  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Plainsboro",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "69803",
      "recalling_firm": "Integra LifeSciences Corp.",
      "address_1": "311 Enterprise Dr",
      "address_2": "N/A",
      "postal_code": "08536-3344",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide, Foreign",
      "recall_number": "Z-0831-2015",
      "product_description": "Integra Licox Brain Tissue Oxygen Monitoring  Integra Licox Recon Tissue Oxygen Monitoring    1) LICOX¿ Complete Brain Probe Kit (comprised of IP1, CC1.P1)  2) LICOX¿ Complete Brain Probe Kit (comprised of IP2, CC1.P1)  3) LICOX¿ Complete Brain Tunnelling Probe Kit (comprised of VK5.2, CC1.P1  4)  LICOX¿ Complete Brain Probe Kit (comprised of IM1, CC1.SB)    Sterile Rx Only  Manufacturer CMS -Gesse;schaft for medizinische Sondentechnik mbH Dorfstrasse 2 24247 <eilkendorf Germany Distributed by Integra NeuroSciences 311 Enterprise Drive Plainsboro, NJ 08536, USA",
      "product_quantity": "2,506 units",
      "reason_for_recall": "Due to a complaint, it was determined that multiple lots of Licox catheters manufactured since October 29, 2013 contained a smart card which was programmed with an inaccurate lkN2 value.",
      "recall_initiation_date": "20141107",
      "center_classification_date": "20141218",
      "termination_date": "20150811",
      "report_date": "20141224",
      "code_info": "CC1   Licox¿ PO2 Microprobe, 200 mm  021213, 040414, 070314, 130114, 240214, 270114,  280314, 300614    CC12   Licox¿ PO2 Microprobe, 300 mm   020614, 040814, 050514, 070414, 070714, 220414    CC1.P1  Licox¿ PMO Combined Probe Containing Oxygen and  Temperature  050614, 060214, 060314, 100414, 100714, 130214,  130314, 170114, 190614, 210814, 260614, 310714    IM3ST  Licox¿ Complete Brain IMC-Probe  Kit:  020614, 030314, 110714, 130114, 140314, 160614,  180714, 200114, 280314    IP1.P  Licox¿ Single Lumen Bolt CMPS  and Temperature Bolt Kit  030714, 060214, 060314, 070814, 080514, 100714,  120614, 130314, 190614, 310714    IP2.P  Licox¿ Double Lumen Bolt CMPS,  Temperature and ICP Monitoring  Kit  030714, 050614, 060214, 060314, 070814, 080514,  090114, 100714, 130314, 140814, 200214, 200414,  260614, 290814    IT2  Licox¿ Tunneling CMPS and  Temperature Monitoring Kit  090114, 100414, 190614, 200214, 260614, 310714"
    }
  ]
}