  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Charleston",
      "state": "SC",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79336",
      "recalling_firm": "Trumpf Medical Systems, Inc.",
      "address_1": "1046 Legrand Blvd",
      "address_2": "N/A",
      "postal_code": "29492-7672",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Distribution to the states of : GA, IL, IN, PA, SC, TX, UT, VA.   and to the countries of :  Austria, Australia, Bangladesh, Belgium, France, Switzerland, Czech Republic, Germany, Denmark, Spain, United Arab Emirates, United Kingdom, Ireland, Italy, Kuwait, Lithuania, Poland, Portugal, Romania, Sweden, Slovenia, Thailand, Turkey.",
      "recall_number": "Z-0824-2018",
      "product_description": "Camera System, Television, Surgical without audio    Intended for transmission and documentation of surgical interventions for the purpose of archiving and teaching.",
      "product_quantity": "212 devices",
      "reason_for_recall": "The radiated emission from the TruVidia\" Wireless Receiver is above the normative accepted limits, in which the radiated emission may interact with other devices in the operating room and cause them not to function properly.",
      "recall_initiation_date": "20171215",
      "center_classification_date": "20180301",
      "termination_date": "20200423",
      "report_date": "20180307",
      "code_info": "Catalog Number 1940747, All serial numbers: 102584474, 102584475, 102584477, 102584478, 102805330, 102813656, 102813657, 102813658, 102813659, 102813660, 102813661, 102813662, 102860647, 102860652, 102863706, 102863707, 102867783, 102867784, 102925734, 102925735, 102925736, 102925737, 102925738, 102925739, 102925740, 102925741, 102925742, 102925743, 102928703, 102932185, 102932186, 102932187, 102932188, 102932189, 102932203, 102932204, 102932205, 102932206, 102932207, 102932208, 102932209, 102932210, 102932211, 102932212, 102932217, 102932218, 102932219, 102932220, 102932221, 102932223, 102934525, 102934526, 102934527, 102934528, 102934529, 102936117, 102939484, 102940382, 102940383, 102940384, 102940386, 102940387, 102940388, 102940389, 102940390, 102940391, 102940408, 102940409, 102940410, 102961390, 102961391, 102961392, 102961393, 102961394, 102961395, 102961396, 102961397, 102961398, 102961399, 102961400, 102961401, 102975336, 102986494, 102986495, 102988480, 102989096, 102989607, 102989608, 102989615, 102990285, 102990286, 102994502, 102994503, 102994504, 102994505, 102994506, 102994525, 102994526, 102994529, 102994530, 102996306, 103005879, 103005880, 103005889, 103005890, 103005891, 103005892, 103005893, 103006636, 103010023, 103010759, 103010760, 103010761, 103010762, 103010763, 103012786, 103012787, 103013644, 103017040, 103017041, 103019900, 103020713, 103022348, 103022349, 103022350, 103022351, 103022352, 103022794, 103024049, 103024050, 103030284, 103031623, 103034106, 103034109, 103038842, 103038843, 103038844, 103038845, 103038846, 103038847, 103038848, 103038849, 103038850, 103038853, 103038854, 103038855, 103038856, 103038857, 103038858, 103038859, 103038860, 103038861, 103038862, 103038863, 103038864, 103038865, 103038866, 103039468, 103042319, 103042320, 103042321, 103042322, 103042925, 103042926, 103046119, 103046120, 103046121, 103046122, 103046123, 103046124, 103046125, 103046126, 103046127, 103046128, 103046129, 103046130, 103048248, 103065061, 103070795, 103070796, 103070797, 103070798, 103079001, 103079053, 103087365, 103087366, 103087367, 103087368, 103087369, 103087370, 103087371, 103090135, 103090136, 103090137, 103091507, 103091508, 103091509, 103091510, 103091511, 103095664, 103101676, 103101826, 103101864, 103101886, 103103841, 103103842, 103103843, 103103844, 103103845, 103104556, 103113059, 103117961, 103118986, 103118987, 103118988, 103118989, 103118990, 103118991, 103118992, 103118993, 103118994, 103118995, 103123759, 103133224, 103134656, 103138822, 103143214, 103149993, 103164298, 103168053, 103168341, 103168431, 103176559, 103176571, 103176574, 103191728, 103191729, 103194419, 103214005, 103214022."
    }
  ]
}