  62130953  62130957  62131030  62131122  62131182  62131183  62131184  62131193  62131198  62131199  62131202  62131203  62131223  62131225  62131247  62131262  62131263  62131275  62131276  62131285  62131292  62132058  62132062  62132069  62132079  62132081  62132095  62132099  62132217  62132218  62132219  62132220  62132221  62132222  62132223  62132224  62132225  62132226  62132227  62132228  62132229  62132230  62132231  62132234  62132235  62132236  62132238  62132241  62132242  62132245  62132246  62132250  62132251  62132252  62132444  62132446  62132447  62132453  62132741  62132863  62132873  62132877  62132888  62132889  62133155  62134261  62134262  62134266  62134267  62134272  62134705  62134744  62135954  62135955  62136375  62136376  62136377  62136378  62136381  62136382  62136566  62137053  62137054  62137509  62137512  62137534  62137535  62137963  62137967  62138237  62138239  62138240  62138251  62138252  62138253  62138254  62138256  62138290  62138291  62138292  62138293  62138294  62138295  62138956  62138957  62139109  62139110  62139111  62139117  62139118  62139119  62139120  62139121  62139122  62139123  62139124  62139125  62139126  62139130  62139131  62139132  62139133  62139134  62139135  62139136  62139137  62139138  62139139  62139144  62139145  62139146  62139147  62139151  62139152  62139153  62139154  62139155  62139156  62139303  62139315  62139319  62140626  62140630  62140631  62140633  62141724  62142289  62142290  62142291  62142302  62142305  62142572  62142582  62142611  62142612  62143143  62143152  62143612  62143613  62143615  62143638  62143639  62143655  62143735  62143736  62143737  62144119  62144122  62144124  62144286  62144289  62144313  62144594  62144595  62144596  62144597  62144598  62144599  62144601  62144602  62144603  62144604  62144605  62144607  62144608  62144610  62144611  62144617  62144618  62144620  62144621  62144622  62144658  62144664  62144666  62144932  62144938  62144946  62145066  62145399  62145762  62145763  62145764  62145765  62146616  62146618  62146619  62146929  62146930  62146931  62146932  62146933  62146934  62146936  62146954  62146955  62146957  62148236  62148237  62148239  62148259  62148750  62148752  62148753  62148754  62148756  62148757  62149005  62149006  62149007  62149008  62149034  62149035  62149036  62149364  62149517  62149519  62149520  62149521  62149532  62149538  62149539  62149549  62149550  62150094  62150114  62150117  62150120  62150489  62150490  62150491  62150523  62150526  62150527  62150528  62150531  62150533  62150534  62150538  62150540  62150557  62150558  62150560  62150562  62150563  62150564  62150566  62150569  62150571  62150572  62150574  62150578  62150628  62150629  62150630  62150631  62150632  62150633  62150634  62150635  62150636  62150637  62150638  62150640  62150642  62150844  62150859  62151228  62151239  62151242  62151761  62151764  62151770  62152206  62152223  62152616  62152641  62152900  62152914  62154037  62154038  62154261  62154264  62154271  62154993  62154995  62154996  62155129  62155131  62155189  62155195  62155273  62155483  62155503  62155504  62155505  62155512  62155513  62155514  62155515  62155516  62155517  62155518  62155520  62155521  62155522  62155523  62155524  62155525  62155526  62155527  62155528  62155530  62155531  62155532  62155533  62155535  62155536  62155537  62155538  62155539  62155540  62155542  62155543  62156805  62156820  62157851  62157852  62157854  62157952  62157953  62157954  62157955  62159257  62159260  62159323  62159324  62159326  62160154  62160155  62160156  62160157  62160158  62160276  62160277  62160330  62160465  62160466  62160467  62160468  62160469  62160470  62160471  62160472  62160473  62160474  62160481  62160482  62160504  62160781  62160782  62160784  62160786  62160788  62161068  62161194  62161195  62161197  62161203  62161207  62161208  62161210  62161212  62161221  62161597  62162196  62162197  62162198  62162206  62162215  62162216  62162218  62162219  62162228  62162747  62162753  62163073  62163098  62163101  62164295  62164301  62164668  62164815  62164861  62164862  62164951  62164964  62164965  62164966  62164969  62164970  62165851  62165874  62165946  62166648  62166650  62166666  62166738  62166750  62167492  62167494  62167519  62167521  62168033  62168035  62168036  62168061  62168063  62168119  62168351  62169035  62170568  62172341  62172587  62176042  62181237  62182673  62199880  62205227  62210351  62217448  62223770  62234304  62238218  62238219  77001340  77001341  77001342  77001343  77001344  77001345  77001346  77001347  77001348  77001349  77001350  77001351  77001352  77001353  77001354  77001355  77001356  77001357  77001359  77001360  77001361"
    }
  ]
}