{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "HIROSHIMA",
      "state": "N/A",
      "country": "Japan",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "69825",
      "recalling_firm": "MITSUBISHI HEAVY INDUSTRIES, LTD., HIROSHIMA MACHINERY WORK",
      "address_1": "6-22, 4-CHOME, KAN-ON-SHIN-MACHI",
      "address_2": "NISHI-KU",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US Distribution to the states of NY, FL, TX and OH.",
      "recall_number": "Z-0821-2015",
      "product_description": "VERO Linear Accelerator System, Model Number MHI-TM2000, MHI Identifier M101-14021, Software Version: 3.1, 3.2 and 3.5.0.",
      "product_quantity": "Domestic: 4 units",
      "reason_for_recall": "Software Anomaly. The VERO MHI-TM2000 Operator Console may provide Patient Positioning System (ExacTrac) with Image Angle Information Used for the 1st Port Image, for a subsequent port image fusion.  As the result, ExacTrac may display a Digitally Reconstructed Radiograph (DRR) image taken from the angle for the 1st port image fusion, rather than the one taken from the angle for the intended port",
      "recall_initiation_date": "20141202",
      "center_classification_date": "20141217",
      "termination_date": "20170626",
      "report_date": "20141224",
      "code_info": "Serial Numbers: 201902, 203901,  203919, and 203924 (US only)"
    }
  ]
}