      "country": "United States",
      "city": "Portage",
      "address_1": "3800 E Centre Ave",
      "reason_for_recall": "It was identified that the volume of air inside the mattress may expand in regions of high altitude which may result in an unintended gap between the plateau of the inflated patient surface and tub. A Health Hazard Evaluation was completed which identified the potential risk of patients moving out of position and rolling over on the mattress. While the highest possible severity of harm associated with this condition is death, there have been no reports of injury or harm. This condition has only presented in regions of high altitude; however, Stryker has chosen to replace all Naera mattresses in the field.",
      "address_2": "",
      "product_quantity": "5915",
      "code_info": "2016000600003  2016000600004  2016000600005  2016000600006  2016000600007  2016000600008  2016000600009  2016000600010  2016000600011  2016000600012  2016000600013  2016000600014  2016000600015  2016000600016  2016000600017  2016000600018  2016000600019  2016000600020  2016000600021  2016000600022  2016000600023  2016000600024  2016000600025  2016000600026  2016000600027  2016000600028  2016000600029  2016000600030  2016000600031  2016000600032  2016000600033  2016000600034  2016000600035  2016000600036  2016000600037  2016000600038  2016000600039  2016000600040  2016000600041  2016000600042  2016000600043  2016000600044  2016000600045  2016000600046  2016000600047  2016000600048  2016000600049  2016000600050  2016000600054  2016000600059  2016000600060  2016000600061  2016000600062  2016000600063  2016000600064  2016000600065  2016000600066  2016000600067  2016000600068  2016000600069  2016000600070  2016000600073  2016000600082  2016000600083  2016000600084  2016000600085  2016000600086  2016000600087  2016000600088  2016000600089  2016000600090  2016000600091  2016000600092  2016000600093  2016000600094  2016000600095  2016000600096  2016000600097  2016000600098  2016000600099  2016000600100  2016000600101  2016000600102  2016000600103  2016000600104  2016000600105  2016000600106  2016000600107  2016000600108  2016000600109  2016000600110  2016000600111  2016000600112  2016000600113  2016000600114  2016000600115  2016000600116  2016000600117  2016000600118  2016000600119  2016000600120  2016000600121  2016000600122  2016000600123  2016000600124  2016000600125  2016000600126  2016000600127  2016000600128  2016000600129  2016000600130  2016000600131  2016000600132  2016000600133  2016000600134  2016000600135  2016000600136  2016000600137  2016000600138  2016000600139  2016000600140  2016000600141  2016000600142  2016000600143  2016000600144  2016000600145  2016000600146  2016000600147  2016000600148  2016000600149  2016000600150  2016000600151  2016000600152  2016000600153  2016000600154  2016000600155  2016000600156  2016000600157  2016000600158  2016000600159  2016000600160  2016000600161  2016000600162  2016000600163  2016000600164  2016000600165  2016000600166  2016000600167  2016000600170  2016000600176  2016000600179  2016000600180  2016000600181  2016000600182  2016000600183  2016000600184  2016000600185  2016000600186  2016000600187  2016000600188  2016000600189  2016000600190  2016000600192  2016000600193  2016000600194  2016000600197  2016000600198  2016000600199  2016000600200  2016000600201  2016000600202  2016000600203  2016000600204  2016000600205  2016000600206  2016000600207  2016000600208  2016000600209  2016000600210  2016000600211  2016000600212  2016000600213  2016000600214  2016000600215  2016000600219  2016000600220  2016000600221  2016000600222  2016000600224  2016000600227  2016000600228  2016000600229  2016000600230  2016000600231  2016000600232  2016000600233  2016000600234  2016000600235  2016000600236  2016000600237  2016000600238  2016000600239  2016000600240  2016000600242  2016000600244  2016000600245  2016000600247  2016000600248  2016000600250  2016000600251  2016000600252  2016000600253  2016000600254  2016000600255  2016000600256  2016000600257  2016000600258  2016000600259  2016000600260  2016000600261  2016000600262  2016000600263  2016000600264  2016000600265  2016000600266  2016000600267  2016000600268  2016000600269  2016000600270  2016000600271  2016000600272  2016000600273  2016000600274  2016000600275  2016000600276  2016000600277  2016000600278  2016000600279  2016000600280  2016000600281  2016000600282  2016000600283  2016000600284  2016000600285  2016000600286  2016000600287  2016000600288  2016000600289  2016000600290  2016000600291  2016000600292  2016000600293  2016000600294  2016000600295  2016000600296  2016000600297  2016000600298  2016000600299  2016000600300  2016000600301  2016000600302  2016000600303  2016000600304  2016000600305  2016000600306  2016000600307  2016000600308  2016000600309  2016000600310  2016000600311  2016000600312  2016000600313  2016000600314  2016000600316  2016000600317  2016000600318  2016000600319  2016000600320  2016000600321  2016000600322  2016000600323  2016000600324  2016000600325  2016000600326  2016000600327  2016000600328  2016000600330  2016000600331  2016000600332  2016000600333  2016000600334  2016000600335  2016000600336  2016000600337  2016000600338  2016000600339  2016000600340  2016000600341  2016000600342  2016000600343  2016000600344  2016000600345  2016000600346  2016000600347  2016000600348  2016000600349  2016000600350  2016000600351  2016000600352  2016000600353  2016000600354  2016000600355  2016000600356  2016000600357  2016000600358  2016000600359  2016000600360  2016000600361  2016000600362  2016000600363  2016000600364  2016000600365  2016000600366  2016000600367  2016000600368  2016000600369  2016000600370  2016000600371  2016000600372  2016000600373  2016000600374  2016000600375  2016000600376  2016000600377  2016000600378  2016000600379  2016000600380  2016000600381  2016000600382  2016000600383  2016000600384  2016000600385  2016000600386  2016000600387  2016000600388  2016000600389  2016000600390  2016000600391  2016000600392  2016000600393  2016000600394  2016000600395  2016000600396  2016000600397  2016000600398  2016000600399  2016000600400  2016000600401  2016000600402  2016000600403  2016000600404  2016000600405  2016000600406  2016000600407  2016000600408  2016000600409  2016000600410  2016000600411  2016000600412  2016000600413  2016000600414  2016000600415  2016000600416  2016000600417  2016000600418  2016000600419  2016000600420  2016000600421  2016000600422  2016000600423  2016000600424  2016000600425  2016000600426  2016000600427  2016000600428  2016000600429  2016000600430  2016000600431  2016000600432  2016000600433  2016000600434  2016000600435  2016000600436  2016000600437  2016000600438  2016000600439  2016000600440  2016000600441  2016000600442  2016000600443  2016000600444  2016000600445  2016000600446  2016000600447  2016000600448  2016000600449  2016000600450  2016000600451  2016000600452  2016000600453  2016000600454  2016000600455  2016000600456  2016000600457  2016000600458  2016000600459  2016000600460  2016000600461  2016000600462  2016000600463  2016000600464  2016000600465  2016000600467  2016000600468  2016000600469  2016000600470  2016000600471  2016000600472  2016000600473  2016000600474  2016000600475  2016000600476  2016000600477  2016000600478  2016000600479  2016000600480  2016000600482  2016000600483  2016000600484  2016000600485  2016000600486  2016000600487  2016000600488  2016000600489  2016000600490  2016000600491  2016000600492  2016000600493  2016000600494  2016000600495  2016000600496  2016000600497  2016000600498  2016000600499  2016000600500  2016000600501  2016000600502  2016000600503  2016000600504  2016000600505  2016000600506  2016000600507  2016000600508  2016000600509  2016000600510  2016000600511  2016000600512  2016000600513  2016000600514  2016000600515  2016000600516  2016000600517  2016000600518  2016000600519  2016000600520  2016000600521  2016000600522  2016000600523  2016000600524  2016000600525  2016000600526  2016000600527  2016000600528  2016000600529  2016000600530  2016000600531  2016000600532  2016000600533  2016000600534  2016000600535  2016000600536  2016000600537  2016000600538  2016000600539  2016000600540  2016000600541  2016000600542  2016000600543  2016000600544  2016000600545  2016000600546  2016000600547  2016000600548  2016000600549  2016000600550  2016000600551  2016000600552  2016000600553  2016000600554  2016000600555  2016000600556  2016000600557  2016000600559  2016000600560  2016000600561  2016000600562  2016000600563  2016000600564  2016000600565  2016000600566  2016000600567  2016000600568  2016000600569  2016000600570  2016000600571  2016000600572  2016000600573  2016000600574  2016000600575  2016000600576  2016000600577  2016000600578  2016000600579  2016000600580  2016000600581  2016000600582  2016000600583  2016000600584  2016000600585  2016000600586  2016000600587  2016000600588  2016000600589  2016000600590  2016000600591  2016000600593  2016000600594  2016000600595  2016000600596  2016000600597  2016000600598  2016000600599  2016000600600  2016000600601  2016000600602  2016000600603  2016000600604  2016000600605  2016000600606  2016000600607  2016000600608  2016000600609  2016000600610  2016000600611  2016000600612  2016000600613  2016000600614  2016000600615  2016000600616  2016000600617  2016000600618  2016000600619  2016000600620  2016000600621  2016000600625  2016000600626  2016000600627  2016000600628  2016000600629  2016000600630  2016000600631  2016000600632  2016000700001  2016000700002  2016000700003  2016000700004  2016000700005  2016000700006  2016000700007  2016000700008  2016000700009  2016000700010  2016000700011  2016000700012  2016000700013  2016000700014  2016000700015  2016000700016  2017000600001  2017000600002  2017000600003  2017000600004  2017000600005  2017000600006  2017000600007  2017000600008  2017000600009  2017000600010  2017000600011  2017000600012  2017000600013  2017000600014  2017000600015  2017000600016  2017000600017  2017000600018  2017000600019  2017000600020  2017000600021  2017000600022  2017000600023  2017000600024  2017000600025  2017000600026  2017000600027  2017000600028  2017000600029  2017000600030  2017000600031  2017000600032  2017000600033  2017000600034  2017000600035  2017000600036  2017000600037  2017000600038  2017000600039  2017000600040  2017000600042  2017000600043  2017000600044  2017000600045  2017000600046  2017000600047  2017000600048  2017000600049  2017000600050  2017000600051  2017000600052  2017000600053  2017000600054  2017000600055  2017000600056  2017000600057  2017000600058  2017000600059  2017000600060  2017000600061  2017000600062  2017000600063  2017000600064  2017000600065  2017000600066  2017000600067  2017000600068  2017000600069  2017000600070  2017000600071  2017000600072  2017000600073  2017000600074  2017000600075  2017000600076  2017000600077  2017000600078  2017000600079  2017000600080  2017000600081  2017000600082  2017000600083  2017000600084  2017000600085  2017000600086  2017000600087  2017000600088  2017000600089  2017000600090  2017000600091  2017000600092  2017000600093  2017000600094  2017000600095  2017000600096  2017000600097  2017000600098  2017000600099  2017000600100  2017000600101  2017000600102  2017000600103  2017000600104  2017000600105  2017000600106  2017000600107  2017000600108  2017000600109  2017000600110  2017000600111  2017000600112  2017000600113  2017000600114  2017000600115  2017000600116  2017000600117  2017000600118  2017000600119  2017000600120  2017000600121  2017000600122  2017000600123  2017000600124  2017000600125  2017000600126  2017000600127  2017000600128  2017000600129  2017000600130  2017000600131  2017000600132  2017000600133  2017000600134  2017000600135  2017000600136  2017000600137  2017000600138  2017000600139  2017000600140  2017000600141  2017000600142  2017000600143  2017000600144  2017000600145  2017000600146  2017000600147  2017000600148  2017000600149  2017000600150  2017000600151  2017000600152  2017000600153  2017000600154  2017000600155  2017000600156  2017000600157  2017000600158  2017000600159  2017000600160  2017000600161  2017000600162  2017000600163  2017000600164  2017000600165  2017000600166  2017000600167  2017000600168  2017000600169  2017000600170  2017000600171  2017000600172  2017000600173  2017000600174  2017000600175  2017000600176  2017000600177  2017000600178  2017000600179  2017000600180  2017000600181  2017000600182  2017000600183  2017000600184  2017000600185  2017000600186  2017000600187  2017000600188  2017000600189  2017000600190  2017000600191  2017000600192  2017000600193  2017000600194  2017000600195  2017000600196  2017000600197  2017000600198  2017000600199  2017000600200  2017000600201  2017000600202  2017000600203  2017000600204  2017000600205  2017000600206  2017000600207  2017000600208  2017000600209  2017000600210  2017000600211  2017000600212  2017000600213  2017000600214  2017000600215  2017000600216  2017000600217  2017000600218  2017000600219  2017000600220  2017000600221  2017000600222  2017000600223  2017000600224  2017000600225  2017000600226  2017000600227  2017000600228  2017000600229  2017000600230  2017000600231  2017000600232  2017000600233  2017000600234  2017000600235  2017000600236  2017000600237  2017000600238  2017000600239  2017000600240  2017000600241  2017000600242  2017000600243  2017000600244  2017000600245  2017000600246  2017000600247  2017000600248  2017000600249  2017000600250  2017000600251  2017000600252  2017000600253  2017000600254  2017000600255  2017000600256  2017000600257  2017000600258  2017000600259  2017000600260  2017000600261  2017000600262  2017000600263  2017000600264  2017000600265  2017000600266  2017000600267  2017000600268  2017000600269  2017000600270  2017000600271  2017000600272  2017000600273  2017000600274  2017000600275  2017000600276  2017000600277  2017000600278  2017000600279  2017000600280  2017000600281  2017000600282  2017000600283  2017000600284  2017000600285  2017000600286  2017000600287  2017000600288  2017000600289  2017000600290  2017000600291  2017000600292  2017000600293  2017000600294  2017000600295  2017000600296  2017000600297  2017000600298  2017000600299  2017000600300  2017000600301  2017000600302  2017000600303  2017000600304  2017000600305  2017000600306  2017000600307  2017000600308  2017000600309  2017000600310  2017000600311  2017000600312  2017000600313  2017000600314  2017000600315  2017000600316  2017000600317  2017000600318  2017000600319  2017000600320  2017000600321  2017000600322  2017000600323  2017000600324  2017000600325  2017000600326  2017000600327  2017000600328  2017000600329  2017000600330  2017000600331  2017000600332  2017000600334  2017000600335  2017000600336  2017000600337  2017000600338  2017000600339  2017000600340  2017000600341  2017000600342  2017000600343  2017000600344  2017000600345  2017000600346  2017000600347  2017000600348  2017000600349  2017000600350  2017000600351  2017000600352  2017000600353  2017000600354  2017000600355  2017000600356  2017000600357  2017000600358  2017000600359  2017000600360  2017000600361  2017000600362  2017000600363  2017000600364  2017000600365  2017000600366  2017000600367  2017000600368  2017000600369  2017000600370  2017000600371  2017000600372  2017000600373  2017000600374  2017000600375  2017000600376  2017000600377  2017000600378  2017000600379  2017000600380  2017000600381  2017000600382  2017000600383  2017000600384  2017000600385  2017000600386  2017000600387  2017000600388  2017000600389  2017000600390  2017000600391  2017000600392  2017000600393  2017000600394  2017000600395  2017000600396  2017000600397  2017000600398  2017000600399  2017000600400  2017000600401  2017000600402  2017000600403  2017000600404  2017000600405  2017000600406  2017000600407  2017000600408  2017000600409  2017000600410  2017000600411  2017000600412  2017000600413  2017000600414  2017000600415  2017000600416  2017000600417  2017000600418  2017000600419  2017000600420  2017000600421  2017000600422  2017000600423  2017000600424  2017000600425  2017000600426  2017000600427  2017000600428  2017000600429  2017000600430  2017000600431  2017000600432  2017000600433  2017000600434  2017000600435  2017000600436  2017000600437  2017000600438  2017000600439  2017000600440  2017000600441  2017000600442  2017000600443  2017000600444  2017000600445  2017000600446  2017000600447  2017000600448  2017000600449  2017000600450  2017000600451  2017000600452  2017000600453  2017000600454  2017000600455  2017000600456  2017000600457  2017000600458  2017000600459  2017000600460  2017000600461  2017000600462  2017000600463  2017000600464  2017000600465  2017000600466  2017000600467  2017000600468  2017000600469  2017000600470  2017000600471  2017000600472  2017000600473  2017000600474  2017000600475  2017000600476  2017000600477  2017000600478  2017000600479  2017000600480  2017000600481  2017000600482  2017000600483  2017000600484  2017000600485  2017000600486  2017000600487  2017000600488  2017000600489  2017000600490  2017000600491  2017000600492  2017000600493  2017000600494  2017000600495  2017000600496  2017000600497  2017000600498  2017000600499  2017000600500  2017000600501  2017000600502  2017000600503  2017000600504  2017000600505  2017000600506  2017000600507  2017000600508  2017000600509  2017000600510  2017000600511  2017000600512  2017000600513  2017000600514  2017000600515  2017000600516  2017000600517  2017000600518  2017000600519  2017000600520  2017000600521  2017000600522  2017000600523  2017000600524  2017000600525  2017000600526  2017000600527  2017000600528  2017000600529  2017000600530  2017000600531  2017000600532  2017000600533  2017000600534  2017000600535  2017000600536  2017000600537  2017000600538  2017000600539  2017000600540  2017000600541  2017000600542  2017000600543  2017000600544  2017000600545  2017000600546  2017000600547  2017000600548  2017000600549  2017000600550  2017000600551  2017000600552  2017000600553  2017000600554  2017000600555  2017000600556  2017000600557  2017000600558  2017000600559  2017000600560  2017000600561  2017000600562  2017000600563  2017000600564  2017000600565  2017000600566  2017000600567  2017000600568  2017000600569  2017000600570  2017000600571  2017000600572  2017000600573  2017000600574  2017000600575  2017000600576  2017000600577  2017000600578  2017000600579  2017000600580  2017000600581  2017000600582  2017000600583  2017000600584  2017000600585  2017000600586  2017000600587  2017000600588  2017000600589  2017000600590  2017000600591  2017000600592  2017000600593  2017000600594  2017000600595  2017000600596  2017000600597  2017000600598  2017000600599  2017000600600  2017000600601  2017000600602  2017000600603  2017000600604  2017000600605  2017000600606  2017000600607  2017000600608  2017000600609  2017000600610  2017000600611  2017000600612  2017000600613  2017000600614  2017000600615  2017000600616  2017000600617  2017000600618  2017000600619  2017000600620  2017000600621  2017000600622  2017000600623  2017000600624  2017000600625  2017000600626  2017000600627  2017000600628  2017000600629  2017000600630  2017000600631  2017000600632  2017000600633  2017000600634  2017000600635  2017000600636  2017000600637  2017000600638  2017000600639  2017000600640  2017000600641  2017000600642  2017000600643  2017000600644  2017000600645  2017000600646  2017000600647  2017000600648  2017000600649  2017000600650  2017000600651  2017000600652  2017000600653  2017000600654  2017000600655  2017000600656  2017000600657  2017000600658  2017000600659  2017000600660  2017000600661  2017000600662  2017000600663  2017000600664  2017000600665  2017000600666  2017000600667  2017000600668  2017000600669  2017000600670  2017000600671  2017000600672  2017000600673  2017000600674  2017000600675  2017000600676  2017000600678  2017000600679  2017000600680  2017000600681  2017000600682  2017000600683  2017000600684  2017000600685  2017000600686  2017000600687  2017000600688  2017000600689  2017000600690  2017000600691  2017000600692  2017000600693  2017000600694  2017000600695  2017000600696  2017000600697  2017000600698  2017000600699  2017000600700  2017000600701  2017000600702  2017000600703  2017000600704  2017000600705  2017000600706  2017000600707  2017000600708  2017000600709  2017000600710  2017000600711  2017000600712  2017000600713  2017000600714  2017000600715  2017000600716  2017000600717  2017000600718  2017000600719  2017000600720  2017000600721  2017000600722  2017000600723  2017000600724  2017000600725  2017000600726  2017000600727  2017000600728  2017000600729  2017000600730  2017000600731  2017000600732  2017000600733  2017000600734  2017000600735  2017000600736  2017000600737  2017000600738  2017000600739  2017000600740  2017000600741  2017000600742  2017000600743  2017000600744  2017000600745  2017000600746  2017000600747  2017000600748  2017000600749  2017000600750  2017000600751  2017000600752  2017000600753  2017000600754  2017000600755  2017000600756  2017000600757  2017000600758  2017000600759  2017000600760  2017000600761  2017000600762  2017000600763  2017000600764  2017000600765  2017000600766  2017000600767  2017000600768  2017000600769  2017000600771  2017000600772  2017000600773  2017000600774  2017000600775  2017000600776  2017000600777  2017000600778  2017000600779  2017000600780  2017000600781  2017000600782  2017000600783  2017000600784  2017000600785  2017000600786  2017000600787  2017000600788  2017000600789  2017000600790  2017000600791  2017000600792  2017000600793  2017000600794  2017000600795  2017000600796  2017000600797  2017000600798  2017000600799  2017000600800  2017000600801  2017000600802  2017000600803  2017000600804  2017000600805  2017000600806  2017000600807  2017000600808  2017000600809  2017000600810  2017000600811  2017000600812  2017000600813  2017000600814  2017000600815  2017000600816  2017000600817  2017000600818  2017000600819  2017000600820  2017000600821  2017000600822  2017000600823  2017000600824  2017000600825  2017000600826  2017000600827  2017000600828  2017000600829  2017000600830  2017000600831  2017000600832  2017000600833  2017000600834  2017000600835  2017000600836  2017000600837  2017000600838  2017000600839  2017000600840  2017000600841  2017000600842  2017000600843  2017000600844  2017000600845  2017000600846  2017000600847  2017000600848  2017000600849  2017000600850  2017000600851  2017000600852  2017000600853  2017000600854  2017000600855  2017000600856  2017000600857  2017000600858  2017000600859  2017000600860  2017000600861  2017000600862  2017000600863  2017000600864  2017000600865  2017000600866  2017000600867  2017000600868  2017000600869  2017000600870  2017000600871  2017000600872  2017000600873  2017000600874  2017000600875  2017000600876  2017000600877  2017000600878  2017000600879  2017000600880  2017000600881  2017000600883  2017000600884  2017000600885  2017000600886  2017000600887  2017000600888  2017000600889  2017000600890  2017000600891  2017000600892  2017000600893  2017000600894  2017000600895  2017000600896  2017000600897  2017000600898  2017000600899  2017000600900  2017000600901  2017000600902  2017000600903  2017000600904  2017000600905  2017000600906  2017000600907  2017000600908  2017000600909  2017000600910  2017000600911  2017000600912  2017000600913  2017000600914  2017000600915  2017000600916  2017000600917  2017000600918  2017000600919  2017000600920  2017000600921  2017000600922  2017000600923  2017000600924  2017000600925  2017000600926  2017000600927  2017000600928  2017000600929  2017000600930  2017000600931  2017000600932  2017000600933  2017000600934  2017000600935  2017000600936  2017000600937  2017000600938  2017000600939  2017000600940  2017000600941  2017000600942  2017000600943  2017000600944  2017000600945  2017000600946  2017000600947  2017000600948  2017000600949  2017000600950  2017000600951  2017000600952  2017000600953  2017000600954  2017000600955  2017000600956  2017000600957  2017000600958  2017000600959  2017000600960  2017000600961  2017000600962  2017000600963  2017000600964  2017000600965  2017000600966  2017000600967  2017000600968  2017000600969  2017000600970  2017000600971  2017000600972  2017000600973  2017000600974  2017000600975  2017000600976  2017000600977  2017000600978  2017000600979  2017000600980  2017000600981  2017000600982  2017000600983  2017000600984  2017000600985  2017000600986  2017000600987  2017000600988  2017000600989  2017000600990  2017000600992  2017000600995  2017000600996  2017000600997  2017000600998  2017000600999  2017000601000  2017000601001  2017000601002  2017000601003  2017000601005  2017000601006  2017000601007  2017000601008  2017000601009  2017000601010  2017000601011  2017000601012  2017000601013  2017000601014  2017000601015  2017000601016  2017000601017  2017000601018  2017000601019  2017000601020  2017000601021  2017000601022  2017000601023  2017000601024  2017000601025  2017000601026  2017000601027  2017000601028  2017000601030  2017000601031  2017000601032  2017000601033  2017000601034  2017000601035  2017000601036  2017000601037  2017000601038  2017000601039  2017000601040  2017000601041  2017000601042  2017000601043  2017000601044  2017000601045  2017000601046  2017000601047  2017000601048  2017000601049  2017000601050  2017000601051  2017000601052  2017000601053  2017000601054  2017000601055  2017000601056  2017000601057  2017000601058  2017000601059  2017000601060  2017000601061  2017000601062  2017000601063  2017000601064  2017000601065  2017000601066  2017000601067  2017000601068  2017000601069  2017000601070  2017000601071  2017000601072  2017000601073  2017000601074  2017000601075  2017000601076  2017000601077  2017000601078  2017000601079  2017000601080  2017000601081  2017000601082  2017000601083  2017000601084  2017000601085  2017000601086  2017000601087  2017000601088  2017000601089  2017000601090  2017000601091  2017000601092  2017000601093  2017000601094  2017000601095  2017000601096  2017000601097  2017000601098  2017000601099  2017000601100  2017000601101  2017000601102  2017000601103  2017000601104  2017000601105  2017000601106  2017000601107  2017000601108  2017000601109  2017000601110  2017000601111  2017000601112  2017000601113  2017000601114  2017000601115  2017000601116  2017000601117  2017000601118  2017000601119  2017000601120  2017000601121  2017000601122  2017000601123  2017000601124  2017000601125  2017000601126  2017000601127  2017000601128  2017000601129  2017000601130  2017000601131  2017000601132  2017000601133  2017000601134  2017000601135  2017000601136  2017000601137  2017000601138  2017000601139  2017000601140  2017000601141  2017000601142  2017000601143  2017000601144  2017000601145  2017000601146  2017000601147  2017000601148  2017000601149  2017000601150  2017000601151  2017000601152  2017000601153  2017000601154  2017000601155  2017000601156  2017000601157  2017000601158  2017000601159  2017000601160  2017000601161  2017000601162  2017000601163  2017000601164  2017000601165  2017000601166  2017000601167  2017000601168  2017000601169  2017000601170  2017000601171  2017000601172  2017000601173  2017000601174  2017000601175  2017000601176  2017000601177  2017000601178  2017000601179  2017000601180  2017000601181  2017000601182  2017000601183  2017000601184  2017000601185  2017000601186  2017000601187  2017000601188  2017000601189  2017000601190  2017000601191  2017000601192  2017000601193  2017000601194  2017000601195  2017000601196  2017000601197  2017000601198  2017000601199  2017000601200  2017000601201  2017000601202  2017000601203  2017000601204  2017000601205  2017000601206  2017000601207  2017000601208  2017000601209  2017000601210  2017000601211  2017000601212  2017000601213  2017000601214  2017000601215  2017000601216  2017000601217  2017000601218  2017000601219  2017000601220  2017000601221  2017000601222  2017000601223  2017000601224  2017000601225  2017000601226  2017000601227  2017000601228  2017000601229  2017000601230  2017000601231  2017000601232  2017000601233  2017000601234  2017000601235  2017000601236  2017000601237  2017000601238  2017000601239  2017000601240  2017000601241  2017000601242  2017000601243  2017000601244  2017000601245  2017000601246  2017000601247  2017000601248  2017000601249  2017000601250  2017000601251  2017000601252  2017000601253  2017000601254  2017000601255  2017000601256  2017000601257  2017000601258  2017000601259  2017000601260  2017000601261  2017000601262  2017000601263  2017000601264  2017000601265  2017000601266  2017000601267  2017000601268  2017000601269  2017000601270  2017000601271  2017000601272  2017000601273  2017000601274  2017000601275  2017000601276  2017000601277  2017000601278  2017000601279  2017000601280  2017000601281  2017000601282  2017000601283  2017000601284  2017000601285  2017000601286  2017000601287  2017000601288  2017000601289  2017000601290  2017000601291  2017000601292  2017000601293  2017000601294  2017000601295  2017000601296  2017000601297  2017000601298  2017000601299  2017000601300  2017000601301  2017000601302  2017000601303  2017000601304  2017000601305  2017000601306  2017000601307  2017000601308  2017000601309  2017000601310  2017000601311  2017000601312  2017000601313  2017000601314  2017000601315  2017000601316  2017000601317  2017000601318  2017000601319  2017000601320  2017000601321  2017000601322  2017000601323  2017000601324  2017000601325  2017000601326  2017000601327  2017000601328  2017000601329  2017000601330  2017000601331  2017000601332  2017000601333  2017000601334  2017000601335  2017000601336  2017000601337  2017000601338  2017000601339  2017000601340  2017000601341  2017000601342  2017000601343  2017000601344  2017000601345  2017000601346  2017000601347  2017000601348  2017000601349  2017000601350  2017000601351  2017000601352  2017000601353  2017000601354  2017000601355  2017000601356  2017000601357  2017000601358  2017000601359  2017000601360  2017000601361  2017000601362  2017000601363  2017000601364  2017000601365  2017000601366  2017000601367  2017000601368  2017000601369  2017000601370  2017000601371  2017000601372  2017000601373  2017000601374  2017000601375  2017000601376  2017000601377  2017000601379  2017000601380  2017000601381  2017000601382  2017000601383  2017000601384  2017000601385  2017000601386  2017000601387  2017000601388  2017000601389  2017000601390  2017000601391  2017000601392  2017000601393  2017000601394  2017000601395  2017000601396  2017000601397  2017000601398  2017000601399  2017000601400  2017000601401  2017000601402  2017000601403  2017000601404  2017000601405  2017000601406  2017000601407  2017000601408  2017000601409  2017000601410  2017000601411  2017000601412  2017000601413  2017000601414  2017000601415  2017000601416  2017000601417  2017000601418  2017000601419  2017000601420  2017000601421  2017000601422  2017000601423  2017000601424  2017000601425  2017000601426  2017000601427  2017000601428  2017000601429  2017000601430  2017000601431  2017000601432  2017000601433  2017000601434  2017000601435  2017000601436  2017000601437  2017000601438  2017000601439  2017000601440  2017000601441  2017000601442  2017000601443  2017000601444  2017000601445  2017000601446  2017000601447  2017000601448  2017000601449  2017000601450  2017000601451  2017000601452  2017000601453  2017000601454  2017000601455  2017000601456  2017000601457  2017000601458  2017000601459  2017000601460  2017000601461  2017000601462  2017000601463  2017000601464  2017000601465  2017000601466  2017000601467  2017000601468  2017000601469  2017000601470  2017000601471  2017000601472  2017000601473  2017000601474  2017000601475  2017000601476  2017000601477  2017000601478  2017000601479  2017000601480  2017000601481  2017000601482  2017000601483  2017000601484  2017000601485  2017000601486  2017000601487  2017000601488  2017000601489  2017000601490  2017000601491  2017000601492  2017000601493  2017000601494  2017000601495  2017000601496  2017000601497  2017000601498  2017000601499  2017000601500  2017000601501  2017000601502  2017000601503  2017000601504  2017000601505  2017000601506  2017000601507  2017000601508  2017000601509  2017000601510  2017000601511  2017000601512  2017000601513  2017000601514  2017000601515  2017000601516  2017000601517  2017000601518  2017000601519  2017000601520  2017000601521  2017000601522  2017000601523  2017000601524  2017000601525  2017000601526  2017000601527  2017000601528  2017000601529  2017000601530  2017000601531  2017000601532  2017000601533  2017000601534  2017000601535  2017000601536  2017000601537  2017000601538  2017000601539  2017000601540  2017000601541  2017000601542  2017000601543  2017000601544  2017000601545  2017000601546  2017000601547  2017000601548  2017000601549  2017000601550  2017000601551  2017000601552  2017000601553  2017000601554  2017000601555  2017000601556  2017000601557  2017000601558  2017000601559  2017000601560  2017000601561  2017000601562  2017000601563  2017000601564  2017000601565  2017000601566  2017000601567  2017000601568  2017000601569  2017000601570  2017000601571  2017000601572  2017000601573  2017000601574  2017000601575  2017000601576  2017000601577  2017000601578  2017000601579  2017000601580  2017000601581  2017000601582  2017000601583  2017000601584  2017000601585  2017000601586  20170",
      "center_classification_date": "20190207",
      "distribution_pattern": "Nationwide distribution. Foreign distribution to Argentina, Australia, Canada, Chile, China, Colombia, Germany, India, Japan, Netherlands, South Africa, Turkey, Saudi Arabia, United Kingdom, and Mexico.",
      "state": "MI",
      "product_description": "Naera Hospital Bassinet, Model No. 4402",
      "report_date": "20190213",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Stryker Medical Division of Stryker Corporation",
      "recall_number": "Z-0813-2019",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "81393",
      "termination_date": "20200429",
      "more_code_info": "00601587  2017000601588  2017000601589  2017000601590  2017000601591  2017000601592  2017000601593  2017000601594  2017000601595  2017000601596  2017000601597  2017000601598  2017000601599  2017000601600  2017000601601  2017000601602  2017000601603  2017000601604  2017000601605  2017000601606  2017000601607  2017000601608  2017000601609  2017000601610  2017000601611  2017000601612  2017000601613  2017000601614  2017000601615  2017000601616  2017000601617  2017000601618  2017000601619  2017000601620  2017000601621  2017000601622  2017000601623  2017000601624  2017000601625  2017000601626  2017000601627  2017000601628  2017000601629  2017000601630  2017000601631  2017000601632  2017000601633  2017000601634  2017000601635  2017000601636  2017000601637  2017000601638  2017000601639  2017000601640  2017000601641  2017000601642  2017000601643  2017000601644  2017000601645  2017000601646  2017000601647  2017000601648  2017000601649  2017000601650  2017000601651  2017000601652  2017000601653  2017000601654  2017000601655  2017000601656  2017000601657  2017000601658  2017000601659  2017000601660  2017000601661  2017000601662  2017000601663  2017000601664  2017000601665  2017000601666  2017000601667  2017000601668  2017000601669  2017000601670  2017000601671  2017000601672  2017000601673  2017000601674  2017000601675  2017000601676  2017000601677  2017000601678  2017000601679  2017000601680  2017000601681  2017000601682  2017000601683  2017000601684  2017000601685  2017000601686  2017000601687  2017000601688  2017000601689  2017000601690  2017000601691  2017000601692  2017000601693  2017000601694  2017000601695  2017000601696  2017000601697  2017000601698  2017000601699  2017000601700  2017000601701  2017000601702  2017000601703  2017000601704  2017000601705  2017000601706  2017000601707  2017000601708  2017000601709  2017000601710  2017000601711  2017000601712  2017000601713  2017000601714  2017000601715  2017000601716  2017000601717  2017000601718  2017000601719  2017000601720  2017000601721  2017000601722  2017000601723  2017000601724  2017000601725  2017000601726  2017000601727  2017000601728  2017000601729  2017000601730  2017000601731  2017000601732  2017000601733  2017000601734  2017000601735  2017000601736  2017000601737  2017000601738  2017000601739  2017000601740  2017000601741  2017000601742  2017000601743  2017000601744  2017000601745  2017000601746  2017000601747  2017000601748  2017000601749  2017000601750  2017000601751  2017000601752  2017000601753  2017000601754  2017000601755  2017000601756  2017000601757  2017000601758  2017000601759  2017000601760  2017000601761  2017000601762  2017000601763  2017000601764  2017000601765  2017000601766  2017000601767  2017000601768  2017000601769  2017000601770  2017000601771  2017000601772  2017000601773  2017000601774  2017000601775  2017000601776  2017000601777  2017000601778  2017000601779  2017000601780  2017000601781  2017000601782  2017000601783  2017000601784  2017000601785  2017000601786  2017000601787  2017000601788  2017000601789  2017000601790  2017000601791  2017000601792  2017000601793  2017000601794  2017000601795  2017000601796  2017000601797  2017000601798  2017000601799  2017000601800  2017000601801  2017000601802  2017000601803  2017000601804  2017000601805  2017000601806  2017000601807  2017000601808  2017000601809  2017000601810  2017000601811  2017000601812  2017000601813  2017000601814  2017000601815  2017000601816  2017000601817  2017000601818  2017000601819  2017000601820  2017000601821  2017000601822  2017000601823  2017000601824  2017000601825  2017000601826  2017000601827  2017000601828  2017000601829  2017000601830  2017000601831  2017000601832  2017000601833  2017000601834  2017000601835  2017000601836  2017000601837  2017000601838  2017000601839  2017000601840  2017000601841  2017000601842  2017000601843  2017000601844  2017000601845  2017000601846  2017000601847  2017000601848  2017000601849  2017000601850  2017000601851  2017000601852  2017000601853  2017000601854  2017000601855  2017000601856  2017000601857  2017000601858  2017000601859  2017000601860  2017000601861  2017000601862  2017000601863  2017000601864  2017000601865  2017000601866  2017000601867  2017000601868  2017000601869  2017000601870  2017000601871  2017000601872  2017000601873  2017000601874  2017000601875  2017000601876  2017000601877  2017000601878  2017000601879  2017000601880  2017000601881  2017000601882  2017000601883  2017000601884  2017000601885  2017000601886  2017000601887  2017000601888  2017000601889  2017000601890  2017000601891  2017000601892  2017000601893  2017000601894  2017000601895  2017000601896  2017000601897  2017000601898  2017000601899  2017000601900  2017000601901  2017000601902  2017000601903  2017000601904  2017000601905  2017000601906  2017000601907  2017000601908  2017000601909  2017000601910  2017000601911  2017000601912  2017000601913  2017000601914  2017000601915  2017000601916  2017000601917  2017000601918  2017000601919  2017000601920  2017000601921  2017000601922  2017000601923  2017000601924  2017000601926  2017000601927  2017000601928  2017000601929  2017000601930  2017000601931  2017000601932  2017000601933  2017000601934  2017000601935  2017000601936  2017000601937  2017000601938  2017000601939  2017000601941  2017000601942  2017000601943  2017000601944  2017000601945  2017000601946  2017000601947  2017000601948  2017000601949  2017000601950  2017000601951  2017000601952  2017000601953  2017000601954  2017000601955  2017000601956  2017000601957  2017000601958  2017000601959  2017000601960  2017000601961  2017000601962  2017000601963  2017000601964  2017000601965  2017000601966  2017000601967  2017000601968  2017000601969  2017000601970  2017000601971  2017000601972  2017000601973  2017000601974  2017000601975  2017000601976  2017000601977  2017000601978  2017000601979  2017000601980  2017000601981  2017000601982  2017000601983  2017000601984  2017000601985  2017000601986  2017000601987  2017000601988  2017000601989  2017000601990  2017000601991  2017000601992  2017000601993  2017000601994  2017000601995  2017000601996  2017000601997  2017000601998  2017000601999  2017000602000  2017000602001  2017000602002  2017000602003  2017000602004  2017000602005  2017000602006  2017000602007  2017000602008  2017000602009  2017000602010  2017000602011  2017000602012  2017000602013  2017000602014  2017000602015  2017000602016  2017000602017  2017000602018  2017000602019  2017000602020  2017000602021  2017000602022  2017000602023  2017000602024  2017000602025  2017000602026  2017000602027  2017000602028  2017000602029  2017000602030  2017000602031  2017000602032  2017000602033  2017000602034  2017000602035  2017000602036  2017000602037  2017000602038  2017000602039  2017000602040  2017000602041  2017000602042  2017000602043  2017000602044  2017000602045  2017000602046  2017000602047  2017000602048  2017000602049  2017000602050  2017000602051  2017000602052  2017000602053  2017000602054  2017000602055  2017000602056  2017000602057  2017000602058  2017000602059  2017000602060  2017000602061  2017000602062  2017000602063  2017000602064  2017000602065  2017000602066  2017000602067  2017000602068  2017000602069  2017000602070  2017000602071  2017000602072  2017000602073  2017000602074  2017000602075  2017000602076  2017000602077  2017000602078  2017000602079  2017000602080  2017000602081  2017000602082  2017000602083  2017000602084  2017000602085  2017000602086  2017000602087  2017000602088  2017000602089  2017000602090  2017000602091  2017000602092  2017000602093  2017000602094  2017000602095  2017000602096  2017000602097  2017000602098  2017000602099  2017000602100  2017000602101  2017000602102  2017000602103  2017000602104  2017000602105  2017000602106  2017000602107  2017000602108  2017000602109  2017000602110  2017000602111  2017000602112  2017000602113  2017000602114  2017000602115  2017000602116  2017000602117  2017000602118  2017000602119  2017000602120  2017000602121  2017000602122  2017000602123  2017000602124  2017000602125  2017000602126  2017000602127  2017000602128  2017000602129  2017000602130  2017000602131  2017000602132  2017000602133  2017000602134  2017000602135  2017000602136  2017000602137  2017000602138  2017000602139  2017000602140  2017000602141  2017000602142  2017000602143  2017000602144  2017000602145  2017000602146  2017000602147  2017000602148  2017000602149  2017000602150  2017000602151  2017000602152  2017000602153  2017000602154  2017000602155  2017000602156  2017000602157  2017000602158  2017000602159  2017000602160  2017000602161  2017000602162  2017000602163  2017000602164  2017000602165  2017000602166  2017000602167  2017000602168  2017000602169  2017000602170  2017000602171  2017000602172  2017000602173  2017000602174  2017000602175  2017000602176  2017000602177  2017000602178  2017000602179  2017000602180  2017000602181  2017000602182  2017000602183  2017000602184  2017000602185  2017000602186  2017000602187  2017000602188  2017000602189  2017000602190  2017000602191  2017000602192  2017000602193  2017000602194  2017000602195  2017000602196  2017000602197  2017000602198  2017000602199  2017000602200  2017000602201  2017000602202  2017000602203  2017000602204  2017000602205  2017000602206  2017000602207  2017000602208  2017000602209  2017000602210  2017000602211  2017000602212  2017000602213  2017000602214  2017000602215  2017000602216  2017000602217  2017000602218  2017000602219  2017000602220  2017000602221  2017000602222  2017000602223  2017000602224  2017000602225  2017000602226  2017000602227  2017000602228  2017000602229  2017000602230  2017000602231  2017000602232  2017000602233  2017000602234  2017000602235  2017000602236  2017000602237  2017000602238  2017000602239  2017000602240  2017000602241  2017000602242  2017000602243  2017000602244  2017000602245  2017000602246  2017000602247  2017000602248  2017000602249  2017000602250  2017000602251  2017000602252  2017000602253  2017000602254  2017000602255  2017000602256  2017000602257  2017000602258  2017000602259  2017000602260  2017000602261  2017000602262  2017000602263  2017000602264  2017000602265  2017000602266  2017000602267  2017000602268  2017000602269  2017000602270  2017000602271  2017000602272  2017000602273  2017000602274  2017000602275  2017000602276  2017000602277  2017000602278  2017000602279  2017000602280  2017000602281  2017000602282  2017000602283  2017000602284  2017000602285  2017000602286  2017000602287  2017000602288  2017000602289  2017000602290  2017000602291  2017000602292  2017000602293  2017000602294  2017000602295  2017000602296  2017000602297  2017000602298  2017000602299  2017000602300  2017000602301  2017000602302  2017000602303  2017000602304  2017000602305  2017000602306  2017000602307  2017000602308  2017000602309  2017000602310  2017000602311  2017000602312  2017000602313  2017000602314  2017000602315  2017000602316  2017000602317  2017000602318  2017000602319  2017000602320  2017000602321  2017000602322  2017000602323  2017000602324  2017000602325  2017000602326  2017000602327  2017000602328  2017000602329  2017000602330  2017000602331  2017000602332  2017000602333  2017000602334  2017000602335  2017000602336  2017000602337  2017000602338  2017000602339  2017000602340  2017000602341  2017000602342  2017000602343  2017000602345  2017000602346  2017000602347  2017000602350  2017000602351  2017000602352  2017000602353  2017000602354  2017000602355  2017000602356  2017000602357  2017000602358  2017000602359  2017000602360  2017000602361  2017000602362  2017000602363  2017000602364  2017000602365  2017000602366  2017000602367  2017000602368  2017000602369  2017000602370  2017000602371  2017000602372  2017000602373  2017000602374  2017000602375  2017000602376  2017000602377  2017000602378  2017000602379  2017000602380  2017000602381  2017000602382  2017000602383  2017000602384  2017000602385  2017000602386  2017000602387  2017000602388  2017000602389  2017000602390  2017000602391  2017000602392  2017000602393  2017000602394  2017000602395  2017000602396  2017000602397  2017000602398  2017000602399  2017000602400  2017000602401  2017000602402  2017000602403  2017000602404  2017000602405  2017000602406  2017000602407  2017000602408  2017000602409  2017000602410  2017000602411  2017000602412  2017000602413  2017000602414  2017000602415  2017000602416  2017000602417  2017000602418  2017000602419  2017000602420  2017000602421  2017000602422  2017000602423  2017000602424  2017000602425  2017000602426  2017000602427  2017000602428  2017000602429  2017000602430  2017000602431  2017000602432  2017000602433  2017000602434  2017000602435  2017000602436  2017000602437  2017000602438  2017000602439  2017000602440  2017000602441  2017000602442  2017000602443  2017000602444  2017000602445  2017000602446  2017000602447  2017000602448  2017000602449  2017000602450  2017000602451  2017000602452  2017000602453  2017000602454  2017000602455  2017000602456  2017000602457  2017000602458  2017000602459  2017000602460  2017000602461  2017000602462  2017000602463  2017000602464  2017000602465  2017000602466  2017000602467  2017000602468  2017000602469  2017000602470  2017000602471  2017000602472  2017000602473  2017000602474  2017000602475  2017000602476  2017000602477  2017000602478  2017000602479  2017000602480  2017000602481  2017000602482  2017000602483  2017000602484  2017000602485  2017000602486  2017000602487  2017000602488  2017000602489  2017000602490  2017000602492  2017000602493  2017000602494  2017000602495  2017000602496  2017000602497  2017000602498  2017000602499  2017000602500  2017000602501  2017000602502  2017000602503  2017000602504  2017000602505  2017000602506  2017000602507  2017000602508  2017000602509  2017000602510  2017000602511  2017000602512  2017000602513  2017000602514  2017000602515  2017000602516  2017000602517  2017000602518  2017000602519  2017000602520  2017000602521  2017000602522  2017000602523  2017000602524  2017000602525  2017000602526  2017000602527  2017000602528  2017000602529  2017000602530  2017000602531  2017000602532  2017000602533  2017000602534  2017000602535  2017000602536  2017000602537  2017000602538  2017000602539  2017000602540  2017000602541  2017000602542  2017000602543  2017000602544  2017000602545  2017000602546  2017000602547  2017000602548  2017000602549  2017000602550  2017000602551  2017000602552  2017000602553  2017000602554  2017000602555  2017000602556  2017000602557  2017000602558  2017000602559  2017000602560  2017000602561  2017000602562  2017000602563  2017000602564  2017000602565  2017000602566  2017000602567  2017000602568  2017000602569  2017000602570  2017000602571  2017000602572  2017000602573  2017000602574  2017000602575  2017000602576  2017000602577  2017000602578  2017000602579  2017000602580  2017000602581  2017000602582  2017000602583  2017000602584  2017000602585  2017000602586  2017000602587  2017000602588  2017000602589  2017000602590  2017000602591  2017000602592  2017000602593  2017000602594  2017000602595  2017000602596  2017000602597  2017000602598  2017000602599  2017000602600  2017000602601  2017000602602  2017000602603  2017000602604  2017000602605  2017000602606  2017000602607  2017000602608  2017000602609  2017000602610  2017000602611  2017000602612  2017000602613  2017000602614  2017000602615  2017000602616  2017000602617  2017000602618  2017000602619  2017000602620  2017000602621  2017000602622  2017000602623  2017000602624  2017000602625  2017000602626  2017000602627  2017000602628  2017000602629  2017000602630  2017000602631  2017000602632  2017000602633  2017000602634  2017000602635  2017000602636  2017000602637  2017000602638  2017000602639  2017000602640  2017000602641  2017000602642  2017000602643  2017000602644  2017000602645  2017000602646  2017000602647  2017000602648  2017000602649  2017000602650  2017000602651  2017000602652  2017000602653  2017000602654  2017000602655  2017000602656  2017000602657  2017000602658  2017000602659  2017000602660  2017000602661  2017000602662  2017000602663  2017000602664  2017000602665  2017000602666  2017000602667  2017000602668  2017000602669  2017000602670  2017000602671  2017000602672  2017000602673  2017000602674  2017000602675  2017000602676  2017000602677  2017000602678  2017000602679  2017000602680  2017000602681  2017000602682  2017000602683  2017000602684  2017000602685  2017000602686  2017000602687  2017000602688  2017000602689  2017000602690  2017000602691  2017000602692  2017000602693  2017000602694  2017000602695  2017000602696  2017000602697  2017000602698  2017000602699  2017000602700  2017000602701  2017000602702  2017000602703  2017000602704  2017000602705  2017000602706  2017000602707  2017000602708  2017000602709  2017000602710  2017000602711  2017000602712  2017000602713  2017000602714  2017000602715  2017000602716  2017000602717  2017000602718  2017000602719  2017000602720  2017000602721  2017000602722  2017000602723  2017000602724  2017000602725  2017000602726  2017000602727  2017000602728  2017000602729  2017000602730  2017000602731  2017000602732  2017000602733  2017000602734  2017000602735  2017000602736  2017000602737  2017000602738  2017000602739  2017000602740  2017000602741  2017000602742  2017000602743  2017000602744  2017000602745  2017000602746  2017000602747  2017000602748  2017000602749  2017000602750  2017000602751  2017000602752  2017000602753  2017000602754  2017000602755  2017000602756  2017000602757  2017000602758  2017000602759  2017000602760  2017000602761  2017000602762  2017000602763  2017000602764  2017000602765  2017000602766  2017000602767  2017000602768  2017000602769  2017000602770  2017000602771  2017000602772  2017000602773  2017000602774  2017000602775  2017000602776  2017000602777  2017000602778  2017000602779  2017000602780  2017000602781  2017000602782  2017000602783  2017000602784  2017000602785  2017000602786  2017000602787  2017000602788  2017000602789  2017000602790  2017000602791  2017000602792  2017000602793  2017000602794  2017000602795  2017000602796  2017000602797  2017000602798  2017000602799  2017000602800  2017000602801  2017000602802  2017000602803  2017000602804  2017000602805  2017000602806  2017000602807  2017000602808  2017000602809  2017000602810  2017000602811  2017000602812  2017000602813  2017000602814  2017000602815  2017000602816  2017000602817  2017000602818  2017000602819  2017000602820  2017000602821  2017000602822  2017000602823  2017000602824  2017000602825  2017000602826  2017000602827  2017000602828  2017000602829  2017000602830  2017000602831  2017000602832  2017000602833  2017000602834  2017000602835  2017000602836  2017000602837  2017000602838  2017000602839  2017000602840  2017000602841  2017000602842  2017000602843  2017000602844  2017000602845  2017000602846  2017000602847  2017000602848  2017000602849  2017000602850  2017000602851  2017000602852  2017000602853  2017000602854  2017000602855  2017000602856  2017000602857  2017000602858  2017000602859  2017000602860  2017000602861  2017000602862  2017000602863  2017000602864  2017000602865  2017000602866  2017000602867  2017000602868  2017000602869  2017000602870  2017000602871  2017000602872  2017000602873  2017000602874  2017000602875  2017000602876  2017000602877  2017000602878  2017000602879  2017000602880  2017000602881  2017000602882  2017000602883  2017000602884  2017000602885  2017000602886  2017000602887  2017000602888  2017000602889  2017000602890  2017000602891  2017000602892  2017000602893  2017000602894  2017000602895  2017000602896  2017000602897  2017000602898  2017000602899  2017000602900  2017000602901  2017000602902  2017000602903  2017000602904  2017000602905  2017000602906  2017000602907  2017000602908  2017000602909  2017000602910  2017000602911  2017000602912  2017000602913  2017000602914  2017000602915  2017000602916  2017000602917  2017000602918  2017000602919  2017000602920  2017000602921  2017000602922  2017000602923  2017000602924  2017000602925  2017000602926  2017000602927  2017000602928  2017000602929  2017000602930  2017000602931  2017000602932  2017000602933  2017000602934  2017000602935  2017000602936  2017000602937  2017000602938  2017000602939  2017000602940  2017000602941  2017000602942  2017000602943  2017000602944  2017000602945  2017000602946  2017000602947  2017000602948  2017000602949  2017000602950  2017000602951  2017000602952  2017000602953  2017000602954  2017000602955  2017000602956  2017000602957  2017000602958  2017000602959  2017000602960  2017000602961  2017000602962  2017000602963  2017000602964  2017000602965  2017000602966  2017000602967  2017000602968  2017000602969  2017000602970  2017000602971  2017000602972  2017000602973  2017000602974  2017000602975  2017000602976  2017000602977  2017000602978  2017000602979  2017000602980  2017000602981  2017000602982  2017000602983  2017000602984  2017000602985  2017000602986  2017000602987  2017000602988  2017000602989  2017000602990  2017000602991  2017000602992  2017000602993  2017000602994  2017000602995  2017000602996  2017000602997  2017000602998  2017000602999  2017000603000  2017000700001  2017000700002  2017000700003  2017000700004  2017000700005  2017000700006  2017000700007  2017000700008  2017000700009  2017000700010  2017000700011  2017000700012  2017000700013  2017000700014  2017000700015  2017000700016  2017000700017  2017000700018  2017000700019  2017000700020  2017000700021  2017000700022  2017000700023  2017000700024  2017000700025  2017000700026  2017000700027  2017000700028  2017000700029  2017000700030  2017000700031  2017000700032  2017000700033  2017000700034  2017000700035  2017000700036  2017000700037  2017000700038  2017000700039  2017000700040  2017000700041  2017000700042  2017000700043  2017000700044  2017000700045  2017000700046  2017000700047  2017000700048  2017000700049  2017000700050  2017000700051  2017000700052  2017000700053  2017000700054  2017000700055  2017000700056  2017000700057  2017000700058  2017000700059  2017000700060  2017000700061  2017000700062  2017000700063  2017000700064  2017000700065  2017000700066  2017000700067  2017000700068  2017000700069  2017000700070  2017000700071  2017000700072  2017000700073  2017000700074  2017000700075  2017000700076  2017000700077  2017000700078  2017000700079  2017000700080  2017000700081  2017000700082  2017000700083  2017000700084  2017000700085  2017000700086  2017000700087  2017000700088  2017000700089  2017000700090  2017000700091  2017000700092  2017000700093  2017000700094  2017000700095  2017000700096  2017000700097  2017000700098  2017000700099  2017000700100  2017000700101  2017000700102  2017000700103  2017000700104  2017000700105  2017000700106  2017000700107  2017000700108  2017000700109  2017000700110  2017000700111  2017000700112  2017000700113  2017000700114  2017000700115  2017000700116  2017000700117  2017000700118  2017000700119  2017000700120  2017000700121  2017000700122  2017000700123  2017000700124  2017000700125  2017000700126  2017000700127  2017000700128  2017000700129  2017000700130  2017000700131  2017000700132  2017000700133  2017000700134  2017000700135  2017000700136  2017000700137  2017000700138  2017000700139  2017000700140  2017000700141  2017000700142  2017000700143  2017000700144  2017000700145  2017000700146  2017000700147  2017000700148  2017000700149  2017000700150  2017000700151  2017000700152  2017000700153  2017000700154  2017000700155  2017000700156  2017000700157  2017000700158  2017000700159  2017000700160  2017000700161  2017000700162  2017000700163  2017000700164  2017000700165  2017000700166  2017000700167  2017000700168  2017000700169  2017000700170  2017000700171  2017000700172  2017000700173  2017000700174  2017000700175  2017000700176  2017000700177  2017000700178  2017000700179  2017000700180  2017000700181  2017000700182  2017000700183  2017000700184  2017000700185  2017000700186  2017000700187  2017000700188  2017000700189  2017000700190  2017000700191  2017000700192  2017000700193  2017000700194  2017000700195  2017000700196  2017000700197  2017000700198  2017000700199  2017000700200  2017000700201  2017000700202  2017000700203  2017000700204  2017000700205  2017000700206  2017000700207  2017000700208  2017000700209  2017000700210  2017000700211  2017000700212  2017000700213  2017000700214  2017000700215  2017000700216  2017000700217  2017000700218  2017000700219  2017000700220  2017000700221  2017000700222  2017000700223  2017000700224  2017000700225  2017000700226  2017000700227  2017000700228  2017000700229  2017000700230  2017000700231  2017000700232  2017000700233  2017000700234  2017000700235  2017000700236  2017000700237  2017000700238  2017000700239  2017000700240  2017000700241  2017000700242  2017000700243  2017000700244  2017000700245  2018000600001  2018000600002  2018000600003  2018000600004  2018000600005  2018000600006  2018000600007  2018000600008  2018000600010  2018000600011  2018000600012  2018000600013  2018000600014  2018000600015  2018000600016  2018000600017  2018000600018  2018000600019  2018000600020  2018000600021  2018000600022  2018000600023  2018000600024  2018000600025  2018000600026  2018000600027  2018000600028  2018000600029  2018000600030  2018000600031  2018000600032  2018000600033  2018000600034  2018000600035  2018000600036  2018000600037  2018000600038  2018000600039  2018000600040  2018000600041  2018000600042  2018000600043  2018000600044  2018000600045  2018000600046  2018000600047  2018000600048  2018000600049  2018000600050  2018000600051  2018000600052  2018000600053  2018000600054  2018000600055  2018000600056  2018000600057  2018000600058  2018000600059  2018000600060  2018000600061  2018000600062  2018000600063  2018000600064  2018000600065  2018000600066  2018000600067  2018000600068  2018000600069  2018000600070  2018000600071  2018000600072  2018000600073  2018000600074  2018000600075  2018000600076  2018000600077  2018000600078  2018000600079  2018000600080  2018000600081  2018000600082  2018000600083  2018000600084  2018000600085  2018000600086  2018000600087  2018000600088  2018000600089  2018000600090  2018000600091  2018000600092  2018000600093  2018000600094  2018000600095  2018000600096  2018000600097  2018000600098  2018000600099  2018000600100  2018000600101  2018000600102  2018000600103  2018000600104  2018000600105  2018000600106  2018000600107  2018000600108  2018000600109  2018000600110  2018000600111  2018000600112  2018000600113  2018000600114  2018000600115  2018000600116  2018000600117  2018000600118  2018000600119  2018000600120  2018000600121  2018000600122  2018000600123  2018000600124  2018000600125  2018000600126  2018000600127  2018000600128  2018000600129  2018000600130  2018000600131  2018000600132  2018000600133  2018000600134  2018000600135  2018000600136  2018000600137  2018000600138  2018000600139  2018000600140  2018000600141  2018000600142  2018000600143  2018000600144  2018000600145  2018000600146  2018000600147  2018000600148  2018000600149  2018000600150  2018000600151  2018000600152  2018000600153  2018000600154  2018000600155  2018000600156  2018000600157  2018000600158  2018000600159  2018000600160  2018000600161  2018000600162  2018000600163  2018000600164  2018000600165  2018000600166  2018000600167  2018000600168  2018000600169  2018000600170  2018000600172  2018000600173  2018000600174  2018000600175  2018000600176  2018000600177  2018000600178  2018000600179  2018000600180  2018000600181  2018000600182  2018000600183  2018000600184  2018000600185  2018000600186  2018000600187  2018000600188  2018000600189  2018000600190  2018000600191  2018000600192  2018000600193  2018000600194  2018000600195  2018000600196  2018000600197  2018000600198  2018000600199  2018000600200  2018000600201  2018000600202  2018000600203  2018000600204  2018000600205  2018000600206  2018000600207  2018000600208  2018000600209  2018000600210  2018000600211  2018000600212  2018000600213  2018000600214  2018000600215  2018000600216  2018000600217  2018000600218  2018000600219  2018000600220  2018000600221  2018000600222  2018000600223  2018000600224  2018000600225  2018000600226  2018000600227  2018000600228  2018000600229  2018000600230  2018000600231  2018000600232  2018000600233  2018000600234  2018000600235  2018000600236  2018000600237  2018000600238  2018000600239  2018000600240  2018000600241  2018000600242  2018000600243  2018000600244  2018000600245  2018000600246  2018000600247  2018000600248  2018000600249  2018000600250  2018000600251  2018000600252  2018000600253  2018000600254  2018000600255  2018000600256  2018000600257  2018000600258  2018000600259  2018000600260  2018000600261  2018000600262  2018000600263  2018000600264  2018000600265  2018000600266  2018000600267  2018000600268  2018000600269  2018000600270  2018000600271  2018000600272  2018000600274  2018000600275  2018000600276  2018000600277  2018000600278  2018000600279  2018000600280  2018000600281  2018000600282  2018000600283  2018000600284  2018000600285  2018000600286  2018000600287  2018000600288  2018000600289  2018000600290  2018000600291  2018000600292  2018000600293  2018000600294  2018000600295  2018000600296  2018000600297  2018000600298  2018000600299  2018000600300  2018000600301  2018000600302  2018000600303  2018000600304  2018000600305  2018000600306  2018000600307  2018000600308  2018000600309  2018000600310  2018000600311  2018000600312  2018000600313  2018000600314  2018000600315  2018000600316  2018000600317  2018000600318  2018000600319  2018000600320  2018000600321  2018000600322  2018000600323  2018000600324  2018000600325  2018000600326  2018000600327  2018000600328  2018000600329  2018000600330  2018000600331  2018000600332  2018000600333  2018000600334  2018000600335  2018000600336  2018000600337  2018000600338  2018000600339  2018000600340  2018000600341  2018000600342  2018000600343  2018000600344  2018000600345  2018000600346  2018000600347  2018000600348  2018000600349  2018000600350  2018000600351  2018000600352  2018000600353  2018000600354  2018000600355  2018000600356  2018000600357  2018000600358  2018000600359  2018000600360  2018000600361  2018000600362  2018000600363  2018000600364  2018000600365  2018000600366  2018000600367  2018000600368  2018000600369  2018000600370  2018000600371  2018000600372  2018000600373  2018000600374  2018000600375  2018000600376  2018000600377  2018000600378  2018000600379  2018000600380  2018000600381  2018000600382  2018000600383  2018000600384  2018000600385  2018000600386  2018000600387  2018000600388  2018000600389  2018000600390  2018000600391  2018000600392  2018000600393  2018000600394  2018000600395  2018000600396  2018000600397  2018000600398  2018000600399  2018000600400  2018000600401  2018000600402  2018000600403  2018000600404  2018000600405  2018000600406  2018000600407  2018000600408  2018000600409  2018000600410  2018000600411  2018000600412  2018000600413  2018000600414  2018000600415  2018000600416  2018000600417  2018000600418  2018000600419  2018000600420  2018000600421  2018000600422  2018000600423  2018000600424  2018000600425  2018000600426  2018000600427  2018000600428  2018000600429  2018000600430  2018000600431  2018000600432  2018000600433  2018000600434  2018000600435  2018000600436  2018000600437  2018000600438  2018000600439  2018000600440  2018000600441  2018000600442  2018000600443  2018000600444  2018000600445  2018000600446  2018000600447  2018000600448  2018000600449  2018000600450  2018000600451  2018000600452  2018000600453  2018000600454  2018000600455  2018000600456  2018000600457  2018000600458  2018000600459  2018000600460  2018000600461  2018000600462  2018000600463  2018000600464  2018000600465  2018000600466  2018000600467  2018000600468  2018000600469  2018000600470  2018000600471  2018000600472  2018000600473  2018000600474  2018000600475  2018000600476  2018000600477  2018000600478  2018000600479  2018000600480  2018000600481  2018000600482  2018000600483  2018000600484  2018000600485  2018000600486  2018000600487  2018000600488  2018000600489  2018000600490  2018000600491  2018000600492  2018000600493  2018000600494  2018000600495  2018000600496  2018000600497  2018000600498  2018000600499  2018000600500  2018000600501  2018000600502  2018000600503  2018000600504  2018000600505  2018000600506  2018000600507  2018000600508  2018000600509  2018000600510  2018000600511  2018000600512  2018000600513  2018000600514  2018000600515  2018000600516  2018000600517  2018000600518  2018000600519  2018000600520  2018000600521  2018000600522  2018000600523  2018000600 524  2018000600525  2018000600526  2018000600527  2018000600528  2018000600529  2018000600530  2018000600531  2018000600532  2018000600533  2018000600534  2018000600535  2018000600536  2018000600537  2018000600538  2018000600539  2018000600540  2018000600541  2018000600542  2018000600543  2018000600544  2018000600545  2018000600546  2018000600547  2018000600548  2018000600549  2018000600550  2018000600551  2018000600552  2018000600553  2018000600554  2018000600555  2018000600556  2018000600557  2018000600558  2018000600559  2018000600560  2018000600561  2018000600562  2018000600563  2018000600564  2018000600565  2018000600566  2018000600567  2018000600568  2018000600569  2018000600570  2018000600571  2018000600572  2018000600573  2018000600574  2018000600575  2018000600576  2018000600577  2018000600578  2018000600579  2018000600580  2018000600581  2018000600582  2018000600583  2018000600584  2018000600585  2018000600586  2018000600587  2018000600588  2018000600589  2018000600590  2018000600591  2018000600592  2018000600593  2018000600594  2018000600595  2018000600596  2018000600597  2018000600598  2018000600599  2018000600600  2018000600601  2018000600602  2018000600603  2018000600604  2018000600605  2018000600606  2018000600607  2018000600608  2018000600609  2018000600611  2018000600612  2018000600613  2018000600614  2018000600615  2018000600616  2018000600617  2018000600618  2018000600619  2018000600620  2018000600621  2018000600622  2018000600623  2018000600624  2018000600625  2018000600626  2018000600627  2018000600628  2018000600629  2018000600630  2018000600631  2018000600632  2018000600633  2018000600635  2018000600636  2018000600637  2018000600638  2018000600639  2018000600640  2018000600641  2018000600642  2018000600643  2018000600644  2018000600645  2018000600646  2018000600647  2018000600648  2018000600649  2018000600650  2018000600651  2018000600652  2018000600653  2018000600654  2018000600655  2018000600656  2018000600657  2018000600658  2018000600659  2018000600660  2018000600661  2018000600662  2018000600663  2018000600664  2018000600665  2018000600666  2018000600667  2018000600668  2018000600669  2018000600670  2018000600671  2018000600672  2018000600673  2018000600674  2018000600675  2018000600676  2018000600677  2018000600678  2018000600679  2018000600680  2018000600681  2018000600682  2018000600683  2018000600684  2018000600685  2018000600686  2018000600687  2018000600688  2018000600689  2018000600690  2018000600691  2018000600692  2018000600693  2018000600694  2018000600695  2018000600696  2018000600697  2018000600699  2018000600700  2018000600701  2018000600702  2018000600703  2018000600704  2018000600705  2018000600706  2018000600707  2018000600709  2018000600710  2018000600711  2018000600712  2018000600713  2018000600714  2018000600715  2018000600716  2018000600717  2018000600718  2018000600719  2018000600720  2018000600721  2018000600722  2018000600723  2018000600724  2018000600725  2018000600726  2018000600727  2018000600728  2018000600729  2018000600730  2018000600731  2018000600732  2018000600733  2018000600734  2018000600735  2018000600736  2018000600737  2018000600738  2018000600739  2018000600740  2018000600741  2018000600742  2018000600743  2018000600744  2018000600745  2018000600746  2018000600747  2018000600748  2018000600749  2018000600750  2018000600751  2018000600752  2018000600753  2018000600754  2018000600755  2018000600756  2018000600757  2018000600758  2018000600759  2018000600760  2018000600761  2018000600762  2018000600763  2018000600764  2018000600765  2018000600766  2018000600767  2018000600768  2018000600769  2018000600770  2018000600771  2018000600772  2018000600773  2018000600774  2018000600775  2018000600776  2018000600777  2018000600778  2018000600780  2018000600781  2018000600782  2018000600783  2018000600784  2018000600785  2018000600786  2018000600787  2018000600788  2018000600789  2018000600790  2018000600791  2018000600792  2018000600793  2018000600794  2018000600795  2018000600796  2018000600797  2018000600798  2018000600799  2018000600800  2018000600801  2018000600802  2018000600803  2018000600804  2018000600805  2018000600806  2018000600807  2018000600808  2018000600809  2018000600810  2018000600811  2018000600812  2018000600813  2018000600814  2018000600815  2018000600816  2018000600817  2018000600818  2018000600819  2018000600820  2018000600821  2018000600822  2018000600823  2018000600824  2018000600825  2018000600826  2018000600827  2018000600828  2018000600829  2018000600830  2018000600831  2018000600832  2018000600833  2018000600834  2018000600835  2018000600836  2018000600837  2018000600838  2018000600839  2018000600840  2018000600841  2018000600842  2018000600843  2018000600844  2018000600845  2018000600846  2018000600847  2018000600848  2018000600849  2018000600850  2018000600851  2018000600852  2018000600853  2018000600854  2018000600855  2018000600856  2018000600857  2018000600858  2018000600859  2018000600860  2018000600861  2018000600862  2018000600863  2018000600864  2018000600865  2018000600866  2018000600867  2018000600868  2018000600869  2018000600870  2018000600871  2018000600872  2018000600873  2018000600874  2018000600875  2018000600876  2018000600877  2018000600878  2018000600879  2018000600880  2018000600881  2018000600882  2018000600883  2018000600884  2018000600885  2018000600886  2018000600887  2018000600888  2018000600889  2018000600890  2018000600891  2018000600892  2018000600893  2018000600894  2018000600895  2018000600896  2018000600897  2018000600898  2018000600899  2018000600901  2018000600902  2018000600903  2018000600904  2018000600905  2018000600906  2018000600907  2018000600908  2018000600909  2018000600910  2018000600911  2018000600912  2018000600913  2018000600914  2018000600915  2018000600916  2018000600917  2018000600918  2018000600919  2018000600920  2018000600921  2018000600923  2018000600924  2018000600925  2018000600926  2018000600927  2018000600928  2018000600929  2018000600930  2018000600931  2018000600932  2018000600933  2018000600934  2018000600935  2018000600936  2018000600937  2018000600938  2018000600939  2018000600940  2018000600941  2018000600942  2018000600943  2018000600944  2018000600945  2018000600946  2018000600947  2018000600948  2018000600949  2018000600950  2018000600952  2018000600953  2018000600954  2018000600955  2018000600956  2018000600957  2018000600958  2018000600959  2018000600960  2018000600961  2018000600962  2018000600963  2018000600964  2018000600965  2018000600966  2018000600967  2018000600968  2018000600969  2018000600970  2018000600971  2018000600972  2018000600973  2018000600974  2018000600975  2018000600976  2018000600977  2018000600978  2018000600979  2018000600980  2018000600981  2018000600982  2018000600983  2018000600984  2018000600985  2018000600986  2018000600987  2018000600988  2018000600989  2018000600990  2018000600991  2018000600992  2018000600993  2018000600994  2018000600995  2018000600996  2018000600997  2018000600998  2018000600999  2018000601000  2018000601001  2018000601002  2018000601003  2018000601004  2018000601005  2018000601006  2018000601007  2018000601008  2018000601009  2018000601010  2018000601011  2018000601012  2018000601013  2018000601014  2018000601015  2018000601016  2018000601017  2018000601018  2018000601019  2018000601020  2018000601021  2018000601022  2018000601023  2018000601024  2018000601025  2018000601026  2018000601027  2018000601028  2018000601029  2018000601030  2018000601031  2018000601032  2018000601033  2018000601034  2018000601035  2018000601036  2018000601037  2018000601038  2018000601039  2018000601040  2018000601041  2018000601042  2018000601043  2018000601044  2018000601045  2018000601046  2018000601047  2018000601048  2018000601049  2018000601050  2018000601051  2018000601052  2018000601053  2018000601054  2018000601055  2018000601056  2018000601057  2018000601058  2018000601059  2018000601060  2018000601061  2018000601062  2018000601063  2018000601064  2018000601065  2018000601066  2018000601067  2018000601068  2018000601069  2018000601070  2018000601071  2018000601072  2018000601073  2018000601074  2018000601075  2018000601076  2018000601077  2018000601078  2018000601079  2018000601080  2018000601081  2018000601082  2018000601083  2018000601084  2018000601085  2018000601086  2018000601087  2018000601088  2018000601089  2018000601090  2018000601091  2018000601092  2018000601093  2018000601094  2018000601095  2018000601096  2018000601097  2018000601098  2018000601099  2018000601100  2018000601101  2018000601102  2018000601103  2018000601104  2018000601105  2018000601106  2018000601107  2018000601108  2018000601109  2018000601110  2018000601111  2018000601112  2018000601113  2018000601114  2018000601115  2018000601116  2018000601117  2018000601118  2018000601119  2018000601120  2018000601121  2018000601122  2018000601123  2018000601124  2018000601125  2018000601126  2018000601127  2018000601128  2018000601129  2018000601130  2018000601131  2018000601132  2018000601133  2018000601134  2018000601135  2018000601136  2018000601137  2018000601138  2018000601139  2018000601140  2018000601141  2018000601142  2018000601143  2018000601144  2018000601145  2018000601146  2018000601147  2018000601148  2018000601149  2018000601150  2018000601151  2018000601152  2018000601153  2018000601154  2018000601155  2018000601156  2018000601157  2018000601158  2018000601159  2018000601160  2018000601161  2018000601162  2018000601163  2018000601164  2018000601165  2018000601166  2018000601167  2018000601168  2018000601169  2018000601170  2018000601171  2018000601172  2018000601173  2018000601174  2018000601175  2018000601176  2018000601177  2018000601178  2018000601179  2018000601180  2018000601181  2018000601182  2018000601183  2018000601184  2018000601185  2018000601186  2018000601187  2018000601188  2018000601189  2018000601190  2018000601191  2018000601192  2018000601193  2018000601194  2018000601195  2018000601196  2018000601197  2018000601198  2018000601199  2018000601200  2018000601201  2018000601202  2018000601203  2018000601204  2018000601205  2018000601206  2018000601207  2018000601208  2018000601209  2018000601210  2018000601211  2018000601212  2018000601213  2018000601214  2018000601215  2018000601216  2018000601217  2018000601218  2018000601219  2018000601220  2018000601221  2018000601222  2018000601223  2018000601224  2018000601225  2018000601226  2018000601227  2018000601228  2018000601229  2018000601230  2018000601231  2018000601232  2018000601233  2018000601234  2018000601235  2018000601236  2018000601237  2018000601238  2018000601239  2018000601240  2018000601241  2018000601242  2018000601243  2018000601244  2018000601245  2018000601246  2018000601247  2018000601248  2018000601249  2018000601250  2018000601251  2018000601252  2018000601253  2018000601254  2018000601255  2018000601256  2018000601257  2018000601258  2018000601259  2018000601260  2018000601261  2018000601262  2018000601263  2018000601264  2018000601265  2018000601266  2018000601267  2018000601268  2018000601269  2018000601270  2018000601271  2018000601272  2018000601273  2018000601274  2018000601275  2018000601276  2018000601277  2018000601278  2018000601279  2018000601280  2018000601281  2018000601282  2018000601283  2018000601284  2018000601285  2018000601286  2018000601287  2018000601288  2018000601289  2018000601290  2018000601291  2018000601292  2018000601293  2018000601294  2018000601295  2018000601296  2018000601297  2018000601298  2018000601299  2018000601300  2018000601301  2018000601302  2018000601303  2018000601304  2018000601305  2018000601306  2018000601307  2018000601308  2018000601309  2018000601310  2018000601311  2018000601312  2018000601313  2018000601314  2018000601315  2018000601316  2018000601317  2018000601318  2018000601319  2018000601320  2018000601321  2018000601322  2018000601323  2018000601324  2018000601325  2018000601326  2018000601327  2018000601328  2018000601329  2018000601330  2018000601331  2018000601332  2018000601333  2018000601334  2018000601335  2018000601336  2018000601337  2018000601338  2018000601340  2018000601341  2018000601342  2018000601343  2018000601344  2018000601345  2018000601346  2018000601347  2018000601348  2018000601349  2018000601350  2018000601351  2018000601352  2018000601353  2018000601354  2018000601355  2018000601356  2018000601357  2018000601358  2018000601359  2018000601360  2018000601361  2018000601362  2018000601363  2018000601364  2018000601365  2018000601366  2018000601368  2018000601369  2018000601370  2018000601371  2018000601372  2018000601373  2018000601374  2018000601375  2018000601376  2018000601377  2018000601378  2018000601379  2018000601380  2018000601381  2018000601382  2018000601383  2018000601384  2018000601385  2018000601386  2018000601387  2018000601388  2018000601389  2018000601390  2018000601391  2018000601392  2018000601393  2018000601394  2018000601395  2018000601396  2018000601397  2018000601398  2018000601399  2018000601400  2018000601401  2018000601402  2018000601403  2018000601404  2018000601405  2018000601406  2018000601407  2018000601408  2018000601409  2018000601410  2018000601411  2018000601412  2018000601413  2018000601414  2018000601415  2018000601416  2018000601418  2018000601419  2018000601420  2018000601422  2018000601442  2018000601443  2018000601444  2018000601445  2018000601446  2018000601447  2018000601448  2018000601449  2018000601451  2018000601452  2018000601455  2018000601456  2018000601457  2018000601458  2018000601459  2018000601460  2018000601461  2018000601463  2018000601464  2018000601465  2018000601466  2018000601467  2018000601468  2018000601469  2018000601470  2018000601471  2018000601472  2018000601473  2018000601474  2018000601475  2018000601476  2018000601477  2018000601478  2018000601479  2018000601480  2018000601481  2018000601482  2018000601483  2018000601484  2018000601485  2018000601490  2018000601491  2018000601492  2018000601493  2018000601494  2018000601495  2018000601496  2018000601497  2018000601498  2018000601499  2018000601500  2018000601501  2018000601502  2018000601503  2018000601504  2018000601505  2018000601506  2018000601507  2018000601508  2018000601509  2018000601510  2018000601511  2018000601512  2018000601513  2018000601514  2018000601515  2018000601516  2018000601517  2018000601518  2018000601519  2018000601520  2018000601521  2018000601522  2018000601523  2018000601524  2018000601525  2018000601526  2018000601527  2018000601528  2018000601529  2018000601530  2018000601531  2018000601532  2018000601533  2018000601534  2018000601535  2018000601536  2018000601537  2018000601538  2018000601539  2018000601540  2018000601541  2018000601542  2018000601543  2018000601544  2018000601545  2018000601546  2018000601547  2018000601548  2018000601549  2018000601550  2018000601551  2018000601552  2018000601553  2018000601554  2018000601555  2018000601556  2018000601557  2018000601558  2018000601559  2018000601560  2018000601561  2018000601562  2018000601563  2018000601564  2018000601565  2018000601566  2018000601567  2018000601568  2018000601569  2018000601570  2018000601571  2018000601572  2018000601573  2018000601574  2018000601575  2018000601576  2018000601577  2018000601578  2018000601579  2018000601580  2018000601581  2018000601582  2018000601583  2018000601584  2018000601585  2018000601586  2018000601587  2018000601588  2018000601589  2018000601590  2018000601591  2018000601592  2018000601593  2018000601594  2018000601595  2018000601596  2018000601597  2018000601598  2018000601599  2018000601600  2018000601601  2018000601602  2018000601603  2018000601604  2018000601605  2018000601606  2018000601607  2018000601608  2018000601609  2018000601610  2018000601611  2018000601612  2018000601613  2018000601614  2018000601615  2018000601616  2018000601617  2018000601618  2018000601619  2018000601620  2018000601621  2018000601622  2018000601623  2018000601624  2018000601625  2018000601626  2018000601627  2018000601628  2018000601629  2018000601630  2018000601631  2018000601632  2018000601633  2018000601634  2018000601635  2018000601636  2018000601637  2018000601638  2018000601639  2018000601640  2018000601641  2018000601642  2018000601643  2018000601644  2018000601645  2018000601646  2018000601647  2018000601648  2018000601649  2018000601650  2018000601651  2018000601652  2018000601653  2018000601654  2018000601655  2018000601656  2018000601657  2018000601658  2018000601659  2018000601660  2018000601661  2018000601662  2018000601663  2018000601664  2018000601665  2018000601666  2018000601667  2018000601668  2018000601669  2018000601670  2018000601671  2018000601672  2018000601673  2018000601674  2018000601675  2018000601676  2018000601677  2018000601678  2018000601679  2018000601680  2018000601681  2018000601682  2018000601683  2018000601684  2018000601685  2018000601686  2018000601687  2018000601688  2018000601689  2018000601690  2018000601691  2018000601692  2018000601693  2018000601694  2018000601695  2018000601696  2018000601697  2018000601698  2018000601699  2018000601700  2018000601701  2018000601702  2018000601703  2018000601704  2018000601705  2018000601706  2018000601707  2018000601708  2018000601709  2018000601710  2018000601711  2018000601712  2018000601713  2018000601714  2018000601715  2018000601716  2018000601717  2018000601718  2018000601719  2018000601720  2018000601721  2018000601722  2018000601723  2018000601724  2018000601725  2018000601726  2018000601727  2018000601728  2018000601729  2018000601730  2018000601731  2018000601732  2018000601733  2018000601734  2018000601735  2018000601736  2018000601737  2018000601738  2018000601739  2018000601740  2018000601741  2018000601742  2018000601743  2018000601744  2018000601745  2018000601746  2018000601747  2018000601748  2018000601749  2018000601750  2018000601751  2018000601752  2018000601753  2018000601754  2018000601755  2018000601756  2018000601757  2018000601758  2018000601759  2018000601760  2018000601761  2018000601762  2018000601763  2018000601764  2018000601765  2018000601766  2018000601767  2018000601768  2018000601769  2018000601770  2018000601771  2018000601772  2018000601773  2018000601774  2018000601775  2018000601776  2018000601777  2018000601778  2018000601779  2018000601780  2018000601781  2018000601782  2018000601783  2018000601784  2018000601785  2018000601786  2018000601787  2018000601788  2018000601789  2018000601790  2018000601791  2018000601792  2018000601793  2018000601794  2018000601795  2018000601796  2018000601797  2018000601798  2018000601799  2018000601800  2018000601802  2018000601803  2018000601804  2018000601805  2018000601806  2018000601807  2018000601808  2018000601809  2018000601811  2018000601813  2018000601814  2018000601815  2018000601816  2018000601817  2018000601818  2018000601820  2018000601821  2018000601822  2018000601825  2018000601826  2018000601827  2018000601828  2018000601829  2018000601830  2018000601831  2018000601832  2018000601833  2018000601834  2018000601835  2018000601836  2018000601837  2018000601838  2018000601839  2018000601840  2018000601841  2018000601842  2018000601843  2018000601844  2018000601845  2018000601846  2018000601847  2018000601848  2018000601849  2018000601850  2018000601851  2018000601852  2018000601853  2018000601854  2018000601855  2018000601856  2018000601857  2018000601858  2018000601859  2018000601860  2018000601861  2018000601862  2018000601863  2018000601864  2018000601865  2018000601866  2018000601867  2018000601868  2018000601869  2018000601870  2018000601871  2018000601872  2018000601873  2018000601874  2018000601875  2018000601876  2018000601877  2018000601878  2018000601879  2018000601880  2018000601881  2018000601882  2018000601883  2018000601885  2018000601886  2018000601887  2018000601888  2018000601889  2018000601890  2018000601891  2018000601893  2018000601894  2018000601895  2018000601896  2018000601897  2018000601898  2018000601899  2018000601900  2018000601901  2018000601902  2018000601903  2018000601904  2018000601906  2018000601907  2018000601908  2018000601909  2018000601910  2018000601911  2018000601912  2018000601914  2018000601915  2018000601916  2018000601917  2018000601918  2018000601919  2018000601921  2018000601922  2018000601925  2018000601927  2018000601928  2018000601929  2018000601930  2018000601932  2018000601933  2018000601934  2018000601935  2018000601936  2018000601937  2018000601939  2018000601940  2018000601941  2018000601942  2018000601943  2018000601944  2018000601945  2018000601946  2018000601947  2018000601948  2018000601949  2018000601950  2018000601951  2018000601952  2018000601953  2018000601954  2018000601955  2018000601956  2018000601957  2018000601958  2018000601959  2018000601960  2018000601961  2018000601962  2018000601963  2018000601964  2018000601965  2018000601966  2018000601967  2018000601968  2018000601970  2018000601971  2018000601972  2018000601975  2018000601976  2018000601977  2018000601978  2018000601979  2018000601980  2018000601981  2018000601982  2018000601983  2018000601984  2018000601985  2018000601986  2018000601987  2018000601988  2018000601989  2018000601990  2018000601991  2018000601992  2018000601993  2018000601994  2018000601995  2018000601996  2018000601997  2018000601998  2018000601999  2018000602000  2018000602001  2018000602002  2018000602003  2018000602004  2018000602005  2018000602006  2018000602008  2018000602009  2018000602010  2018000602011  2018000602012  2018000602013  2018000602014  2018000602015  2018000602016  2018000602017  2018000602018  2018000602019  2018000602020  2018000602021  2018000602022  2018000602023  2018000602024  2018000602025  2018000602026  2018000602027  2018000602028  2018000602029  2018000602030  2018000602031  2018000602033  2018000602035  2018000602037  2018000602094  2018000700001  2018000700002  2018000700003  2018000700004  2018000700005  2018000700006  2018000700007  2018000700008  2018000700009  2018000700010  2018000700011  2018000700012  2018000700013  2018000700014  2018000700015  2018000700016  2018000700017  2018000700018  2018000700019  2018000700020  2018000700021  2018000700022  2018000700023  2018000700024  2018000700025  2018000700026  2018000700027  2018000700028  2018000700029  2018000700030  2018000700031  2018000700032  2018000700033  2018000700034  2018000700035  2018000700037  2018000700038  2018000700039  2018000700040  2018000700041  2018000700042  2018000700043  2018000700044  2018000700045  2018000700046  2018000700048  2018000700049  2018000700051  2018000700052  2018000700053  2018000700054  2018000700055  2018000700058  2018000700059  2018000700060  2018000700061  2018000700062  2018000700063  2018000700064  2018000700065  2018000700066  2018000700067  2018000700068  2018000700069  2018000700070  2018000700071  2018000700072  2018000700073  2018000700074  2018000700075    ***added 5/13/19***  2018000602521  2018000602522  2018000602523  2018000602524  2018000602525  2018000602526  2018000602527  2018000602528  2018000602529  2018000602530  2018000602531  2018000602532  2018000602533  2018000602534  2018000602535  2018000602536  2018000602537  2018000602538  2018000602539  2018000602540  2018000602541  2018000602542  2018000602543  2018000602544  2018000602545  2018000602546  2018000602547  2018000602548  2018000602549  2018000602550  2018000602551  2018000602552  2018000602553  2018000602554  2018000602555  2018000602556  2018000602557  2018000602558  2018000602559  2018000602560  2018000602561  2018000602562  2018000602563  2018000602564  2018000602565  ***Updated 9/9/19***  2017000600882",
      "recall_initiation_date": "20181010",
      "postal_code": "49002-5826",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}