  "results": [
    {
      "status": "Ongoing",
      "city": "Branchburg",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89643",
      "recalling_firm": "Roche Molecular Systems, Inc.",
      "address_1": "1080 US Highway 202 S",
      "address_2": "N/A",
      "postal_code": "08876-3733",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution in the states of CO, IL, IN, KY, MI, NC, and TX.  The countries of Cyprus, Spain, and Hong Kong.",
      "recall_number": "Z-0828-2022",
      "product_description": "cobas LiatSystem, respiratory virus panel nucleic acid assay system, Material Number 07341920190",
      "product_quantity": "14 units",
      "reason_for_recall": "Roche has identified a small number of cobas Liat analyzer units that have consistently high noise levels (background signal) originating from the amber detector (used for Influenza B detection). This noise increases the probability of false positive Influenza B results.",
      "recall_initiation_date": "20211116",
      "center_classification_date": "20220328",
      "report_date": "20220406",
      "code_info": "UDI: 07613336100097; Serial Numbers: 16333, 16454, 16688, 16695, 16871, 16914, 17048, 17266, 19924, 18094, 18323, 18357, 17116, 17086"
    }
  ]
}