{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Boulder",
      "address_1": "6135 Gunbarrel Ave",
      "reason_for_recall": "Covidien is correcting labeling to clarify the operational life of the oxygen sensor (O2 sensor), PN 4-072214-00 used in the Puritan Bennett\" 840, and PN, G-062010-00, used in Puritan Bennett\" 740, and760 ventilator systems.  The Operator\u0019s Manual for PB840 states a nominal life of one year and the Addendum incorrectly states 2 years.",
      "address_2": "",
      "product_quantity": "7,959",
      "code_info": "PB700 Series Operator\u0019s Manual: PN 10066984 & PN G-061874-00 and PB700 Series Service Manual: PN 10070389",
      "center_classification_date": "20140123",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) and Internationally to: Albania, Argentina, Australia, Bahrain, Belgium, Belize, Bermuda Bolivia, Bosnia and Herzegovina, Brazil, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Croatia, Ecuador, Egypt, El Salvador, Finland, France, French Polynesia, Georgia, Germany, Greece, Guam, Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Lebanon, Libya, Lithuania, Macedonia, The former Yugoslav, Malaysia, Mexico, Moldova, Republic of, Montenegro, Morocco, Netherlands, Nicaragua, Northern Mariana Islands, Oman, Panama, Paraguay, Peru, Philippines, Poland, Puerto Rico, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Switzerland, Syria, Taiwan, Thailand, Tunisia, Turkey, Uganda, Ukrainian SSR, United Arab Emirates (UAE) United Kingdom, Uruguay, Venezuela and Yugoslavia.",
      "state": "CO",
      "product_description": "Puritan Bennett 700 Series Ventilators (PB740 and PB760) PB700 Series Operator\u0019s Manual: PN 10066984 & PN G-061874-00 and PB700 Series Service Manual: PN 10070389     The intended use of the oxygen sensor is to provide the oxygen concentration of the delivered gas at the patient wye (inspiratory-expiratory interface). The oxygen sensor does not control the oxygen level or the flow of gases. The intended use of the devices is as critical care ventilators that provide continuous mechanical ventilation for patients.",
      "report_date": "20140129",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Nellcor Puritan Bennett Inc. (dba Covidien LP)",
      "recall_number": "Z-0822-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "67163",
      "termination_date": "20141124",
      "more_code_info": "",
      "recall_initiation_date": "20130125",
      "postal_code": "80301-3214",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}