{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Utica",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "64248",
      "recalling_firm": "ConMed Corporation",
      "address_1": "525 French Road",
      "address_2": "N/A",
      "postal_code": "13502",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) and the countries of China, Dubai and Indonesia.",
      "recall_number": "Z-0826-2013",
      "product_description": "ELECTROLASE HYFRECATOR(R) ELECTRODES - SHARP, REF/Catalog Number 7-100-8BX, Disposable Hyfrecator(R) Tips, Sterile, Rx Only --- CONMED CORPORATION USA.    Intended to be used as active RF electrodes for desiccation, fulguration and coagulation.",
      "product_quantity": "Domestic Distribution:  97,450 units; Foreign Distribution: 6,800 units",
      "reason_for_recall": "Some devices had broken through the seal of the sterile pouch.  ConMed confirmed instances where the pouch seal was compromised on the affected products.",
      "recall_initiation_date": "20130128",
      "center_classification_date": "20130215",
      "termination_date": "20170411",
      "report_date": "20130227",
      "code_info": "Beginning Lot Code 120628X, Ending Lot Code 120917X. [Lot codes on boxes and packaging contain a lot code in the following format: 120917X where 12 represents the year (2012), 09 represents the month (September), 17 represents the day of the month, and X represents the manufacturing shift code (1, 2, 3, 4,5,6 or 9)]."
    }
  ]
}