  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73142",
      "recalling_firm": "Zimmer Biomet, Inc.",
      "address_1": "1800 W Center St",
      "address_2": "N/A",
      "postal_code": "46580-2304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US (nationwide) and Internationally to   ARGENTINA               AUSTRALIA  AUSTRIA  BARBADOS  BELGIUM  BERMUDA  BOLIVIA  BRAZIL  CARIBBEAN  CHILE  CHINA  COLOMBIA  COST RICA  DENMARK  DOMINICAN REPUBLIC  DORAL FL  ECUADOR  EL SALVADOR  ENGLAND  FRANCE  GERMANY  GUATEMALA  HONDURAS  HONG KONG  INDIA  ISRAEL  ITALY  JAMAICA  JAPAN  KOREA  MALAYSIA  MEXICO  NEW ZEALAND  NICARAGUA  PANAMA  PERU  SINGAPORE  SPAIN  SWEDEN  SWITZERLAND  TAIWAN  THAILAND  VENEZUELA  WEST INDIES  Canary Islands  Croatia  Cyprus  Czech Republic  Canada",
      "recall_number": "Z-0831-2016",
      "product_description": "NexGen CR-FLEX Option FEM  NexGen CR-FLEX Precoat FEM  NexGen CR-FLEX Porous FEM  prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer/metal/polymer  various sizes  \"\" This device is indicated for patients with severe knee pain and disability due to:   Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis.   Collagen disorders, and/or avascular necrosis of the femoral condyle.   Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy.   Moderate valgus, varus, or flexion deformities.   The salvage of previously failed surgical attempts or for a knee in which satisfactory stability in flexion cannot be obtained at the time of surgery. \" CR-Flex and LPS-Flex porous coated femoral and tibial baseplate components may be used cemented or uncemented (biological fixation). CR-Flex HA/TCP porous coated femoral and tibial baseplate components may only be used uncemented. All other femoral, tibial baseplate and all-polyethylene patella components are indicated for cemented use only. \" Specific uses with CR-Flex or LPS-Flex femorals:   Provides increased flexion capability for patients who have both the flexibility and desire to increase their flexion range.   The CR-Flex femoral, when used with 90-prefix NexGen CR articular surfaces, 00-prefix 10, 12 or 14 mm or 90-prefix 17 or 20 mm Prolong¿ Highly Crosslinked Polyethylene CR articular surfaces, is designed for use with a functional posterior cruciate ligament and when load bearing range of motion (ROM) is expected to be less than or equal to 155 degrees.   The LPS-Flex femoral, when used with the LPS-Flex articular surfaces, is designed for use with both cruciate ligaments excised and when load bearing ROM is expected to be less than or equal to 155 degrees. \" Specific Uses with NexGen LPS or LCCK femorals:   The LPS femoral is designed for use with both cruciate ligaments excised and when load bearing ROM is expected to be less than or equal to 130 degrees.   The usage of the LCCK femoral is the same as the LPS with the exception that femoral bone loss can be accommodated by femoral stem extensions and augmentation. No varus/valgus constraint or other stability other than that provided by the NexGen LPS is provided when the LCCK femoral component is used with the LPS-Flex articular surfaces.\"",
      "product_quantity": "42064",
      "reason_for_recall": "LDPE bag containing the implant adheres to the highly polished implant surface.",
      "recall_initiation_date": "20160111",
      "center_classification_date": "20160222",
      "termination_date": "20170523",
      "report_date": "20160302",
      "code_info": "Item numbers:  00-5950-012-01  00-5950-012-02  00-5950-013-01  00-5950-013-02  00-5950-013-05  00-5950-013-06  00-5950-014-01  00-5950-014-02  00-5950-014-05  00-5950-014-06  00-5950-015-01  00-5950-015-02  00-5950-015-05  00-5950-015-06  00-5950-016-01  00-5950-016-02  00-5950-016-05  00-5950-016-06  00-5950-017-01  00-5950-017-02  00-5950-017-05  00-5950-017-06  00-5952-012-01  00-5952-012-02  00-5952-013-01  00-5952-013-02  00-5952-013-05  00-5952-013-06  00-5952-014-01  00-5952-014-02  00-5952-014-05  00-5952-014-06  00-5952-015-01  00-5952-015-02  00-5952-015-05  00-5952-015-06  00-5952-016-01  00-5952-016-02  00-5952-016-05  00-5952-016-06  00-5952-017-01  00-5952-017-02  00-5952-017-05  00-5952-017-06  00-5956-012-01  00-5956-012-02  00-5956-013-01  00-5956-013-02  00-5956-013-05  00-5956-013-06  00-5956-014-01  00-5956-014-02  00-5956-014-05  00-5956-014-06  00-5956-015-01  00-5956-015-02  00-5956-015-05  00-5956-015-06  00-5956-016-01  00-5956-016-02  00-5956-016-05  00-5956-016-06  00-5956-017-01  00-5956-017-02  00-5956-017-05  00-5956-017-06    Lot numbers:  11003331  11003479  11004184  11004653  11004819  11004864  11004891  11004997  11005134  11005150  11005197  11005206  11005279  11005343  11005424  11005628  11005673  11005698  11005702  11005794  11005844  11005855  11005857  11005897  11005937  11005939  11005982  11006036  11006046  11006061  11006062  11006063  11006070  11006106  11006198  11006211  11006214  11006272  11006310  11006376  11006438  11006476  11006477  11006485  11006525  11006589  11006590  11006594  11006651  11006717  11006799  11006813  11006891  11006943  11006976  11007101  11007410  11007427  11400110  11400113  11400131  11400138  11400146  11400150  11400155  11400160  11400162  11400169  11400174  11400179  11400180  11400181  11400183  11400195  11400206  11400231  11400236  11400241  11400250  11400251  11400257  11400258  11400260  11400281  11400293  11400355  11400361  11400365  11400369  11400370  11400371  11400372  11400376  11400377  11400378  11400379  11400384  11400388  11400390  11400391  11400393  11400394  11400397  11400403  11400404  11400406  11400409  11400413  11400414  11400422  11400423  11400437  11400438  11400439  11400444  11400445  11400446  11400450  11400477  11400481  11400484  11400497  11400504  61748853  61766417  61772531  61777556  61782487  61782488  61782489  61782491  61782593  61782594  61782596  61782598  61782599  61782600  61782602  61789017  61789018  61789020  61789021  61789022  61789023  61789629  61789630  61789631  61789632  61789633  61789634  61795536  61795537  61795538  61795539  61795540  61795541  61795542  61795543  61795544  61795545  61795943  61795944  61795946  61795947  61795948  61795949  61795950  61795951  61795952  61800450  61800452  61801074  61801075  61801076  61801077  61801078  61801079  61801081  61801082  61801083  61801084  61802103  61802104  61802105  61803568  61804260  61804261  61804262  61804263  61805665  61805666  61805667  61805668  61805669  61805670  61806280  61806281  61807830  61807832  61808801  61808802  61808805  61808806  61808807  61808809  61810190  61810191  61810192  61810193  61810194  61810195  61810196  61810199  61810201  61810202  61810497  61810698  61810821  61810822  61810823  61810824  61810825  61810826  61810827  61810828  61811781  61811782  61811795  61813671  61813672  61813673  61813674  61814763  61814764  61815416  61815417  61815418  61815419  61815429  61815504  61815505  61815506  61815507  61815508  61815509  61815713  61815715  61816390  61816391  61816393  61816417  61816418  61817751  61817752  61817753  61817754  61817756  61819043  61819050  61819401  61819402  61819403  61819404  61819405  61819931  61821133  61821417  61821419  61822047  61822048  61822049  61822050  61822051  61822052  61822053  61822346  61822347  61822348  61822349  61822457  61822458  61822459  61822460  61822943  61822952  61822954  61823884  61823886  61824489  61824490  61824491  61824492  61824493  61824494  61824495  61824511  61824512  61824513  61824514  61824515  61824516  61824517  61824535  61824536  61824541  61824787  61826929  61826930  61826932  61826935  61827934  61827939  61827941  61829063  61829064  61829835  61830367  61830368  61830370  61830373  61831034  61831036  61831038  61831039  61831040  61831041  61831042  61831570  61831572  61831573  61831574  61831575  61831576  61831577  61832283  61832288  61832290  61832306  61832307  61833593  61833601  61833602  61833603  61834858  61834860  61835329  61835331  61835332  61835334  61835335  61835336  61835392  61835394  61835827  61835828  61835829  61835830  61835831  61835832  61835833  61835932  61835939  61835948  61836556  61836557  61836558  61836559  61836560  61836979  61836980  61836981  61836982  61836983  61836984  61837326  61839708  61839710  61839711  61839712  61839714  61839715  61839716  61840462  61840463  61840464  61840466  61840467  61840468  61840469  61840470  61840471  61840472  61841242  61841243  61841244  61841276  61841277  61841278  61841280  61841281  61842347  61842349  61842350  61842365  61843127  61844037  61844038  61844040  61844233  61844570  61844571  61844572  61844586  61844587  61844589  61844894  61844899  61844901  61844903  61844904  61844905  61844936  61844937  61844941  61844943  61844944  61845988  61845989  61845990  61845991  61846111  61846112  61846113  61846114  61846115  61846116  61846117  61846118  61846309  61846310  61846311  61847661  61847662  61847664  61847665  61848284  61848285  61848286  61848287  61848288  61848289  61848479  61848480  61848481  61848482  61848483  61848484  61849041  61849042  61849043  61849044  61849232  61850214  61850216  61850217  61850218  61850219  61850220  61850221  61850222  61850244  61850919  61850920  61850924  61850925  61850926  61852859  61852860  61853371  61853372  61853383  61853384  61853385  61853386  61854814  61854816  61854817  61854820  61854821  61854823  61855455  61855456  61855457  61855458  61855472  61855473  61855474  61855475  61855476  61855477  61855478  61855479  61855956  61855957  61855962  61855963  61855964  61857411  61857412  61857413  61857414  61857415  61857416  61857417  61857418  61857419  61857893  61857894  61858474  61858476  61859220  61859221  61859225  61859227  61859229  61860213  61860214  61860215  61860216  61860217  61860219  61860222  61860223  61860224  61861356  61861357  61861358  61861585  61861586  61861587  61861588  61861706  61861707  61861708  61861709  61861710  61861711  61861712  61861713  61861714  61863950  61864876  61864882  61864883  61864884  61865913  61865914  61866629  61866630  61866632  61866633  61866634  61866635  61866636  61866637  61866638  61866639  61866640  61866641  61866707  61866708  61866709  61867171  61867604  61867607  61867608  61867609  61867613  61867614  61867615  61868861  61868862  61868864  61868865  61868867  61868869  61868870  61869849  61869850  61869851  61869852  61869853  61869854  61869875  61870727  61871023  61871025  61871026  61871027  61871028  61871642  61871647  61871648  61871649  61871650  61872334  61872335  61872340  61872341  61873436  61873437  61873438  61873442  61873443  61873445  61873446  61874464  61874465  61875571  61875572  61875573  61875595  61875597  61876142  61876143  61876144  61876145  61876146  61876147  61876148  61876149  61876150  61876151  61876152  61876278  61876280  61876914  61876915  61876919  61876920  61876921  61876937  61878273  61880102  61880103  61881039  61881041  61881042  61881665  61881666  61884329  61884330  61884331  61884332  61884333  61884334  61884335  61884336  61884979  61884980  61884981  61884982  61884983  61884984  61884985  61884986  61884987  61884988  61884989  61885001  61885002  61885010  61885012  61885014  61885015  61885016  61885039  61885040  61885041  61885042  61885043  61885044  61885045  61885046  61885047  61885176  61885177  61885770  61885771  61885772  61885773  61885774  61885775  61885776  61885777  61888031  61888032  61888033  61888034  61888035  61888036  61888193  61888194  61888195  61888196  61888197  61888198  61888199  61888201  61888210  61888211  61888231  61888232  61888531  61893885  61893887  61893888  61893890  61893891  61893892  61893893  61893894  61893895  61893896  61893897  61893898  61893899  61894388  61894389  61894390  61894391  61894392  61894393  61894394  61894395  61894532  61894533  61894534  61894535  61894536  61895310  61895311  61895312  61895314  61895315  61895316  61895317  61895318  61895320  61895321  61897288  61898488  61898489  61898490  61898498  61898499  61898501  61898825  61899430  61899431  61899432  61899433  61899434  61899437  61899484  61899485  61899486  61899487  61899488  61899489  61899490  61899491  61899492  61900461  61901219  61903963  61903964  61903965  61903966  61903967  61903968  61904103  61904104  61904105  61904106  61904107  61904108  61904109  61904110  61904111  61904194  61904195  61904196  61904197  61904922  61905268  61906066  61906067  61906383  61906385  61906388  61907442  61907449  61907450  61907452  61907453  61908288  61908290  61908291  61909489  61910196  61910197  61910198  61910201  61910203  61910204  61910205  61910206  61910434  61910439  61910440  61910441  61910442  61910443  61910444  61910498  61910499  61910501  61910502  61910503  61910504  61910505  61910506  61910507  61910866  61910867  61910868  61910869  61910872  61912357  61915305  61915306  61915308  61915312  61915314  61915315  61915470  61915926  61915927  61915928  61915929  61915930  61915931  61915932  61915933  61915934  61915966  61915967  61915968  61915969  61915970  61915971  61915972  61915973  61915974  61916097  61916098  61916148  61916153  61917755  61917757  61917759  61918728  61918973  61918976  61918977  61920443  61920444  61920445  61920446  61920449  61921238  61921239  61921240  61921241  61921242  61921243  61921244  61921245  61921246  61921275  61921276  61921277  61921278  61921279  61921280  61921281  61921282  61921283  61921449  61921450  61922251  61922253  61922254  61922255  61922256  61923567  61923569  61923570  61923571  61926129  61926133  61926134  61926135  61927770  61927771  61927772  61927773  61927774  61927775  61927776  61927963  61927964  61927966  61927967  61927968  61928587  61928588  61928589  61928591  61928592  61928594  61928595  61929405  61929406  61929407  61930155  61930156  61932292  61932293  61933271  61933272  61933273  61933274  61933275  61933276  61933695  61933699  61933708  61933835  61933837  61934958  61934962  61934964  61934967  61935330  61935331  61935332  61935333  61935334  61935335  61935336  61935337  61935338  61935653  61935654  61935655  61936612  61936614  61936615  61936671  61936672  61936673  61936674  61936675  61936777  61936779  61937289  61937291  61937292  61937293  61937294  61937756  61937757  61937759  61937760  61937761  61938441  61938712  61938713  61939596  61939597  61939598  61939599  61939601  61939602  61939603  61939604  61939605  61939722  61939723  61939766  61939768  61939798  61939799  61939803  61939804  61940229  61940230  61940231  61940233  61940235  61940957  61940958  61940962  61940964  61940966  61940967  61941365  61941581  61941587  61942848  61942849  61942851  61943189  61943190  61943191  61946489  61946491  61946492  61946493  61946613  61946871  61947011  61947012  61947013  61947021  61947193  61947194  61948136  61948137  61949475  61949476  61949477  61949479  61949480  61949517  61949518  61949519  61949520  61949522  61949523  61949524  61949525  61949526  61949527  61949528  61949704  61949705  61949706  61949707  61949708  61951109  61951110  61951496  61951497  61951498  61951548  61951549  61951550  61951551  61951552  61951553  61951554  61951555  61951556  61951557  61951558  61951559  61951560  61951561  61951698  61951704  61952881  61953098  61953102  61953105  61953106  61954546  61955210  61955218  61955219  61955292  61955293  61955562  61955563  61955566  61955567  61955884  61956826  61956840  61956841  61958021  61958212  61958213  61958214  61958215  61958216  61958217  61958218  61958219  61958220  61958297  61958301  61958319  61958323  61958753  61958757  61959174  61959175  61960716  61961810  61961811  61961813  61963680  61963681  61963762  61963763  61963764  61963765  61963766  61963767  61963896  61964130  61966299  61966300  61967198  61967199  61968131  61968132  61968133  61968134  61968135  61968136  61968626  61968629  61968631  61968632  61968633  61968634  61968635  61968872  61968874  61971729  61972736  61972739  61973236  61973237  61973238  61973240  61974023  61974028  61974040  61974041  61974042  61974124  61974753  61974927  61974928  61974931  61974933  61974934  61974935  61974936  61974937  61974956  61974958  61977553  61977554  61978288  61978524  61979131  61979132  61979133  61979135  61979777  61979779  61980315  61980316  61980576  61982242  61982244  61982459  61982762  61982763  61982764  61982765  61982769  61982770  61982771  61982772  61982773  61982774  61982776  61982778  61982779  61982780  61983394  61983395  61983396  61983397  61983398  61983399  61983401  61984233  61985026  61985027  61985028  61986179  61986186  61986190  61986419  61986421  61986501  61986703  61986704  61986705  61986855  61986856  61987060  61987061  61987416  61987417  61987418  61987419  61987750  61987943  61987944  61987945  61987946  61987947  61987949  61987950  61988142  61988187  61988189  61988450  61989679  61990283  61990284  61990638  61990916  61990917  61990920  61991758  61992038  61992055  61992808  61992809  61994788  61994789  61994790  61994791  61994792  61994793  61994855  61994858  61994903  61994904  61994905  61994906  61994907  61994908  61994909  61995680  61995681  61995682  61996946  61997341  61997342  61997343  61998625  61998828  61998830  61999558  61999988  62000757  62000758  62000759  62000760  62000761  62000762  62000763  62000764  62000765  62000766  62000782  62000971  62000972  62000973  62001342  62001343  62001344  62001345  62001346  62001347  62001348  62002029  62002042  62002043  62002069  62005654  62005655  62005656  62005657  62005658  62005659  62005660  62005661  62005720  62005721  62005722  62005723  62005724  62005725  62005726  62005883  62005887  62005888  62005889  62005950  62005951  62005952  62006991  62006998  62006999  62007001  62007002  62007003  62007694  62008271  62008272  62008273  62008274  62008285  62008286  62009113  62009115  62009303  62009316  62010328  62010342  62010343  62011265  62011274  62011276  62012069  62012070  62012073  62012074  62012075  62012076  62012077  62012078  62012079  62012080  62012081  62012657  62012659  62012668  62012669  62013414  62013415  62013416  62013417  62013418  62013419  62013420  62013422  62013487  62013622  62013623  62014075  62014090  62014092  62015650  62015651  62015838  62015840  62015847  62015848  62015849  62015850  62015864  62015865  62015866  62015867  62016800  62016801  62016818  62018072  62018073  62018144  62018145  62018556  62018559  62019309  62019310  62019311  62019324  62019325  62019326  62019490  62019493  62019494  62019495  62019957  62019986  62019990  62019993  62019994  62020213  62020214  62020215  62020803  62020805  62020806  62020807  62022037  62022991  62022992  62022993  62023789  62023790  62023792  62023793  62023794  62024353  62024377  62024903  62024904  62024958  62025573  62025651  62025652  62025653  62025654  62025655  62025656  62025657  62025710  62025711  62025713  62025714  62025715  62025716  62025717  62026093  62026096  62026804  62026807  62027196  62028034  62030344  62030589  62030591  62030592  62030635  62030645  62030647  62030648  62031943  62031944  62031945  62031951  62031952  62031953  62031954  62031955  62031956  62031957  62032077  62032080  62032081  62032084  62032085  62032555  62032556  62032557  62032558  62032559  62032560  62032561  62032562  62032563  62033087  62034611  62036732  62036733  62038016  62038686  62038687  62038688  62038689  62038690  62038691  62038692  62038693  62038694  62038695  62038696  62038697  62038698  62038701  62038702  62038703  62038733  62038734  62038735  62038843  62038844  62038845  62038846  62038847  62038848  62038849  62038850  62038851  62039188  62039968  62039972  62040701  62040752  62040753  62040754  62040755  62040771  62041501  62043148  62043149  62043299  62043304  62043307  62043309  62043311  62043970  62043972  62043977  62043978  62044615  62044616  62044617  62045329  62045330  62045331  62046016  62046020  62046048  62046049  62046050  62046053  62046054  62046055  62046056  62046057  62046058  62046059  62046060  62046061  62046062  62046063  62046077  62046078  62046079  62046080  62046081  62046082  62046226  62046228  62051192  62051193  62051194  62051195  62052092  62052103  62052105  62052664  62052665  62052671  62052715  62052716  62054068  62054069  62054073  62054082  62054084  62054122  62054123  62054124  62054125  62054126  62054127  62054128  62054129  62054130  62054131  62054132  62054133  62054134  62054169  62054170  62054550  62054552  62054553  62054554  62054555  62054556  62054557  62054558  62054559  62054580  62054886  62054912  62055045  62055046  62055047  62055075  62055076  62055078  62055079  62056241  62056242  62056243  62056244  62056265  62056266  62056267  62056268  62056270  62056271  62056272  62056273  62056778  62056779  62056781  62056881  62056882  62056883  62056884  62056885  62056886  62056887  62056888  62057245  62057251  62057257  62058202  62058203  62058213  62058214  62058215  62058216  62058217  62058218  62058775  62058776  62058777  62059378  62059379  62060530  62060541  62060543  62061643  62061878  62061879  62061925  62061926  62061927  62061928  62061929  62061930  62061931  62061932  62061934  62061935  62061936  62061937  62061938  62061939  62061940  62061941  62061942  62061943  62061944  62061945  62062032  62062033  62062034  62062035  62062036  62062037  62062228  62062524  62063269  62063272  62064371  62064378  62064889  62064890  62064899  62065380  62065396  62066545  62066546  62066547  62066759  62066809  62066811  62067347  62067348  62067357  62067358  62067880  62067887  62067897  62067916  62067938  62067939  62067962  62067963  62067964  62067965  62067966  62067967  62067968  62067969  62067970  62067971  62067972  62067990  62067991  62067992  62067993  62067994  62067995  62067996  62068143  62068144  62068145  62068146  62068147  62068148  62068149  62068495  62068496  62069336  62069337  62069338  62069339  62070263  62070269  62070270  62070828  62070829  62070831  62071679  62072211  62072515  62073106  62073107  62073108  62074245  62074253  62074260  62074277  62074517  62074524  62074528  62074766  62074767  62074768  62074769  62074770  62074771  62074772  62074773  62074774  62074775  62074776  62074777  62074778  62074875  62074876  62074877  62074878  62074879  62075008  62075009  62075018  62077019  62077291  62077292  62077293  62077858  62077859  62077860  62077861  62078823  62079820  62079821  62079858  62080320  62080564  62080565  62080566  62080567  62080568  62080569  62080570  62080571  62080572  62080573  62080574  62080575  62080592  62080593  62080594  62080595  62083128  62083129  62084434  62084435  62084438  62085528  62085946  62085950  62085952  62086013  62086014  62086358  62086766  62086767  62086768  62086769  62086770  62086772  62086773  62086774  62086775  62086776  62086796  62086797  62086798  62086799  62086800  62086801  62088406  62088642  62088875  62088876  62088877  62088878  62088880  62090141  62090392  62090393  62090396  62090399  62091413  62091445  62091455  62091544  62091545  62092080  62092254  62092261  62092266  62092267  62092268  62092359  62092360  62092361  62092362  62092363  62092364  62092365  62092427  62092428  62092429  62092430  62095226  62095237  62095238  62096008  62096009  62096620  62096624  62096706  62096707  62096708  62096709  62096710  62096711  62098056  62098058  62098059  62098061  62098063  62098283  62098518  62099036  62101528  62101557  62101853  62101854  62102060  62102062  62102063  62102778  62102781  62103214  62103215  62103528  62103532  62103533  62103996  62103997  62103998  62103999  62104001  62104002  62104003  62104144  62104145  62104146  62104147  62104148  62104149  62104851  62104857  62105276  62105360  62105361  62105362  62105369  62105370  62105716  62105717  62107948  62109164  62109665  62110234  62110583  62110584  62110585  62110586  62110587  62110588  62110589  62110606  62110607  62110608  62110609  62110610  62110798  62110813  62111948  62111949  62112677  62113001  62113004  62114024  62114097  62114438  62115962  62117115  62117116  62117117  62117118  62117119  62117120  62117145  62117148  62118575  62118577  62119165  62119166  62119547  62120695  62121196  62121231  62121615  62121617  62121618  62122757  62122862  62122864  62124280  62124281  62124284  62124817  62125122  62125143  62125781  62126113  62126116  62126734  62128164  62128165  62128166  62128167  62128168  62128169  62128170  62128243  62128244  62128245  62129495  62131187  62131226  62131227  62131250  62131251  62131842  62131847  62132084  62132085  62132109  62132174  62132175  62132176  62132534  62132543  62134710  62134732  62134746  62135962  62135965  62135968  62137061  62137513  62137514  62138257  62138677  62138680  62138682  62139379  62139380  62139381  62139382  62139383  62139384  62139385  62139971  62142596  62142615  62143160  62143164  62143176  62143618  62143619  62143621  62143622  62144130  62144134  62144291  62144314  62144315  62145118  62145119  62145120  62145121  62145122  62145123  62145402  62145403  62145766  62145767  62145818  62145819  62145820  62147814  62147816  62148261  62148758  62148997  62149009  62149011  62149012  62149038  62149369  62149474  62149524  62149525  62149542  62150028  62150029  62150730  62150731  62150732  62150733  62150734  62150736  62150738  62152643  62152923  62152924  62152925  62152930  62154040  62154042  62154071  62154073  62154273  62154278  62154960  62154961  62154962  62154997  62154999  62155266  62155269  62155271  62155390  62155391  62155682  62156808  62157838  62160336  62160789  62161198  62161199  62161214  62161215  62161593  62162201  62162203  62162222  62162223  62162225  62162234  62162763  62162769  62163075  62163076  62163102  62164313  62166744  62166745  62167499  62167506  62167509  62167510  62168042  62168043  62168046  62168050  62168051  62168120  62168121  62168122  62168123  62168124  61824516R  61836560R  62025657R",
      "more_code_info": ""
    }
  ]
}