      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73142",
      "recalling_firm": "Zimmer Biomet, Inc.",
      "address_1": "1800 W Center St",
      "address_2": "N/A",
      "postal_code": "46580-2304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US (nationwide) and Internationally to   ARGENTINA               AUSTRALIA  AUSTRIA  BARBADOS  BELGIUM  BERMUDA  BOLIVIA  BRAZIL  CARIBBEAN  CHILE  CHINA  COLOMBIA  COST RICA  DENMARK  DOMINICAN REPUBLIC  DORAL FL  ECUADOR  EL SALVADOR  ENGLAND  FRANCE  GERMANY  GUATEMALA  HONDURAS  HONG KONG  INDIA  ISRAEL  ITALY  JAMAICA  JAPAN  KOREA  MALAYSIA  MEXICO  NEW ZEALAND  NICARAGUA  PANAMA  PERU  SINGAPORE  SPAIN  SWEDEN  SWITZERLAND  TAIWAN  THAILAND  VENEZUELA  WEST INDIES  Canary Islands  Croatia  Cyprus  Czech Republic  Canada",
      "recall_number": "Z-0830-2016",
      "product_description": "LPS-FLEX GSF OPT   prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer/metal/polymer  Various sizes¿  Product Usage\t\"\" This device is indicated for patients with severe knee pain and disability due to:   Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis.   Collagen disorders, and/or avascular necrosis of the femoral condyle.   Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy.   Moderate valgus, varus, or flexion deformities.   The salvage of previously failed surgical attempts or for a knee in which satisfactory stability in flexion cannot be obtained at the time of surgery. \" The CR-Flex GSF and LPS-Flex GSF uncoated (Option) and precoat styles are intended for cemented use only. \" The CR-Flex GSF and LPS-Flex GSF porous coated femoral components may be used cemented or uncemented (biological fixation). \" Specific uses with CR-Flex GSF or LPS-Flex GSF femorals:   Provides increased flexion capability for patients who have both the flexibility and desire to increase their flexion range.   The CR-Flex GSF femoral, when used with 90-prefix NexGen CR articular surfaces, 00-prefix 10, 12 or 14 mm or 90-prefix 17 or 20 mm Prolong¿ Highly Crosslinked Polyethylene CR articular surfaces, or with Gender Solutions Natural-Knee Flex Congruent articular surfaces, is designed for use with a functional posterior cruciate ligament and when load bearing range of motion (ROM) is expected to be less than or equal to 155 degrees.   The LPS-Flex GSF femoral, when used with LPS-Flex articular surfaces, is designed for use with both cruciate ligaments excised and when load bearing ROM is expected to be less than or equal to 155 degrees.\"",
      "product_quantity": "84396",
      "reason_for_recall": "LDPE bag containing the implant adheres to the highly polished implant surface.",
      "recall_initiation_date": "20160111",
      "center_classification_date": "20160222",
      "termination_date": "20170523",
      "report_date": "20160302",
      "code_info": "Item numbers:  00-5764-013-51  00-5764-013-52  00-5764-014-51  00-5764-014-52  00-5764-015-51  00-5764-015-52  00-5764-016-51  00-5764-016-52  00-5764-017-51  00-5764-017-52    Lots:  11003937  11004287  11004298  11004337  11004544  11004612  11004769  11004797  11004818  11005062  11005066  11005155  11005276  11005598  11005680  11005705  11005748  11005900  11005931  11005959  11005980  11006190  11006737  11006764  11006827  11006840  11006882  11006922  11006986  11007165  11007217  11007272  11007323  11007651  11007849  11007988  11008186  11008197  11008221  11008231  11008311  11400099  11400104  11400119  11400120  11400122  11400125  11400126  11400134  11400152  11400165  11400170  11400187  11400203  11400215  11400218  11400244  11400247  11400270  11400271  11400274  11400284  11400339  11400359  11400360  11400364  11400381  11400417  11400418  11400427  11400428  11400448  11400449  11400451  11400452  11400455  11400456  11400457  11400464  11400482  11400488  11400489  11400490  11400492  11400499  11400500  11400506  11400510  11400526  11400532  11400550  11400553  61772478  61777279  61777283  61777284  61782390  61782391  61782393  61782394  61782395  61789567  61789568  61789569  61789570  61789572  61789574  61789575  61789576  61789577  61789578  61792783  61793532  61793549  61793551  61793552  61793554  61794709  61794711  61794712  61794730  61795524  61795525  61795526  61795527  61795528  61795529  61795530  61795531  61795532  61795533  61795534  61795535  61796363  61796380  61796381  61796382  61796461  61796463  61796474  61796476  61797741  61797742  61797743  61797744  61797745  61797746  61797747  61797748  61797749  61797757  61797758  61797759  61797760  61798618  61798619  61798620  61798621  61798622  61798623  61798625  61798637  61798638  61798639  61800345  61800346  61800347  61800348  61800349  61800350  61800351  61800352  61800353  61800354  61800355  61800356  61800357  61800358  61800603  61800604  61800605  61800606  61800607  61800608  61800609  61800628  61800629  61802093  61802095  61802096  61802097  61802113  61802114  61802115  61802116  61802117  61803347  61803374  61803379  61803557  61803558  61803575  61803576  61803577  61803578  61803579  61803580  61803581  61803582  61803583  61803584  61803585  61803586  61803587  61804248  61804249  61804250  61804251  61804252  61804254  61804256  61804257  61804258  61804259  61804278  61804279  61804280  61804281  61804282  61804283  61804284  61804285  61804728  61804729  61804730  61804748  61804749  61804750  61804751  61804752  61805095  61805096  61805097  61805098  61805099  61805100  61805101  61805102  61805103  61805104  61805105  61805106  61805107  61805108  61805109  61805110  61805111  61805112  61805113  61806269  61806270  61806272  61806277  61806295  61806296  61807824  61807825  61807826  61807827  61807829  61807840  61807841  61807842  61807843  61807844  61807845  61807852  61807853  61807854  61807855  61807856  61808797  61808798  61808799  61809582  61809584  61809736  61809737  61809738  61809739  61809740  61809741  61809742  61809743  61810427  61810428  61810429  61810430  61810431  61810432  61810786  61810787  61810788  61810789  61810790  61810791  61810792  61810793  61810794  61810795  61810796  61810798  61811788  61811789  61811790  61811791  61811792  61811793  61812441  61813680  61813681  61813682  61813683  61813684  61813695  61813696  61813697  61814505  61814506  61814752  61814753  61814755  61814756  61814757  61814771  61814772  61814774  61814775  61814776  61815420  61815421  61815422  61815423  61815424  61815425  61815426  61815427  61815428  61815702  61815703  61815704  61815705  61815721  61815724  61815725  61815726  61816384  61816387  61816388  61816403  61816404  61816406  61816407  61816550  61816675  61816678  61817747  61817748  61817769  61817770  61817774  61819047  61819048  61819066  61819069  61819070  61819078  61819079  61819080  61819226  61819228  61819230  61819233  61819234  61819235  61819236  61819237  61819249  61819252  61819256  61819261  61819264  61819275  61819406  61819407  61819408  61819948  61819949  61819950  61819951  61819952  61820659  61820660  61820661  61820662  61820663  61820664  61820665  61820666  61820733  61820734  61821128  61821129  61821145  61821148  61821149  61821150  61821410  61821412  61821413  61821414  61821430  61821431  61821432  61821433  61821434  61821436  61821437  61821438  61822035  61822036  61822037  61822038  61822039  61822040  61822041  61822042  61822043  61822044  61822045  61822046  61822550  61822551  61822940  61822941  61822942  61822961  61822962  61822978  61822980  61822981  61822982  61823899  61823901  61823902  61823903  61823904  61823968  61824453  61824454  61824455  61824456  61824457  61824458  61824459  61824460  61824461  61824462  61824463  61824464  61824773  61824774  61824775  61824776  61824791  61824792  61824794  61824795  61824937  61825250  61825254  61825256  61825267  61825268  61825269  61825270  61826924  61826925  61826926  61826927  61826945  61826946  61826947  61826948  61827928  61827929  61827930  61827931  61827932  61827950  61829072  61829073  61829074  61829075  61829077  61829142  61829690  61829747  61829857  61829858  61829860  61830382  61830383  61830384  61830385  61830386  61830387  61831075  61831076  61831077  61831078  61831079  61831080  61831081  61831084  61831086  61831087  61831088  61832265  61832266  61832267  61832268  61832269  61832277  61832278  61832280  61832281  61832282  61832297  61832301  61832302  61832303  61832446  61832447  61832455  61832456  61832457  61833580  61833581  61833589  61833590  61833591  61833592  61833611  61833612  61833613  61833620  61833621  61833623  61834862  61834864  61834865  61834866  61834867  61834879  61834880  61834881  61834882  61835001  61835349  61835351  61835352  61835353  61835363  61835364  61835365  61835366  61835846  61835847  61835850  61835851  61835852  61836952  61836953  61836954  61836955  61836956  61836957  61836958  61836959  61836960  61836961  61837337  61837338  61837339  61837340  61837341  61837342  61837359  61837684  61838153  61838160  61838164  61838205  61838206  61838207  61838211  61838230  61838767  61838768  61838769  61838772  61838773  61839695  61839698  61839699  61839703  61839704  61839717  61839718  61839719  61839720  61839721  61839722  61839723  61840453  61840454  61840479  61840483  61840484  61840485  61840486  61840487  61840488  61840588  61840589  61842367  61842368  61843142  61843143  61843985  61843986  61843987  61843988  61844044  61844045  61844046  61844047  61844048  61844049  61844241  61844243  61844245  61844246  61844594  61844595  61844596  61844597  61844915  61844916  61844917  61844918  61844919  61844920  61844923  61844924  61844925  61844926  61844958  61844959  61844961  61844962  61844963  61844964  61845040  61845045  61845970  61845971  61845972  61845973  61846272  61846273  61846274  61846275  61846276  61846277  61846278  61846279  61846280  61846283  61846284  61846285  61846286  61846380  61847671  61847672  61847673  61847674  61847675  61847676  61847681  61847682  61847712  61847713  61848303  61848304  61848305  61848306  61848307  61848308  61848309  61848310  61848311  61848312  61848385  61848386  61848387  61849058  61849059  61849060  61849061  61849062  61849063  61849064  61849065  61850233  61850234  61850235  61850236  61850237  61850238  61850239  61850240  61850241  61850939  61850940  61850941  61850942  61850943  61850944  61850945  61850946  61851734  61851735  61851736  61851737  61851738  61851739  61851740  61851741  61851742  61852865  61852866  61852867  61852868  61852869  61852870  61852873  61852874  61853391  61853392  61853395  61853396  61853397  61853398  61853399  61853401  61854798  61854799  61854801  61854803  61854804  61854805  61854806  61854807  61854808  61854831  61854832  61855459  61855460  61855461  61855462  61855463  61855464  61855940  61855942  61855943  61855944  61855945  61855946  61855947  61855949  61855950  61855972  61856753  61856754  61856756  61856757  61856758  61856759  61856760  61857133  61857134  61857135  61857136  61857137  61857138  61857139  61857140  61857875  61857876  61857877  61857878  61857879  61857880  61857908  61857911  61858489  61858490  61858491  61858492  61858493  61858494  61858495  61858496  61859241  61859242  61859243  61859246  61859247  61859248  61859249  61860196  61860197  61860198  61860199  61860201  61860202  61860226  61860227  61861371  61861372  61861373  61861374  61861375  61861589  61861590  61861591  61861592  61861593  61861594  61862148  61862149  61862150  61862151  61862152  61862153  61862154  61862155  61862156  61862157  61862753  61862756  61862763  61863956  61863957  61863962  61863963  61863964  61863965  61863966  61863967  61863968  61863970  61864893  61864894  61864895  61864896  61864897  61864898  61864899  61864901  61865928  61865929  61865930  61865931  61865932  61865933  61865934  61865935  61865936  61865957  61865958  61865959  61865960  61865961  61865962  61865963  61865964  61865965  61865966  61866577  61866578  61866579  61866580  61866581  61866582  61866583  61866584  61866585  61866586  61866587  61866726  61866728  61866729  61866730  61866731  61866735  61866736  61866737  61867172  61867173  61867174  61867618  61867619  61867620  61867621  61867623  61867624  61867625  61867626  61867627  61867628  61867629  61867630  61867631  61867632  61868879  61868880  61868881  61868882  61868883  61868884  61868885  61868888  61869865  61869866  61869867  61869868  61869870  61869871  61869872  61870743  61870744  61870745  61870746  61870747  61870748  61870749  61870750  61870751  61870752  61870753  61870754  61871074  61871075  61871076  61871077  61871078  61871079  61871080  61871081  61871082  61871083  61871084  61871085  61871086  61871088  61871089  61871090  61871091  61871092  61871093  61871094  61871534  61871535  61871536  61871537  61871538  61871539  61871540  61871543  61871544  61871545  61871546  61871547  61871555  61871556  61871652  61871653  61871654  61871655  61871656  61871657  61871658  61871659  61871660  61871661  61871662  61871663  61871664  61871665  61871666  61871667  61871668  61871669  61871670  61871671  61872346  61872347  61872348  61872349  61872350  61872351  61872352  61872353  61872354  61872482  61872483  61873452  61873453  61873454  61873455  61873456  61873457  61873458  61873459  61873460  61873461  61873462  61874479  61874480  61874482  61874483  61874484  61874485  61874486  61875455  61875456  61875457  61875584  61875585  61875586  61875587  61875588  61875589  61875590  61875591  61875592  61875593  61875610  61876296  61876297  61876927  61876928  61876930  61876932  61876933  61876934  61876935  61876936  61878277  61878278  61878279  61878282  61878283  61878284  61878287  61878288  61878289  61878290  61878291  61878292  61879234  61879236  61879237  61879239  61879240  61879241  61880093  61880094  61880095  61880110  61880391  61880392  61881031  61881033  61881054  61881055  61881056  61881057  61881656  61881658  61881688  61881689  61881690  61881693  61883278  61883279  61883282  61884199  61884201  61884396  61884398  61884836  61884837  61884838  61884839  61884840  61884841  61884842  61884843  61884844  61884845  61884846  61884847  61884848  61885150  61885152  61885153  61885154  61885155  61885162  61885163  61885164  61885184  61885760  61885762  61885763  61885764  61885765  61885783  61885784  61885785  61885787  61885788  61886275  61886276  61886277  61886278  61888052  61888053  61888054  61888556  61888557  61888558  61888559  61888560  61888561  61888562  61888563  61888564  61889070  61889071  61889175  61889177  61889179  61890012  61890014  61891917  61891918  61891920  61891924  61891926  61891928  61892384  61892385  61892386  61892388  61892718  61892719  61893245  61893246  61893248  61893249  61893250  61894376  61894377  61894378  61894379  61894380  61894381  61894382  61894383  61894384  61894385  61894386  61894387  61894542  61894543  61894544  61895332  61895333  61895334  61896053  61896054  61896055  61896057  61896060  61896061  61896062  61897296  61897297  61897299  61897302  61897303  61898508  61898509  61898512  61898513  61898514  61898515  61898516  61899439  61899440  61899441  61899442  61899443  61899444  61899445  61899446  61899447  61899448  61899449  61899450  61899647  61899649  61899650  61900473  61901227  61901229  61901230  61901231  61902965  61902967  61902969  61904005  61904006  61904007  61904008  61904009  61904010  61904011  61904012  61904013  61904014  61904015  61904016  61904937  61904938  61904940  61905203  61905205  61905206  61905295  61905296  61905297  61905298  61906072  61906074  61906077  61906391  61906392  61907460  61907461  61907465  61907466  61907468  61908308  61908310  61908311  61908312  61909353  61909354  61909355  61910173  61910174  61910175  61910176  61910177  61910178  61910179  61910180  61910181  61910182  61910183  61910184  61910218  61910219  61910220  61910221  61910222  61910223  61910885  61910886  61910887  61912340  61912341  61912342  61912343  61912344  61912345  61912348  61912349  61912351  61912366  61912367  61912368  61912369  61912370  61913177  61913178  61913179  61913180  61913181  61913182  61913183  61913184  61913185  61913186  61913205  61913206  61913207  61913208  61914445  61914446  61914447  61914448  61914449  61914450  61914454  61914455  61914462  61914464  61914465  61914466  61914467  61914469  61914470  61914471  61914472  61914473  61915326  61915330  61915331  61915333  61915914  61915915  61915916  61915917  61915918  61915919  61915920  61915921  61915922  61915923  61915924  61915925  61916103  61916104  61916168  61916169  61916170  61917768  61917769  61917770  61918699  61918701  61918702  61918703  61918736  61918737  61918738  61918988  61918989  61918990  61920434  61920435  61920459  61920460  61920461  61920462  61920463  61920465  61920466  61920467  61920468  61920469  61920470  61921224  61921225  61921226  61921227  61921228  61921229  61921230  61921231  61921232  61921233  61921234  61921235  61921236  61921237  61921437  61921438  61921439  61921440  61921441  61921442  61921461  61921462  61921463  61921464  61921465  61921466  61921467  61921468  61921469  61921470  61921471  61921475  61921476  61922246  61922247  61922272  61922273  61922274  61922275  61922277  61922278  61922279  61924471  61924473  61924474  61924475  61924476  61924477  61926149  61926150  61926151  61926152  61926154  61926155  61926157  61927701  61927702  61927703  61927704  61927705  61927706  61927707  61927708  61927709  61927710  61927783  61927784  61927785  61927978  61927980  61927982  61927983  61927984  61927985  61927987  61929418  61929420  61929421  61929423  61929425  61929426  61930165  61930166  61930167  61930168  61931688  61931689  61931690  61931691  61931692  61931693  61931694  61931695  61931696  61931697  61931698  61932304  61932305  61932306  61932308  61932309  61932310  61932312  61932313  61933259  61933260  61933261  61933262  61933263  61933264  61933265  61933266  61933267  61933268  61933305  61933306  61933815  61933817  61933820  61933845  61933847  61933848  61933849  61933851  61933853  61934980  61934981  61934982  61934983  61934984  61934985  61934986  61934987  61935998  61936002  61936003  61936535  61936536  61936537  61936538  61936539  61936540  61936664  61936665  61936811  61936812  61936813  61936814  61936815  61936816  61936818  61936819  61936820  61936821  61936822  61937302  61937303  61937813  61937814  61937815  61937816  61937817  61937818  61937819  61937820  61937821  61937822  61938592  61938593  61938594  61938595  61938596  61938599  61938600  61938979  61938982  61939192  61939193  61939194  61939195  61939196  61939197  61939199  61939201  61939202  61939560  61939563  61939564  61939565  61939811  61939812  61939813  61939814  61939815  61940974  61940975  61941345  61941346  61941600  61941601  61941602  61943055  61943057  61943058  61943060  61943101  61943198  61943199  61943200  61943201  61943202  61944651  61944652  61944653  61944654  61944903  61944904  61944905  61944907  61944908  61944909  61944910  61944911  61944912  61946150  61946152  61946155  61946156  61946157  61946451  61946452  61946453  61946457  61946458  61946460  61946469  61946470  61946471  61946473  61946506  61946512  61946517  61946526  61946528  61946531  61946733  61946734  61946735  61946736  61946955  61946956  61946968  61946969  61946970  61946971  61946972  61946973  61946993  61946994  61946995  61946996  61946997  61946998  61946999  61947001  61947002  61947016  61947017  61947024  61947201  61947202  61947203  61947204  61947205  61947207  61947208  61947209  61947215  61947216  61947217  61947219  61948142  61948143  61948144  61948145  61948146  61948148  61948149  61949443  61949444  61949445  61949446  61949447  61949449  61949715  61949716  61949717  61949718  61949721  61949722  61949723  61949726  61949727  61949728  61950810  61950811  61950813  61950814  61950815  61950816  61950817  61951488  61951489  61951490  61951491  61951492  61951493  61951715  61951716  61951717  61952889  61952890  61952893  61952894  61952948  61952949  61952950  61952951  61952952  61952953  61953108  61953110  61953111  61953112  61954532  61954533  61954550  61954909  61954915  61954917  61954926  61954929  61954960  61954974  61955449  61955450  61955451  61955557  61955561  61955568  61955569  61955575  61955576  61955577  61955578  61956810  61956820  61956822  61956831  61956832  61956833  61956834  61956835  61956836  61956842  61956843  61957815  61957817  61957818  61957825  61957829  61957833  61957892  61957893  61957894  61958013  61958014  61958015  61958016  61958017  61958018  61958020  61958032  61958033  61958228  61958229  61958230  61958231  61958232  61958270  61958272  61958280  61958282  61958285  61958288  61958328  61958330  61958747  61958749  61958750  61958751  61958752  61958766  61960705  61960726  61960728  61960729  61960731  61960733  61960734  61960735  61961823  61961824  61961825  61961826  61961827  61961828  61961829  61961830  61962885  61962886  61962887  61963693  61963696  61963753  61963754  61963755  61963756  61963757  61963758  61963759  61963760  61963906  61963909  61963910  61964183  61964190  61964435  61964436  61964437  61964447  61964448  61964449  61964452  61965500  61965507  61965509  61965575  61965583  61965584  61965585  61966176  61966230  61966231  61966238  61966249  61966250  61966251  61966663  61966664  61967191  61967195  61967202  61967203  61967204  61967214  61967894  61967895  61967896  61967897  61967906  61967908  61967910  61968123  61968124  61968125  61968126  61968127  61968128  61968129  61968130  61970694  61971130  61971131  61971132  61971166  61971168  61971172  61971173  61971175  61971176  61971179  61971183  61971192  61971737  61971738  61971743  61971744  61971745  61971746  61971747  61972177  61972196  61972197  61972751  61972752  61972753  61972758  61973073  61973083  61973260  61973261  61973262  61973263  61974059  61974060  61974113  61974138  61974139  61974140  61974143  61974144  61974145  61974747  61974748  61974749  61974750  61974751  61974752  61976052  61976058  61976332  61976333  61976335  61976336  61976338  61976339  61976341  61976423  61976424  61976425  61976426  61976428  61976429  61977397  61977564  61977568  61977570  61977571  61977572  61978304  61978306  61978307  61978546  61978547  61978548  61978549  61978550  61979140  61979141  61979142  61979144  61979787  61979788  61979789  61979792  61979794  61980328  61980341  61980342  61980590  61980591  61980593  61982321  61982322  61982329  61982330  61982475  61982479  61982954  61982955  61982956  61982957  61982958  61982959  61984189  61984191  61984194  61984196  61984213  61984215  61984217  61984220  61984247  61984248  61984249  61984893  61984894  61985012  61985014  61985034  61985035  61985036  61985039  61985040  61985041  61986098  61986424  61987067  61987393  61987394  61987395  61987396  61987397  61987398  61987399  61987400  61987401  61987402  61987403  61987404  61987405  61988204  61988207  61989669  61989670  61989671  61990930  61990931  61990933  61990936  61992042  61992043  61992044  61992045  61992066  61992068  61992986  61992987  61992988  61992989  61992990  61992991  61993965  61993969  61993971  61993975  61994226  61994227  61994228  61994229  61994230  61994231  61994232  61994233  61994234  61994235  61994237  61994239  61994240  61994241  61994242  61994243  61995666  61995668  61995690  61995691  61995692  61995693  61995695  61995697  61996340  61996341  61996352  61996354  61996912  61996913  61996914  61996915  61996961  61996963  61997332  61997333  61997348  61997349  61997355  61997654  61997656  61997657  61998604  61998606  61998609  61998611  61998653  61998654  61998655  61998659  61998660  61998795  61998819  61998821  61998822  61998823  61998824  61998835  61999547  61999548  61999549  61999550  61999566  61999567  61999568  61999569  61999570  62000055  62000070  62000086  62000124  62000129  62000428  62000429  62000430  62000431  62000432  62000433  62000434  62000435  62000436  62000437  62000438  62000439  62000440  62000441  62000442  62000443  62000980  62000981  62000982  62002046  62002047  62002048  62002049  62002074  62002075  62002076  62003184  62003371  62003521  62003526  62003782  62003784  62003785  62003786  62003787  62003788  62003789  62003790  62004881  62004883  62004898  62004900  62004989  62004990  62004991  62004992  62005008  62005009  62005012  62005013  62005014  62005638  62005639  62005640  62005641  62005642  62005643  62005644  62005645  62005646  62005647  62005648  62005649  62005946  62005947  62005960  62005961  62005962  62005966  62006636  62006640  62006644  62006993  62006994  62007009  62007010  62007011  62007012  62008262  62008263  62008264  62008265  62008266  62008267  62008268  62008283  62008284  62008766  62008767  62008786  62008977  62008978  62008979  62009297  62009312  62009313  62009317  62009318  62009319  62010340  62010341  62010345  62010346  62011155  62011158  62011160  62011346  62011347  62011552  62011553  62011554  62011572  62011573  62011574  62011575  62011580  62011581  62011582  62011583  62012057  62012058  62012059  62012060  62012061  62012062  62012063  62012064  62012065  62012066  62012067  62012068  62012358  62012362  62012397  62012398  62012652  62012653  62012665  62012666  62013466  62013467  62013571  62013659  62014070  62014071  62014094  62014760  62014765  62014770  62014776  62015861  62015862  62016813  62016815  62016820  62016821  62017268  62017269  62017270  62017271  62017273  62017636  62017637  62017891  62017899  62017900  62017901  62017902  62018153  62018154  62018155  62018157  62018579  62018581  62018798  62018799  62018800  62018833  62018834  62018835  62018836  62018837  62018838  62019507  62019509  62019512  62019513  62020277  62020801  62020802  62020814  62020815  62020816  62020819  62021892  62021943  62022078  62022089  62022102  62022103  62022293  62022315  62022318  62022331  62022332  62022333  62022334  62022338  62022339  62022985  62022988  62023303  62023305  62023308  62023331  62023335  62023336  62023805  62024578  62024579  62024585  62024587  62024588  62024964  62024966  62024967  62025682  62025683  62025684  62025685  62025686  62026101  62026102  62026105  62027208  62027209  62027211  62027212  62027213  62028435  62028458  62029871  62030270  62030455  62030475  62030478  62030577  62030579  62030581  62030609  62030610  62030611  62030612  62030614  62030616  62030618  62030619  62030629  62030630  62030642  62030643  62031968  62031970  62031971  62031972  62031973  62031974  62031975  62031976  62032078  62032091  62032092  62032093  62032094  62032095  62032625  62032627  62032991  62032995  62033081  62033082  62033083  62033095  62033096  62033743  62034621  62034622  62034625  62035848  62035849  62035853  62035857  62036139  62036149  62036150  62036151  62036152  62036153  62036154  62036155  62036156  62036665  62036667  62036759  62038025  62038026  62038028  62038029  62038030  62038031  62038768  62038769  62038770  62038771  62038772  62038773  62039075  62039077  62039088  62039090  62039172  62039182  62039183  62039184  62039185  62039943  62039949  62039952  62039995  62040745  62040746  62040747  62040774  62042057  62042328  62042329  62042330  62043156  62043157  62043320  62043321  62045333  62045334  62045335  62045787  62045788  62045789  62045790  62045791  62045792  62045793  62045794  62046046  62046047  62046240  62046241  62046242  62046245  62046927  62046930  62047518  62047532  62047533  62047534  62047535  62047536  62048286  62048287  62048288  62048435  62048447  62048448  62048450  62049384  62049385  62049386  62049405  62049406  62049414  62050075  62050076  62050077  62050089  62050090  62050092  62050093  62050094  62051188  62051189  62051190  62051208  62051209  62051210  62051211  62052006  62052009  62052022  62052131  62052133  62052134  62052136  62052660  62052681  62052682  62052683  62052685  62052686  62052687  62052688  62052689  62052690  62052724  62052725  62052775  62054071  62054072  62054090  62054091  62054092  62054093  62054094  62054096  62054097  62054098  62054105  62054106  62054107  62054108  62054109  62054110  62054111  62054112  62054113  62054119  62054902  62054904  62054907  62054908  62054909  62054913  62054914  62055071  62055074  62055083  62055084  62055085  62055086  62055597  62055598  62055599  62055600  62055601  62055602  62056205  62056206  62056207  62056208  62056209  62056210  62056211  62056212  62056214  62056215  62056216  62056217  62056218  62056219  62056220  62056223  62056785  62056786  62056787  62056788  62056789  62056790  62057284  62057290  62058205  62058226  62058227  62058228  62058229  62058230  62058851  62058858  62059391  62059392  62059393  62059394  62059395  62059396  62059397  62059399  62060483  62060484  62060506  62060538  62060539  62060540  62060544  62060545  62061666  62061667  62061668  62061669  62061670  62061676  62061685  62061686  62061874  62061875  62061876  62061892  62061893  62061894  62061895  62061896  62061897  62061898  62061899  62061901  62061902  62061903  62061904  62061905  62061906  62061907  62061908  62061909  62061910  62061911  62061912  62062226  62062227  62062233  62062234  62062235  62062238  62062239  62062560  62062565  62062574  62062579  62063250  62063251  62063252  62063747  62063748  62063749  62063750  62063751  62063752  62063753  62063761  62063768  62063772  62063778  62063779  62063781  62063978  62063979  62063980  62063981  62064412  62064419  62064488  62064499  62064507  62064904  62064905  62064906  62064907  62064908  62065591  62065633  62065638  62065642  62065643  62065645  62065992  62065993  62065996  62066467  62066470  62066479  62066480  62066553  62066554  62066555  62066556  62066795  62066796  62066799  62066802  62066816  62066825  62066826  62066827  62067355  62067356  62067955  62067981  62067982  62067983  62067984  62067985  62067986  62067987  62067988  62067989  62068514  62068530  62068535  62068536  62068558  62068570  62068571  62068572  62068573  62069349  62069350  62069352  62069353  62069354  62069355  62069356  62069357  62069358  62070273  62070838  62070839  62070840  62070841  62070842  62070843  62070844  62071258  62071260  62071262  62071897  62072218  62072222  62072526  62072527  62072528  62072531  62072533  62073117  62073118  62073122  62074511  62074512  62074514  62074607  62074608  62074609  62074610  62074611  62074614  62074615  62074616  62074617  62074756  62074757  62074758  62074759  62074760  62074761  62074762  62074763  62074764  62074765  62075004  62075005  62075006  62075007  62075010  62075011  62075944  62075948  62076159  62076164  62076313  62076314  62076315  62076316  62076330  62076331  62076332  62076333  62076335  62076336  62076337  62076338  62076339  62077023  62077024  62077025  62077282  62077304  62077307  62077308  62077310  62077871  62077909  62077933  62078032  62078039  62078834  62078835  62078836  62079822  62079823  62079863  62079864  62080342  62080343  62080344  62080349  62080352  62080353  62080552  62080553  62080554  62080555  62080556  62080557  62080558  62080559  62080560  62080561  62080783  62080828  62080829  62081531  62081532  62081533  62081536  62081537  62081538  62081543  62081871  62081872  62081873  62081874  62081875  62081876  62081877  62081878  62081879  62081880  62081881  62083082  62083083  62083140  62083141  62083142  62083143  62083144  62083145  62083146  62083147  62083148  62083149  62083150  62083660  62083661  62083662  62084276  62084427  62084428  62084445  62084446  62084448  62084450  62084455  62084458  62084471  62085545  62085546  62085548  62085557  62085959  62085960  62085961  62085962  62085963  62085965  62085966  62086405  62086407  62086408  62086409  62086410  62086412  62086414  62086593  62086594  62086617  62086618  62086619  62086620  62086621  62086622  62086641  62086642  62086645  62086646  62086647  62086648  62086747  62086748  62086749  62086750  62086751  62086752  62086753  62086754  62086755  62086756  62087002  62087003  62087004  62087939  62088861  62088890  62088891  62088892  62088895  62088896  62088897  62088898  62089827  62089830  62089835  62089840  62089844  62089852  62089855  62090172  62090372  62090406  62090407  62090408  62090409  62090410  62091049  62091053  62091555  62091556  62091557  62091558  62091563  62091564  62091565  62091566  62091860  62091861  62091863  62092101  62092102  62092274  62092275  62092276  62092279  62092280  62092291  62092292  62092297  62092350  62092351  62092352  62092353  62092354  62092355  62092356  62092519  62092520  62092521  62092532  62092533  62092534  62092540  62092541  62092542  62092543  62093692  62093697  62093702  62094377  62094387  62094392  62094394  62094395  62095361  62095364  62095370  62095377  62095396  62095401  62095405  62096016  62096019  62096021  62096022  62096023  62096025  62096026  62096630  62096716  62096717  62096718  62096719  62097839  62098070  62098071  62098072  62098074  62098075  62098076  62098566  62098568  62098782  62098783  62098785  62098801  62098803  62098835  62098846  62098849  62098967  62099001  62099003  62099008  62099009  62099058  62099059  62099060  62099263  62099265  62099277  62099785  62101705  62101707  62102072  62102073  62102074  62102078  62102079  62102080  62102312  62102739  62102756  62102759  62102814  62102824  62102828  62102862  62102863  62102864  62103210  62103211  62103212  62103219  62103220  62103222  62103223  62103224  62103270  62103271  62103272  62103986  62103987  62103988  62103989  62103990  62103991  62103992  62103993  62103994  62103995  62104838  62104839  62104840  62105266  62105268  62105269  62105272  62105273  62105274  62105279  62105281  62105728  62105744  62105745  62105746  62105747  62105748  62105749  62105750  62105751  62105752  62105753  62105754  62105755  62106705  62106706  62107938  62107940  62107941  62107953  62107954  62107958  62109176  62109178  62109179  62109180  62109181  62109748  62109749  62109750  62110240  62110242  62110618  62110619  62110620  62110622  62110623  62110624  62110625  62110627  62110628  62110629  62110630  62110631  62110632  62110802  62110821  62110823  62110824  62110826  62111574  62111575  62111576  62111639  62111839  62111840  62111946  62111947  62111954  62111957  62111958  62111959  62111985  62111986  62111987  62112036  62112037  62112038  62112039  62112762  62112974  62112975  62112985  62112995  62112996  62112997  62112998  62112999  62113017  62113018  62113019  62113020  62113021  62114031  62114033  62114038  62114095  62114096  62114390  62114396  62114462  62114469  62114475  62114478  62114482  62115853  62115854  62115855  62115856  62115958  62115959  62116028  62116030  62116031  62116032  62116033  62116034  62116087  62116088  62116089  62116090  62117130  62117131  62117132  62117133  62117134  62117135  62117136  62117137  62117138  62117140  62117290  62117291  62117340  62117341  62117342  62117343  62117344  62117345  62117346  62117347  62118628  621186",
      "more_code_info": "29  62118630  62118631  62118632  62118633  62118634  62118635  62118636  62118637  62118640  62118654  62118660  62119499  62119501  62120511  62120514  62120531  62121207  62121208  62121209  62121210  62121211  62121212  62121236  62121237  62121238  62121239  62121783  62121785  62121805  62121806  62121807  62122584  62122592  62122608  62122618  62122623  62122624  62122663  62122664  62122665  62122666  62122667  62122668  62122674  62122712  62122713  62122715  62122716  62122717  62122718  62122872  62122873  62122874  62122875  62122876  62122877  62122878  62123099  62123107  62123894  62123895  62123896  62123897  62123898  62123899  62123901  62123902  62124877  62124878  62124879  62126103  62126129  62126130  62126819  62126821  62127095  62127096  62127098  62127122  62127123  62127126  62127127  62127490  62127529  62127534  62127540  62127558  62128212  62128213  62128214  62128215  62128216  62128218  62128219  62128220  62128221  62128368  62128370  62128371  62128381  62128383  62128384  62129238  62129241  62129244  62129245  62129246  62129248  62129374  62129413  62129418  62129423  62129428  62129498  62129500  62129503  62129609  62129620  62129643  62129824  62129826  62129827  62129885  62131190  62131191  62131258  62131259  62131260  62131272  62131273  62131274  62132090  62132091  62132093  62132114  62132117  62132203  62132204  62132205  62132206  62132207  62132738  62132739  62132868  62132870  62132871  62132884  62132885  62133193  62133201  62133830  62133831  62133832  62133836  62133837  62133838  62133839  62133840  62133841  62133842  62133845  62133846  62133847  62133848  62133849  62133856  62133857  62133858  62133859  62134317  62134320  62134324  62134327  62134715  62134716  62134717  62134729  62134730  62134752  62135974  62135976  62135983  62135988  62135993  62136389  62136390  62136391  62136392  62137125  62137138  62137160  62137163  62137168  62137175  62137466  62137468  62137470  62137522  62137524  62137525  62137526  62137527  62137542  62137543  62137544  62137545  62137546  62137547  62137888  62137889  62137892  62137997  62137998  62138247  62138248  62138263  62138264  62138330  62138332  62138668  62138671  62138672  62138673  62138674  62139309  62139664  62139674  62139680  62139792  62139803  62139811  62139969  62139970  62139977  62139978  62139981  62140670  62140671  62140672  62140673  62140674  62141730  62141731  62141735  62142379  62142533  62142534  62142568  62142602  62142603  62142606  62142607  62142621  62142622  62142626  62143322  62143325  62143629  62143632  62143633  62143648  62143649  62144297  62144298  62144299  62144304  62144305  62144308  62144320  62144321  62144323  62144324  62144325  62145077  62145078  62145079  62145080  62145081  62145084  62145087  62145409  62145410  62145411  62145412  62145413  62145774  62145775  62145776  62145777  62145778  62145779  62145780  62145781  62146911  62146918  62146920  62146945  62146946  62146947  62146948  62146950  62146951  62146952  62146953  62147823  62147824  62147825  62147826  62147828  62147829  62148248  62148249  62148250  62148251  62148252  62148268  62148269  62148270  62148271  62148779  62148780  62148783  62148788  62148790  62148792  62148794  62148801  62148804  62148808  62149017  62149018  62149019  62149020  62149022  62149046  62149047  62149049  62149050  62149051  62149378  62149379  62149383  62149390  62149471  62149472  62149473  62149530  62149531  62149541  62149548  62151249  62151250  62151251  62152336  62152342  62152414  62152415  62152416  62152625  62152626  62152627  62152631  62152648  62152964  62152966  62152973  62152974  62152975  62154049  62154052  62154064  62154077  62154080  62154291  62154292  62154293  62154295  62154966  62154967  62154968  62155003  62155688  62156815  62156827  62156851  62157845  62157859  62157860  62157951  62157964  62157965  62157966  62157967  62157968  62158325  62158326  62159296  62159297  62159336  62159337  62159338  62159339  62159340  62159877  62159881  62159883  62160288  62160289  62160291  62160292  62160343  62160345  62160896  62160902  62160909  62160910  62160911  62161067  62161189  62161190  62161191  62161192  62161193  62161219  62161220  62162212  62162213  62162214  62162217  62162226  62162227  62162605  62162606  62162776  62162778  62162779  62162781  62162782  62162784  62163084  62163085  62163086  62163087  62163089  62163091  62163110  62163111  62163112  62163113  62163114  62164374  62164377  62164383  62164388  62164399  62164409  62164411  62164677  62164678  62164679  62164680  62164682  62164833  62164834  62164837  62164840  62164847  62164848  62164873  62164874  62164875  62164876  62165871  62165872  62165942  62165943  62165944  62166603  62166604  62166605  62166663  62166664  62168358  62168359  62168361  62168673  62168682  62172719  62176148  62182519  62199841  62199842  62205213  62205214  62205215  62210169  62210170  62210171  62217831  62217832  62217833  62223531  62223532  62223533  62223534  62223587  62223588  62223589  62223590  62223591  62223592  62223593  62223594  62223595  62223596  62223597  62223598  62223600  62223601  62223602  62223603  62223604  62223605  62230197  62230203  62230205  62230208  62230209  62230210  62230211  62238456  62238458  62245232  62259927  62259928  77001138  77001139  77001320  77001321  77001322  77001323  77001324  77001325"
    }
  ]
}