  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Memphis",
      "address_1": "1450 E Brooks Rd",
      "reason_for_recall": "The package for catalog item, a LEGION Hemi Stepped Tibial Wedge, 5 MM Size 5-6 LT-LAT/ RT-MDL, should include both a screw-on hemi stepped tibial wedge and two tibial wedge screws to affix the wedge to the tibial baseplate. A complaint was received indicating that the screws packaged with the wedge were  too long.",
      "address_2": "",
      "product_quantity": "8 units",
      "code_info": "14KM16422",
      "center_classification_date": "20160217",
      "distribution_pattern": "Distributed in the states of TN, WI, OH, and KS, and in the countries of Belgium, Germany, and United Kingdom.",
      "state": "TN",
      "product_description": "LEGION(TM) HEMI STEPPED, TIBIAL SCREW-ON WEDGE, REF 71421147, QTY 1, 5 MM, SIZE 5-6 LT-LAT / RT-MDL;  Tibial knee prosthesis, augment.",
      "report_date": "20160224",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Smith & Nephew, Inc.",
      "recall_number": "Z-0816-2016",
      "initial_firm_notification": "E-Mail",
      "product_type": "Devices",
      "event_id": "73164",
      "termination_date": "20160525",
      "more_code_info": "",
      "recall_initiation_date": "20160118",
      "postal_code": "38116-1804",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}