  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Newark",
      "address_1": "500 Gbc Dr Ms 514",
      "reason_for_recall": "Under extremely rare circumstances, a patient result that had been previously disabled may be released without proper review. (Disabled Test in CentraLink System or Aptio System May Become Enabled in CentraLink System.)",
      "address_2": "PO BOX 6101",
      "product_quantity": "40",
      "code_info": "CentraLink\" Data Management System V14.0.4, V14.0.5 or V14.0.8  Catalog Numbers: 10810210, 10811596, 10814296, 10814298, 10814877,  10815474.",
      "center_classification_date": "20140123",
      "distribution_pattern": "US Distribution in the states of: MN, NJ, and NV.",
      "state": "DE",
      "product_description": "Siemens CentraLink\" Data Management System V14.0.4, V14.0.5 or V14.0.8    The CentraLink system software is a network solution provider and multi-system data manager for the instruments and lab automation systems (LAS) within the lab. The CentraLink software consolidates data from all connected instruments so that an operator can review and edit patient and quality control results from a single location.",
      "report_date": "20140129",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc.",
      "recall_number": "Z-0823-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "66694",
      "termination_date": "20141215",
      "more_code_info": "",
      "recall_initiation_date": "20131030",
      "postal_code": "19702-2466",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}