  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84438",
      "recalling_firm": "GE Healthcare, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: US (Nationwide) and countries of: Albania, Algeria, Argentina, Australia, Bahrain, Bangladesh, Belgium, Bolivia, Brazil, Canada, Chile, China, Colombia, Croatia and Hrvatska, Czech Republic, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia, Iraq, Israel, Italy, Jamaica, Japan, Jordan, Kenya, Kuwait, Latvia, Lebanon, Lithuania, Malaysia, Maldives, Mexico, Myanmar, Nepal, Netherlands, New Zealand, Nigeria, Oman, Paraguay, Peru, Philippines, Poland, Portugal, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom, Uruguay, and Viet Nam.",
      "recall_number": "Z-0814-2020",
      "product_description": "Carestation 620/650/650c A2    Product Usage: anesthesia systems are intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients (pediatric and adult.",
      "product_quantity": "280 units",
      "reason_for_recall": "GE Healthcare has become aware that there is a potential for a loose cable connection inside specific manufactured anesthesia devices. This would cause a loss of mechanical ventilation and the system will provide high priority audio and visual alarms. Loss of mechanical ventilation could lead to hypoxia if the clinician does not intervene.",
      "recall_initiation_date": "20191125",
      "center_classification_date": "20200123",
      "termination_date": "20231130",
      "report_date": "20200129",
      "code_info": "CARESTATION 620 A2, Model number 1012-9620-002, No UDI, OUS only  Mfg. Lot or Serial #  SM618370003WA  SM618370004WA  SM618370005WA  SM618370006WA  SM618370010WA  SM618370070WA  SM618370071WA  SM618380001WA  SM618380002WA  SM618380008WA  SM618380009WA  SM618380010WA  SM618380011WA  SM618450033WA  SM618480061WA  SM618500014WA  SM619110035WA  SM618360022WA  SM618360027WA  SM618360028WA  SM618370020WA  SM618370021WA  SM618370022WA  SM618370023WA  SM618370051WA  SM618370052WA  SM618370053WA  SM618370056WA  SM618360007WA  SM619030013WA  SM619030014WA  SM619030015WA  SM619030016WA  SM619040001WA  SM619150001WA  SM618430007WA  SM618440004WA  SM618480064WA  SM618390031WA  SM619100010WA  SM618420021WA  SM618420022WA  SM618420023WA  SM618420024WA  SM618480063WA  SM619070003WA  SM619170051WA  SM618460048WA  SM619070001WA  SM619070002WA  SM618410001WA  SM618410002WA  SM618500015WA  SM618450034WA  SM618380007WA  SM619030007WA  SM619110044WA  SM619120001WA  SM619135003WA  SM618440035WA  SM619110011WA  SM618360011WA  SM618450013WA  SM619090011WA  SM618460062WA  SM618350016WA  SM618460049WA  SM618460050WA  SM618460051WA  SM618480036WA  SM618380017WA  SM618380018WA  SM618500004WA  SM618500005WA  SM618430008WA  SM618450012WA  SM618490017WA  SM618490018WA  SM619030012WA  SM619135004WA  SM618380014WA  SM618370073WA  SM619120002WA  SM619120003WA  SM619120004WA  SM618430005WA  SM618490010WA  SM618490013WA  SM618490014WA  SM618380020WA  SM618500001WA  SM618500002WA  SM618360009WA  SM618360013WA  SM618370024WA  SM618500016WA  SM618500017WA  SM618500018WA  SM618500019WA  SM618500020WA  SM618500021WA  SM618380003WA  SM618380004WA  SM618370072WA  SM618370074WA  SM618490015WA  SM618490016WA  SM618490019WA  SM618490020WA  SM618490025WA  SM618490026WA  SM618500003WA  SM618370014WA  SM618370015WA  SM618370016WA  SM618370017WA  SM618370018WA  SM618370019WA  SM618370013WA  SM619080002WA  SM618420020WA  SM619080022WA  SM618510004WA  SM618510005WA  SM618510006WA  SM619020020WA  SM619090008WA  SM619090009WA  SM619090010WA  SM619090012WA  SM619090013WA  SM618370054WA  SM618390009WA  SM618390010WA  SM619110002WA  SM619110003WA  SM619110004WA  SM619110006WA  SM619110007WA  SM619110009WA  SM619110012WA  SM619110014WA  SM619230009WA  SM619110005WA  SM619110008WA  SM619110013WA  SM619110015WA  SM619110021WA  SM619110022WA  SM619110023WA  SM619200048WA  SM619200049WA  SM619210001WA  SM619210002WA  SM618340078WA  SM619110024WA  SM619040006WA  SM618370068WA  SM618370069WA  SM619110018WA  SM619110019WA  SM619110020WA  SM619170001WA  SM619110016WA  SM619110017WA  SM618390025WA  SM618390026WA  SM618390027WA  SM618390028WA  SM618390029WA  SM618390030WA  SM618470001WA  SM618470024WA  SM618470025WA  SM618470026WA  SM618490024WA  SM618510003WA  SM618510009WA  SM619100004WA  SM619110026WA  SM619110027WA  SM618380005WA  SM618380006WA  SM619100011WA  SM618360008WA  SM618360010WA  SM618360012WA  SM618360029WA  SM618360030WA  SM618370055WA  SM619170002WA  SM619170003WA  SM618380039WA  SM618370001WA  SM618370002WA  SM618380013WA  SM618460039WA  SM618380012WA  SM618380019WA  SM618370012WA  SM619110001WA  SM619170052WA  SM619200046WA  SM619200047WA  SM619220001WA  SM619240008WA  SM619240009WA  SM619240010WA  SM619240011WA  SM619150007WA  SM618480035WA    CARESTATION 650 A2, Model Number 1012-9650-002, No UDI, OUS only  Mfg. Lot or Serial #  SM718450028WA  SM718450029WA  SM718450030WA  SM718450031WA  SM718450032WA  SM718450033WA  SM718340013WA  SM718340014WA  SM718340015WA  SM718340019WA  SM718340020WA  SM718370012WA  SM718410018WA  SM718510028WA  SM718510029WA  SM718490037WA  SM718380010WA  SM718380011WA  SM718380006WA  SM718380004WA  SM718490029WA  SM718490030WA  SM718490032WA  SM718490033WA  SM718490034WA  SM718490035WA  SM718340018WA  SM718360037WA  SM718360038WA  SM718360042WA  SM718360043WA  SM718470055WA  SM718490024WA  SM718490025WA  SM718490026WA  SM718490027WA  SM718490028WA  SM718370013WA  SM719070010WA  SM719070011WA  SM719070012WA  SM719070013WA  SM719070014WA  SM719070015WA  SM718380002WA  SM718380003WA  SM719230023WA  SM719230024WA  SM719230025WA  SM718490036WA  SM718370074WA  SM718370075WA  SM719240003WA  SM718370031WA  SM718370032WA  SM718370033WA  SM718370073WA  SM718470056WA  SM718510030WA  SM719110001WA  SM719110003WA  SM719110004WA  SM718380005WA  SM719135001WA  SM719135002WA  SM718470011WA    CARESTATION 650C A2, Model Number 1012-9655-002, No UDI, OUS only  Mfg. Lot or Serial #  SM818370001WA  SM818480001WA  SM818480005WA",
      "more_code_info": ""
    }
  ]
}