      "status": "Terminated",
      "city": "Telford",
      "state": "PA",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73241",
      "recalling_firm": "Draeger Medical, Inc.",
      "address_1": "3135 Quarry Rd",
      "address_2": "N/A",
      "postal_code": "18969-1042",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution including Puerto Rico.",
      "recall_number": "Z-0827-2016",
      "product_description": "Evita Infinity V500 Ventilator; Babylog Infinity VN500 Ventilator.  Ventilation.",
      "product_quantity": "2553",
      "reason_for_recall": "The battery capacity of the optional PS500 power supply unit for the Infinity (ACS) Workstation Critical and Neonatal Care was rapidly and unexpectedly reduced, despite using the most recent power supply firmware version 1.50 released in November 2015 (recall Z-0436-2016). The batteries were prematurely discharged, even though an adequate battery charge status was displayed on the device.",
      "recall_initiation_date": "20160204",
      "center_classification_date": "20160225",
      "termination_date": "20200401",
      "report_date": "20160302",
      "code_info": "Catalog #s 8416400 and 8417400",
      "more_code_info": ""
    }
  ]
}